Abstract
Effects of stroke therapies area highly time dependent but onset‐to‐treatment times for recanalizing treatment are mostly beyond optimal time windows. Mobile stroke units shorten time to treatment and improve functional outcome. However, costs and logistical demands preclude their widespread use. There is a pressing need for alternative prehospital diagnostic tools that enable early, on‐site treatment and facilitate targeted transport in patients needing interventional therapies. This systematic review aimed at examining the current landscape of noncomputed tomography‐based technologies for prehospital stroke diagnosis. We conducted a literature search of original research papers on predetermined technologies and through systematic scouting on PubMed and included papers in English found through MEDLINE and Embase, published until October 2025. Relevant papers not identified by the systematic search but known to the authors were included and indicated as such. The final review included 105 publications evaluating innovations in microwave, transcranial eddy current damping, volumetric impedance phase shift spectroscopy, ultrasound, magnetic resonance imaging, electroencephalography, near‐infrared spectroscopy, blood biomarkers, telemedicine, and artificial intelligence. Each technology's potentials, limitations, and clinical feasibility for enhancing prehospital diagnosis and therapeutic and monitoring utility were assessed. No single technology provides capabilities similar to noncomputed tomography in terms of detection and differentiation of ischemic and hemorrhagic stroke. A combination of ultrasound and near‐infrared spectroscopy may complement each other in hemorrhage detection. Electroencephalography shows most promise in identifying large vessel occlusion. Telemedicine can be used for reliable prehospital stroke assessment. Future research should target both the improvement of individual technologies as well as combining different modalities.
Keywords: artificial intelligence, diagnosis, prehospital, stroke, technology, telemedicine, ultrasound
Subject Categories: Diagnostic Testing
Nonstandard Abbreviations and Acronyms
- ECD
eddy current damping
- GFAP
glial fibrillary acidic protein
- ICH
intracerebral hemorrhage
- LVO
large vessel occlusion
- MSU
mobile stroke unit
- MWI
microwave imaging
- NIRS
near‐infrared spectroscopy
Reperfusion therapies like intravenous thrombolysis and mechanical thrombectomy significantly improve functional outcomes after acute ischemic stroke, but their effectiveness relies on swift treatment. 1 , 2 Studies have shown a dramatic improvement in outcomes when treatment occurs within 60 minutes of onset, the “Golden Hour,” 3 yet median onset to treatment times for in‐hospital thrombolysis typically exceed 100 minutes. Mobile stroke units (MSUs) with onboard computed tomography (CT) scanners were developed to shift stroke workup to the prehospital setting, and recent studies confirm that MSUs not only shorten time to treatment but also improve outcomes. 4 This has resulted in a class 1 recommendation in the latest guideline of the American Heart/Stroke Association. 5 However, MSUs are costly, pose logistical challenges, and rely on precise dispatching, limiting their scalability particularly outside of metropolitan areas. Hence, stroke diagnostic technologies designed for usage in normal ambulances could leverage the prehospital treatment concept for a much higher proportion of patients with stroke in the field.
Although mechanical thrombectomy remains an in‐hospital procedure at specialized centers, delays are common due to secondary transfers from facilities without endovascular capabilities. Ambulance‐based stroke workup with vessel imaging could identify patients with large‐vessel occlusions (LVOs) and route them directly to mechanical thrombectomy‐capable centers. Prehospital diagnosis of stroke subtype and vessel occlusion could also allow hospitals to be prenotified, speeding up treatment, enabling enhanced thrombolytic therapy (eg, sonothrombolysis), and facilitating hemostatic treatment for patients with intracerebral hemorrhage (ICH). This systematic review outlines the current status of non‐CT‐based stroke diagnostics for prehospital use complemented by critical appraisals from researchers involved in these technological developments.
METHODS
Search Strategy
In order to determine whether important emerging technologies had been missed in our initial scouting, we conducted a systematic search using the terms (“Stroke”[Mesh] OR stroke OR “Ischemic Stroke”[Mesh] OR “Ischemic Stroke” OR “Hemorrhagic Stroke”[Mesh] OR “Hemorrhagic Stroke” OR “Cerebral Hemorrhage”[Mesh] OR “Cerebral Hemorrhage”) AND (“pre‐hospital” OR prehospital OR “Emergency Medical Services”[Mesh] OR “emergencies”[Mesh] OR “ambulances”[Mesh] OR ambulance) AND (“Stroke/diagnosis”[Mesh] OR “Stroke/diagnostic imaging”[Mesh]) in PubMed. Headlines were screened for technologies and approaches. Next, we conducted a systematic review by entering the comprehensive search terms for every technology in PubMed and Embase (Supplemental Material 1). Databases were last consulted on October 30, 2025. Literature not identified through the systematic search but already known by the authors was also considered. For each technology, 2 authors independently reviewed titles and abstracts for eligibility. Discrepancies were adjudicated by an independent referee. Finally, full texts were reviewed by 2 independent reviewers.
Selection Criteria
After removal of duplicates, our initial search identified 7609 database entries. Studies were included in the final review if they were original articles, written in English language, describing technologies and processes for prehospital stroke diagnosis. The final analysis included 136 studies (Figure1). All technologies were reviewed according to predefined criteria (Table1).
Figure 1. Preferred Reporting Items for Systematic Reviews and Meta‐Analyses flow diagram for article identification.

Searches were conducted in Medline via PubMed and Embase. Additional sources were articles already known to the authors.
Table 1.
Technology Rating Matrix
| Microwave imaging | ECD sensors | VIPS | EEG | MRI | NIRS | Biomarkers | Ultrasound | Telemedicine | AI | |
|---|---|---|---|---|---|---|---|---|---|---|
| Needed space and weight | ++ | ++ | ++ | ++ | − | ++ | ++ | ++ | + | ++ |
| Ability to differentiate stroke and stroke mimics | + | − | + | + | ++ | − | + | − | ++ | ++ |
| Arterial vessel imaging | − | − | − | − | − | − | − | ++ | − | − |
| Perfusion imaging | − | − | − | − | + | + | − | + | +* | +* |
| (Indirect) evaluation of large vessel occlusion | + | + | + | + | + | + | + | ++ | ++ | + |
| Detection of deep intracerebral hematoma | + | + | + | − | + | − | + | ++ | − | + |
| Detection of superficial hematoma | + | + | + | − | + | ++ | + | − | − | + |
| Spatial resolution | + | + | − | + | ++ | + | − | ++ | + | + |
| Blind spots in supratentorial cavity | − | − | − | + | − | + | − | + | − | n.a. |
| Blind spots in posterior fossa | + | + | − | ++ | + | ++ | − | ++ | − | − |
| Susceptibility to motion artifacts | − | − | − | ++ | + | − | − | + | − | − |
| Potential therapeutic utility | − | − | − | − | − | − | − | + | + | + |
| Potential monitoring utility | + | − | − | + | − | + | + | + | ++ | + |
− indicates Week; +, Middle; ++, Strong; AI, artificial intelligence; ECD, eddy current damping; EEG, electroencephalography; MRI, magnetic resonance imaging; n.a., not applicable; NIRS, near‐infrared spectroscopy; and VIPS, volumetric impedance phase shift spectroscopy.
By indirect assessment using clinical severity.
RESULTS
Microwave Imaging
Microwave Imaging (MWI), also termed electromagnetic brain scanning, is a portable and low‐cost diagnostic tool that uses tissue dielectric variations to differentiate between healthy and abnormal tissues. 6 Initially used for breast cancer detection, MWI has shown potential for stroke diagnosis. 7 , 8 , 9 , 10 MWI works by transmitting microwave signals through the brain using distributed antennas around the head for enhanced sensitivity. 6 , 7 , 8 , 9 , 10 , 11 Each antenna serves as a transmitter one at a time, and the remaining antennas act as receivers. 8 The scattered wave pattern resulting from variations in dielectric contrast are analyzed to differentiate between healthy tissue and pathological substrates. 7 , 8 , 9 , 11
Two main approaches, microwave tomography and ultra‐wideband radar imaging, are employed. 6 Microwave tomography reconstructs tissue morphology and electromagnetic characteristics but requires solving complex inverse problems and a training phase. 6 Ultra‐wideband radar imaging locates scattering points, generating energy maps. 6 Computational advances in the past 5 to 10 years have made it clinically viable. 7 , 8
Four MWI devices studies had published scientific evaluations 12 : EMTensor (EMTensor, Inc., Newport Beach, CA, and EMTensor, GmbH, Vienna, Austria), EMvision (EMvision Medical Devices, Ltd., Brisbane, Australia), StrokefinderTM MD100 (Medfield Diagnostics AB, Göteborg, Sweden), 12 , 13 and TES HT100 (B&B Srl, Naples, Italy). 14
Ljungqvist et al. demonstrated MWI's effectiveness in detecting subdural hemorrhage, achieving an area under the curve (AUC) of 0.94 with 100% sensitivity and 75% specificity. 9 Persson et al. studied 9 patients with ICH and 11 patients with acute ischemic stroke and showed MWI could distinguish between the 2 with an AUC of 0.88. In a subsequent phase of the study, it correctly differentiated between ICH and IS (AUC, 0.85) and ICH and healthy volunteers (AUC, 0.87). 8 Abbosh et al. studied 50 patients with ischemic or hemorrhagic stroke and indicated promising results in detecting, localizing, and classifying strokes. 15
Montemarani et al. examined 23 patients with mild traumatic brain injury after hospital arrival and compared the results of a 5‐minute portable TES HT 100 device analysis with findings in cranial CT. After exclusion of 3 patients with storage error, they found complete concordance with regard to any brain lesion (14 patients with and 6 patients without brain lesion). No information was provided for the type or location of the brain lesion. 14
Transcranial Eddy Current Damping Sensors
Transcranial eddy current damping (ECD) sensing is a fairly recent, experimental approach to locating, distinguishing hemorrhagic and ischemic strokes. 16
The method uses a solenoid magnet to induce eddy currents in the radio‐frequency range within a tissue, whose magnitude depends on electric conductivity, determined by the tissue's composition. The resulting electromagnetic field reduces the inductance of the coil, dampening resonance in an oscillator circuit that is measured in real time.
The ECD device consists of a handheld probe containing ≥1 coils of different sizes and inductances, an integrated circuit that drives the coil and detect inductance changes, and a microcontroller. Data are collected by moving the probe over the head, transmitting the measurements to a computer for 2‐dimensional or 3‐dimensional mapping. Larger coils provide greater sensing depth but lower spatial resolution. ECD identifies lesions by sensing irregularities in tissue conductivity: hemorrhages increase conductivity due to ionic concentration (not restricted by cell membrane) in the hematoma, and ischemic lesions show reduced conductivity.
A study using an anthropomorphic head phantom achieved centimeter resolution in simulated hemorrhages of 30 to 50 mL, requiring 2.43 minutes for a complete scan. 16
A clinical study with eight subjects (3 hemorrhagic strokes, 2 ischemic, 3 healthy) demonstrated accurate lesion localization and differentiation among the stroke types within an average scan time of 15 minutes. 17 The investigators were blind with respect to the location and type of lesion, and all lesions were correctly localized and identified. The method showed greater sensitivity to hemorrhages than ischemic lesions. Further phantom tests estimated a sensing depth of up to 4.97 cm, potentially covering most of the human cranium. However, ECD's ability to differentiate intracerebral hematomas from other fluid‐filled structures like ventricles, arachnoid cysts, or lesions deeper than 5 cm remains uncertain. In a recent phantom study, the addition of machine learning (ML) further improved classification metrics. 18
Volumetric Impedance Phase Shift Spectroscopy
Volumetric impedance shift spectroscopy leverages bioimpedance when the distribution of fluids between compartments shifts in tissue affected by stroke. An array of low energy radio waves is sent from the back of a portable, helmet‐like device to the front, thereby being modified when passing through brain tissue. 19 In a study with 252 patients and healthy controls, a volumetric impedance phase shift spectroscopy device was able to differentiate subjects with severe stroke from the other participants with an AUC of 0.93 (95% CI, 0.89–0.96). 19
Electroencephalography and Somatosensory Evoked Potential
Electroencephalography (EEG) and somatosensory evoked potential have been addressed for prehospital stroke detection in various studies. EEG parameters tested included theta/alpha ratio, relative theta power, weighted phase lag index, 20 , 21 delta/alpha ratio, (delta + theta)/(alpha + beta) ratio, pairwise‐derived brain symmetry index, 22 fast frequency amplitudes, changes in relative delta power, or indices derived from more complex algorithms. 23 Regarding somatosensory evoked potentials, 1 study used complex automated algorithms. 23
Most studies focused on LVO detection, with only 1 conducted prehospital. 21 Although some studies explicitly restricted the examined area to the anterior circulation, 20 , 21 , 23 even in the remaining studies, no LVO was detected in the posterior circulation. 24
Accuracy varied: Sergot et al. reported 80% sensitivity and specificity using a portable EEG/somatosensory evoked potential device. 23 The ELECTRA (EEG Controlled Triage in the Ambulance for Acute Ischemic Stroke) study showed in patients with analyzable EEG measurements (212/311 [68%]) an AUC of 0.80 for the theta/alpha ratio, with 50% sensitivity and 83% specificity. The Brain Symmetry Index in the delta band had an AUC of 0.91, with 80% sensitivity and 93% specificity. 21 In an emergency department (ED) study by the same group, combining theta power and weighted phase lag index yielded 100% sensitivity and 84% specificity for LVO detection in 65 patients with suspected stroke (ischemic or hemorrhagic) or confirmed LVO. 20
Other studies focused on stroke detection per se, and the definition of stroke varied. Wilkinson et al. tested a consumer wireless EEG system in patients with subacute ischemic stroke against healthy controls and stroke mimics. 22 Delta/alpha ratio and (delta + theta)/(alpha + beta) ratio increased with stroke severity with significant differences between large strokes and both small strokes and controls. A model differentiating between minor strokes/controls and moderate/severe strokes had a 76% accuracy with a sensitivity of 63% and a specificity of 86%.
Michelson et al. compared EEG to CT in and patients with acute ischemic stroke and ICH and stroke mimics (migraine and syncope). Sensitivity of stroke detection was 91.7%, specificity 50.4%. Eighty percent of patients with ischemic stroke with normal CT findings were correctly identified as stroke by EEG at the time of CT scan, whereas no differentiation between ischemic and hemorrhagic stroke was attempted. 25
Portable EEG devices, especially dry electrode cap systems, reduced setup times to around 5 minutes, with 1‐hour training sufficient for examiner‐independent results. 20 , 21 , 22 , 23 One study conducted in the prehospital setting in patients with neurological symptoms reported easy application and EEG recordings usable for clinical interpretation but not specifically for stroke diagnosis. 26
The primary limitation of all studies using EEG is the high incidence of poor signal quality recordings, which consequently results in a high dropout rate. Numbers ranged between 10% and 34%. 20 , 23 , 27 This was particularly true for women, patients with long hair, and those with LVOs, who comprised the target population for the studies. 20 , 21 None of the studies differentiated LVO from hemorrhagic stroke, precluding early recombinant tissue plasminogen activator administration in prehospital settings. Deep brain regions remain challenging for accurate detection.
Portable Magnetic Resonance Imaging
Magnetic resonance imaging (MRI) is a noninvasive technology that uses a magnetic field and radiofrequency pulses to create detailed images of the brain including the posterior fossa, which is challenging for many other technologies. MRI is highly sensitive for detecting acute ischemic stroke, with diffusion‐weighted imaging showing changes within minutes of arterial occlusion. However, traditional MRI machines require large magnetic fields, advanced infrastructure, and cooling with cryogens. Recent developments have led to portable MRI scanners with ultra‐low field strength (<0.1 T), which are compact (140×90 cm), lightweight (650 kg), and operate on standard electrical outlets without the need for cryogens. 28 , 29 , 30 These devices are considerably cheaper, making them potentially more accessible even for low‐income communities in the future. 31 , 32 Portable MRI is currently most promising in critical care settings, enabling bedside scans for mechanically ventilated patients. However, caution is warranted in patients with pacemakers, cochlear implants, or deep brain stimulators. 33
For prehospital stroke care, portable MRI faces 2 main challenges; long image acquisition times and limited image quality. Current ultra‐low field scanners do not support perfusion imaging or time‐of‐flight angiography, essential for identifying patients needing endovascular therapy. Other low‐field (eg, 0.5 T) MR scanners with time‐of‐flight angiography capability are currently not portable (≈2000 kg). 32 , 34
Few studies have examined the use of portable MRI in stroke care. One prospective single‐center study in 2019 to 2020 included 50 critically ill patients with various conditions including ischemic stroke (n=9), hemorrhagic stroke (n=12), brain tumor, and brain injury. Portable MRI findings largely matched conventional imaging (CT or MRI), except for 1 missed subarachnoid hemorrhage, which could have had serious implications for stroke treatment. 33 In a subsequent study involving 144 individuals (56 ICH, 48 acute ischemic stroke), portable MRI showed 80.4% sensitivity for ICH, missing 11 cases. The lack of T2* and susceptibility‐weighted imaging likely contributed to this limitation. Specificity was high at 96.6%, with 3 patients misclassified as having ICH. 29 More recently, portable MRI was tested in 50 patients with subacute ischemic stroke. The diffusion‐weighted imaging sequence, critical for hyperacute stroke detection, had modest quality, showing 86% sensitivity. Five patients with small infarcts (4–10 mm) were missed. T2‐W and fluid‐attenuated inversion recovery sequences had higher sensitivity at 98% and 100%, respectively. 35 This portable MRI system (Swoop, Hyperfine, Inc., Guilford, CT) was the first to gain Food and Drug Administration approval in August 2020. It was successfully integrated into a modified cargo van, showing similar imaging quality to static systems. 31
Image acquisition times of 7 to 10 minutes are too slow for the hyperacute stroke setting. Sequences with long acquisition time increase the risk of motion artifacts, especially in agitated patients with stroke. Several teams are working to improve acquisition times and image quality using deep learning and undersampling. 36 , 37 Rigorous clinical studies are needed to assess whether faster acquisition approaches do not increase false negative and false positive findings.
Near‐Infrared Spectroscopy and Other Optical Technologies
Near‐infrared spectroscopy (NIRS) is a noninvasive, easy‐to‐apply technique that has been used for years to observe cerebral oxygenation by measuring the absorption of 690 to 830 nm light by erythrocytes approximately in a depth of up to 2 to 3 cm. NIRS has the potential capability to detect superficial cranial bleeds, such as subdural and superficial intracerebral hematomas or subarachnoid hemorrhages over the brain convexity. Depending on the device used, blood and blood oxygenation may be computed in 1 or in several locations simultaneously. The devices can be battery operated and are portable to a varying degree, making them potentially attractive for prehospital diagnostics. The few human studies have focused largely on traumatic intracranial hemorrhage and not on stroke. 38 , 39 , 40 , 41 , 42 , 43
In 1 study by Gopinath et al., hemispheric differences were observed in 167 traumatic brain injury cases, suggesting potential to identify hemorrhage, although controls were absent, leaving sensitivity and specificity unverified. 38 In a multicenter study of 365 trauma patients, NIRS showed 88% sensitivity and 90.7% specificity for intracranial hematomas >3.5 mL and within 2.5 cm from the skull's surface. However, for all hematoma locations, sensitivity fell to 68.7% with the same specificity. 39
Moreau et al. studied regional NIRS changes in 5 patients with stroke versus cadavers and healthy controls, observing differences that suggested NIRS may monitor stroke evolution, though data quality issues arose. 41
Muehlschlegel et al. evaluated interhemispheric asymmetry in 9 patients after ischemic stroke compared with controls, proposing NIRS might detect hemispheric microvascular hemodynamic asymmetry relevant to prehospital stroke assessment. 42
Similar to NIRS, 1 study used an optical device that illuminates tissue with high‐intensity laser light and measures speckle contrast to monitor cerebral blood flow. 44 In a cohort of 135 patients, the device was able to detect LVO with a sensitivity of 78.9% (95% CI, 65.5%–88.0%) and a specificity of 84.3% (95% CI, 74.7%–90.7%). 44
In summary, modern NIRS devices are portable and user friendly, yet high‐quality evidence for prehospital stroke diagnosis remains limited. Most studies did not address ischemic stroke, and applicability to early stroke is unclear. The studies are based on small case series, few provide matched control groups, and due to a significant risk of bias and imprecision, the level of current evidence is weak. Specificity and sensitivity in stroke identification have not been investigated and many other questions like device resilience to motion and other artifacts are yet to be addressed.
Blood‐Based Biomarkers
Innovative approaches like integrating blood‐based biomarkers in ambulances offer potential in prehospital stroke diagnosis. Despite many years of research, no biomarkers have become standard in clinical practice, but advancements in proteomics, metabolomics, transcriptomics, and genomics have led to new candidate biomarkers. 45
A proof‐of‐concept study from Helsinki showed that obtaining blood samples in ambulances could save around 30 minutes compared with in‐hospital sampling, with point‐of‐care measurements enabling earlier treatment decisions. 46
An MSU study found several elevated neuroinflammatory factors in patients with stroke only minutes after onset, supporting further exploration of ultra‐early immune response in stroke. Although promising, the sample size was small, and further studies are needed. 47
Attempts to differentiate stroke from mimics using biomarkers have largely been unsuccessful. In fact, a 21‐biomarker panel selected among previous studies and from the literature showed that the studied biomarkers were not sufficient for an accurate differential diagnosis of stroke. 48 Copeptin, a general outcome predictor, did not show discriminatory value between stroke and stroke mimics in the prehospital setting. 49 Recently, small noncoding RNAs have been identified as potential candidates; however, RNA sequencing is not yet feasible in the prehospital setting. 50 Symmetric dimethylarginine shows certain promise in severe stroke patients, but only in late sampling windows. 51
Among the most validated markers for prehospital use, GFAP (glial fibrillary acidic protein) shows potential in identifying ICH. Studies in hospital settings demonstrated GFAP's high sensitivity and specificity for diagnosing ICH. In a prehospital setting, Rozanski et al. collected plasma samples from ICH and ischemic stroke patients and measured GFAP using a fully automated immunoanalyzer platform, reporting a sensitivity of 36% and a specificity 100% for identifying ICH using a GFAP cutoff of 0.29 ng/mL, suggesting its potential for treatment initiation in patients with ICH. 52
Another study using ELISA reported that GFAP accurately (AUC, 0.86) differentiated ICH from ischemic stroke and mimics, regardless of age, severity, or time elapsed since onset, with GFAP being correlated to ICH volume. 53 Another prehospital study showed that serial prehospital GFAP measurements improved ultra‐early differentiation of stroke subtypes, while there was no additional value of the biomarker TAU in ICH identification. 54
Studies using a point‐of‐care device for GFAP measurements demonstrated similar predictive values, streamlining prehospital measurements. 55 , 56
In the Treat‐NASPP (Treat‐Norwegian Acute Stroke Prehospital Project), serum GFAP levels were measured in 299 patients with suspected stroke and onset ≤4.5 hours. 57 A decision tree combining GFAP serum levels with a prehospital stroke score was developed to differentiate between ICH, ischemic LVO/non‐LVO, and stroke mimics. This approach first identified patients with a low stroke probability, then used GFAP to detect ICH with 25% sensitivity and 100% specificity.
In ED studies, the diagnostic accuracy improved by adding ischemia‐related biomarkers like RBP‐4 (retinol binding protein 4) or NT‐proBNP (N‐terminal pro‐B‐type natriuretic peptide) to GFAP. If these findings are replicated in ambulances and rapid GFAP tests are developed, biomarker‐based stroke workups could enable targeted triage of patients with ICH (eg, directing them to hospitals with neurosurgery) and potentially allow for prehospital ICH treatment studies.
In a retrospective study, endogenous tissue‐type plasminogen activator levels were lower in patients with hemorrhagic stroke, and their inclusion improved classification accuracy when combined with prehospital stroke scales and vital parameters. 58
The ongoing BIOSHIP (Potential Use of a POCT [Point‐of‐Care Test] for the Direct Transfer of Stroke Patients From the Ambulance to the Thrombectomy Center) study (https://clinicaltrials.gov/study/NCT05833945) is evaluating a rapid point‐of‐care test for FABP (fatty acid binding protein) and NT‐proBNP in ambulances to detect LVO patients. 59 The addition of D‐dimer to FABP and NT‐proBNP might become a promising LVO biomarker panel. 60
Although blood‐based biomarkers in ambulances show promise for stroke diagnosis, challenges remain, including assay standardization, device availability, and emergency medical services (EMS) personnel training.
Transcranial Ultrasound
Transcranial ultrasound is a promising tool for stroke detection, primarily focusing on LVO and intracranial hemorrhage detection. 61 , 62 , 63 , 64 , 65 , 66 , 67 , 68
Studies show that experienced stroke neurologists using transcranial color‐coded duplex sonography can detect LVO, especially in the middle cerebral artery, with sensitivity up to 91% and specificity up to 98%. 63 , 64 , 66 Most studies used microbubble contrast agents, improving sensitivity and reducing insufficient bone window rates. Connolly et al. reported only 6% insufficiency using ultrasound contrast agent (SonoVue), 63 whereas typical rates of insufficient bone windows are around 20% in patients with stroke. 69 One case report used mobile brain perfusion ultrasound, a technique creating hemispheric perfusion curves following ultrasound contrast injection, to detect middle cerebral artery occlusion in a patient with acute stroke. 70
Antipova et al. reported a sensitivity of 63% for hemorrhage detection in a small cohort (10/18 patients). 62 A more recent study identified intracerebral hemorrhages in 87.2%. 71 Localization of the hemorrhages might be more critical than size. This is especially true for cortical and deep infratentorial hemorrhages, albeit Kukulska‐Pawluczuk et al. were able to detect all hemorrhages located in the brainstem and cerebellum. 71
Ultrasound examinations are quick, with times between 2 and 6 minutes depending on scope (eg, vessel focus, contrast use). 63 , 65 , 66
With its easy portability, devices can be used at the patient's side, in ambulances, or even helicopters. Perfusion imaging was not addressed in these studies, but it could enhance future stroke diagnostics.
Ultrasound is examiner dependent, which limits its use on ambulances at the scene. Connolly et al. observed significant examination time differences between novice and expert sonographers. 63 Experience also likely affects detection accuracy for hemorrhages and occlusions. This could be addressed by having trained specialists evaluate ultrasound images via telemedicine, with relatively novice ambulance staff handling the scanning. In a study conducted by Antipova et al., novice ultrasound users received a 3‐hour training and were afterwards able to perform simple B‐Mode and transcranial color‐coded duplex sonography examinations on healthy volunteers with satisfactory results. 61 Mort et al. demonstrated a multichannel communication device that livestreams ultrasound images to hospital‐based experts, though broadband coverage remains a challenge in rural areas. 72 Automated or robotic assisted transcranial color‐coded duplex sonography devices may be an alternative to facilitate the use of ultrasound in normal ambulances. The biggest challenge remains the reliable exclusion of intracranial bleeding, which is necessary for the administration of systemic thrombolysis therapy. Smaller, superficial, and infratentorial hemorrhages are particularly challenging for ultrasound‐detection. Experimental procedures, such as the application of ultrasonic shear wave elastography, could possibly improve the negative predictive value in the future. 73 , 74
Telemedicine
Telemedicine using video examination enables reliable remote assessment of stroke related neurological deficits. 75 , 76 , 77 , 78 Given the time‐dependent effects of intravenous thrombolysis, telemedicine‐based prehospital assessments have been proposed as a means of quality improvement since the early times of telemedicine stroke care. 79 Initially, limited bandwidth bore the risk of slow or unstable data transmissions. 79 , 80 , 81 With the advent of 4G networks, remote examination before or during EMS transports became more reliable and multichannel connectivity solutions further improved performance. 80 , 82 , 83
Technical failures during video examination have been reported in most studies, ranging between 60% 81 when only 3G was available to only 7% in a recent report from New Zealand. 84
Telemedicine has also been used for transmitting data from CT imaging 85 on an MSU and from other diagnostic devices such as ultrasound. 72
Several studies investigated the feasibility and reliability of simplified neurological assessments 86 and examinations without assistance in driving ambulances, 87 often using standardized scenarios 86 or actors simulating acute stroke. 77 , 81 , 82 , 87 , 88 , 89 Usually, comparisons between onsite and remote assessments performed during ambulance transport 81 , 82 , 90 showed moderate (weighted kappa 0.4–0.5) 89 or moderate to good (weighted Ƙ=0.69) 81 reliability.
Real‐life reliability has been tested only on MSUs comparing onsite assessment by the MSU neurologist and telestroke examination. 91 Interrater reliability was high for the National Institutes of Health Stroke Scale sum scores (Ƙ=0.86). Generalizability was limited by the fact that neurological assessments were conducted at the scene and not en route.
Satisfaction rates of prehospital telemedicine users were rated high (>80%) 92 and audio/video quality rating improved with higher bandwidth available in 4G and 5G telecommunication standards. 77 , 81 , 82 , 91
Studies report varying impacts of prehospital telestroke on process times. In Germany, early video assessments by nonneurologists in real patients with stroke showed no significant differences in pre‐ and in‐hospital times. 80 Duration of neurologists' telemedical assessment ranged between 8 and 19 minutes for the examination at the scene in MSUs 85 , 91 and 18 minutes during transport. 89 Hospital prenotification via paramedic‐based stroke assessment and checklist documentation was associated with 4 minutes shorter door‐to‐needle times and higher percentage of door‐to‐needle times ≤20 or 60 minutes in the German Stroke Angel project. 93 Telephone‐based assessment by neurologists or purpose‐trained residents was associated with no difference in onset‐to‐needle times for intravenous thrombolysis but a 69‐minute shorter onset to puncture time for thrombectomy compared with a preimplementation historical cohort in Stockholm. 94
The implementation of a joint paramedic and remote physician video examination on ambulances was associated with time saving in hospital‐arrival‐to‐CT time of 5.9 minutes. 95 A 17‐minute shorter door‐to‐needle time and 87‐minute shorter onset‐to‐groin time compared with “standard telestroke” was reported after extension of the telemedicine network using telemedicine in EMS assessment in South Carolina. 96
Decision accuracy has generally been reported as high in several studies. 82 , 84 , 85 , 91 , 97
Paramedic‐based stroke assessment and telephone‐based consultation by hospital stroke physicians showed 87% accuracy in predicting LVO stroke in the Stockholm Stroke Triage System.
A cluster randomized trial in New Zealand showed better accuracy in identification of reperfusion candidates (intravenous thrombolysis and endovascular thrombectomy) by neurologist‐guided telestroke assessment en route to hospital (100%) than a large‐vessel score performed by paramedics (71%). 84
In an adjusted comparison with historical controls, modified Rankin Scale score 0 to 1 in thrombectomy‐eligible patients significantly improved after implementation of the Stockholm Stroke Triage Project. 98
Only a small proportion of disagreements regarding stroke and nonstroke diagnosis was reported when comparing onsite and remote (video‐based) assessment of acute patients with suspected stroke on an MSU by neurologists. 91 However, a German study found that remote, nonneurologist assessments were 91% accurate for stroke diagnosis, there was a 68% misdiagnosis rate for epileptic seizures, frequently misidentified as stroke. 99
Artificial Intelligence
The increasing availability of digital neuromonitoring, new sensor, and wearable technologies as well as computer power may offer opportunities to improve prehospital management of acute stroke. Artificial intelligence methods play a crucial role in this process in order to extract the relevant information from these data throughout the stroke care pathway, including diagnosis, consultation, patient transfer, treatment and rehabilitation. 100 Artificial intelligence methods based on parameters from various sensor systems, including accelerometers, gyroscopes, magnetometers, computer vision systems, audio, pressure sensors, mechanical and optical keyboard press sensors, and microwaves have been explored for stroke detection and management. These technologies aim to detect biometric markers such as speech, motor skills, gait, and sensory changes to identify early signs of stroke. 7 , 101 , 102 , 103
Artificial intelligence can also improve prehospital triage by automated speech recognition software. 104 ML models have demonstrated significantly better prediction performance in identifying patients with stroke in EDs using factors like presenting symptoms, triage level, medical history, vital signs, and presence of atrial fibrillation. 100 , 102 , 105 , 106 More recently, decision‐support models have also been proposed in the form of smartphone applications to optimize EMS stroke triage, potentially reducing hospital arrival times. 107
Different modalities and artificial intelligence approaches have been investigated for their potential to detect and classify stroke symptoms. Video‐based approaches showed promise of automatically assessing orofacial stroke impairments. 108 Other motor symptoms may also be detected using sensors and ML. One validation study assessed motor weakness and graded National Institutes of Health Stroke Scale and Medical Research Council scores of 4 limbs similar to clinical examinations performed by medical staff. 103 Specifically, the proposed system composed of a measuring unit with wearable sensors and a grading unit trained with ML was able to grade National Institutes of Health Stroke Scale scores with an accuracy of 83.3% and an AUC of 0.912.
Artificial intelligence models have also been employed to differentiate between ischemic stroke and stroke mimics, such as using artificial neural networks in the ED to recognize acute ischemic stroke. 109 Similarly, ML models using National Institutes of Health Stroke Scale components have been successful in predicting LVOs in the ED. 100
Although progress has been made in applying artificial intelligence methods for prehospital stroke management, challenges include attaining sufficient specificity and sensitivity, determining the best strategies to link automated stroke symptom detection to automated notification for help and validating findings in external and larger patient cohorts.
CONCLUSIONS
This systematic review on non‐CT‐based technologies highlights a broad spectrum of prehospital tools for stroke diagnosis. Although these tools aim to facilitate early diagnosis and streamline the triage process, a critical analysis indicates that many technologies may provide results only by complimentary use, suggesting a multimodal approach to optimizing prehospital stroke care. Although we conducted a comprehensive review of the technologies, their heterogeneity precluded meta‐analysis of the data.
MWI offers a portable, possibly low‐cost method for stroke diagnosis and the potential of distinguishing between ischemic and hemorrhagic strokes. Early studies show promise, but further research is needed to validate its accuracy for prehospital use. Portable EEG devices have shown moderate to high accuracy in prehospital LVO detection. Quick application times and tolerability are clear advantages, but limitations include high dropout rates due to bad signal quality. The very limited capability to distinguish ischemic from hemorrhagic strokes precludes application to the use of recombinant tissue plasminogen activator in prehospital settings. NIRS potentially offers a straightforward approach for detecting superficial cranial hemorrhages, though its applicability for deeper hemorrhages and ischemic strokes is constrained. NIRS's limited specificity could be mitigated by combining it with other biomarkers or ultrasound, potentially allowing a layered approach to hemorrhage detection. A downside of many of the discussed technologies is their inability to produce visual images; radiologists and neurologists are accustomed to making diagnostic decisions based on actual imaging rather than on numerical data alone.
Portable MRI, in contrast, shows potential for stroke imaging, allowing bedside examinations in critically ill patients. However, long image acquisition times can delay treatment, and current systems for potential prehospital use lack essential capabilities like perfusion imaging and time‐of‐flight angiography. Additionally, concerns about motion artifacts and modest quality of diffusion‐weighted imaging limit its diagnostic applicability.
Transcranial ultrasound holds significant promise particularly for LVO detection, although examiner dependence is a limitation. Telemedicine or robotic devices may alleviate this issue. Further research is needed on hemorrhage detection, but ultrasound's portability, low cost, and speed make it a strong candidate for broad use. Combining it with NIRS could potentially mitigate its weaknesses in superficial hemorrhage detection and offer almost complete coverage for ICH exclusion. 110
ECD is still in an experimental phase but shows promise as a portable, low‐cost and safe method to localize and characterize stroke lesions. Current data are insufficient to evaluate diagnostic accuracy and further research is therefore needed to determine ECD's effectiveness across stroke types, sizes, and locations and to assess the impact of operator skills on scan quality.
Although only a single study has evaluated a volumetric impedance phase shift spectroscopy device, its portability and high accuracy in detecting severe strokes make it a promising tool for prehospital triage.
Telemedicine has proven to be effective for remote stroke assessment, with high diagnostic accuracy, improved process times, and growing reliability due to advancements in telecommunication. The continued enhancement of connectivity enables remote use of new diagnostic modalities with high bandwidth requirements. Combined with robotic technologies for automated examinations, it might allow for widespread use of specialized assessments like transcranial color‐coded duplex sonography.
New technologies will provide an added value in prehospital stroke management only if they exceed the accuracy of prehospital stroke scales such as the Rapid Arterial Occlusion Evaluation E or the Los Angeles Motor Scale scale–either alone or in conjunction with such scales. This has been shown for telemedicine and in a small study for volumetric impedance phase shift spectroscopy but not yet for other technologies, although biomarkers such as GFAP can obviously identify intracerebral hemorrhages better than clinical assessments.
Future research should prioritize the clinical integration of these tools, especially by leveraging advances in ML, deep learning, and telecommunication technologies. Before new technologies are implemented in routine emergency care, they need to be thoroughly tested in the prehospital setting, including evaluations of their effects on process quality (eg, times from EMS arrival to start of treatment) and clinical outcomes. These innovations may significantly affect stroke survival and recovery by accelerating prehospital stroke diagnosis and enabling targeted, timely treatment.
Sources of Funding
This review did not receive any specific grant from funding agencies in the public, commercial, or not‐for‐profit sectors.
Disclosures
Heinrich J. Audebert declares having received consultancy and speaker honoraria from Bayer Healthcare, Boehringer Ingelheim, Bristol Myers Squibb, Novo Nordisk, Eli Lilly and Sanofi. He has received institutional grants from the German Research Foundation, the German Federal Ministry of Education and research, the Stiftung Deutsche Schlaganfall‐Hilfe, and the German Innovation Fund. During the preparation of the current work, he served as president of the PREhospital Stroke Organisation. Joan Montaner holds the patent “Method for selecting a patient for a reperfusion therapy” EP19382384.6 and is cofounder and shareholder of ABCDx. Sean I. Savitz has served in the following roles as an employee of the institution UTHealth: as a site investigator in clinical trials sponsored by industry companies Brain.Q, IPSEN, Neurastasis, for which UTHealth receives payments on the basis of clinical trial contracts; as an investigator on clinical trials or supported by National Institutes of Health grants; as an investigator in the National Institutes of Health funded Clinical and Translational Science Awards Program; as a principal investigator or coinvestigator on National Institutes of Health‐funded grants in basic science and clinical research, as a principal investigator in state funded projects under the auspices of the LoneStar Stroke Consortium. In his capacity as a UTHealth employee, Sean I. Savitz has provided consulting services on behalf of UTHealth to Lumosa, Neurexcell, Abbvie, Neurastasis, ArunA, Ceribell, AI‐Stroke, CerebralAI, Bayer, CellProthera, DE Shaw, APT Therapeutics, and Pikralida. All compensation from such consulting arrangements has been paid to UTHealth. Guillaume Turc declares having received lecture fees from Guerbet France, consultancy fees from NeuroLogica Corporation and AI‐Stroke. All other authors declare no relevant conflicts of interest related to this work.
Supporting information
Data S1. Supplemental Material
PRISMA 2020 Checklist
Acknowledgments
We are grateful to Manuel Olma for assistance with designing the search strategy. Automated assistive writing technologies were used for word count reduction. Open Access funding enabled and organized by Projekt DEAL.
This article was sent to Fadar Oliver Otite, MS, SM, Associate Editor, for review by expert referees, editorial decision, and final disposition.
Supplemental Material is available at https://www.ahajournals.org/doi/suppl/10.1161/JAHA.125.045588
For Sources of Funding and Disclosures, see page 10.
References
- 1. Lees KR, Bluhmki E, von Kummer R, Brott TG, Toni D, Grotta JC, Albers GW, Kaste M, Marler JR, Hamilton SA, et al. Time to treatment with intravenous alteplase and outcome in stroke: an updated pooled analysis of ECASS, ATLANTIS, NINDS, and EPITHET trials. Lancet. 2010;375:1695–1703. doi: 10.1016/S0140-6736(10)60491-6 [DOI] [PubMed] [Google Scholar]
- 2. Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, Davalos A, Majoie CB, van der Lugt A, de Miquel MA, et al. Endovascular thrombectomy after large‐vessel ischaemic stroke: a meta‐analysis of individual patient data from five randomised trials. Lancet. 2016;387:1723–1731. doi: 10.1016/S0140-6736(16)00163-X [DOI] [PubMed] [Google Scholar]
- 3. Mackey J, Yamal JM, Parker SA, Silnes K, Rajan SS, Jacob AP, Wang M, Singh N, Jones WJ, Spokoyny I, et al. Golden hour treatment with tPA (tissue‐type plasminogen activator) in the BEST‐MSU study. Stroke. 2023;54:415–425. doi: 10.1161/STROKEAHA.122.039821 [DOI] [PubMed] [Google Scholar]
- 4. Turc G, Hadziahmetovic M, Walter S, Churilov L, Larsen K, Grotta JC, Yamal JM, Bowry R, Katsanos AH, Zhao H, et al. Comparison of mobile stroke unit with usual care for acute ischemic stroke management: a systematic review and meta‐analysis. JAMA Neurol. 2022;79:281–290. doi: 10.1001/jamaneurol.2021.5321 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Prabhakaran S, Gonzalez NR, Zachrison KS, Adeoye O, Alexandrov AW, Ansari SA, Chapman S, Czap AL, Dumitrascu OM, Ishida K, et al. 2026 guideline for the early management of patients with acute ischemic stroke: a guideline from the American Heart Association/American Stroke Association. Stroke. 2026;57(8):e316‐e436. doi: 10.1161/STR.0000000000000513 [DOI] [PubMed] [Google Scholar]
- 6. Ricci E dDS, Cianca E, Rossi T, Diomedi M. PCA‐based artifact removal algorithm for stroke detection using UWB radar imaging. Med Biol Eng Comput. 2017;55:909–921. doi: 10.1007/s11517-016-1568-8 [DOI] [PubMed] [Google Scholar]
- 7. Fhager A, Candefjord S, Elam M, Persson M. 3D simulations of intracerebral hemorrhage detection using broadband microwave technology. Sensors (Basel). 2019;19:3482. doi: 10.3390/s19163482 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Persson MFA, Trefná HD, Yu Y, McKelvey T, Pegenius G, Karlsson JE, Elam M. Microwave‐based stroke diagnosis making global prehospital thrombolytic treatment possible. IEEE Trans Biomed Eng. 2014;61:2806–2817. doi: 10.1109/tbme.2014.2330554 [DOI] [PubMed] [Google Scholar]
- 9. Ljungqvist JCS, Persson M, Jönsson L, Skoglund T, Elam M. Clinical evaluation of a microwave‐based device for detection of traumatic intracranial hemorrhage. J Neurotrauma. 2017;34:2176–2182. doi: 10.1089/neu.2016.4869 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Mobashsher ATAA. On‐site rapid diagnosis of intracranial hematoma using portable multi‐slice microwave imaging system. Sci Rep. 2016;6:37620. doi: 10.1038/srep37620 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Xu JCJ, Yu W, Zhang H, Wang F, Zhuang W, Yang J, Bai Z, Xu L, Sun J, Jin G, et al. Noninvasive and portable stroke type discrimination and progress monitoring based on a multichannel microwave transmitting‐receiving system. Sci Rep. 2020;10:21647. doi: 10.1038/s41598-020-78647-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. W KB. Non‐invasive sensor technology for prehospital stroke diagnosis: current status and future directions. Int J Stroke. 2019;14:592–602. [DOI] [PubMed] [Google Scholar]
- 13. Martinez‐Gutierrez JCCR, Hirsch JA, Leslie‐Mazwi T. Technological innovation for prehospital stroke triage: ripe for disruption. J Neurointerv Surg. 2019;11:1085–1090. doi: 10.1136/neurintsurg-2019-014902 [DOI] [PubMed] [Google Scholar]
- 14. Montemerani S, Fabrizi C, Sacchi C, Belperio A, Moriani L, Pacchi S, Garofalo C, Sbrana G, Venezia D, Zanobetti M, et al. Reliability evaluation of temnography for early detection of intracranial lesions in mild traumatic brain injury patient: a preliminary report of a new portable, non‐invasive device. Open Access Emerg Med. 2024;16:313–321. doi: 10.2147/OAEM.S475328 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Abbosh A, Bialkowski K, Guo L, Al‐Saffar A, Zamani A, Trakic A, Brankovic A, Bialkowski A, Zhu G, Cook D, et al. Clinical electromagnetic brain scanner. Sci Rep. 2024;14:5760. doi: 10.1038/s41598-024-55360-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Shahrestani S, Strickland BA, Bakhsheshian J, Mack WJ, Toga AW, Sanossian N, Tai YC, Zada G. Transcranial eddy current damping sensors for detection and imaging of hemorrhagic stroke: feasibility in benchtop experimentation. Neurosurg Focus. 2021;51:E15. doi: 10.3171/2021.4.FOCUS21121 [DOI] [PubMed] [Google Scholar]
- 17. Shahrestani S, Zada G, Chou TC, Toy B, Yao B, Garrett N, Sanossian N, Brunswick A, Shang KM, Tai YC. Noninvasive transcranial classification of stroke using a portable eddy current damping sensor. Sci Rep. 2021;11:10297. doi: 10.1038/s41598-021-89735-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Shen H, Ghodsi SM, Fixman B, Ghodsi B, Azarraga K, Prasad N, Shahrestani S, Sanossian N, Zada G, Tai YC. Machine learning enhanced noninvasive transcranial stroke detection using a portable eddy current damping sensor. iScience. 2025;28:114130. doi: 10.1016/j.isci.2025.114130 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Kellner CP, Sauvageau E, Snyder KV, Fargen KM, Arthur AS, Turner RD, Alexandrov AV. The VITAL study and overall pooled analysis with the VIPS non‐invasive stroke detection device. J Neurointerv Surg. 2018;10:1079–1084. doi: 10.1136/neurintsurg-2017-013690 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. van Meenen LCC, van Stigt MN, Marquering HA, Majoie C, Roos Y, Koelman J, Potters WV, Coutinho JM. Detection of large vessel occlusion stroke with electroencephalography in the emergency room: first results of the ELECTRA‐STROKE study. J Neurol. 2022;269:2030–2038. doi: 10.1007/s00415-021-10781-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. van Stigt MN, Groenendijk EA, van Meenen LCC, van de Munckhof A, Theunissen M, Franschman G, Smeekes MD, van Grondelle JAF, Geuzebroek G, Siegers A, et al. Prehospital detection of large vessel occlusion stroke with EEG. Neurology. 2023;101:e2522–e2532. doi: 10.1212/WNL.0000000000207831 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Wilkinson CM, Burrell JI, Kuziek JWP, Thirunavukkarasu S, Buck BH, Mathewson KE. Predicting stroke severity with a 3‐min recording from the muse portable EEG system for rapid diagnosis of stroke. Sci Rep. 2020;10:18465. doi: 10.1038/s41598-020-75379-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Sergot PB, Maza AJ, Derrick BJ, Smith LM, Berti LT, Wilcox MR, Kesinger MR, Peacock WF; Group ES . Portable neuromonitoring device detects large vessel occlusion in suspected acute ischemic stroke. Stroke. 2021;52:1437–1440. doi: 10.1161/STROKEAHA.120.031225 [DOI] [PubMed] [Google Scholar]
- 24. Erani F, Zolotova N, Vanderschelden B, Khoshab N, Sarian H, Nazarzai L, Wu J, Chakravarthy B, Hoonpongsimanont W, Yu W, et al. Electroencephalography might improve diagnosis of acute stroke and large vessel occlusion. Stroke. 2020;51:3361–3365. doi: 10.1161/STROKEAHA.120.030150 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Michelson EA, Hanley D, Chabot R, Prichep LS. Identification of acute stroke using quantified brain electrical activity. Acad Emerg Med. 2015;22:67–72. doi: 10.1111/acem.12561 [DOI] [PubMed] [Google Scholar]
- 26. Guterman EL, Mercer MP, Wood AJ, Amorim E, Kleen JK, Gerard D, Kellison C, Yamashita S, Auerbach B, Joshi N, et al. Evaluating the feasibility of prehospital point‐of‐care EEG: the prehospital implementation of rapid EEG (PHIRE) study. J Am Coll Emerg Physicians Open. 2024;5:e13303. doi: 10.1002/emp2.13303 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Lohi S, Jakala P, Kurola J, Tuunanen P, Westeren‐Punnonen S, Muraja‐Murro A, Kalviainen R, Mervaala E. Feasibility of recording EEG in the ambulance using a portable, wireless EEG recording system. PLoS One. 2025;20:e0327415. doi: 10.1371/journal.pone.0327415 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Nakagomi M, Kajiwara M, Matsuzaki J, Tanabe K, Hoshiai S, Okamoto Y, Terada Y. Development of a small car‐mounted magnetic resonance imaging system for human elbows using a 0.2 T permanent magnet. J Magn Reson. 2019;304:1–6. doi: 10.1016/j.jmr.2019.04.017 [DOI] [PubMed] [Google Scholar]
- 29. Mazurek MH, Cahn BA, Yuen MM, Prabhat AM, Chavva IR, Shah JT, Crawford AL, Welch EB, Rothberg J, Sacolick L, et al. Portable, bedside, low‐field magnetic resonance imaging for evaluation of intracerebral hemorrhage. Nat Commun. 2021;12:5119. doi: 10.1038/s41467-021-25441-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Cooley CZ, McDaniel PC, Stockmann JP, Srinivas SA, Cauley SF, Sliwiak M, Sappo CR, Vaughn CF, Guerin B, Rosen MS, et al. A portable scanner for magnetic resonance imaging of the brain. Nat Biomed Eng. 2021;5:229–239. doi: 10.1038/s41551-020-00641-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Deoni SCL, Medeiros P, Deoni AT, Burton P, Beauchemin J, D'Sa V, Boskamp E, By S, McNulty C, Mileski W, et al. Development of a mobile low‐field MRI scanner. Sci Rep. 2022;12:5690. doi: 10.1038/s41598-022-09760-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Bhat SS, Fernandes TT, Poojar P, da Silva Ferreira M, Rao PC, Hanumantharaju MC, Ogbole G, Nunes RG, Geethanath S. Low‐field MRI of stroke: challenges and opportunities. J Magn Reson Imaging. 2021;54:372–390. doi: 10.1002/jmri.27324 [DOI] [PubMed] [Google Scholar]
- 33. Sheth KN, Mazurek MH, Yuen MM, Cahn BA, Shah JT, Ward A, Kim JA, Gilmore EJ, Falcone GJ, Petersen N, et al. Assessment of brain injury using portable, low‐field magnetic resonance imaging at the bedside of critically ill patients. JAMA Neurol. 2020;78:41–47. doi: 10.1001/jamaneurol.2020.3263 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Stainsby JA, Harris CT, Bindseil GA, Wiens CN, Beatty PJ, Curtis AT. High‐performance diffusion imaging on a 0.5T system. Proc. Intl. Soc. Mag. Reson. Med; 2019:27. International Society for Magnetic Resonance in Medicine (ISMRM). [Google Scholar]
- 35. Yuen MM, Prabhat AM, Mazurek MH, Chavva IR, Crawford A, Cahn BA, Beekman R, Kim JA, Gobeske KT, Petersen NH, et al. Portable, low‐field magnetic resonance imaging enables highly accessible and dynamic bedside evaluation of ischemic stroke. Sci Adv. 2022;8:eabm3952. doi: 10.1126/sciadv.abm3952 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Lau V, Xiao L, Zhao Y, Su S, Ding Y, Man C, Wang X, Tsang A, Cao P, Lau GKK, et al. Pushing the limits of low‐cost ultra‐low‐field MRI by dual‐acquisition deep learning 3D superresolution. Magn Reson Med. 2023;90:400–416. doi: 10.1002/mrm.29642 [DOI] [PubMed] [Google Scholar]
- 37. Man C, Lau V, Su S, Zhao Y, Xiao L, Ding Y, Leung GKK, Leong ATL, Wu EX. Deep learning enabled fast 3D brain MRI at 0.055 tesla. Sci Adv. 2023;9:eadi9327. doi: 10.1126/sciadv.adi9327 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Gopinath SP, Robertson CS, Contant CF, Narayan RK, Grossman RG, Chance B. Early detection of delayed traumatic intracranial hematomas using near‐infrared spectroscopy. J Neurosurg. 1995;83:438–444. doi: 10.3171/jns.1995.83.3.0438 [DOI] [PubMed] [Google Scholar]
- 39. Robertson CS, Zager EL, Narayan RK, Handly N, Sharma A, Hanley DF, Garza H, Maloney‐Wilensky E, Plaum JM, Koenig CH, et al. Clinical evaluation of a portable near‐infrared device for detection of traumatic intracranial hematomas. J Neurotrauma. 2010;27:1597–1604. doi: 10.1089/neu.2010.1340 [DOI] [PubMed] [Google Scholar]
- 40. Ayaz H, Izzetoglu M, Izzetoglu K, Onaral B, Ben Dor B. Early diagnosis of traumatic intracranial hematomas. J Biomed Opt. 2019;24:1–10. doi: 10.1117/1.JBO.24.5.051411 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Moreau F, Yang R, Nambiar V, Demchuk AM, Dunn JF. Near‐infrared measurements of brain oxygenation in stroke. Neurophotonics. 2016;3:031403. doi: 10.1117/1.NPh.3.3.031403 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Muehlschlegel S, Selb J, Patel M, Diamond SG, Franceschini MA, Sorensen AG, Boas DA, Schwamm LH. Feasibility of NIRS in the neurointensive care unit: a pilot study in stroke using physiological oscillations. Neurocrit Care. 2009;11:288–295. doi: 10.1007/s12028-009-9254-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Shah J, Solanki S, Chandra A, Laljibhai Shivabhai M, Gwalani P. Evaluating performance of a near‐infrared‐spectroscopy‐based and machine‐learning‐powered bio‐optical sensitivity parameters in identifying intracranial hemorrhages in TBI across different age‐groups. Brain Inj. 2024;38:1227–1235. doi: 10.1080/02699052.2024.2381056 [DOI] [PubMed] [Google Scholar]
- 44. Favilla CG, Baird GL, Grama K, Konecky S, Carter S, Smith W, Gitlevich R, Lebron‐Cruz A, Yodh AG, McTaggart RA. Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke. J Neurointerv Surg. 2025;17:388–393. doi: 10.1136/jnis-2024-021536 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Montaner J, Ramiro L, Simats A, Tiedt S, Makris K, Jickling GC, Debette S, Sanchez JC, Bustamante A. Multilevel omics for the discovery of biomarkers and therapeutic targets for stroke. Nat Rev Neurol. 2020;16:247–264. doi: 10.1038/s41582-020-0350-6 [DOI] [PubMed] [Google Scholar]
- 46. Mattila OS, Harve H, Pihlasviita S, Ritvonen J, Sibolt G, Pystynen M, Strbian D, Curtze S, Kuisma M, Tatlisumak T, et al. Ultra‐acute diagnostics for stroke: large‐scale implementation of prehospital biomarker sampling. Acta Neurol Scand. 2017;136:17–23. doi: 10.1111/ane.12687 [DOI] [PubMed] [Google Scholar]
- 47. Kowalski RG, Ledreux A, Violette JE, Neumann RT, Ornelas D, Yu X, Griffiths SG, Lewis S, Nash P, Monte AA, et al. Rapid activation of neuroinflammation in stroke: plasma and extracellular vesicles obtained on a Mobile stroke unit. Stroke. 2023;54:e52–e57. doi: 10.1161/STROKEAHA.122.041422 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Bustamante A, Lopez‐Cancio E, Pich S, Penalba A, Giralt D, Garcia‐Berrocoso T, Ferrer‐Costa C, Gasull T, Hernandez‐Perez M, Millan M, et al. Blood biomarkers for the early diagnosis of stroke: the stroke‐chip study. Stroke. 2017;48:2419–2425. doi: 10.1161/STROKEAHA.117.017076 [DOI] [PubMed] [Google Scholar]
- 49. Wendt M, Ebinger M, Kunz A, Rozanski M, Waldschmidt C, Weber JE, Winter B, Koch PM, Nolte CH, Hertel S, et al. Copeptin levels in patients with acute ischemic stroke and stroke mimics. Stroke. 2015;46:2426–2431. doi: 10.1161/STROKEAHA.115.009877 [DOI] [PubMed] [Google Scholar]
- 50. Nguyen TTM, van der Bent ML, Wermer MJH, van den Wijngaard IR, van Zwet EW, de Groot B, Quax PHA, Kruyt ND, Nossent AY. Circulating tRNA fragments as a novel biomarker class to distinguish acute stroke subtypes. Int J Mol Sci. 2020;22:135. doi: 10.3390/ijms22010135 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Pihlasviita S, Mattila OS, Nukarinen T, Kuisma M, Harve‐Rytsala H, Ritvonen J, Sibolt G, Curtze S, Strbian D, Pystynen M, et al. Plasma symmetric dimethylarginine as a metabolite biomarker of severe acute ischemic stroke. Front Neurol. 2024;15:1472424. doi: 10.3389/fneur.2024.1472424 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Rozanski M, Waldschmidt C, Kunz A, Grittner U, Ebinger M, Wendt M, Winter B, Bollweg K, Villringer K, Fiebach JB, et al. Glial fibrillary acidic protein for prehospital diagnosis of intracerebral hemorrhage. Cerebrovasc Dis. 2017;43:76–81. doi: 10.1159/000453460 [DOI] [PubMed] [Google Scholar]
- 53. Luger S, Jaeger HS, Dixon J, Bohmann FO, Schaefer J, Richieri SP, Larsen K, Hov MR, Bache KG, Foerch C, et al. Diagnostic accuracy of glial fibrillary acidic protein and ubiquitin carboxy‐terminal hydrolase‐L1 serum concentrations for differentiating acute intracerebral hemorrhage from ischemic stroke. Neurocrit Care. 2020;33:39–48. doi: 10.1007/s12028-020-00931-5 [DOI] [PubMed] [Google Scholar]
- 54. Mattila OS, Ashton NJ, Blennow K, Zetterberg H, Harve‐Rytsala H, Pihlasviita S, Ritvonen J, Sibolt G, Nukarinen T, Curtze S, et al. Ultra‐early differential diagnosis of acute cerebral ischemia and hemorrhagic stroke by measuring the prehospital release rate of GFAP. Clin Chem. 2021;67:1361–1372. doi: 10.1093/clinchem/hvab128 [DOI] [PubMed] [Google Scholar]
- 55. Kalra LP, Zylyftari S, Blums K, Barthelmes S, Baum H, Meckel S, Heilgeist A, Luger S, Foerch C. Rapid diagnosis of intracerebral hemorrhage in patients with acute stroke by measuring prehospital GFAP levels on a point‐of‐care device (DETECT). Neurology. 2025;105:e213823. doi: 10.1212/WNL.0000000000213823 [DOI] [PubMed] [Google Scholar]
- 56. Zylyftari S, Luger S, Blums K, Barthelmes S, Humm S, Baum H, Meckel S, Braun J, Lichy G, Heilgeist A, et al. GFAP point‐of‐care measurement for prehospital diagnosis of intracranial hemorrhage in acute coma. Crit Care. 2024;28:109. doi: 10.1186/s13054-024-04892-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Jaeger HS, Tranberg D, Larsen K, Valentin JB, Blauenfeldt RA, Luger S, Bache KG, Gude MF. Diagnostic performance of glial fibrillary acidic protein and prehospital stroke scale for identification of stroke and stroke subtypes in an unselected patient cohort with symptom onset <4.5 h. Scand J Trauma Resusc Emerg Med. 2023;31:1. doi: 10.1186/s13049-022-01065-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Jauquet M, Gagnepain P, La Porte E, Thiebaut AM, Rochey A, Legros H, Laine B, Berthelot M, Roussel V, Montaner J, et al. Endogenous tPA levels: a biomarker for discriminating hemorrhagic stroke from ischemic stroke and stroke mimics. Ann Clin Transl Neurol. 2024;11:2877–2890. doi: 10.1002/acn3.52197 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Study of the potential use of a rapid point‐of‐care device for the direct transfer of stroke patients from the ambulance to the thrombectomy center. Hospital Universitario Virgen M, Hospitales Universitarios Virgen del R, Hospital Universitario T, Hospital Universitario P, Hospital Clínico Universitario Lozano B, Hospital Miguel S, Hospital Universitario Royo V, Hospital General San J, Hospital San Pedro de L, Hospital Clínico Universitario de V2023.
- 60. Ramos‐Pachon A, Lopez‐Cancio E, Bustamante A, Perez de la Ossa N, Millan M, Hernandez‐Perez M, Garcia‐Berrocoso T, Cardona P, Rubiera M, Serena J, et al. D‐dimer as predictor of large vessel occlusion in acute ischemic stroke. Stroke. 2021;52:852–858. doi: 10.1161/STROKEAHA.120.031657 [DOI] [PubMed] [Google Scholar]
- 61. Antipova D, Eadie L, MacAden AS, Wilson P. Training novice users to assess stroke aetiology in remote settings using transcranial ultrasound: pilot study. Rural Remote Health. 2021;21:5618. doi: 10.22605/RRH5618 [DOI] [PubMed] [Google Scholar]
- 62. Antipova D, Eadie L, Makin S, Shannon H, Wilson P, Macaden A. The use of transcranial ultrasound and clinical assessment to diagnose ischaemic stroke due to large vessel occlusion in remote and rural areas. PLoS One. 2020;15:e0239653. doi: 10.1371/journal.pone.0239653 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Connolly F, Rohl JE, Guthke C, Wengert O, Valdueza JM, Schreiber SJ. Emergency room use of “fast‐track” ultrasound in acute stroke: an observational study. Ultrasound Med Biol. 2019;45:1103–1111. doi: 10.1016/j.ultrasmedbio.2019.01.001 [DOI] [PubMed] [Google Scholar]
- 64. Herzberg M, Boy S, Holscher T, Ertl M, Zimmermann M, Ittner KP, Pemmerl J, Pels H, Bogdahn U, Schlachetzki F. Prehospital stroke diagnostics based on neurological examination and transcranial ultrasound. Crit Ultrasound J. 2014;6:3. doi: 10.1186/2036-7902-6-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65. Holscher T, Schlachetzki F, Zimmermann M, Jakob W, Ittner KP, Haslberger J, Bogdahn U, Boy S. Transcranial ultrasound from diagnosis to early stroke treatment. 1. Feasibility of prehospital cerebrovascular assessment. Cerebrovasc Dis. 2008;26:659–663. doi: 10.1159/000166844 [DOI] [PubMed] [Google Scholar]
- 66. Schlachetzki F, Herzberg M, Holscher T, Ertl M, Zimmermann M, Ittner KP, Pels H, Bogdahn U, Boy S. Transcranial ultrasound from diagnosis to early stroke treatment: part 2: prehospital neurosonography in patients with acute stroke: the Regensburg stroke mobile project. Cerebrovasc Dis. 2012;33:262–271. doi: 10.1159/000334667 [DOI] [PubMed] [Google Scholar]
- 67. Lindsey BD, Nicoletto HA, Bennett ER, Laskowitz DT, Smith SW. Simultaneous bilateral real‐time 3‐d transcranial ultrasound imaging at 1 MHz through poor acoustic windows. Ultrasound Med Biol. 2013;39:721–734. doi: 10.1016/j.ultrasmedbio.2012.11.019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Thorpe SG, Thibeault CM, Canac N, Wilk SJ, Devlin T, Hamilton RB. Decision criteria for large vessel occlusion using transcranial doppler waveform morphology. Front Neurol. 2018;9:847. doi: 10.3389/fneur.2018.00847 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69. Baumgartner RW, Mattle HP, Schroth G. Assessment of >/=50% and <50% intracranial stenoses by transcranial color‐coded duplex sonography. Stroke. 1999;30:87–92. doi: 10.1161/01.str.30.1.87 [DOI] [PubMed] [Google Scholar]
- 70. Kilic M, Wendl C, Wilfling S, Olmes D, Linker RA, Schlachetzki F. Acute middle cerebral artery occlusion detection using mobile non‐imaging Brain perfusion ultrasound‐first case. J Clin Med. 2022;11:3384. doi: 10.3390/jcm11123384 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71. Kukulska‐Pawluczuk B, Ksiazkiewicz B, Nowaczewska M. Imaging of spontaneous intracerebral hemorrhages by means of transcranial color‐coded sonography. Eur J Radiol. 2012;81:1253–1258. doi: 10.1016/j.ejrad.2011.02.066 [DOI] [PubMed] [Google Scholar]
- 72. Mort A, Eadie L, Regan L, Macaden A, Heaney D, Bouamrane MM, Rushworth G, Wilson P. Combining transcranial ultrasound with intelligent communication methods to enhance the remote assessment and management of stroke patients: framework for a technology demonstrator. Health Informatics J. 2016;22:691–701. doi: 10.1177/1460458215580353 [DOI] [PubMed] [Google Scholar]
- 73. Ertl M, Raasch N, Hammel G, Harter K, Lang C. Transtemporal investigation of Brain parenchyma elasticity using 2‐D shear wave elastography: definition of age‐matched Normal values. Ultrasound Med Biol. 2018;44:78–84. doi: 10.1016/j.ultrasmedbio.2017.08.1885 [DOI] [PubMed] [Google Scholar]
- 74. Ertl M, Woeckel M, Maurer C. Differentiation between ischemic and hemorrhagic strokes—a pilot study with transtemporal investigation of Brain parenchyma elasticity using ultrasound shear wave elastography. Ultraschall Med. 2021;42:75–83. doi: 10.1055/a-1248-2022 [DOI] [PubMed] [Google Scholar]
- 75. Alghamdi I, Brunton L, Ashton C, Jenkins DA, Parry‐Jones AR. Prehospital video triage of suspected stroke patients in Greater Manchester: pilot project report. BMJ Open Qual. 2025;14(1):e002954. doi: 10.1136/bmjoq-2024-002954 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Kanchayawong P, Aramvanitch K, Yuksen C, Trakulsrichai S, Sricharoen P, Suwatcharangkoon S, Sirintaranont P, Keandoungchun J, Nuanprom P, Jenpanitpong C, et al. Real‐time telemedical oversight improves prehospital stroke metrics: a five‐year cohort study. Arch Acad Emerg Med. 2025;13:e57. doi: 10.22037/aaemj.v13i1.2693 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77. Lee HW, Ko YC, Tang SC, Hsieh MJ, Tsai LK, Chiang WC, Jeng JS, Ma MH. Prehospital neurologic assessment using mobile phones: comparison between neurologists and emergency physicians. J Formos Med Assoc. 2024;123:1144–1148. doi: 10.1016/j.jfma.2024.06.016 [DOI] [PubMed] [Google Scholar]
- 78. Harahsheh E, English SW Jr, Demaerschalk BM, Barrett KM, Freeman WD. Telemedicine‐enabled ambulances for prehospital acute stroke management: a pilot study. Mayo Clin Proc Digit Health. 2024;2:533–541. doi: 10.1016/j.mcpdig.2024.08.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. LaMonte MP, Cullen J, Gagliano DM, Gunawardane R, Hu P, Mackenzie C, Xiao Y, Brain AT. TeleBAT: mobile telemedicine for the brain attack team. J Stroke Cerebrovasc Dis. 2000;9:128–135. doi: 10.1053/jscd.2000.5867 [DOI] [PubMed] [Google Scholar]
- 80. Bergrath S, Reich A, Rossaint R, Rortgen D, Gerber J, Fischermann H, Beckers SK, Brokmann JC, Schulz JB, Leber C, et al. Feasibility of prehospital teleconsultation in acute stroke—a pilot study in clinical routine. PLoS One. 2012;7:e36796. doi: 10.1371/journal.pone.0036796PONE-D-11-23563 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81. Liman TG, Winter B, Waldschmidt C, Zerbe N, Hufnagl P, Audebert HJ, Endres M. Telestroke ambulances in prehospital stroke management: concept and pilot feasibility study. Stroke. 2012;43:2086–2090. doi: 10.1161/STROKEAHA.112.657270 [DOI] [PubMed] [Google Scholar]
- 82. Winter B, Wendt M, Waldschmidt C, Rozanski M, Kunz A, Geisler F, Grittner U, Kaczmarek S, Ebinger M, Audebert HJ. 4G versus 3G‐enabled telemedicine in prehospital acute stroke care. Int J Stroke. 2019;14:620–629. doi: 10.1177/1747493019830303 [DOI] [PubMed] [Google Scholar]
- 83. Johansson A, Esbjornsson M, Nordqvist P, Wiinberg S, Andersson R, Ivarsson B, Moller S. Technical feasibility and ambulance nurses' view of a digital telemedicine system in pre‐hospital stroke care—a pilot study. Int Emerg Nurs. 2019;44:35–40. doi: 10.1016/j.ienj.2019.03.008 [DOI] [PubMed] [Google Scholar]
- 84. Scott IM, Manoczki C, Swain AH, Ranjan A, McGovern MG, Shyrell Tyson AL, Hyslop MC, Punter MM, Ranta A. Prehospital telestroke vs paramedic scores to accurately identify stroke reperfusion candidates: a cluster randomized controlled trial. Neurology. 2022;99:e2125–e2136. doi: 10.1212/WNL.0000000000201104 [DOI] [PubMed] [Google Scholar]
- 85. Itrat A, Taqui A, Cerejo R, Briggs F, Cho SM, Organek N, Reimer AP, Winners S, Rasmussen P, Hussain MS, et al. Telemedicine in prehospital stroke evaluation and thrombolysis: taking stroke treatment to the doorstep. JAMA Neurol. 2016;73:162–168. doi: 10.1001/jamaneurol.2015.3849 [DOI] [PubMed] [Google Scholar]
- 86. Gonzalez MA, Hanna N, Rodrigo ME, Satler LF, Waksman R. Reliability of prehospital real‐time cellular video phone in assessing the simplified National Institutes of Health Stroke Scale in patients with acute stroke: a novel telemedicine technology. Stroke. 2011;42:1522–1527. doi: 10.1161/STROKEAHA.110.600296 [DOI] [PubMed] [Google Scholar]
- 87. Van Hooff RJ, De Smedt A, De Raedt S, Moens M, Marien P, Paquier P, De Keyser J, Brouns R. Unassisted assessment of stroke severity using telemedicine. Stroke. 2013;44:1249–1255. doi: 10.1161/STROKEAHA.111.680868 [DOI] [PubMed] [Google Scholar]
- 88. Wu TC, Nguyen C, Ankrom C, Yang J, Persse D, Vahidy F, Grotta JC, Savitz SI. Prehospital utility of rapid stroke evaluation using in‐ambulance telemedicine: a pilot feasibility study. Stroke. 2014;45:2342–2347. doi: 10.1161/STROKEAHA.114.005193 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Chapman Smith SN, Brown PC, Waits KH, Wong JS, Bhatti MS, Toqeer Q, Ricks JV, Stockner ML, Habtamu T, Seelam J, et al. Development and evaluation of a user‐centered mobile telestroke platform. Telemed J E Health. 2019;25:638–648. doi: 10.1089/tmj.2018.0044 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90. Eadie L, Regan L, Mort A, Shannon H, Walker J, MacAden A, Wilson P. Telestroke assessment on the move: prehospital streamlining of patient pathways. Stroke. 2015;46:e38–e40. doi: 10.1161/STROKEAHA.114.007475 [DOI] [PubMed] [Google Scholar]
- 91. Geisler F, Kunz A, Winter B, Rozanski M, Waldschmidt C, Weber JE, Wendt M, Zieschang K, Ebinger M, Audebert HJ. Telemedicine in prehospital acute stroke care. J Am Heart Assoc. 2019;8:e011729. doi: 10.1161/jaha.118.011729 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92. Barrett KM, Pizzi MA, Kesari V, TerKonda SP, Mauricio EA, Silvers SM, Habash R, Brown BL, Tawk RG, Meschia JF, et al. Ambulance‐based assessment of NIH stroke scale with telemedicine: a feasibility pilot study. J Telemed Telecare. 2017;23:476–483. doi: 10.1177/1357633X16648490 [DOI] [PubMed] [Google Scholar]
- 93. Eder PA, Laux G, Rashid A, Kniess T, Haeusler KG, Shammas L, Griewing B, Hofmann S, Stangl S, Wiedmann S, et al. Stroke angel: effect of telemedical prenotification on in‐hospital delays and systemic thrombolysis in acute stroke patients. Cerebrovasc Dis. 2021;50:1–9. doi: 10.1159/000514563 [DOI] [PubMed] [Google Scholar]
- 94. Berglund A, Svensson L, Sjostrand C, von Arbin M, von Euler M, Wahlgren N, Engerstrom L, Hojeberg B, Kall TB, Mjornheim S, et al. Higher prehospital priority level of stroke improves thrombolysis frequency and time to stroke unit: the Hyper Acute STroke Alarm (HASTA) study. Stroke. 2012;43:2666–2670. doi: 10.1161/STROKEAHA.112.652644 [DOI] [PubMed] [Google Scholar]
- 95. Bilotta M, Sigal AP, Shah A, Martin A, Schlappy DA, Sorensen G, Barbera C. A novel use of prehospital telemedicine to decrease door to computed tomography results in acute strokes. J Healthc Qual. 2020;42:264–268. doi: 10.1097/JHQ.0000000000000229 [DOI] [PubMed] [Google Scholar]
- 96. Al Kasab S, Almallouhi E, Grant C, Hewitt D, Hewitt J, Baki M, Sabatino P, Jones D, Holmstedt CA. Telestroke consultation in the emergency medical services unit: a novel approach to improve thrombolysis times. J Stroke Cerebrovasc Dis. 2021;30:105710. doi: 10.1016/j.jstrokecerebrovasdis.2021.105710 [DOI] [PubMed] [Google Scholar]
- 97. Bowry R, Parker SA, Yamal JM, Hwang H, Appana S, Rangel‐Gutierrez N, Wu TC, Rajan SS, Grotta JC. Time to decision and treatment with tPA (tissue‐type plasminogen activator) using telemedicine versus an onboard neurologist on a Mobile stroke unit. Stroke. 2018;49:1528–1530. doi: 10.1161/strokeaha.117.020585 [DOI] [PubMed] [Google Scholar]
- 98. Keselman B, Berglund A, Ahmed N, Bottai M, von Euler M, Holmin S, Laska AC, Mathe JM, Sjostrand C, Eriksson EE, et al. The Stockholm stroke triage project: outcomes of endovascular thrombectomy before and after triage implementation. Stroke. 2022;53:473–481. doi: 10.1161/STROKEAHA.121.034195 [DOI] [PubMed] [Google Scholar]
- 99. Quadflieg LTM, Beckers SK, Bergrath S, Brockert AK, Schroder H, Sommer A, Brokmann JC, Rossaint R, Felzen M. Comparing the diagnostic concordance of tele‐EMS and on‐site‐EMS physicians in emergency medical services: a retrospective cohort study. Sci Rep. 2020;10:17982. doi: 10.1038/s41598-020-75149-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100. Rios Delgado M, Reig Rosello G, Riera‐Lopez N, Vivancos JA, Ayala JL. Early stroke detection through machine learning in the prehospital setting. Front Cardiovasc Med. 2025;12:1629853. doi: 10.3389/fcvm.2025.1629853 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101. Bat‐Erdene BO, Saver JL. Automatic acute stroke symptom detection and emergency medical systems alerting by mobile health technologies: a review. J Stroke Cerebrovasc Dis. 2021;30:105826. doi: 10.1016/j.jstrokecerebrovasdis.2021.105826 [DOI] [PubMed] [Google Scholar]
- 102. Sung SF, Hung LC, Hu YH. Developing a stroke alert trigger for clinical decision support at emergency triage using machine learning. Int J Med Inform. 2021;152:104505. doi: 10.1016/j.ijmedinf.2021.104505 [DOI] [PubMed] [Google Scholar]
- 103. Park E, Lee K, Han T, Nam HS. Automatic grading of stroke symptoms for rapid assessment using optimized machine learning and 4‐limb kinematics: clinical validation study. J Med Internet Res. 2020;22:e20641. doi: 10.2196/20641 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Scholz ML, Collatz‐Christensen H, Blomberg SNF, Boebel S, Verhoeven J, Krafft T. Artificial intelligence in emergency medical services dispatching: assessing the potential impact of an automatic speech recognition software on stroke detection taking the Capital Region of Denmark as case in point. Scand J Trauma Resusc Emerg Med. 2022;30:36. doi: 10.1186/s13049-022-01020-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105. Uchida K, Kouno J, Yoshimura S, Kinjo N, Sakakibara F, Araki H, Morimoto T. Development of machine learning models to predict probabilities and types of stroke at prehospital stage: the Japan Urgent Stroke Triage Score Using Machine Learning (JUST‐ML). Transl Stroke Res. 2022;13:370–381. doi: 10.1007/s12975-021-00937-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106. Yoshida Y, Hayashi Y, Shimada T, Hattori N, Tomita K, Miura RE, Yamao Y, Tateishi S, Iwadate Y, Nakada TA. Prehospital stroke‐scale machine‐learning model predicts the need for surgical intervention. Sci Rep. 2023;13:9135. doi: 10.1038/s41598-023-36004-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107. Nogueira RG, Silva GS, Lima FO, Yeh YC, Fleming C, Branco D, Yancey AH, Ratcliff JJ, Wages RK, Doss E, et al. The FAST‐ED app: a smartphone platform for the field triage of patients with stroke. Stroke. 2017;48:1278–1284. doi: 10.1161/STROKEAHA.116.016026 [DOI] [PubMed] [Google Scholar]
- 108. Bandini A, Green JR, Richburg BD, Yunusova Y. Automatic Detection of Orofacial Impairment in Stroke. Paper/Poster presented at: Interspeech. 2018.
- 109. Abedi V, Goyal N, Tsivgoulis G, Hosseinichimeh N, Hontecillas R, Bassaganya‐Riera J, Elijovich L, Metter JE, Alexandrov AW, Liebeskind DS, et al. Novel screening tool for stroke using artificial neural network. Stroke. 2017;48:1678–1681. doi: 10.1161/STROKEAHA.117.017033 [DOI] [PubMed] [Google Scholar]
- 110. Freitag E, Erdur H, Khalil AA, Harmel P, Kaffes M, Schmitz CH, Weber JE, Audebert HJ. Potential of transcranial ultrasound‐ and near‐infrared spectroscopy‐based acute stroke imaging for decision‐making on intravenous thrombolysis treatment. Front Neurol. 2025;16:1499821. doi: 10.3389/fneur.2025.1499821 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data S1. Supplemental Material
PRISMA 2020 Checklist