Surgical treatment of atlantoaxial dislocation using a novel posterior atlantoaxial screw–rod fixation system: a retrospective controlled study

Xiangyang Ma 1,2,*,✉, Xiaobao Zou 1,*, Junlin Chen 1,2,*, Xinzhao Huang 1,2,*, Haozhi Yang 1, Rencai Ma 1,3, Zexing Chen 1,2
PMCID: PMC13747653  PMID: 41844060

Abstract

Study Design

We developed a novel posterior atlantoaxial screw–rod (PASR) fixation system to manage atlantoaxial dislocation (AAD) and compared it with a conventional PASR fixation system. The novel system was designed to simplify the fixation procedure and achieve direct reduction of irreducible AAD without the need for surgical release.

Purpose

To perform a comparative assessment of clinical and radiological outcomes between our novel PASR fixation system and a conventional PASR fixation system.

Overview of Literature

The current literature provides limited evidence on the efficacy of this novel PASR fixation system for managing irreducible AAD.

Methods

A novel PASR fixation system was designed to address the limitations of a conventional PASR fixation system, and it included C1 reduction screws, C2 fulcrum screws, and antirotation rods. From June 2016 to June 2021, 39 patients with AAD underwent a single posterior approach. Of these 39 patients, 21 were treated with the conventional PASR fixation system and 18 were treated with the novel PASR fixation system. The imaging and clinical data before and after the surgery were retrospectively analyzed.

Results

All 39 patients were successfully treated, and there were no vascular or neurological complications. Both groups demonstrated significant improvements in the atlantodental interval (ADI), Japanese Orthopaedic Association (JOA) score, visual analog scale pain score, and Neck Disability Index at the final follow-up. The operative time was shorter, postoperative ADI was better, and JOA score at follow-up was higher in the novel PASR group than in the conventional PASR group. No cases of implant loosening or screw fracture were observed. All patients achieved complete anatomical reduction postoperatively, and there was radiographic evidence of solid bony fusion within 12 months.

Conclusions

A posterior approach with our novel PASR fixation system is safe and effective for the treatment of AAD. Compared with a conventional PASR fixation system, our novel PASR fixation system can simplify the operation and provide a better reduction effect. This novel fixation system is expected to facilitate the procedure of atlantoaxial fixation and achieve reduction of irreducible AAD without release.

Keywords: Atlantoaxial dislocation, Posterior approach surgery, Internal fixation, Posterior atlantoaxial screw-rod fixation system

Graphical Abstract

graphic file with name asj-2025-0341f6.webp

Introduction

Posterior screw–rod fixation has become the preferred method of atlantoaxial fixation with the advent of C1 lateral mass screws [1,2]. In addition, the emergence of multiple techniques of posterior screw fixation for C1 and C2 has further expanded the versatility of posterior atlantoaxial fixation, allowing greater flexibility in screw configuration and expanding its clinical applicability [3,4]. Reduction and stabilization represent the primary surgical objectives for atlantoaxial dislocation (AAD) [5]. In posterior screw–rod fixation, bilateral screw placement in C1 and C2 offers procedural efficiency. The posterior atlantoaxial screw–rod (PASR) fixation system provides not only rigid stabilization but also intrinsic reduction capacity through mechanical distraction and compression.

At present, more attention is being paid to the research of screw fixation techniques for the improvement of the PASR fixation system [6–9], while less attention is being paid to the improvement of instruments. Given the maturation of atlantoaxial screw fixation techniques, current scientific efforts should be redirected to the development of instrumentation that facilitates a more simplified and effective surgical reduction of AAD. Posterior C1–C2 screw–rod fixation is currently the standard approach for atlantoaxial stabilization. Nevertheless, significant limitations persist when applied to AAD reduction surgery. First, an inadequate reduction force frequently leads to intraoperative reduction failure. Second, prebending deformation causes rod rotation within screw heads, compromising directional stability and weakening the reduction effect. Third, in the case of severe dislocation, the gap between C1 and C2 is quite large, which makes it difficult to insert the rod into the screw groove. To address these shortcomings, we designed a novel PASR fixation system. This study compared surgical results between our modified PASR fixation system and a conventional PASR fixation system and evaluated the preliminary clinical efficacy of this novel system. To our knowledge, this is the first study to report the instrument improvement of the PASR fixation system for AAD treatment.

Materials and Methods

Ethics statement

This study was approved by the Institutional Review Board (IRB) of the General Hospital of Southern Theater Command ((IRB approval no., NZLLKZ2025256).

Study design and patient selection

From June 2016 to June 2021, a total of 39 consecutive patients with AAD were recruited at the General Hospital of Southern Theater Command. Irreducible AAD is defined as a dislocation between the atlas (C1) and axis (C2) that persists despite flexion–extension dynamic radiographs and a trial of in-line skull traction.

All participants underwent PASR fixation. They were randomized into two treatment groups using stratified randomization that accounted for age (≥60 years vs.<60 years), sex, and dislocation type (reducible vs. irreducible). Within each stratum, randomization sequences were generated in a 1:1 allocation ratio through IBM SPSS ver. 19.0 (IBM Corp., Armonk, NY, USA) by an independent researcher not involved in surgical procedures or data collection. Of the 39 patients, 18 were treated with the novel PASR fixation system (novel PASR group) and 21 were treated with the conventional PASR fixation system (conventional PASR group). All surgeries were performed by the same surgeon (X.Y.M.). Demographic characteristics (age and sex), type of dislocation (reducible or irreducible), disease duration, and follow-up time were compared between the groups.

Introduction of the novel PASR fixation system

Our novel PASR fixation system included C1 reduction screws, C2 fulcrum screws, and antirotation rods (WEGO, Weihai, China). The C1 reduction screw, also known as a long-arm screw, was designed in reference to a thoracolumbar long-arm screw and had two types of polyaxial and uniaxial screws (Fig. 1A, B). The C2 fulcrum screw was designed to increase the bottom of the screw groove by 2 or 4 mm, and it had polyaxial and uniaxial screws (Fig. 1A, B). One end of the antirotation rod was designed as a “D-shaped” structure about 15 mm long, which was inserted into the C2 screw groove for automatic antirotation, and the other end was a cylindrical structure about 35 mm long, which could be cut according to the required length during the operation (Fig. 1A, B). Comparison of reduction performance between the conventional and novel PASR fixation systems is presented in Fig. 1C and D. The operation procedure and reduction principle of the new instruments are shown in Fig. 2A–G.

Fig. 1.

Fig. 1

(A, B) The construction of the novel posterior atlantoaxial screw–rod (PASR) systems including C1 reduction screw (a), C2 fulcrum screw (b) and anti-rotation rod (c) with “O” end in the head (C1) and “D” end in the caudal (C2). The “D” end can completely match between nut and groove for automatic anti-rotation function. The “O” structure in conventional rod cannot completely match between nut and groove without anti-rotation function. (C, D) Comparison of reduction performance between conventional and novel PASR systems. The abutment height of C2 fulcrum screw (H2) is higher than that of conventional C2 screw (H1). After tightening the nuts of C1 screw, the reduction distance of novel PASR system (H4) is greater than that of conventional PASR system (H3).

Fig. 2.

Fig. 2

The reduction steps of atlantoaxial dislocation (AAD) using novel posterior atlantoaxial screw–rod (PASR) system on computer-aided models. (A) C1 reduction screws and C2 fulcrum screws were inserted, then the anti-rotation rods were placed and the “D” ends were locked into C2 fulcrum screw grooves. (B) The nuts of C1 reduction screws were gradually tightened to complete the reduction. (C) The extended arms of C1 reduction screws were broken. The intraoperative reduction procedure of AAD using novel PASR system. (D) The unilateral C1 reduction screw and C2 fulcrum screw were implanted, and the “D” end of anti-rotation rod was inserted into C2 fulcrum screw groove and locked. (E) The fluoroscopy image before reduction. (F) The nut of C1 reduction screw was tightened to complete the reduction. (G) The fluoroscopy image after reduction.

Surgical procedure

After general anesthesia, the patient was placed in the prone position with moderate cervical flexion. Skull traction (4 kg) was applied to stabilize the head. A longitudinal incision of about 6–8 cm was made in the posterior middle of the neck from the occipital to the C2 spinous process. The occipital bone, the posterior arch of C1, and the C2 lateral mass were exposed by subperiosteal dissection, and the cervical semispinalis cervicis muscle attached to the C2 spinous process was retained.

A pedicle screw or a lateral mass screw was used for C1 fixation, with a pedicle screw being preferred. If the C1 posterior arch was too thin and less than 2.0 mm in height, a C1 lateral mass screw was inserted according to the technique described by Goel and Laheri [1] and Harms et al. [2]; otherwise, a C1 pedicle screw was used according to the technique described by Ma et al. [10]. The newly designed C1 reduction screws were used in place of all C1 screws in the novel PASR group, and conventional C1 screws were used in the conventional PASR group. In the novel and conventional PASR groups, reduction was typically achieved by sequentially tightening nuts along a prebending rod. The adequacy of reduction was then verified using intraoperative X-ray imaging.

A C2 pedicle screw was preferred for C2 fixation. If anatomical variation made it impossible to safely implant C2 pedicle screws, C2 pars screws or C2 translaminar screws were used instead [11,12]. The newly designed C2 fulcrum screws were used in place of all C2 screws in the novel PASR group, and C2 fulcrum screws with a bottom height of 2 or 4 mm were selected according to the degree of dislocation in preoperative imaging. For severe dislocation, 4-mm screws were selected to provide greater reduction force, and for mild-to-moderate dislocation, 2-mm screws were used. Conventional C2 screws were used in the conventional PASR group.

After two antirotation rods were trimmed, the “D-shaped” end was inserted into the C2 fulcrum screw groove, and the cylindrical part was placed into the C1 reduction screw groove (Fig. 2A–E). The nuts of the C2 fulcrum screws were tightened, and the direction of the rods was automatically fixed. Then, the nuts of the C1 reduction screws were gradually tightened to complete the reduction (Fig. 2B, F, G). After good reduction was confirmed by intraoperative fluoroscopy, the long arms of the C1 reduction screws were broken (Fig. 2C). In the conventional PASR group, two prebending conventional cylindrical rods were used. Subsequently, autografts taken from the iliac crest were grafted.

Data collection and postoperative outcomes

The surgical time and estimated blood loss (EBL) were recorded. The atlantodental interval (ADI) and atlantoaxial angle (AAA) were measured before and after the surgery to assess the reduction. Japanese Orthopaedic Association (JOA) scores were used to assess neurological function. The Visual Analog Scale (VAS) was used to assess the degree of occipital neck pain. The Neck Disability Index (NDI) was used to evaluate cervical spine function.

Patients were followed up at 3, 6, 9, and 12 months, and then once per year or as needed. Surgical complications were documented after the surgery and during the follow-up. Cervical X-ray imaging and computed tomography (CT) scans were performed at each follow-up for the assessment of device failure, stability, and bone fusion.

Statistical analysis

Data were analyzed statistically with the Mann-Whitney U test, using IBM SPSS 19.0 software (IBM Corp.). The significance threshold was set at p<0.05.

Results

Patient demographics

Of the 39 patients, 18 were enrolled in the novel PASR group and 21 were enrolled in the conventional PASR group. There were no significant differences between groups with respect to age (p=0.603), gender (p=0.433), type of dislocation (p=0.197), disease duration (p=0.132), and follow-up time (p=0.124). The clinical and demographical characteristics for all participants are presented in Table 1.

Table 1.

Clinical and demographical characteristics of study participants

Characteristic Novel PASR group Conventional PASR group p-value
No. of cases 18 21
Age (yr) 36.8±20.1 39.7±15.8 0.603
Gender 0.433
 Male 9 (50.0) 13 (61.9)
 Female 9 (50.0) 8 (38.1)
Type of dislocation 0.197
 Reducible dislocation 16 (88.9) 20 (95.2)
 Irreducible dislocation 2 (11.1) 1 (4.8)
Disease duration (mo) 7.5±5.5 10.6±5.2 0.132
Follow-up time (mo) 19.1±7.0 23.8±10.7 0.124

Values are presented as mean±standard deviation or number (%).

PASR, posterior atlantoaxial screw-rod.

Surgical outcomes

The surgery was performed successfully in all 39 patients. The surgical time was significantly lower in the novel PASR group than in the conventional PASR group (p<0.01), and there was no significant difference in EBL between the groups (p>0.05) (Table 2). No vertebral artery or spinal cord injury occurred during the operations in both groups.

Table 2.

Comparisons of clinical data before and after surgery between the two groups

Items Novel PASR group Conventional PASR group p-valuea)
Surgical time (min) 120.6±25.2 164.5±36.4 0.000
Estimated blood loss (mL) 113.9±61.4 103.1±51.6 0.631
Atlantodental interval (mm)
 Before operation 5.7±1.1 6.0±1.5 0.334
 Final follow-up 1.8±0.8 2.3±0.7 0.022
 p-valueb) 0.000 0.000
Atlantoaxial angle (°)
 Before operation 84.6±13.2 84.0±11.6 0.805
 Final follow-up 103.6±6.6 101.7±6.1 0.449
 p-valueb) 0.000 0.000
Japanese Orthopaedic Association score
 Before operation 12.6±1.1 11.9±1.8 0.566
 Final follow-up 16.6±0.6 16.0±1.0 0.024
 p-valueb) 0.000 0.000
Visual Analog Scale score
 Before operation 3.3±0.8 3.0±1.0 0.121
 Final follow-up 0.2±0.4 0.2±0.5 0.913
 p-valueb) 0.000 0.000
Neck Disability Index (%)
 Before operation 25.8±6.7 23.6±7.3 0.411
 Final follow-up 5.8±5.5 4.5±5.5 0.442
 p-valueb) 0.000 0.000
Bone fusion time (mo) 7.5±3.1 7.7±3.1 0.861

Values are presented as mean±standard deviation unless otherwise stated.

PASR, posterior atlantoaxial screw-rod.

a)

By independent t-test.

b)

By paired t-test.

Clinical outcomes

There were no differences in preoperative data between the two groups (p>0.05). In both groups, the ADI, AAA, JOA score, VAS score, and NDI were significantly better at the last follow-up assessment than at the preoperative assessment (p<0.01). There were no significant differences in the AAA, VAS score, and NDI at the final follow-up between the two groups (p>0.05). However, at the final follow-up, the ADI was significantly lower in the novel PASR group than in the conventional PASR group (p<0.05), while the JOA score was significantly higher in the novel PASR group than in the conventional PASR group (p<0.05). All patients achieved bone fusion, which was confirmed on CT scans, and there was no difference in fusion time between the two groups (p>0.05) (Table 2). No instrument loosening or breakage, or loss of reduction was noted on X-ray imaging and CT scans during the follow-up (Figs. 3–5).

Fig. 3.

Fig. 3

Case 1, a 22-year-old man with secondary reducible atlantoaxial dislocation (AAD) caused by old odontoid fracture, treated with the novel posterior atlantoaxial screw–rod. (A–C) The X-rays prior to surgery showed reducible AAD with odontoid fracture. (D) Sagittal computed tomography (CT) scan showed evidence of old odontoid fracture. (E) Sagittal magnetic resonance imaging T2 weighted imaging showed slight spinal compression. (F–H) The postoperative X-rays and sagittal CT scan showed ideal reduction and internal fixation. (I, J) Lateral X-ray and sagittal CT scan from the 6-month follow-up showed satisfying sequence and bone fusion.

Fig. 4.

Fig. 4

Case 2, a 16-year-old man with irreducible atlantoaxial dislocation (AAD) treated with the novel posterior atlantoaxial screw–rod. (A–C) The preoperative X-rays showed AAD. (D) Lateral X-ray after skull traction showed no reduction of AAD. (E, F) Sagittal computed tomography (CT) scan and sagittal magnetic resonance imaging (MRI) T2 weighted imaging showed severe dislocation with compression on the ventral medullary. (G–J) The postoperative X-rays, sagittal CT scan, and sagittal MRI T2 weighted imaging showed effective reduction, good internal fixation, and decompression. (K, L) Lateral X-ray and sagittal CT scan from the 12-month follow-up showed stabilization of C1–C2 with bone fusion.

Fig. 5.

Fig. 5

Case 3, a 45-year-old woman with irreducible atlantoaxial dislocation (AAD) treated with conventional posterior atlantoaxial screw–rod. (A–C) The X-rays prior to surgery showed irreducible AAD. (D, E) Sagittal computed tomography (CT) scan and magnetic resonance imaging (MRI) showed severe dislocation with compression on the ventral medullary. (F–H) The postoperative X-rays, sagittal CT scan, and MRI showed uncompleted reduction and still compression on the ventral medullary. (I, J) Lateral X-ray and sagittal CT scan from the 12-month follow-up showed satisfying sequence and bone fusion.

Discussion

This study evaluated the clinical outcomes of a novel PASR fixation system and compared it with the conventional PASR fixation system in the treatment of AAD. All patients treated with the novel system achieved stable fixation, sustained radiological correction, and better clinical benefit, indicating its safety and effectiveness in the treatment of AAD.

Our findings showed that the novel system significantly shortened the operative time, which was attributable to its more efficient reduction process and improved reduction capability. Although there was no difference in the AAA between the two study groups at the last follow-up, the ADI was lower in the novel PASR group than in the conventional PASR group, indicating that this novel PASR fixation system has a superior reduction capacity and helps to relieve cord compression. This may also explain why the JOA score at the final follow-up was higher in the novel PASR group than in the conventional PASR group. Furthermore, successful operative treatment of two cases with irreducible AAD without anterior or posterior soft tissue release by only using this novel PASR fixation system suggests that this new approach has a great reduction ability.

The advantages of the novel PASR fixation system are associated with several key design elements. The C1 reduction screw, with an extended arm, allows gradual sliding and reduction during nut tightening, avoiding the need for additional tools. The C2 multiaxial fulcrum screw and monoaxial fulcrum screw, designed with a raised groove (2 or 4 mm), increase the vertical difference between the C1 and C2 screws, providing greater leverage for reduction (Fig. 1B). It is proposed that the reduction ability of a monoaxial screw is greater than that of a multiaxial screw. Therefore, for severe dislocation, a monoaxial fulcrum screw would be preferred. In addition, the antirotation rod with a “D-shaped” end prevents rod rotation during tightening, eliminating the need for auxiliary instruments in a narrow surgical field and reducing the risk of spinal cord injury. The cylindrical design of the other end of the rod makes it easy to place the rod into the groove of the atlas screw, and it can be automatically attached and anastomosed. Collectively, these features enable simultaneous reduction and fixation, which can simplify the surgical procedure and improve reduction efficiency compared with conventional systems.

Compared with earlier fixation techniques, the novel PASR fixation system offers additional benefits. The Magerl screw technique, once the gold standard, provides strong biomechanical stability but requires preoperative reduction, has limited applicability in patients with a short neck or obesity, and carries the risks of vertebral artery injury and radiation exposure [13–15]. Conventional PASR fixation systems partially address these issues [16], but they have not been specifically designed for the unique anatomical and biomechanical challenges of the atlantoaxial region. Conventional posterior cervical screws are all short-arm screws [17]. When the anterior dislocation of C1 is severe, it is very inconvenient and dangerous to use the approach involving a lifting screw and press rod to assist the operation in a narrow space. The C2 screw is the reduction fulcrum in the PASR fixation system, and theoretically, a higher fulcrum is associated with a better reduction effect [18,19]. To enhance the reduction capability, Zhou et al. [8] used pars screws with the entry point placed as inferiorly as possible to increase the height difference between the C1 and C2 screws. Clinical applications have shown that the reduction effect of the pars screw is better; however, due to the shorter length of the channel, it may lack sufficient compressive strength when reduction resistance is excessive or in the presence of osteoporosis. The novel PASR fixation system builds on these foundations and introduces structural innovations tailored to the treatment of AAD, thereby enhancing safety, efficiency, and reduction success rates. The system has the ability to achieve automatic reduction through the tightening of nuts into sliding C1 screw heads (the C1 screws have long heads), the presence of C2 screws with a higher abutment level, and the presence of antirotation rods.

This study has some limitations. The study was retrospective and involved a relatively small sample size, which may limit the generalizability of the results. Furthermore, biomechanical testing and longer-term follow-up were not performed. Future studies should include larger prospective cohorts and biomechanical validation to further confirm the advantages of this system and clarify its indications.

In summary, the novel PASR fixation system provides a safe and effective alternative to conventional fixation for AAD. Its ability to achieve automatic reduction during fixation shortens the operative time, enhances reduction, and may improve neurological recovery, particularly in cases of irreducible AAD.

Conclusions

We developed a novel PASR fixation system to treat AAD. Compared with the conventional PASR fixation system, the novel fixation system could better simplify the atlantoaxial fixation procedure and achieve atlantoaxial dislocation reduction without additional surgical release.

Key Points.

  • A novel posterior atlantoaxial screw–rod (PASR) fixation system consisting of C1 reduction screws, C2 fulcrum screws and antirotation rods was designed for atlantoaxial dislocation (AAD).

  • Compared with conventional system, PASR system achieves direct reduction of irreducible AAD without surgical release and simplifies the fixation procedure.

  • The novel PASR fixation system is safe and effective for AAD treatment with no vascular or neurological complications, implant loosening or screw fracture observed, and it can improve reduction efficiency and neurological recovery, especially for irreducible AAD cases.

Footnotes

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This work was supported by fund of the Guangzhou Municipal Science and Technology Project (No. 201803010046) and the National Natural Science Foundation of China (No. 82272582).

Author Contributions

Conceptualization: XYM. Data curation: XBZ. Formal analysis: XBZ. Funding acquisition: XYM. Investigation: XBZ, JLC, XZH, HZY. Methodology: XBZ, JLC, XZH, HZY. Project administration: RCM, ZXC. Resources: RCM, ZXC. Software: RCM, ZXC. Supervision: XYM, XBZ. Validation: XYM, XBZ. Visualization: XYM, XBZ. Writing–original draft: XYM, XBZ, JLC, XZH, HZY. Writing–review and editing: XYM, XBZ. Final approval of the manuscript: all authors.

References

  • 1.Goel A, Laheri V. Plate and screw fixation for atlanto-axial subluxation. Acta Neurochir (Wien) 1994;129:47–53. doi: 10.1007/bf01400872. [DOI] [PubMed] [Google Scholar]
  • 2.Harms J, Melcher RP. Posterior C1–C2 fusion with polyaxial screw and rod fixation. Spine (Phila Pa 1976) 2001;26:2467–71. doi: 10.1097/00007632-200111150-00014. [DOI] [PubMed] [Google Scholar]
  • 3.Bhowmick DA, Benzel EC. Posterior atlantoaxial fixation with screw-rod constructs: safety, advantages, and shortcomings. World Neurosurg. 2014;81:288–9. doi: 10.1016/j.wneu.2012.10.024. [DOI] [PubMed] [Google Scholar]
  • 4.Huang DG, Hao DJ, He BR, et al. Posterior atlantoaxial fixation: a review of all techniques. Spine J. 2015;15:2271–81. doi: 10.1016/j.spinee.2015.07.008. [DOI] [PubMed] [Google Scholar]
  • 5.Yin QS, Wang JH. Current trends in management of atlantoaxial dislocation. Orthop Surg. 2015;7:189–99. doi: 10.1111/os.12196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Jin GX, Wang H, Li L, Cui SQ, Duan JZ. C1 posterior arch crossing screw fixation for atlantoaxial joint instability. Spine (Phila Pa 1976) 2013;38:E1397–404. doi: 10.1097/brs.0b013e3182a40869. [DOI] [PubMed] [Google Scholar]
  • 7.Takeshima Y, Shigematsu H, Konishi K, Nakagawa I, Motoyama Y, Nakase H. Posterior atlantoaxial fusion using a C2 transverse foramen-penetrating screw: a technical note. Surg Neurol Int. 2017;8:262. doi: 10.4103/sni.sni_57_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Zhou S, Yuan B, Liu W, Tang Y, Chen X, Jia L. Three-dimensional reduction method with a modified C2 isthmus screw in irreducible atlantoaxial dislocation: a technical note. BMC Surg. 2021;21:324. doi: 10.1186/s12893-021-01321-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Shao M, Dai Y, Zhu W, Yu J, Lyu F. Bicortical short C2 pars screw fixation for high-riding vertebral artery provided sufficient biomechanical stability: a finite element study. Spine (Phila Pa 1976) 2022;47:369–75. doi: 10.1097/brs.0000000000004141. [DOI] [PubMed] [Google Scholar]
  • 10.Ma XY, Yin QS, Wu ZH, Xia H, Liu JF, Zhong SZ. Anatomic considerations for the pedicle screw placement in the first cervical vertebra. Spine (Phila Pa 1976) 2005;30:1519–23. doi: 10.1097/01.brs.0000168546.17788.49. [DOI] [PubMed] [Google Scholar]
  • 11.Punyarat P, Buchowski JM, Klawson BT, Peters C, Lertudomphonwanit T, Riew KD. Freehand technique for C2 pedicle and pars screw placement: is it safe? Spine J. 2018;18:1197–203. doi: 10.1016/j.spinee.2017.11.010. [DOI] [PubMed] [Google Scholar]
  • 12.Ma XY, Yin QS, Wu ZH, Xia H, Riew KD, Liu JF. C2 anatomy and dimensions relative to translaminar screw placement in an Asian population. Spine (Phila Pa 1976) 2010;35:704–8. doi: 10.1097/brs.0b013e3181bb8831. [DOI] [PubMed] [Google Scholar]
  • 13.Silveri CP, Vaccaro AR. Posterior atlantoaxial fixation: the Magerl screw technique. Orthopedics. 1998;21:455–9. doi: 10.3928/0147-7447-19980401-15. [DOI] [PubMed] [Google Scholar]
  • 14.Meyer B, Kuhlen D. Atlantoaxial fusion: transarticular screws versus screw-rod constructs. World Neurosurg. 2013;80:516–7. doi: 10.1016/j.wneu.2012.11.008. [DOI] [PubMed] [Google Scholar]
  • 15.Xu R, Nadaud MC, Ebraheim NA, Yeasting RA. Morphology of the second cervical vertebra and the posterior projection of the C2 pedicle axis. Spine (Phila Pa 1976) 1995;20:259–63. doi: 10.1097/00007632-199502000-00001. [DOI] [PubMed] [Google Scholar]
  • 16.Kuroki H, Rengachary SS, Goel VK, Holekamp SA, Pitkanen V, Ebraheim NA. Biomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation. Neurosurgery. 2005;56:151–9. doi: 10.1227/01.neu.0000144838.01478.35. [DOI] [PubMed] [Google Scholar]
  • 17.Ma XY, Yin QS, Wu ZH, et al. C1 pedicle screws versus C1 lateral mass screws: comparisons of pullout strengths and biomechanical stabilities. Spine (Phila Pa 1976) 2009;34:371–7. doi: 10.1097/brs.0b013e318193a21b. [DOI] [PubMed] [Google Scholar]
  • 18.Qiu F, Zou XB, Xu XL, Jiang WC, Liu GC, Ma XY. A biomechanical comparison of crossed and parallel rod configurations in atlantoaxial internal fixation. Eur Spine J. 2021;30:576–84. doi: 10.1007/s00586-020-06655-6. [DOI] [PubMed] [Google Scholar]
  • 19.Lehman RA, Dmitriev AE, Helgeson MD, Sasso RC, Kuklo TR, Riew KD. Salvage of C2 pedicle and pars screws using the intralaminar technique: a biomechanical analysis. Spine (Phila Pa 1976) 2008;33:960–5. doi: 10.1097/brs.0b013e31816c915b. [DOI] [PubMed] [Google Scholar]
  • 20.Zhu C, Wang J, Wu Z, Ma X, Ai F, Xia H. Management of pediatric patients with irreducible atlantoaxial dislocation: transoral anterior release, reduction, and fixation. J Neurosurg Pediatr. 2019;24:323–9. doi: 10.3171/2019.4.peds1928. [DOI] [PubMed] [Google Scholar]

Articles from Asian Spine Journal are provided here courtesy of Korean Society of Spine Surgery