Optimizing Leadless Atrial Pacemaker Performance Through Strategic Reprogramming

A Quality Improvement Initiative

Amy Kleinhans a,∗, Heather M Ross a,b, Fantasia Hudson a, Rahul N Doshi a,c
PMCID: PMC13749539  PMID: 42747358

Abstract

Background

Although implantation strategies for dual-chamber leadless pacemakers are well documented, long-term management, particularly concerning battery longevity, remains under observation. Atrial leadless pacemaker battery life is shorter than that of its ventricular counterpart in real-world observations without clear guidelines established for postimplantation programming.

Project Rationale

We undertook a quality improvement initiative in our outpatient device clinic to address battery longevity in leadless pacemakers via programming approaches targeting 6 optimization areas. A step-by-step algorithm was developed for device clinic staff.

Project Summary

Twenty patients underwent algorithm-driven reprogramming by device clinic staff, with the average battery life increasing by 1.41 years to 5.48 years, a 33.7% improvement. Repeat interrogation at 90 days showed consistent longevity readings (5.4 years).

Take-Home Messages

Dual-chamber leadless pacemakers require refined programming strategies. Implementing a structured algorithm in an outpatient device clinic enabled clinic staff to tailor settings for each patient, resulting in a substantial increase in battery life.

Key Words: Aveir, battery, device management, leadless, longevity, pacemaker, programming

Visual Summary

graphic file with name ga1.webp

Background

In 2023, the U.S. Food and Drug Administration cleared the first dual-chamber leadless pacemaker (LP).1, 2, 3 Battery longevity is a central consideration for dual-chamber LP systems, which require independent optimization of the atrial LP (aLP) and ventricular LP (vLP) components. As adoption of dual-chamber LP systems grows, it is imperative to refine clinical management strategies.

Take-Home Messages

  • •

    Dual-chamber leadless pacemakers require refined programming strategies.

  • •

    Implementing a structured algorithm in an outpatient device clinic enabled clinic staff to tailor settings for each patient, resulting in a substantial increase in battery life.

In our practice, aLP battery life was often 40% to 50% shorter than that of the vLP at 6 to 12 months postimplant, falling below the manufacturer's labeled longevity projections.4,5 Disproportionate aLP battery depletion compared with vLP is due to 4 main factors. First, aLP battery capacity (174 mAh) is smaller than that of the vLP (241 mAh).4 Second, aLP impedance trends lower, requiring higher current to deliver the programmed voltage. Third, the implant-to-implant (i2i) communication mechanism that provides atrioventricular (AV) synchrony places additional demand on the battery.6 Fourth, in patients with atrial tachyarrhythmias, the aLP is required to communicate more frequently, further depleting the battery. Although it is not possible to modify the inherent impedance or battery capacity at the point of care, several programming adjustments were explored to optimize battery longevity.

Project Rationale

We identified 6 modifiable programming patterns impactful to aLP battery longevity. Of these, 5 represented transvenous programming norms, including: 1) use of dual chamber pacing mode with rate response in patients with sinus node dysfunction (SND) and minimal ventricular pacing requirements; 2) use of a standard base rate of 60 beats/min; 3) application of a 2-fold safety margin for outputs; 4) standardization of pulse width to 0.4 milliseconds; and 5) lack of sensitivity adjustments to prevent unnecessary mode switching (MS). In addition, i2i strength was being programmed to achieve 100% throughput bidirectionally despite safety targets of ≥80% in prioritized directions and ≥60% in nonprioritized directions.7

Based on these observations, we implemented a standardized aLP reprogramming algorithm to guide device clinic staff of different training levels to optimize battery life. Six areas of possible programming changes were targeted in the algorithm: 1) voltage; 2) pulse width; 3) i2i communication; 4) lower rate limits (LRLs); 5) sensitivity; and 6) use of atrial pacing mode with rate response (AAI(R)) (Figure 1).

Figure 1.

Figure 1

Step-by-Step Atrial Leadless Pacemaker Reprogramming Algorithm

Decision pathway developed for device clinic staff of all training levels, guiding sequential optimization of 6 programming domains (pacing voltage, pulse width, implant-to-implant [i2i] communication strength, lower rate limit, atrial sensitivity, and AAI(R) mode) to extend atrial leadless pacemaker battery longevity while preserving atrioventricular synchrony and patient safety. AAI(R) = atrial pacing mode with rate response; MSE = mode switch events; RA = right atrial; RTC = return to clinic; RV = right ventricle; SND = sinus node dysfunction; SSS = sick sinus syndrome; VP = ventricular pacing; WCL = Wenkebach cycle length.

Our objectives were to assess adoption of the new programming strategy across clinic staff of all training levels, document programming changes for each patient, and quantify associated battery life impacts based on the aLP longevity estimate.

Project Description

All device clinic staff received virtual training on the updated protocol (Figure 1), which was then applied to all dual-chamber LP patients during their next scheduled visit. Patients were re-evaluated 2 weeks after algorithm-guided reprogramming to assess device usage and patient tolerance (per clinic standards), and again in 3 months to evaluate the durability of battery longevity estimates. All battery measurements were made in dual chamber pacing mode with rate response for consistency. Statistical significance of projected battery gains was assessed using paired t-tests to compare estimates at baseline to 14 and 90 days.

Project Deliverables

The algorithm was well received and applied consistently by clinic staff, noting variable education needs for nursing vs technician staff. Technician staff required additional in-person training on the programming algorithm. Descriptive statistics for the first 21 patients are presented in Table 1. One patient relocated out of state and was omitted from the 3-month follow-up data detailed below and presented in Table 2.

Table 1.

Patient Demographics and Baseline Characteristics (N = 21)

Age (y) 73
Male, n (%) 13 (61)
Female, n (%) 8 (39)
Indication for pacing, n (%)
 Sick sinus syndrome 13 (62)
 Atrioventricular block 2 (9.5)
 Tachycardia-bradycardia syndrome 2 (9.5)
 Mixed indication 4 (19)
Average atrial pacing, % 42.6
Atrial mode switch, n (%) 15 (71.4)
 Appropriate 9 (60)
 Inappropriate 6 (40)
Atrial lead impedance (Ω) 302.4
Ventricular lead impedance (Ω) 567.6
Average baseline battery longevity (y) 4.1

Table 2.

Device Battery Gain by Programming Strategy

Strategy n Baseline Estimate (y) Updated Estimate (y) Battery Gain (y) (%) P Value
AAI Mode 3 5.3 ± 2.85 10.1 ± 5.46 4.8 (90.6) —
Voltage adjustment 16 3.97 ± 1.64 4.7 ± 1.75 0.73 (18.39) <0.01
Pulse width adjustment 12 4.88 ± 1.75 5.12 ± 1.77 0.24 (4.92) <0.01
i2i adjustment 14 4.33 ± 1.71 4.99 ± 1.79 0.66 (15.24) <0.01
Lower rate limit adjustment 9 4.8 ± 1.74 5.39 ± 1.82 0.59 (12.3) <0.01

AAI = atrial pacing mode; i2i = implant-to-implant.

The dash denotes no P value was calculated for AAI pacing because this mode was not represented in the final study population, despite providing the greatest projected battery longevity.

Voltage and pulse width

Per manufacturer recommendations, pacing outputs were programmed to ≤1.25 V at ≤1.0 milliseconds. A safety margin of 0.5 V was applied for nondependent patients and 1 V for dependent patients, rather than the conventional 2-fold safety margin. Using this strategy, the voltage was successfully decreased in 16 (76.2%) patients with a mean decrease in voltage by 0.9 V, gaining 0.73-year battery longevity (18.4%, P < 0.01). If the target output of ≤1.25 V was not achieved, pulse width was optimized toward 0.2 milliseconds as tolerated based on standard threshold testing. Twelve patients reached pulse width targets, resulting in a battery longevity gain of 0.24 years (4.92%, P < 0.01).

i2i communication

Fourteen patients had i2i strength adjusted with a target i2i setting of ≤4. The average i2i decrease was 1.6, gaining 0.66-year battery longevity (15.24%, P < 0.01). Throughput was consistent at both 14-day and 3-month evaluations and remained within safety targets based on prioritized directionality with adjustments (≥80% in prioritized directions and ≥60% in nonprioritized directions).7

Lower rate limit

Unless contraindicated, LRLs were adjusted to 50 beats/min. Two patients maintained an LRL of 70 beats/min for symptom management (premature ventricular contractions and cardioinhibitory syncope, respectively), 10 patients presented with an LRL of 50 beats/min at baseline, and 9 patients had the LRL adjusted from 60 to 50 beats/min. One patient had the LRL returned to 60 beats/min due to fatigue. Estimated battery longevity gain for this strategy at 14 days was 0.59 years (12.3%, P < 0.01).

Sensitivity

Fifteen patients had MS events (burdens <1%-12%). MS rates were set to 180 beats/min, with sensitivity set to 0.75 mV if appropriate to minimize far-field oversensing. For 6 patients (40%) determined to be oversensing, sensitivity was further adjusted to minimize inappropriate MS. Battery gain assessment for this strategy proved challenging due to low MS burden.

AAI(R) mode

Three patients initially qualified for AAI(R) modes with pacing indications of SND and 1:1 conduction when paced at 110 beats/min. Nine additional patients had AV delays extended by 100 milliseconds for re-evaluation at 3 months. Of these, 3 qualified for AAI(R) pacing at the 90-day visit, and 4 were noted to be candidates for future AAI+VVI programming. This dual-mode strategy could maintain atrial-based pacing with ventricular backup while minimizing i2i comunciation, thereby extending device longevity. Reprogramming to AAI(R) yielded the greatest benefit in projected battery longevity, from a mean of 5.3 years to 10.1 years (90.6% increase). The small sample size prevented inferential statistical analysis.

Project Outcome, Impact, and Future Directions

Across the entire cohort (N = 21), tailored programming changes yielded a mean battery longevity gain of 1.38 years, representing a 33.7% increase (P < 0.001) (Table 3). Decreasing voltage offered the greatest benefit with 0.56 years gained (40.6% of total battery gain), followed by i2i optimization gaining 0.44 years (31.9%), LRL adjustments gaining 0.24 years (17.4%), and pulse width adjustments gaining 0.14 years (10.1%) (Figure 2). The 14-day battery assessment was durable at the 3-month follow-up, with a mean battery life of 5.39 ± 1.41 years, representing appropriate battery depletion (Figure 3).

Table 3.

Cumulative Battery Improvement in DDD Mode (N = 21)

Baseline battery estimate (y) 4.1 ± 1.64
Updated battery estimate (y) 5.48 ± 1.72
Total battery gain (y) (%) 1.38 (33.7)
Contribution by strategy (y) (%)
 Voltage adjustment 0.56 (40.6)
 Pulse width adjustment 0.14 (10.1)
 i2i adjustment 0.44 (31.9)
 Lower rate limit adjustment 0.24 (17.4)

DDD = dual-chamber pacing mode; i2i = implant-to-implant.

Figure 2.

Figure 2

Contribution of Each Programming Strategy to Total Projected Battery Longevity Gain

Mean per-strategy gains across the cohort (N = 21): voltage adjustments 0.56 years (40.6%), i2i optimization 0.44 years (31.9%), lower rate limit adjustments 0.24 years (17.4%), and pulse width adjustment 0.14 years (10.1%). i2i = implant-to-implant; LRL = lower rate limit.

Figure 3.

Figure 3

Durability of Projected Battery Longevity Estimates

Comparison of atrial leadless pacemaker battery longevity estimates at the 14-day and 90-day evaluations, demonstrating that early gains were sustained, with a mean projected battery life of 5.39 ± 1.41 years at 3 months, consistent with appropriate battery depletion. All estimates were obtained in dual-chamber pacing mode with rate response mode for consistency.

No single programming change was applied to every patient, reinforcing the need for tailored device programming. Other than a single complaint of fatigue from a patient after the LRL was dropped to 50 beats/min, programming changes were well tolerated. Based on this experience, our device clinic decided to retain the aLP programming algorithm.

Discussion

For patients who qualify, AAI(R) mode during MS offered substantial battery gain. However, this was limited to patients with low right ventricular pacing burden and required staff training to evaluate Wenckebach cycle length before AAI use. The now available ability to program the vLP in a backup VVI mode would increase AAI(R) programming without risk of asystole and is widely used in our clinic.

In patients contraindicated for AAI(R), optimizing pacing voltage appeared to offer the most battery preservation. Based on the algorithmic approach of performing battery estimates after each change was made, the benefit of subsequent algorithmic strategies may be underestimated as battery gains would be cumulative.

Lowering the signal strength of i2i communication offered the next greatest benefit. Despite the use of conductive technology, which exerts lower battery drain than other forms of communication, we found battery longevity benefits if i2i was programmed to ≤4. i2i throughput testing was performed in seated and supine positions. Special attention was paid to the right atrium-to-right ventricle direction, as it contributes to aLP battery use. Patients with known AV nodal disease were programmed with a target right atrium-to-right ventricle throughput of >80%.

The LRL was comfortably set to 50 beats/min for most patients without any repercussions. Two patients at 70 beats/min for symptom management and 1 patient with fatigue at an LRL of 50 beats/min were noted previously.

MS episodes negatively affected aLP battery longevity. The only patient who had a decline in battery estimates despite programming changes had a high (12%) appropriate MS burden. MS episodes not only increase the use of the aLP battery but also make longevity estimates less accurate. Therefore, patients with a history of atrial arrhythmias may require more frequent in-office evaluation before remote monitoring availability.

Although this quality improvement project is useful to inform longitudinal practices regarding chronic programming-based management of dual-chamber LPs, these findings are limited by the small number of cases in a single setting and a short follow-up duration of 3 months. A larger prospective longitudinal multicenter cohort study may offer additional insights regarding programming optimization for a variety of conditions beyond SND (eg, high-degree AV block), and chronic implications with programming changes that might decrease degree of AV synchrony, such as decreasing atrial sensitivity or i2i strength.

Conclusions

Dual-chamber LPs require refined programming strategies. Implementing a structured algorithm enabled clinic staff to tailor settings to each patient, resulting in a substantial increase in battery life estimates. These findings underscore the importance of empowering device clinic staff with the knowledge and training on emerging implanted device technologies to critically apply algorithmic programming approaches for personalized device management.

Visual Summary.

Visual Summary

Optimizing Leadless Atrial Pacemaker Performance Through Strategic Reprogramming: a Quality Improvement Initiative

The graphic summarizes the clinical problem (accelerated aLP battery depletion relative to the ventricular leadless pacemaker), the intervention (a standardized 6-domain reprogramming algorithm applied by device clinic staff to 21 dual-chamber leadless pacemaker patients, reassessed at 14 and 90 days), and the outcome (mean projected battery longevity gain of 1.38 years, a 33.7% increase [P < 0.001], sustained at 90-day follow-up). AAI(R) = atrial pacing mode with rate response; i2i = implant-to-implant.

Funding Support and Author Disclosures

Dr Doshi has received consulting fees, steering committee participation, and clinical research support from Abbott Medical. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.

Footnotes

The authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’ institutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information, visit the Author Center.

References

  • 1.Ali E. FDA approves first leadless dual-chamber pacing system; a breakthrough in therapy of bradycardia. Int J Surg Open. 2023;58 [Google Scholar]
  • 2.Toon L., Paisey J., Roberts P.R. The dual chamber leadless pacemaker: a game changer for bradycardia management? Expert Rev Cardiovasc Ther. 2024;22(7):285–288. doi: 10.1080/14779072.2024.2375331. [DOI] [PubMed] [Google Scholar]
  • 3.Knops R.E., Reddy V.Y., Ip J.E., Doshi R., et al. A dual-chamber leadless pacemaker. N Engl J Med. 2023;388(25):2360–2370. doi: 10.1056/NEJMoa2300080. [DOI] [PubMed] [Google Scholar]
  • 4.Abbott AveirTM leadless pacemaker model LSP112V and delivery catheter model LSCD111 instructions for use. 2021. https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150035D.pdf
  • 5.Hindricks G., Doshi R., Defaye P., et al. Six-month electrical performance of the first dual-chamber leadless pacemaker. Heart Rhythm. 2024;12:1929–1938. doi: 10.1016/j.hrthm.2024.04.091. [DOI] [PubMed] [Google Scholar]
  • 6.Cantillon D.J., Gambhir A., Banker R., et al. Wireless communication between paired leadless pacemakers for dual-chamber synchrony. Circ Arrhythmia Electrophysiol. 2022;15(7) doi: 10.1161/CIRCEP.122.010909. [DOI] [PubMed] [Google Scholar]
  • 7.Neugebauer F., Noti F., Van Gool S., et al. Leadless atrioventricular synchronous pacing in an outpatient setting: early lessons learned on factors affecting atrioventricular synchrony. Heart Rhythm. 2021;19(5):748–756. doi: 10.1016/j.hrthm.2021.12.025. [DOI] [PubMed] [Google Scholar]

Articles from JACC Case Reports are provided here courtesy of Elsevier

RESOURCES