PaO2/FiO2 and SpO2/FiO2 ratios revisited: Useful, misleading, or both?

Prashant Nasa 1,2, Denise Battaglini 3,4, Deven Juneja 5, Marcus J Schultz 6,7,8,10, Jozef Kesecioglu 9,⁎
PMCID: PMC13728404  PMID: 42979078

Abstract

The ratio of arterial partial pressure of oxygen to the inspired oxygen fraction (PaO2/FiO2) serves as a cornerstone metric for the management of patients with acute respiratory distress syndrome (ARDS). This ratio is integral to the Berlin definition of ARDS and is used for severity classification, prognostication, and guiding therapeutic interventions. Furthermore, the PaO2/FiO2 ratio has been employed in landmark trials to ascertain patients’ eligibility for interventions such as prone positioning and extracorporeal membrane oxygenation. The modifications of the Berlin definition, the Kigali modification, and the 2023 New Global definition suggested using the ratio of peripheral oxygen saturation to FiO2 (SpO2/FiO2) as a surrogate for the PaO2/FiO2 ratio when arterial blood gas is inaccessible. Although these oxygenation ratios are valuable for timely recognition and categorizing ARDS, as well as informing clinical decision-making, several significant limitations limit their efficacy, such as dependence on ventilator settings and FiO2. The isolated use of these oxygenation ratios may not accurately monitor the evolution of the pathology and response to interventions. Additionally, the arbitrary cutoffs established for these oxygenation metrics frequently result in overlapping categories rather than distinct subgroups of ARDS. Finally, pulse oximetry measurements can be compromised by factors such as patient motion, perfusion variability, dyshemoglobinemia, and skin pigmentation-related bias, which may obscure hypoxemia. Considering these challenges, recent research has explored the trajectory of the PaO2/FiO2 ratio relative to static measurements and the reclassification of ARDS severity after 24–48 h for prognostication. Alternative metrics such as the oxygenation index, oxygenation saturation index, and (PaO2 ×10)/(FiO2 × PEEP) ratio (which incorporates positive end-expiratory pressure) have demonstrated greater ability to predict patients’ outcomes. This review summarizes the evidence base, clinical applications, and pitfalls of PaO2/FiO2 and SpO2/FiO2 ratios in ARDS and potential alternatives, particularly in the context of diagnosis, categorization, and management of ARDS.

Keywords: Acute respiratory distress syndrome, Partial pressure of oxygen to fraction of inspired oxygen (PaO2/FiO2), Peripheral oxygen saturation to fraction of inspired oxygen (SpO2/FiO2), Oxygenation ratios, Pulse oximetry, Acute hypoxemic respiratory failure

Introduction

Assessment of gas exchange and hypoxemia remains a cornerstone of managing critically ill and perioperative patients. For decades, the oxygenation ratio, described as the partial pressure of oxygen (PaO2) to the fraction of inspired oxygen (FiO2) (PaO2/FiO2 ratio), has guided clinical decisions, diagnosis, severity classification, research categorization, and guideline development.[1] Originally conceived for stratification of acute respiratory distress syndrome (ARDS), it is now widely used for treatment escalation, resource allocation, and prognostication of other forms of respiratory failure, far beyond its original scope.[2] Yet mounting evidence questions this reliance.

Another oxygenation ratio in which oxygen saturation from a pulse oximeter (SpO2) replaces PaO2, SpO2/FiO2 ratio, has emerged as a noninvasive and feasible alternative, especially in resource-limited settings.[3] The indices such as PaO2/FiO2 and SpO2/FiO2 ratios offer a simple, widely applicable, and clinically useful approach for rapid severity stratification and guiding therapeutic interventions, including in nonspecialist settings. However, oxygenation composite metrics incorporating lung mechanics and gas exchange offer potentially superior insights. Meanwhile, advances in monitoring, such as lung ultrasound and electrical impedance tomography, challenge the assumption that a single-point oxygenation measurements adequately capture the complexity of respiratory failure.[4]

This review explores the evolution, evidence, and limitations of oxygenation ratios and delves into the emerging oxygenation metrics. As healthcare systems increasingly prioritize resource efficiency and precision-based interventions, the time has come to assess whether our traditional approaches to oxygenation assessment still serve modern clinical practice, or whether a shift toward more comprehensive, precise strategies is needed to optimize patient management and outcomes.

History of Oxygenation Ratios

The first description of ARDS provided by David Ashbaugh in 1967[5] emphasized the hypoxia that was refractory to conventional oxygen therapy. In 1988, Murray et al.[6] proposed a Lung Injury Score as an expanded definition that incorporated key pathophysiological features of ARDS. This framework incorporated the PaO2/FiO2 ratio as a component reflecting the severity of lung injury, alongside other parameters such as positive end-expiratory pressure (PEEP), respiratory system compliance, and chest radiographic findings. Furthermore, while a PEEP was included in the Lung Injury Score, it is important to note that the PaO2/FiO2 ratio was scored independently of the PEEP level. The American-European Consensus Conference delineated acute lung injury and ARDS using PaO2/FiO2 ratios of ≤300 and <200 mmHg, respectively.[7] Notably, this definition also failed to specify ventilator settings, particularly minimum PEEP and FiO2 levels, at the time of measurement. Ferguson et al.,[8] proposed a minimum PEEP of 10 cmH2O for the assessment of oxygenation in the first Delphi definition.

Villar et al.,[9] prospectively evaluated different combinations of PEEP and FiO2 settings in a cohort of patients with ARDS and demonstrated that ventilator settings of PEEP ≥ 10 cmH2O and FiO2 ≥ 0.5 on day 1, were associated with significant difference in intensive care unit (ICU) mortality across categories of acute lung injury, ARDS, and acute hypoxemic respiratory failure (AHRF).[9] In a separate multicenter prospective study, the severity categorization of ARDS based on PaO2/FiO2 measurements with a PEEP ≥ 10 cmH2O and FiO2 ≥ 0.5 on day 1 was associated with significant differences in mortality across categories.[10]

Subsequently, the Berlin definition reduced the minimum level of PEEP to 5 cmH2O for PaO2/FiO2 ratio measurement on the basis of feasibility.[11] Additionally, the PaO2/FiO2 ratio was used to stratify ARDS into mild (201 to ≤300 mmHg), moderate (101 to ≤200 mmHg), and severe (≤100 mmHg) categories. The thresholds for the PaO2/FiO2 ratio in the New Global definition remain unchanged.[12]

The SpO2/FiO2 ratio was proposed by Rice et al.,[13] as an alternative oxygenation ratio for the diagnosis and follow-up of ARDS patients in situations where arterial blood gas (ABG) is unavailable.[13] The Kigali modification of the Berlin definition of ARDS has also reported a comparative efficacy of the SpO2/FiO2 ratio relative to the PaO2/FiO2 ratio in diagnosing ARDS, based on a single-center study, conducted in a resource-constrained setting of Rwanda.[14] The new Global definition incorporated SpO2/FiO2 as an alternative to PaO2/FiO2 ratio in resource-limited settings, for the diagnosis of hypoxia and for classification of severity of ARDS, provided the SpO2 is ≤97%.[10]

Evidence on Oxygenation Ratios

The PaO2/FiO2 ratio for defining hypoxemia in ARDS was formally incorporated into the American-European Consensus Conference definition through expert consensus.[7] The Berlin definition used a Delphi-based consensus process involving an international panel to draft a definition and to propose a PaO2/FiO2 ratio-based categorization of ARDS. The definition and severity categories were validated through a meta-analysis of patient-level datasets from four multicentric clinical trials involving over 4000 patients. The mortality rates, median duration of ventilation, and ventilator-free days were inversely proportional to the PaO2/FiO2 ratios. The thresholds for PaO2/FiO2 ratios were merely developed through consensus for practical considerations in grading outcomes, including mortality, along a continuum, rather than relying on strict physiological inflection points. Therefore, it may seem reasonable to think that the PaO2/FiO2 thresholds are predictive of mortality in ARDS. However, the area under the receiver operating characteristic curve (AUROC) for the Berlin definition was only 0.577 (95% confidence intervals [CI]: 0.561 to 0.593).[11] Moreover, the Berlin definition does not clearly specify the optimal timing for severity assessment using PaO2/FiO2 ratios (e.g., at admission, within the first 24 h or at the worst point within the first 48 h), nor the extent of ventilator optimization before measurement. This lack of standardization may influence the classification of ARDS severity, as oxygenation indices can vary considerably depending on ventilator settings and the timing of assessment.[9,10]

A few clinical trials and guidelines have attempted to categorize ARDS using a single PaO2/FiO2 cutoff of 150 mmHg, distinguishing mild-moderate from moderate-severe disease.[[15], [16], [17]] Notably, several landmark ARDS trials, including those evaluating neuromuscular blockade and prone positioning, have used a PaO2/FiO2 threshold of 150 mmHg for patient selection.[15,16] Moderate-severe ARDS (PaO2/FiO2 <150 mmHg) had worse physiological and anatomical derangements (such as peak pressure, arterial partial pressure of carbon dioxide [PaCO2], pH, heavier lungs, and greater inhomogeneity), greater lung recruitability and higher mortality risk compared with mild-moderate ARDS (PaO2/FiO2 150 to ≤300 mmHg).[18] Yet, the accuracy of PaO2/FiO2 <150 mmHg for predicting mortality was low with an AUROC of 0.56 (95%CI: 0.51 to 0.61).[19]

The SpO2/FiO2 ratio derived from pulse oximetry possesses inherent advantages such as non-invasiveness, continuous, ease of availability, and affordability and is a fair surrogate of PaO2, with SpO2/FiO2 ratios of 315 and 235 corresponding to PaO2/FiO2 ratios of 300 mmHg (91% sensitivity, 56% specificity) and 200 mmHg (86% sensitivity, 85% specificity).[20] However, the relationship between SpO2 and PaO2 is nonlinear and becomes inaccurate above 96% because of the flat portion of oxyhemoglobin curve.[13] In a large mixed cohort with AHRF including pneumonia, Rice et al.,[13] found that SpO2/FiO2=235 corresponded to PaO2/FiO2=200, and SpO2/FiO2=315 to PaO2/FiO2=300, The study reported a strong correlation between SpO2/FiO2 ratio and PaO2/FiO2 ratio, demonstrating a linear relationship described by the equation[13]:

SpO2/FiO2=64+0.84(PaO2/FiO2)

However, the misclassification of ARDS severity persisted in about 30% of cases. The authors emphasized that FiO2 standardization is critical for reliable interpretation.[13] The correlation was further supported by a secondary analysis of the National Heart, Lung, and Blood Institute ARDS Network trials.[[21], [22], [23]] Additionally, secondary analysis of prospective observational study on COVID-19-related ARDS indicated a good correlation between SpO2/FiO2 or PaO2/FiO2 ratios in predicting patient outcomes.[24] Moreover, imputed SpO2/FiO2 values corresponding to PaO2/FiO2 ratios have also been validated within the Sequential Organ Failure Assessment (SOFA) score and have been shown to predict clinical outcomes[25] (Table 1).

Table 1.

Adult studies showing prediction of PaO2/FiO2 from SpO2/FiO2 ratios.

Authors Patient population Sample size Correlation between SpO2/FiO2 and PaO2/FiO2 Statistical results
Rice et al.[13] ARDS patients under MV 672 for derivation
and 402 for validation
SpO2/FiO2=235 indicated PaO2/FiO2=200
SpO2/FiO2=315 indicated PaO2/FiO2=300
NA
Tripathi et al.[26] Patients on IMV as a part of general anesthesia 2754 SpO2/FiO2=206 indicated PaO2/FiO2=300
SpO2/FiO2=180 indicated PaO2/FiO2=200
NA
Namendys-Silva et al.[27] All ICU patients 232 SpO2/FiO2 ≤ 512 indicating PaO2/FiO2 ≤ 400
SpO2/FiO2 ≤ 357 indicating PaO2/FiO2 ≤ 300
SpO2/FiO2 ≤ 214 indicating PaO2/FiO2 ≤ 200
SpO2/FiO2 ≤ 89 indicating PaO2/FiO2 ≤ 100
NA
Serpa Neto et al.[28] Patients with sepsis 260 SpO2/FiO2=154 indicated PaO2/FiO2=100
SaO2/FiO2=241 indicated PaO2/FiO2=300
NA
Schmidt et al.[29] All patients under MV 3767 SpO2/FiO2=295 indicated PaO2/FiO2 ≤ 300 99% sensitivity and 9.9% specificity
Bass et al.[30] Patients with ARF under MV 77 SpO2/FiO2 ≤ 315 indicated PaO2/FiO2 ≤ 300

SpO2/FiO2 ≤ 235 indicated PaO2/FiO2 ≤ 200
Sensitivity of 83% and specificity of 50%
Sensitivity of 70% and specificity of 90%
Bashar et al.[31] ICU post-operative after CABG 175 for derivation
and 358 for validation
SpO2/FiO2=311 indicated PaO2/FiO2=300 Sensitivity of 90% and specificity of 80%
Babu et al.[32] Respiratory failure admitted in surgical ICU 300 SpO2/FiO2=285 indicated PaO2/FiO2=200

SpO2/FiO2=323 indicated PaO2/FiO2=300
Sensitivity of 73% and specificity of 77%
Sensitivity of 72% and specificity of 73%
Kumar et al.[33] ARF admitted in COVID ICU 80 SpO2/FiO2=219 indicated PaO2/FiO2=200
SpO2/FiO2=299 indicated PaO2/FiO2=300
NA
Gandhi et al.[34] Respiratory failure admitted in COVID ICU 100 SpO2/FiO2=205 indicated PaO2/FiO2=200
SpO2/FiO2=290 indicated PaO2/FiO2=300
NA
Choudhury et al.[35] COVID-19 patients admitted to ICU 396 SpO2/FiO2=395 indicated PaO2/FiO2=100

SpO2/FiO2=405 indicated PaO2/FiO2=200

SpO2/FiO2=411 indicated PaO2/FiO2=300
Sensitivity of 66.7% and specificity of 73.5%
Sensitivity of 71.8% and specificity of 70.8%
Sensitivity of 65% and specificity of 80%
Alur et al.[36] ICU patients with AHRF 125 SpO2/FiO2=252 indicated PaO2/FiO2=200
SpO2/FiO2=321 indicated PaO2/FiO2=300
Sensitivity of 69% and specificity of 95%

AHRF, Acute hypoxic respiratory failure; ARDS, Acute respiratory distress syndrome; ARF, Acute respiratory failure; CABG, Coronary artery bypass grafting; ICU, Intensive care unit; MV, Mechanical ventilation; PaO2/FiO2, Partial pressure of arterial oxygen by fraction of inspired oxygen; SpO2/FiO2, Oxygen saturation by fraction of inspired oxygen.

Oxygenation Ratios Beyond ARDS

The PaO2/FiO2 ratio serves as a gold standard for assessing the severity of hypoxemia for ARDS. However, clinicians increasingly use the PaO2/FiO2 ratio and ARDS severity thresholds (e.g., PaO2/FiO2 ≤300 or ≤200 mmHg) in other contexts such as pneumonia, sepsis, perioperative care, and determining the need for respiratory supports like high-flow nasal oxygen (HFNO), noninvasive ventilation (NIV). Moreover, the severity of hypoxemia has been shown to be an independent predictor of mortality in critically ill patients, regardless of the underlying etiology.[37] Furthermore, SpO2/FiO2 ratio is used as an alternative to PaO2/FiO2 ratio in many of these conditions, where ABG analysis access is limited.

Pneumonia

The PaO2/FiO2 ratio has been used in patients with community-acquired pneumonia (CAP) to determine the severity of hypoxemia, to assess the need for respiratory support, and to predict the outcomes.[38] Clinical risk scores such as SMART-COP, which includes the PaO2/FiO2 ratio, are valuable in predicting the severity of CAP and the need for respiratory or vasopressor support.[39] The oxygenation indices can also support clinical decision-making by helping to guide the escalation of respiratory support and monitor response to therapy. Furthermore, worsening hypoxemia, as reflected by declining PaO2/FiO2 ratios, may identify patients at increased risk of progression to ARDS, particularly when invasive ventilation becomes necessary.[40] Collectively, these observations highlight the importance of oxygenation indices not only in severity assessment but also in anticipating disease trajectory in pneumonia.

The noninvasive nature of SpO2/FiO2 made it valuable, especially where ABGs were limited. In adult pneumonia patients, both COVID-19 and non-COVID-19 related, contemporaneous SpO2/FiO2 and PaO2/FiO2 ratios correlate moderately to strongly, but clinically relevant dispersion limits their interchangeability.[33,41] However, wide prediction intervals shown in other settings may reduce their standalone reliability. One multicenter study of patients with COVID-19 pneumonia across wards and ICUs demonstrated that SpO2/FiO2 could be used as a surrogate of PaO2/FiO2 ratio in patients and could be highly useful in resource-limited settings during this COVID-19 pandemic. However, the initial SpO2/FiO2 ratio on admission could not indicate the need for invasive ventilation.[42]

In pneumonia (COVID-19 and non-COVID-19), SpO2/FiO2 provides a practical, noninvasive screening and trending tool. It correlates with PaO2/FiO2 but with a risk of clinically relevant misclassification at decision thresholds. For research eligibility, prognosis, or ventilator management, the PaO2/FiO2 ratio is preferable due to greater precision. For triage and early recognition of deterioration, the SpO2/FiO2 ratio is acceptable, with the caveat that SpO2 >97% blunts its reliability due to a plateau in the oxyhemoglobin curve.

Sepsis and septic shock

The PaO2/FiO2 ratio is important for the diagnosis of sepsis as a variable of the SOFA score. The cutoffs of the PaO2/FiO2 ratio used in SOFA were also similar to those for severity grading of ARDS, with a higher score for a low PaO2/FiO2. However, the thresholds for the PaO2/FiO2 ratio were changed in the SOFA-2 score (Table 2). Although, the accuracy of predicting mortality for overall SOFA–2 (AUROC 0.79, 95%CI: 0.76 to 0.81) was similar to SOFA (AUROC 0.77, 95%CI: 0.74 to 0.81), the revised cutoffs using 10 kPa (75 mmHg) intervals for PaO2/FiO2 ratio, outperformed the respiratory SOFA score (AUROC 0.615, 95%CI: 0.607 to 0.622 vs. AUROC 0.610, 95%CI: 0.603 to 0.618).[[43], [44], [45]] Additionally, lower PaO2/FiO2 (<300 mmHg) ratios have been shown to be independently associated with increased mortality in patients with sepsis.[46] In a retrospective analysis of MIMIC-IV database, the relationship between PaO2/FiO2 ratio and mortality was U-shaped and PaO2/FiO2 ratios between 183 and 219 mmHg were found to be associated with a lower risk of 28-day mortality.[47]

Table 2.

Comparison of respiratory scoring criteria (PaO2/FiO2) between SOFA and SOFA‑2.

Score SOFA SOFA-2
0 ≥400 mmHg (53.3 kPa) >300 mmHg (>40.0 kPa) or SpO2/FiO2 >300 mmHg
1 <400 (53.3) ≤300 mmHg (≤40.0 kPa) or SpO2/FiO2 ≤300 mmHg
2 <300 mmHg (40.0 kPa) ≤225 mmHg (≤30.0 kPa) or SpO2/FiO2 ≤250 mmHg
3 <200 mmHg (26.7 kPa) with respiratory support ≤150 mmHg (≤20.0 kPa) and advanced ventilator support or SpO2/FiO2 ≤200 mmHg
4 <100 mmHg (13.3 kPa) with respiratory support ≤75 mmHg (≤10.0 kPa) and advanced ventilator support or ECMO or SpO2/FiO2 ≤120 mmHg

ECMO, Extracorporeal membrane oxygenation; SaO2/FiO2, Oxygen saturation by fraction of inspired oxygen; SOFA, Sepsis-related organ failure assessment.

SOFA-2 score also incorporates corresponding SpO2/FiO2 ratios when ABG is unavailable. Peripheral perfusion abnormalities associated with worsening shock, due to vasoconstriction or low pulsatility, could be a major concern. Although a low SpO2/FiO2 ratio at ICU admission in patients with sepsis and septic shock was associated with increased mortality risk, especially the lower values; the correlation with the PaO2/FiO2 ratio was only moderate (r = 0.48).[28] In another study in the emergency department, pulse oximetry overestimated arterial oxygen saturation (SaO2) measured by ABG in patients with severe sepsis or septic shock, with larger errors at lower saturations.[48] Hence, in sepsis, PaO2/FiO2 is more reliable when obtainable. If SpO2/FiO2 has to be used as a noninvasive alternative, SpO2 should be kept <97%, and near decision thresholds should be interpreted cautiously.[49]

Pulmonary embolism

The PaO2/FiO2 ratio has been used for risk stratification in patients with pulmonary embolism and for identifying patients with a higher risk of deterioration.[50] The PaO2/FiO2 ratio <265 mmHg has been identified as an increased risk of in-hospital death, though other studies suggest slightly different cutoffs.[51] Furthermore, the PaO2/FiO2 ratio has been used to estimate the severity of pulmonary artery obstruction, as it is negatively correlated with the pulmonary artery computed tomography obstruction index. A lower PaO2/FiO2 ratio corresponds to a more severe obstruction.[52] Although the PaO2/FiO2 ratio may correlate with severity in pulmonary embolism, oxygenation parameters alone do not play a pivotal role in therapeutic decision-making, which is primarily guided by hemodynamic status, right ventricular function, and cardiac biomarkers.

Perioperative and anesthesia settings

In the operating room and postanesthetic care unit, FiO2 delivery fluctuates rapidly, and lung mechanics change with anesthesia and recruitment. Intra or postoperative oxygenation is confounded by high FiO₂, variable PEEP, atelectasis, and recruitment manoeuvres—factors that can acutely swing PaO2/FiO2 without reflecting intrinsic lung pathology. These factors limit the reliability of both PaO2/FiO2 and SpO2/FiO2 thresholds derived from resting or ventilated ARDS populations. Although perioperative studies exist for PaO2/FiO2 ratio,[53] or comparing oxygen reserve index vs. SpO₂ for hypoxemia detection, they generally support pulse oximetry for trend monitoring, but not for applying ARDS–based cutoffs.[54]

Oxygenation Ratios with Respiratory Support

Respiratory support modalities, such as HFNO, continuous positive airway pressure, bilevel positive airway pressure, NIV, invasive mechanical ventilation, or long–term home ventilation, are often used in the management of respiratory failure. The use of PaO2/FiO2 and SpO2/FiO2 ratios with these modalities is influenced by the settings and PEEP levels.

Estimation of FiO2 during conventional oxygen therapy can be challenging, particularly when oxygen is delivered via standard face masks or nasal canula. FiO2 is commonly estimated using conversion tables based on oxygen flow rates. However, the actual delivered FiO2 can vary considerably when estimated by these equations, depending on factors such as the patient’s inspiratory flow, breathing pattern, and mask fit, and consequently exaggerate the severity of hypoxemia, calculated by PaO2/FiO2 or SpO2/FiO2 ratios.[55] This methodological limitation should be considered when interpreting studies comparing different noninvasive respiratory support strategies.

High-flow nasal oxygen

In HFNO, FiO2 delivery is more stable at higher flows but still affected by mouth opening, patient inspiratory demand, respiratory rate, and nasal prong fit. Moreover, the physiological benefits of HFNO are due to flow-dependent positive nasopharyngeal pressure, expiratory resistance (PEEP), thereby reducing respiratory rate and effort. However, the benefits are dependent on the patients’ ability to close their mouths while on HFNO.[[56], [57], [58]] Thus variable FiO2 and PEEP delivery may perturb both PaO2/FiO2 and SpO2/FiO2 ratios, independently of underlying disease severity. This leads to greater heterogeneity and overestimation of ARDS severity.[59] Consequently, PaO2/FiO2 ratio thresholds established in invasively ventilated patients with ARDS cannot be directly extrapolated to this population. Nevertheless, the oxygenation indices retain clinical relevance under HFNO, as lower values remain associated with worse outcomes, reflecting more severe hypoxemia and disease burden. When ABGs are sparse, SpO2/FiO2 trends and composite indices that incorporate respiratory rate (ROX index=[SpO2/FiO2]/respiratory rate) may be useful, but absolute cutoffs beyond ARDS (e.g., SpO2/FiO2 235/315) translate poorly; device- and patient-specific FiO₂ variability widens error bands.

Noninvasive ventilation

NIV introduces additional uncertainty concerning the FiO₂ delivery due to potential leaks and air entrainment at the interface. These factors can significantly lower the delivered FiO2 and thereby distorting the oxygenation ratios. Both bench and clinical studies have demonstrated that the presence of leaks can result in a considerable reduction in FiO2 and risk overestimation of oxygenation status.[55,60] In the post hoc analysis of the LUNG SAFE study, the severity assessment by PaO2/FiO2 ratio on NIV was associated with greater intensity of ventilatory support and higher ICU mortality compared to patients on invasive ventilation, particularly among those with a PaO2/FiO2 <150 mmHg.[61] Nevertheless, SpO2/FiO2-based rules yield important prognostic information in patients with mixed-etiology AHRF on NIV, with a low SpO2/FiO2 ratio is indicative of higher likelihood of NIV failure.[62]

Limitations and Challenges with Oxygenation Ratios

The ongoing pursuit of an ideal oxygenation matrix in the context of AHRF and ARDS remains a critical area of research. Each index has unique advantages and limitations, and its application may be influenced not only by its inherent limitations but also by a multitude of patient and institutional factors. Consequently, it is imperative to judiciously evaluate these factors when designing or interpreting the results of any clinical trial or translating the evidence at the bedside for decision–making.

Traditionally, the PaO2/FiO2 ratio has been used for the diagnosis, management, and prognostication of ARDS.[2,11] For most clinical situations, the PaO2/FiO2 ratio remains the gold standard in the management of AHRF. The PaO2/FiO2 ratio cutoff for categorizing ARDS was, however, challenged by many dataset analyses, especially the threshold of 300 mmHg.[63,64] A recent updated meta-analysis of 28 clinical trials involving 38,270 patients reported a lower accuracy of the PaO2/FiO2 ratio in predicting mortality for patients with ARDS and recommended that the ratio may not be suitable for use as a sole clinical prognostic tool. They reported an AUROC of 0.60 (95%CI: 0.58 to 0.64) for PaO2/FiO2 of 100 mmHg, with a pooled sensitivity and specificity of 44.8% (95%CI: 38.1 to 51.7%), and 70.6% (95%CI: 65.9 to 74.9%), respectively. In contrast, for a PaO2/FiO2 ratio of 200 mmHg, the AUROC was 0.64 (95%CI: 0.60 to 0.69), with pooled sensitivity and specificity of 83.9% (95%CI: 78.9 to 87.8%) and 26.1% (95%CI: 20.8 to 32.1%), respectively.[65]

The PaO2/FiO2 ratio is affected by various factors such as switching from non-invasive to invasive mechanical ventilation and ventilatory settings, for example, PEEP, FiO2, mean airway pressure, and minute ventilation. An important limitation of the PaO2/FiO2 ratio is its nonlinear relationship with FiO2, whereby changes in FiO2 can lead to disproportionate and sometimes misleading changes in the PaO2/FiO2 ratio, independent of true alterations in pulmonary gas exchange. This phenomenon reflects the complex interplay between ventilation–perfusion mismatch, shunt fraction, and oxygen–hemoglobin dissociation curve.[66] As a result, identical PaO2/FiO2 ratios may represent different degrees of lung injury at varying FiO2 levels, limiting their reliability as a standalone marker of disease severity. The nonlinear PaO2/FiO2–FiO2 relationship underscores the need to interpret the PaO2/FiO2 ratio within the broader clinical and ventilatory context. The dependence on FiO2 may also diminish the utility of these indices, as FiO2 can be influenced by various factors, such as mode of oxygen delivery (nasal prongs, face mask, HFNO, NIV, or invasive mechanical ventilation) and ventilatory settings.[49]

Moreover, PaO2/FiO2 requires arterial sampling, which is invasive, painful, and may not be readily available, especially in resource-limited settings. It also requires technical expertise, and its processing requires special apparatus, which is expensive, needs maintenance, and is not widely available. It may not be possible to store and transport ABG samples, as this may affect PaO2 readings.[2]

The SpO2/FiO2 ratio has been suggested as a potential surrogate and may be particularly useful in situations where ABG sampling is unavailable or not feasible.[11,67] The seminal ARDS work that popularized SpO2/FiO2 cutoffs cautioned about its use at high saturations.[13] The correlation between SpO2/FiO2 and PaO2/FiO2 ratios has been evaluated in several recent studies.[[13], [28], [29], [33], [68], [69]] Despite numerous studies, the correlation between PaO2/FiO2 and SpO2/FiO2 ratios, is moderate to strong.[[26], [32], [34], [70]] Moreover, the oxyhemoglobin curve plateaus above SpO2 >96%, and imputation of PaO2/FiO2 from SpO2/FiO2 may become inaccurate.[71] Therefore, different SpO2/FiO2 thresholds have been identified across studies, and different equations have been proposed for imputing PaO2/FiO2 from SpO2/FiO2.[26,36,72] The challenges exist in using SpO2/FiO2 and PaO2/FiO2 interchangeably, especially at higher FiO2, and in monitoring the severity and progression of ARDS.[68,70,73] Imprecision of the SpO2 measurement and these equations may lead to misclassification of ARDS based on the SpO2/FiO2 ratio, which could be as high as 33% of patients. Moreover, the accuracy does not improve with dynamic monitoring of the SpO2/FiO2 ratio.[72] Therefore, panelists from a recent international Delphi study on defining and subphenotyping ARDS expressed reservations regarding the routine implementation of the SpO2/FiO2 ratio, and indicated the need for further validation before its acceptance in future ARDS definitions.[73]

Furthermore, several other factors may also limit the clinical applicability of the SpO2/FiO2 ratio.[2,49] The SpO2/FiO2 ratio is dependent on pulse oximetry readings, which can be influenced by various factors such as anemia, skin color, carbon monoxide intoxication, methemoglobinemia, certain dyes and pigments (e.g., nail polish), hyperbilirubinemia, hypoperfusion, high venous pressure, hypoxemia, and hyperoxia.[49] Moreover, pulse oximetry was associated with greater occult hypoxemia and can artifactually inflate the SpO2/FiO2 ratio, particularly in darker-skinned patients.[74] The hidden hypoxia was particularly higher among Black individuals and those of African American ethnic background.[75] Moreover, this overestimation of oxygen saturation translates into health disparities with delayed treatment and higher readmission to the hospital among patients with dark-skin pigmentation.[76] Even excessive patient movements can impact the accuracy of pulse oximetry readings. Given the common presence of these factors in critically ill patients, the validity of pulse oximetry readings may be compromised.

Beyond oxygenation ratios

Although oxygenation indices are surrogates of lung injury severity and are valuable in clinical decision-making, their interpretation is influenced by the complex interplay of ventilation–perfusion mismatch, shunt fraction, and the level and type of respiratory support provided. As a result, patients with markedly different pathophysiological profiles may have similar PaO2/FiO2 ratios, thereby complicating risk stratification and limiting treatment precision.[66] Despite the ease of use of oxygenation ratios, their limitations are becoming increasingly recognized. Several additional or complementary metrics have been made available to support a more comprehensive evaluation of respiratory function, potentially better capturing the interactions among oxygenation, lung mechanics, and ventilator settings. We have categorized these indices into: those readily available at bedside and do not require advanced ventilatory maneuvers; those requiring specific ventilatory maneuvers/calculations, and those requiring additional devices.

Indices not Requiring Specific Ventilator Maneuvers

PaO2/FiO2 ratio does not account for airway pressures, and thus, investigators included PEEP as the (PaO2 ×10) / (FiO2 × PEEP) ratio,[77] or mean airway pressure as an oxygenation factor,[78] as more integrative metrics. Both indices have been shown to be superior in predicting severity and patient outcomes.

Oxygenation factor=PaO2/FiO2/Paw, where Paw is the mean airway pressure.

Similarly, oxygenation index (Paw × 100 × FiO2/PaO2) and surrogate oxygen saturation index (Paw × 100 × FiO2/SpO2) incorporating mean airway pressure have been found to be superior predictors of outcomes than traditional oxygenation ratios.[79,80]

The dynamic trajectory of the PaO2/FiO2 ratio or reclassification of severity after 24–48 h has also been found to be a better predictor of prognosis, response to interventions such as PEEP, and patient outcomes than static measurements.[[81], [82], [83]]

Indices Requiring Specific Ventilatory Maneuvers or Calculations

Lung compliance and elastance, calculated from plateau pressure and tidal volume, reflect the respiratory system’s mechanical properties. Lower compliance and higher elastance frequently suggest more severe lung parenchymal involvement or atelectasis, and are independently linked with death.[84] Combining oxygenation indices and compliance (for example, the “oxygenation-compliance index”) may aid in the identification of patients with recruitable lung tissue or different physiological phenotypes.[85,86] For example, two patients with equal PaO2/FiO2 ratios may differ significantly in compliance—one with recruitable atelectasis, the other with fixed fibrotic change—implying very different responses to PEEP or prone positioning (Figure 1).

Figure 1.

Figure 1: dummy alt text

Relationship between oxygenation indices and respiratory system compliance to define physiological phenotypes in ARDS. The figure illustrates the relationship between oxygenation and static compliance, highlighting that patients with similar PaO2/FiO2 ratios may have markedly different respiratory mechanics, reflecting distinct underlying pathophysiology. Three representative phenotypes are shown: PaO2/FiO2 with preserved compliance (recruitable atelectasis), suggesting potentially recruitable lung units despite impaired oxygenation; low PaO2/FiO2 with low compliance (fibrotic or nonrecruitable lung), consistent with severe, poorly recruitable lung injury; and preserved PaO2/FiO2 with high compliance, representing mild or resolving disease. Integrating oxygenation and compliance offers a more comprehensive, physiology-based assessment of ARDS, enabling better phenotyping and more individualized ventilatory management than either parameter alone.

ARDS, Acute respiratory distress syndrome; PaO2/FiO2, Partial pressure of oxygen to fraction of inspired oxygen.

Similarly, the dead space fraction (Vd/Vt) measures ventilation efficiency and pulmonary perfusion. Elevated dead space can be caused by alveolar overdistension, microvascular obstruction, or thrombotic complications, none of which are reflected in oxygenation measurements. Increased Vd/Vt has been repeatedly linked to poor outcomes in ARDS and may act as a surrogate for pulmonary vascular dysfunction, supplementing oxygenation-based measurements.[87,88] Surrogates, such as the ventilatory ratio (VR)[89] or the PaCO2/end-tidal carbon dioxide (etCO2) gradient, can be used at the bedside when direct measures of dead space are unavailable.[[90], [91], [92]]

The shunt fraction is the proportion of cardiac output that perfuses nonventilated regions and directly quantifies the severity of intrapulmonary shunt. Although perfect estimation necessitates invasive sampling or numerous gas measurements, simple calculations using alveolar gas equations can approximate the shunt fraction. Integrating shunt, dead space, and compliance data may assist in distinguishing “low-compliance, high-shunt” from “high-compliance, high-dead-space” phenotypes, which respond differently to PEEP, recruitment, or prone positioning.[92,93]

Recent research has examined composite measures that combine oxygenation and mechanics. The “oxygenation–ventilation index,” “mechanical power,” and “stress index” take into account pressure–volume relationships, respiratory rate, and driving pressure to measure both the efficiency and potential harm caused by mechanical ventilation. These multidimensional metrics better reflect the interactions among gas exchange, mechanics, and ventilator settings, improving predictive power and therapy personalization.[94,95]

Indices Requiring Additional Devices

Beyond traditional physiological parameters, innovative monitoring technologies provide novel insights into regional and dynamic variation. Electrical impedance tomography (EIT) provides for continuous bedside monitoring of regional ventilation and, with newer algorithms, perfusion distribution. EIT can assess regional compliance and detect overdistension or collapse during PEEP titration, bridging the gap between global indices and regional physiology. Lung ultrasound and quantitative CT help to characterize aeration and recruitability, while newer techniques such as functional lung MRI or dual-energy CT allow for noninvasive measurement of ventilation–perfusion mismatch and shunt percentage. Integrating these imaging modalities with standard measurements has the potential to reinvent severity classification, shifting away from static thresholds (such as PaO2/FiO2 cutoffs) toward physiologically grounded multidimensional metrics.[4]

Despite these encouraging developments, major issues remain. Many of these measurements necessitate specialized equipment, are operator-dependent, or lack standardized thresholds. Furthermore, their interpretation is complicated by ventilatory settings, hemodynamics, and metabolic state. To translate these intricate factors into effective bedside algorithms, it is essential to conduct large, prospective cohort studies that are integrated with patient outcomes. Notably, many of these indices either incorporate the PaO2/FiO2 ratio or complement these indices to support more precise, physiology-guided management. Therefore, the purpose is not to replace the PaO2/FiO2 ratio, but to contextualize it within a broader framework that includes both gas-exchange efficiency and respiratory system mechanics.

Research Priorities and the Way Forward

The limitations of current oxygenation metrics demand a systematic research agenda to develop more nuanced, context-specific approaches to respiratory assessment and care. Future investigations must move beyond simple oxygenation ratios validation toward a comprehensive evaluation of how oxygenation measures should be integrated into modern clinical practice.[96]

The thresholds for PaO2/FiO2 ratio were established decades ago from relatively small datasets that do not reflect contemporary patient populations, ventilation strategies, or outcomes. Large-scale analyses of modern ICU databases could identify cutoffs that better align with current mortality patterns, length of stay, and resource utilization. Machine learning techniques, applied across diverse healthcare systems, might reveal that optimal thresholds vary substantially by age, comorbidity burden, or underlying pathophysiology.[97,98] Importantly, such efforts should prioritize clinically meaningful outcomes over historical precedent, recognizing that different thresholds may more accurately predict treatment response, recovery trajectory, or the need for escalation of care.

Equally important is the recognition that universal cutoffs may be fundamentally flawed. HFNO, for example, produces unique physiological conditions in which traditional PaO2/FiO2 or SpO2/FiO2 thresholds poorly predict treatment failure or the need for intubation. Systematic studies are urgently required to evaluate oxygenation metrics across the full spectrum of respiratory support modalities, from conventional oxygen therapy and NIV to HFNO and invasive ventilation. These investigations should aim to establish context-specific thresholds that reflect the distinct physiological states created by each intervention. Patient-specific factors must also be considered. The optimal oxygenation target for an elderly patient with multiple comorbidities may differ substantially from that for a young, otherwise healthy patient with isolated respiratory failure (Figure 2).

Figure 2.

Figure 2: dummy alt text

Research priorities for oxygenation metrics.

EHR, Electronic health record; ML, Machine learning; PEEP, Positive end-expiratory pressure; PaO2/FiO2, Partial pressure of oxygen to fraction of inspired oxygen; SpO2/FiO2, Capillary oxygen saturation to fraction of inspired oxygen;

*Standardized measurement includes ventilator settings, FiO2, PEEP, flow rate, patient position, time relative to intervention, vasopressors, and skin pigmentation.

†Subphenotyping using longitudinal trajectories of PaO2/FiO2 or SpO2/FiO2 ratios over static measurement.

‡Integration, automatic extraction, and validation of PaO2/FiO2 and SpO2/FiO2 from EHRs for alerts, dashboards, and research phenotyping.

§Physiological confounders include effect of pH, arterial partial pressure of carbon dioxide, temperature, hemoglobin, 2,3-diphosphoglycerate (DPG), dyshemoglobinemias and cardiac output.

Beyond refining thresholds, future research should move away from single-parameter assessments. Composite indices that capture the complexity of respiratory failure hold greater promise. Metrics combining oxygenation with measures of lung mechanics, such as PEEP, mean airway pressure, lung compliance and driving pressure, gas exchange efficiency, including dead space fraction, hemodynamic parameters, and inflammatory markers, could yield a more accurate and clinically useful assessment. Integration of routinely available ventilator–derived data, such as pressure-volume loops, flow patterns, or indices of work of breathing, with oxygenation measures could enhance bedside evaluation without adding invasive monitoring.[99] Emerging technologies, including lung ultrasound and EIT, further expand opportunities for composite metric development. Studies linking these modalities with traditional oxygenation indices may provide more reliable insight into regional lung function, recruitment potential, and respiratory reserve.

Crucially, research priorities must extend beyond the development of new metrics to their implementation in clinical practice. Understanding how different oxygenation assessment strategies affect clinician decisions, workflow efficiency, and patient comfort represents a major knowledge gap. Studies assessing the cognitive load imposed by various monitoring approaches and their influence on provider satisfaction are necessary for successful adoption.

The ultimate future of oxygenation assessment may lie not in refining individual metrics but in developing intelligent, context-aware systems that integrate multiple data streams to provide personalized respiratory evaluation. Such systems should account for patient-specific factors, treatment context, and resource constraints while maintaining the simplicity required for widespread clinical use.

In summary, the path forward involves a shift from reliance on simple, universal ratios toward sophisticated, adaptable approaches that reflect the complexity of respiratory failure. Progress will depend on combining modern analytics, advanced monitoring, and pragmatic implementation research to ensure that innovations translate into real-world benefit. The ultimate goal is to have oxygenation assessment strategies that improve patient outcomes while reducing the burden on healthcare systems in an era of increasingly personalized, resource–conscious medicine.

Conclusions

Oxygenation ratios provide simple and rapid estimates of lung injury, enabling bedside assessment of ARDS severity and prognostication. They are instrumental in screening, triage, and trend monitoring for standardized interventions. However, both metrics, PaO2/FiO2 and SpO2/FiO2 ratios, can be misleading and are highly influenced by the ventilatory settings such as PEEP, mean airway pressure, and FiO2. The SpO2/FiO2 ratio is a promising alternative to the PaO2/FiO2 ratio, especially in resource–limited settings, but physiologic nonlinearity, FiO₂ variability, and oximetry bias limit precision around decision thresholds. There is a need for research on the validation of oxygenation ratios in diverse clinical settings, modes of oxygen delivery, and integration with other oxygenation indices, including pulmonary mechanics. Although oxygenation ratios are valuable screening and monitoring tools, they should be interpreted within the broader clinical and physiological context rather than as standalone markers of gas exchange and disease severity.

CRediT authorship contribution statement

Prashant Nasa: Writing – review & editing, Writing – original draft, Data curation, Conceptualization. Denise Battaglini: Writing – review & editing, Writing – original draft, Conceptualization. Deven Juneja: Writing – review & editing, Writing – original draft, Conceptualization. Marcus J. Schultz: Writing – review & editing, Writing – original draft, Supervision, Conceptualization. Jozef Kesecioglu: Writing – review & editing, Writing – original draft, Supervision, Methodology, Conceptualization.

Conflict of Interest

All the authors, Prashant Nasa, Denise Battaglini, Deven Juneja, Marcus J. Schultz, and Jozef Kesecioglu declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. Denise Battaglini and Jozef Kesecioglu are Editorial Board Members for this journal and were not involved in the editorial review or the decision to publish this article.

Managing editor: Jingling Bao

References

  • 1.Gattinoni L., Vassalli F., Romitti F. Benefits and risks of the P/F approach. Intensive Care Med. 2018;44(12):2245–2247. doi: 10.1007/s00134-018-5413-4. [DOI] [PubMed] [Google Scholar]
  • 2.Reddy M., Kulkarni M., Kanakalakshmi S.T., Shenoy L., KrishnaBhat R.R. Non-invasive SpO2/FiO2 ratio (SFR) as surrogate for PaO2/FiO2 ratio (PFR): a scoping review. J Crit Care Med. 2025;11(3):221–232. doi: 10.2478/jccm-2025-0024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Rangappa R. A game changer for ARDS? Unraveling the potential of the SF ratio. Indian J Crit Care Med. 2024;28(3):191–192. doi: 10.5005/jp-journals-10071-24678. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Battaglini D., Schultz M.J., Puentes G.A.C., Marini J.J., Rocco P.R.M. Imaging and pulmonary function techniques in ARDS diagnosis and management: current insights and challenges. Crit Care. 2025;29(1) doi: 10.1186/s13054-025-05520-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Ashbaugh D.G., Bigelow D.B., Petty T.L., Levine B.E. Acute respiratory distress in adults. Lancet. 1967;290(7511):319–323. doi: 10.1016/s0140-6736(67)90168-7. [DOI] [PubMed] [Google Scholar]
  • 6.Murray J.F., Matthay M.A., Luce J.M., Flick M.R. An expanded definition of the adult respiratory distress syndrome. Am Rev Respir Dis. 1988;138(3):720–723. doi: 10.1164/ajrccm/138.3.720. [DOI] [PubMed] [Google Scholar]
  • 7.Bernard G.R., Artigas A., Brigham K.L., Carlet J., Falke K., Hudson L., et al. The American-European Consensus Conference on ARDS. Definitions, mechanisms, relevant outcomes, and clinical trial coordination. Am J Respir Crit Care Med. 1994;149(3 Pt 1):818–824. doi: 10.1164/ajrccm.149.3.7509706. [DOI] [PubMed] [Google Scholar]
  • 8.Ferguson N.D., Davis A.M., Slutsky A.S., Stewart T.E. Development of a clinical definition for acute respiratory distress syndrome using the Delphi technique. J Crit Care. 2005;20(2):147–154. doi: 10.1016/j.jcrc.2005.03.001. [DOI] [PubMed] [Google Scholar]
  • 9.Villar J., Pérez-Méndez L., López J., Belda J., Blanco J., Saralegui I., et al. An early PEEP/FIO2 trial identifies different degrees of lung injury in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med. 2007;176(8):795–804. doi: 10.1164/rccm.200610-1534OC. [DOI] [PubMed] [Google Scholar]
  • 10.Villar J., Pérez-Méndez L., Blanco J., Añón J.M., Blanch L., Belda J., et al. A universal definition of ARDS: the PaO2/FiO2 ratio under a standard ventilatory setting—a prospective, multicenter validation study. Intensive Care Med. 2013;39(4):583–592. doi: 10.1007/s00134-012-2803-x. [DOI] [PubMed] [Google Scholar]
  • 11.Ranieri V.M., Rubenfeld G.D., Thompson B.T., Ferguson N.D., Caldwell E., Fan E., et al. Acute respiratory distress syndrome: the Berlin definition. JAMA. 2012;307(23):2526–2533. doi: 10.1001/jama.2012.5669. [DOI] [PubMed] [Google Scholar]
  • 12.Matthay M.A., Arabi Y., Arroliga A.C., Bernard G., Bersten A.D., Brochard L.J., et al. A new global definition of acute respiratory distress syndrome. Am J Respir Crit Care Med. 2024;209(1):37–47. doi: 10.1164/rccm.202303-0558WS. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Rice T.W., Wheeler A.P., Bernard G.R., Hayden D.L., Schoenfeld D.A., Ware L.B., et al. Comparison of the SpO2/FIO2 ratio and the PaO2/FIO2 ratio in patients with acute lung injury or ARDS. Chest. 2007;132(2):410–417. doi: 10.1378/chest.07-0617. [DOI] [PubMed] [Google Scholar]
  • 14.Riviello E.D., Kiviri W., Twagirumugabe T., Mueller A., Banner-Goodspeed V.M., Officer L., et al. Hospital incidence and outcomes of the acute respiratory distress syndrome using the modification of the Berlin definition. Am J Respir Crit Care Med. 2016;193(1):52–59. doi: 10.1164/rccm.201503-0584OC. [DOI] [PubMed] [Google Scholar]
  • 15.Papazian L., Forel J.M., Gacouin A., Penot-Ragon C., Perrin G., Loundou A., et al. Neuromuscular blockers in early acute respiratory distress syndrome. N Engl J Med. 2010;363(12):1107–1116. doi: 10.1056/NEJMoa1005372. [DOI] [PubMed] [Google Scholar]
  • 16.Guérin C., Reignier J., Richard J.C., Beuret P., Gacouin A., Boulain T., et al. Prone positioning in severe acute respiratory distress syndrome. N Engl J Med. 2013;368(23):2159–2168. doi: 10.1056/NEJMoa1214103. [DOI] [PubMed] [Google Scholar]
  • 17.Grasselli G., Calfee C.S., Camporota L., Poole D., Amato M.B.P., Antonelli M., et al. ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies. Intensive Care Med. 2023;49(7):727–759. doi: 10.1007/s00134-023-07050-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Maiolo G., Collino F., Vasques F., Rapetti F., Tonetti T., Romitti F., et al. Reclassifying acute respiratory distress syndrome. Am J Respir Crit Care Med. 2018;197(12):1586–1595. doi: 10.1164/rccm.201709-1804OC. [DOI] [PubMed] [Google Scholar]
  • 19.Sehgal I.S., Agarwal R., Dhooria S., Prasad K.T., Muthu V., Aggarwal A.N. Risk stratification of acute respiratory distress syndrome using a PaO2: FiO2 threshold of 150 mmHg: a retrospective analysis from an Indian intensive care unit. Lung India. 2020;37(6):473–478. doi: 10.4103/lungindia.lungindia_146_20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Wick K.D., Matthay M.A., Ware L.B. Pulse oximetry for the diagnosis and management of acute respiratory distress syndrome. Lancet Respir Med. 2022;10(11):1086–1098. doi: 10.1016/S2213-2600(22)00058-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Acute Respiratory Distress Syndrome Network. RG Brower, Matthay M.A., Morris A., Schoenfeld D., Thompson B.T., et al. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med. 2000;342(18):1301–1308. doi: 10.1056/NEJM200005043421801. [DOI] [PubMed] [Google Scholar]
  • 22.Brower R.G., Lanken P.N., MacIntyre N., Matthay M.A., Morris A., Ancukiewicz M., et al. Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome. N Engl J Med. 2004;351(4):327–336. doi: 10.1056/NEJMoa032193. [DOI] [PubMed] [Google Scholar]
  • 23.Chen W., Janz D.R., Shaver C.M., Bernard G.R., Bastarache J.A., Ware L.B. Clinical characteristics and outcomes are similar in ARDS diagnosed by oxygen saturation/FiO2 ratio compared with Pao2/FiO2 ratio. Chest. 2015;148(6):1477–1483. doi: 10.1378/chest.15-0169. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Roozeman J.P., Mazzinari G., Serpa Neto A., Hollmann M.W., Paulus F., Schultz M.J., et al. Prognostication using SpO2/FiO2 in invasively ventilated ICU patients with ARDS due to COVID-19 – insights from the PRoVENT-COVID study. J Crit Care. 2022;68:31–37. doi: 10.1016/j.jcrc.2021.11.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Pandharipande P.P., Shintani A.K., Hagerman H.E., St Jacques P.J., Rice T.W., Sanders N.W., et al. Derivation and validation of Spo2/Fio2 ratio to impute for Pao2/Fio2 ratio in the respiratory component of the Sequential Organ Failure Assessment score. Crit Care Med. 2009;37(4):1317–1321. doi: 10.1097/CCM.0b013e31819cefa9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Tripathi R.S., Blum J.M., Rosenberg A.L., Tremper K.K. Pulse oximetry saturation to fraction inspired oxygen ratio as a measure of hypoxia under general anesthesia and the influence of positive end-expiratory pressure. J Crit Care. 2010;25(3) doi: 10.1016/j.jcrc.2010.04.009. 542.e9–542e13. [DOI] [PubMed] [Google Scholar]
  • 27.Namendys-Silva S.A., Silva-Medina M.A., Vásquez-Barahona G.M., Baltazar-Torres J.A., Rivero-Sigarroa E., Fonseca-Lazcano J.A., et al. Application of a modified sequential organ failure assessment score to critically ill patients. Braz J Med Biol Res. 2013;46(2):186–193. doi: 10.1590/1414-431x20122308. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Serpa Neto A., Cardoso S.O., Ong D.S., Espósito D.C., Pereira V.G., Manetta J.A., et al. The use of the pulse oximetric saturation/fraction of inspired oxygen ratio for risk stratification of patients with severe sepsis and septic shock. J Crit Care. 2013;28(5):681–686. doi: 10.1016/j.jcrc.2013.04.005. [DOI] [PubMed] [Google Scholar]
  • 29.Schmidt M.F., Gernand J., Kakarala R. The use of the pulse oximetric saturation to fraction of inspired oxygen ratio in an automated acute respiratory distress syndrome screening tool. J Crit Care. 2015;30(3):486–490. doi: 10.1016/j.jcrc.2015.02.007. [DOI] [PubMed] [Google Scholar]
  • 30.Bass C.M., Sajed D.R., Adedipe A.A., West T.E. Pulmonary ultrasound and pulse oximetry versus chest radiography and arterial blood gas analysis for the diagnosis of acute respiratory distress syndrome: a pilot study. Crit Care. 2015;19(1) doi: 10.1186/s13054-015-0995-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Bashar F.R., Vahedian-Azimi A., Farzanegan B., Goharani R., Shojaei S., Hatamian S., et al. Comparison of non-invasive to invasive oxygenation ratios for diagnosing acute respiratory distress syndrome following coronary artery bypass graft surgery: a prospective derivation-validation cohort study. J Cardiothorac Surg. 2018;13(1) doi: 10.1186/s13019-018-0804-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Babu S., Abhilash K.P., Kandasamy S., Gowri M. Association between SpO2/FiO2 rat–io and PaO2/FiO2 ratio in different modes of oxygen supplementation. Indian J Crit Care Med. 2021;25(9):1001–1005. doi: 10.5005/jp-journals-10071-23977. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Kumar A., Aggarwal R., Khanna P., Kumar R., Singh A.K., Soni K.D., et al. Correlation of the SpO2/FiO2 (S/F) ratio and the PaO2/FiO2 (P/F) ratio in patients with COVID-19 pneumonia. Med Intensiva (Engl Ed) 2022;46(7):408–410. doi: 10.1016/j.medine.2021.10.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Gandhi V., Trivedi L., Mulani K., Patel S. Comparison of SpO2/FiO2 (Sf) ratio and PaO2/FiO2 (Pf) ratio in mechanically ventilated patients with ARDS in Covid ICU for prognosis of ARDS: a prospective, observational study. Res J Med Sci. 2023;17:12–16. doi: 10.59218/makrjms.2023.11.12.16. [DOI] [Google Scholar]
  • 35.Choudhury A., Singh M., Jaswal A., Mustafi S.M., Kohli S. Correlation between invasive and noninvasive oxygen indices in critically ill COVID patients with ARDS: a retrospective analysis. J Assoc Chest Physicians. 2024;12(4):160–166. doi: 10.4103/jacp.jacp_23_24. [DOI] [Google Scholar]
  • 36.Alur T.R., Iyer S.S., Shah J.N., Kulkarni S., Jedge P., Patil V. A prospective observational study comparing oxygen saturation/fraction of inspired oxygen ratio with partial pressure of oxygen in arterial blood/fraction of inspired oxygen ratio among critically ill patients requiring different modes of oxygen supplementation. Indian J Crit Care Med. 2024;28(3):251–255. doi: 10.5005/jp-journals-10071-24652. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.SRLF Trial Group Hypoxemia in the ICU: prevalence, treatment, and outcome. Ann Intensive Care. 2018;8(1):82. doi: 10.1186/s13613-018-0424-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Hapsari B.A.D., Roxanne O.G., Reviono R., Adhiputri A., Apriningsih H., Rahayu N.I., et al. P/F ratio is a better predictor for non-invasive ventilation failure and length of stay in patients with community-acquired pneumonia. Jurnal Respirasi. 2024;10(1):42–49. doi: 10.20473/jr.v10-I.1.2024.42-49. [DOI] [Google Scholar]
  • 39.Charles P.G., Wolfe R., Whitby M., Fine M.J., Fuller A.J., Stirling R., et al. SMART-COP: a tool for predicting the need for intensive respiratory or vasopressor support in community-acquired pneumonia. Clin Infect Dis. 2008;47(3):375–384. doi: 10.1086/589754. [DOI] [PubMed] [Google Scholar]
  • 40.Coudroy R., Frat J.P., Boissier F., Contou D., Robert R., Thille A.W. Early identification of acute respiratory distress syndrome in the absence of positive pressure ventilation: implications for revision of the berlin criteria for acute respiratory distress syndrome. Crit Care Med. 2018;46(4):540–546. doi: 10.1097/CCM.0000000000002929. [DOI] [PubMed] [Google Scholar]
  • 41.Sanz F., Dean N., Dickerson J., Jones B., Knox D., Fernández-Fabrellas E., et al. Accuracy of PaO2/FiO2 calculated from SpO2 for severity assessment in ED patients with pneumonia. Respirology. 2015;20(5):813–818. doi: 10.1111/resp.12560. [DOI] [PubMed] [Google Scholar]
  • 42.Bonaventura A., Mumoli N., Mazzone A., Colombo A., Evangelista I., Cerutti S., et al. Correlation of SpO2/FiO2 and PaO2/FiO2 in patients with symptomatic COVID-19: an observational, retrospective study. Intern Emerg Med. 2022;17(6):1769–1775. doi: 10.1007/s11739-022-02981-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Shankar-Hari M., Phillips G.S., Levy M.L., Seymour C.W., Liu V.X., Deutschman C.S., et al. Developing a new definition and assessing new clinical criteria for septic shock: for the third international consensus definitions for sepsis and septic shock (Sepsis-3) JAMA. 2016;315(8):775–787. doi: 10.1001/jama.2016.0289. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Ranzani O.T., Singer M., Salluh J.I.F., Shankar-Hari M., Pilcher D., Berger-Estilita J., et al. Development and validation of the Sequential Organ Failure Assessment (SOFA)-2 score. JAMA. 2025 doi: 10.1001/jama.2025.20516. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Pölkki A., Reinikainen M., Rochwerg B., Jung C., Sendagire C., Chaudhuri D., et al. Optimal cutoffs for the ratio of arterial oxygen partial pressure to inspired oxygen fraction in categorizing respiratory impairment severity in organ failure scores. Acta Anaesthesiol Scand. 2026;70(1) doi: 10.1111/aas.70137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Li T., Hu W.Q., Li X., Zhang J.P., Tan L.Z., Yu L.X., et al. Prognostic value of PaO2/FiO2, SOFA and D-dimer in elderly patients with sepsis. J Int Med Res. 2022;50(6) doi: 10.1177/03000605221100755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Bi H., Liu X., Chen C., Chen L., Liu X., Zhong J., et al. The PaO2/FiO2 is independently associated with 28-day mortality in patients with sepsis: a retrospective analysis from MIMIC-IV database. BMC Pulm Med. 2023;23(1) doi: 10.1186/s12890-023-02491-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Wilson B.J., Cowan H.J., Lord J.A., Zuege D.J., Zygun D.A. The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010;10:9. doi: 10.1186/1471-227X-10-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Carvalho E.B., Leite T.R.S., Sacramento R.F.M., Nascimento PRLD, Samary C.D.S., Rocco P.R.M., et al. Rationale and limitations of the SpO2/FiO2 as a possible substitute for PaO2/FiO2 in different preclinical and clinical scenarios. Rev Bras Ter Intensiva. 2022;34(1):185–196. doi: 10.5935/0103-507X.20220013-pt. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Li Y., Du X., Zhou L., Guan J., Sun Y., Sun X., et al. Use of the PaO2/FiO2 ratio in acute pulmonary embolism: a simple and reliable parameter to predict the risk stratification. J Thorac Dis. 2025;17(7):4681–4688. doi: 10.21037/jtd-24-1821. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Pires M., Santos J.M., Neto V., Correia J., Ferreira G., Costa Cabral J., et al. A new ratio with PaO2/FiO2 and pulmonary arterial systolic pressure in the prognosis of intermediate high risk pulmonary embolism. Eur Heart J. 2021;42(Supplement_1) doi: 10.1093/eurheartj/ehab724.1929. [DOI] [Google Scholar]
  • 52.Tortum F., Tekin E., Gur A., Kerget B., Kasali K. Use of the PaO2/FiO2 ratio in pulmonary embolism: evaluation of its correlation with pulmonary arterial computed tomography obstruction index. Acad Radiol. 2023;30(5):893–899. doi: 10.1016/j.acra.2022.06.004. [DOI] [PubMed] [Google Scholar]
  • 53.Esteve F., Lopez-Delgado J.C., Javierre C., Skaltsa K., Carrio M.L., Rodríguez-Castro D., et al. Evaluation of the PaO2/FiO2 ratio after cardiac surgery as a predictor of outcome during hospital stay. BMC Anesthesiol. 2014;14 doi: 10.1186/1471-2253-14-83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Saraçoğlu K.T., Arslan G., Saraçoğlu A., Sezen Ö, Ratajczyk P., Gaszynski T. Oxygen reserve index vs. peripheral oxygen saturation for the prediction of hypoxemia in morbidly obese patients: a prospective observational study. BMC Anesthesiol. 2024;24(1):367. doi: 10.1186/s12871-024-02755-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Coudroy R., Frat J.P., Girault C., Thille A.W. Reliability of methods to estimate the fraction of inspired oxygen in patients with acute respiratory failure breathing through non-rebreather reservoir bag oxygen mask. Thorax. 2020;75(9):805–807. doi: 10.1136/thoraxjnl-2020-214863. [DOI] [PubMed] [Google Scholar]
  • 56.Hultström M., Hellkvist O., Covaciu L., Fredén F., Frithiof R., Lipcsey M., et al. Limitations of the ARDS criteria during high-flow oxygen or non-invasive ventilation: evidence from critically ill COVID-19 patients. Crit Care. 2022;26(1):55. doi: 10.1186/s13054-022-03933-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Li J., Albuainain F.A., Tan W., Scott J.B., Roca O., Mauri T. The effects of flow settings during high-flow nasal cannula support for adult subjects: a systematic review. Crit Care. 2023;27(1):78. doi: 10.1186/s13054-023-04361-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Vieira F., Bezerra F.S., Coudroy R., Schreiber A., Telias I., Dubo S., et al. High flow nasal cannula compared to continuous positive airway pressure: a bench and physiological study. J Appl Physiol. 2022;132(6):1580. doi: 10.1152/japplphysiol.00416.2021. 2022. [DOI] [PubMed] [Google Scholar]
  • 59.van der Ven FLIM, Valk C.M.A., Blok S., Brouwer M.G., Go D.M., Lokhorst A., et al. Broadening the Berlin definition of ARDS to patients receiving high-flow nasal oxygen: an observational study in patients with acute hypoxemic respiratory failure due to COVID-19. Ann Intensive Care. 2023;13(1):64. doi: 10.1186/s13613-023-01161-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Goutorbe P., Daranda E., Asencio Y., Esnault P., Prunet B., Bordes J., et al. Leaks can dramatically decrease FiO2 on home ventilators: a bench study. BMC Res Notes. 2013;6 doi: 10.1186/1756-0500-6-282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Bellani G., Laffey J.G., Pham T., Madotto F., Fan E., Brochard L., et al. Noninvasive ventilation of patients with acute respiratory distress syndrome. Insights from the LUNG SAFE study. Am J Respir Crit Care Med. 2017;195(1):67–77. doi: 10.1164/rccm.201606-1306OC. [DOI] [PubMed] [Google Scholar]
  • 62.Spada C., Gandhi R., Patel S.R., Nuccio P., Weinhouse G.L., Lee P.S. Oxygen saturation/fraction of inspired oxygen ratio is a simple predictor of noninvasive positive pressure ventilation failure in critically ill patients. J Crit Care. 2011;26(5):510–516. doi: 10.1016/j.jcrc.2010.08.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Matthay M.A., Thompson B.T., Ware L.B. The Berlin definition of acute respiratory distress syndrome: should patients receiving high-flow nasal oxygen be included? Lancet Respir Med. 2021;9(8):933–936. doi: 10.1016/S2213-2600(21)00105-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Ruan S.Y., Huang C.T., Chang H.T., Liu W.L., Wang W.J., Tseng Y.T., et al. Construct validity of PaO2/FiO2 ratios in defining acute respiratory distress syndrome. Am J Respir Crit Care Med. 2022;205(3):364–366. doi: 10.1164/rccm.202108-1924LE. [DOI] [PubMed] [Google Scholar]
  • 65.Yoshimura S., Hashimoto K., Shono Y., Tamura T., Uchimido R., Ando K., et al. Predictive value of the PaO2/FIO2 ratio for mortality in patients with acute respiratory distress syndrome: a systematic review and meta-analysis. Intern Med. 2025;64(13):1955–1964. doi: 10.2169/internalmedicine.4292-24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Karbing D.S., Kjaergaard S., Smith B.W., Espersen K., Allerød C., Andreassen S., et al. Variation in the PaO2/FiO2 ratio with FiO2: mathematical and experimental description, and clinical relevance. Crit Care. 2007;11(6):R118. doi: 10.1186/cc6174. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Khemani R.G., Rubin S., Belani S., Leung D., Erickson S., Smith L.S., et al. Pulse oximetry vs. PaO2 metrics in mechanically ventilated children: Berlin definition of ARDS and mortality risk. Intensive Care Med. 2015;41(1):94–102. doi: 10.1007/s00134-014-3486-2. [DOI] [PubMed] [Google Scholar]
  • 68.Brown S.M., Duggal A., Hou P.C., Tidswell M., Khan A., Exline M., et al. Nonlinear imputation of PaO2/FIO2 from SpO2/FIO2 among mechanically ventilated patients in the ICU: a prospective, observational study. Crit Care Med. 2017;45(8):1317–1324. doi: 10.1097/CCM.0000000000002514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Brown S.M., Grissom C.K., Moss M., Rice T.W., Schoenfeld D., Hou P.C., et al. Nonlinear imputation of Pao2/Fio2 from Spo2/Fio2 among patients with acute respiratory distress syndrome. Chest. 2016;150(2):307–313. doi: 10.1016/j.chest.2016.01.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Ding X.Y., Chen H. Beyond concordance: differential prognostic value of SpO2/FiO2 and PaO2/FiO2 ratios in ARDS severity classification. Crit Care. 2025;29(1) doi: 10.1186/s13054-025-05754-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Severinghaus J.W. Simple, accurate equations for human blood O2 dissociation computations. J Appl Physiol. 1979;46(3):599–602. doi: 10.1152/jappl.1979.46.3.599. [DOI] [PubMed] [Google Scholar]
  • 72.Erlebach R., Pale U., Beck T., Markovic S., Seric M., David S., et al. Limitations of SpO2/FiO2-ratio for classification and monitoring of acute respiratory distress syndrome-an observational cohort study. Crit Care. 2025;29(1):82. doi: 10.1186/s13054-025-05317-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Nasa P., Bos L.D., Estenssoro E., van Haren F.M.P., Neto A.S., Rocco P.R.M., et al. Defining and subphenotyping ARDS: insights from an international Delphi expert panel. Lancet Respir Med. 2025;13(7):638–650. doi: 10.1016/S2213-2600(25)00115-8. [DOI] [PubMed] [Google Scholar]
  • 74.Fawzy A., Wu T.D., Wang K., Robinson M.L., Farha J., Bradke A., et al. Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19. JAMA Intern Med. 2022;182(7):730–738. doi: 10.1001/jamainternmed.2022.1906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Reep C.A.T., Fleuren L.M., Heunks L., Wils E.J. Racial disparities in pulse oximetry, in COVID-19 and ICU settings. Crit Care Explor. 2024;6(8) doi: 10.1097/CCE.0000000000001132. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Fawzy A., Wu T.D., Wang K., Sands K.E., Fisher A.M., Arnold Egloff S.A., et al. Clinical outcomes associated with overestimation of oxygen saturation by pulse oximetry in patients hospitalized with COVID-19. JAMA Netw Open. 2023;6(8) doi: 10.1001/jamanetworkopen.2023.30856. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Palanidurai S., Phua J., Chan Y.H., Mukhopadhyay A. P/FP ratio: incorporation of PEEP into the PaO2/FiO2 ratio for prognostication and classification of acute respiratory distress syndrome. Ann Intensive Care. 2021;11(1):124. doi: 10.1186/s13613-021-00908-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.El-Khatib M.F., Bouakl I.J., Ayoub C.M., Chatburn R.L., Farhat H., Msheik M., et al. Comparison of the oxygenation factor and the oxygenation ratio in subjects with ARDS. Respir Care. 2020;65(12):1874–1882. doi: 10.4187/respcare.07669. [DOI] [PubMed] [Google Scholar]
  • 79.DesPrez K., McNeil J.B., Wang C., Bastarache J.A., Shaver C.M., Ware L.B. Oxygenation saturation index predicts clinical outcomes in ARDS. Chest. 2017;152(6):1151–1158. doi: 10.1016/j.chest.2017.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Vadi S.M., Sanwalka N., Suthar D. Oxygenation index, oxygen saturation index vs PaO2/FiO2 *PEEP: a secondary analysis of OXIVA-CARDS study. Indian J Crit Care Med. 2024;28(10):917–922. doi: 10.5005/jp-journals-10071-24808. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81.Bai Y., Chen S., Yang H., Huang X., Xia J., Zhan Q. Dynamic oxygenation subgroup bringing new insights in ARDS: more predictive of outcomes and response to PEEP than static PaO2/FiO2. Thorax. 2025;80(9):594–603. doi: 10.1136/thorax-2024-222360. [DOI] [PubMed] [Google Scholar]
  • 82.Valda Toro P.L., Willmore A., Wu N.E., Delucchi K.L., Jauregui A., Sinha P., et al. Rapidly improving ARDS differs clinically and biologically from persistent ARDS. Crit Care. 2024;28(1):132. doi: 10.1186/s13054-024-04883-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Schultz M.J., Battaglini D., Nasa P. Classifying ARDS by dynamic or static oxygenation impairment or are other approaches needed for greater value. Thorax. 2025;80(9):592–593. doi: 10.1136/thorax-2025-223490. [DOI] [PubMed] [Google Scholar]
  • 84.Gattinoni L., Chiumello D., Carlesso E., Valenza F. Bench-to-bedside review: chest wall elastance in acute lung injury/acute respiratory distress syndrome patients. Crit Care. 2004;8(5):350–355. doi: 10.1186/cc2854. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Panwar R., Madotto F., Laffey J.G., van Haren F.M.P. Compliance phenotypes in early acute respiratory distress syndrome before the COVID-19 pandemic. Am J Respir Crit Care Med. 2020;202(9):1244–1252. doi: 10.1164/rccm.202005-2046OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Zhao H., Yao T., Liu H. Combined lung compliance and oxygenation dynamics predict high-flow nasal cannula failure in acute respiratory distress syndrome: a retrospective cohort study. Am J Transl Res. 2025;17(6):4175–4186. doi: 10.62347/ELGT2750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87.Kallet R.H., Alonso J.A., Pittet J.F., Matthay M.A. Prognostic value of the pulmonary dead-space fraction during the first 6 days of acute respiratory distress syndrome. Respir Care. 2004;49(9):1008–1014. [PubMed] [Google Scholar]
  • 88.Charron C., Repesse X., Bouferrache K., Bodson L., Castro S., Page B., et al. PaCO2 and alveolar dead space are more relevant than PaO2/FiO2 ratio in monitoring the respiratory response to prone position in ARDS patients: a physiological study. Crit Care. 2011;15(4):R175. doi: 10.1186/cc10324. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 89.Jiang L., Chen H., Chang W., Sun Q., Yuan X., Wu Z., et al. Time-varying intensity of ventilatory inefficiency and mortality in patients with acute respiratory distress syndrome. Ann Intensive Care. 2025;15(1):6. doi: 10.1186/s13613-025-01427-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Siegel E.R., Zhuo H., Sinha P., Papolos A.I., Ni S.A., Vessel K., et al. Ventilatory ratio is a valuable prognostic indicator in an observational cohort of patients with ARDS. Respir Care. 2022;67(9):1075–1081. doi: 10.4187/respcare.09854. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 91.Jiang L., Chen H., Xie J., Liu L., Yang Y. Prognostic value of time-varying dead space estimates in mechanically ventilated patients with acute respiratory distress syndrome. J Intensive Med. 2024;4(2):187–193. doi: 10.1016/j.jointm.2023.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92.Morales-Quinteros L., Schultz M.J., Bringué J., Calfee C.S., Camprubí M., Cremer O.L., et al. Estimated dead space fraction and the ventilatory ratio are associated with mortality in early ARDS. Ann Intensive Care. 2019;9(1):128. doi: 10.1186/s13613-019-0601-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 93.Blanch L., López-Aguilar J., Lucangelo U. Dead space in acute respiratory distress syndrome: more than a feeling! Crit Care. 2016;20(1) doi: 10.1186/s13054-016-1381-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94.Paudel R., Trinkle C.A., Waters C.M., Robinson L.E., Cassity E., Sturgill J.L., et al. Mechanical power: a new concept in mechanical ventilation. Am J Med Sci. 2021;362(6):537–545. doi: 10.1016/j.amjms.2021.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95.Coppola S., Caccioppola A., Froio S., Formenti P., De Giorgis V., Galanti V., et al. Effect of mechanical power on intensive care mortality in ARDS patients. Crit Care. 2020;24(1) doi: 10.1186/s13054-020-02963-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96.Al-Husinat L., Khamis R., Azzam S., Trdeh M., Jouryyeh B., Al Sharie S., et al. Beyond the PaO2/FiO2 ratio: rethinking ARDS severity through the Lens of physiology. Ann Intens Care. 2026;16:100057. doi: 10.1016/j.aicoj.2026.100057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97.Sayed M., Riaño D., Villar J. Novel criteria to classify ARDS severity using a machine learning approach. Crit Care. 2021;25(1) doi: 10.1186/s13054-021-03566-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Lv J., Chen J., Liu M., Dai X., Deng W. Machine learning-based prognostic prediction model of pneumonia-associated acute respiratory distress syndrome. Front Med (Lausanne) 2025;12 doi: 10.3389/fmed.2025.1582426. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99.Nasa P., Juneja D., Schultz M.J. Defining the acute respiratory distress syndrome - are you a splitter or a lumper? The choice matters more than you think. Indian J Crit Care Med. 2025;29(7):551–553. doi: 10.5005/jp-journals-10071-25016. [DOI] [PMC free article] [PubMed] [Google Scholar]