Abstract
Helicobacter pylori (H. pylori) is one of the main causes of gastritis and gastric ulcers. Therefore, early detection of H. pylori is of great significance for the primary prevention of gastric cancer. Taking advantage of the trans-cleavage ability of Cas effector proteins, we developed a novel CRISPR/Cas12a-mediated electrochemical sensor for the detection of the vacAs1 gene of H. pylori. In this system, gold nanoparticles (AuNPs) were first electrodeposited on the gold electrode, which not only served to immobilize nonspecific ssDNA but also improved the conductivity of the electrode, thereby enhancing the sensitivity of the sensor. By combining recombinase-assisted amplification (RAA) with the trans-cleavage activity of CRISPR/Cas12a, changes in the amount of ssDNA were induced, further altering electronic transfer on the electrode surface. The developed CRISPR/Cas12a-based electrochemical biosensor exhibited a wider detection range (5.4 × 10−19−5.4 × 10−13 M) and a low detection limit (9.5 × 10−20 M). Moreover, the biosensor showed good recovery for the detecting H. pylori in spiked stool and artificial saliva samples. In addition, the biosensor exhibited good specificity and stability, suggesting that it may serve as a potential tool for the rapid, stable, and efficient detecting of H. pylori.
Keywords: CRISPR/Cas, electrochemical biosensor, Helicobacter pylori, trans-cleavage, vacA gene
1. Introduction
Helicobacter pylori (H. pylori) is a microaerophilic, Gram-negative bacterium that represents the primary cause of chronic gastritis and gastric ulcers. Its infection is closely related to the occurrence and development of gastric cancer, gastric mucosal-related lymphoma, and other related diseases (Chen et al., 2026; Arif et al., 2026; Son et al., 2025), posing a substantial public health and economic burden to society. In 1994, the World Health Organization (WHO) classified H. pylori as a Class I carcinogen. It is estimated that over half of the global population is infected with H. pylori (Song Q. et al., 2026). To date, humans are the only known reservoir for H. pylori transmission, and the bacterium has been detected in gastric and intestinal secretions, saliva, and feces of infected individuals (Ma et al., 2026; Huang et al., 2026). Notably, studies have shown that H. pylori strains carrying specific virulence genes are more likely to progress to gastric cancer than other strains (Okoye et al., 2026; Mbuthia et al., 2026). Among these virulence genes, vacA, especially the subtype s1, is closely associated with the occurrence of gastric cancer (Liu and Ji, 2026; Song X. et al., 2026; Wang et al., 2026). Therefore, the vacAs1 genotype may serve as a potential marker for the diagnosis of H. pylori infection.
Detection methods for H. pylori can be broadly classified as invasive and non-invasive. Invasive methods are time-consuming and expensive, and are therefore not suitable for mass screening (Zan et al., 2026). Among non-invasive methods, the urea breath test (UBT) involves radiation exposure and requires fasting as well as withdrawal of gastric drugs taken within a week (Singla et al., 2026). In addition, serological antibodies against H. pylori may persist for a long time after eradication, and antibody tests may remain positive for at least 6 months following eradication therapy (Skrebinska et al., 2022). Since H. pylori primarily colonizes gastric mucosal epithelial cells, its metabolic products and bacterial components are shed with the renewal of gastric mucosa, excreted into the gastrointestinal tract, and finally excreted through stool. Therefore, H. pylori can be detected in stool, and stool antigen testing offers the advantages of rapidity and simplicity (Fang et al., 2026). However, false-negative results may occur when the antigen concentration in the sample is low. Consequently, novel detection technologies for H. pylori are needed to meet the clinical diagnostic demands.
For these reasons, we proposed a CRISPR-based electrochemical sensor to detect the vacAs1 gene of H. pylori, aiming to overcome the limitations of traditional methods. Clustered regularly interspaced short palindromic repeats (CRISPR) are DNA sequences found in the genomes of prokaryotes such as bacteria and archaea. Cas12a is an endonuclease related to CRISPR system. In recent years, owing to their ability to accurately identify and cleave nucleic acid targets, CRISPR/Cas12a systems have shown great potential for the rapid detection of pathogenic microorganisms (Gao and Liu, 2026). In the CRISPR/Cas12a system, when a target sequence is present, Cas12a can non-specifically degrade the labeled single-stranded DNA (ssDNA) reporters through crRNA-guided trans-cleavage activity (Zhu et al., 2026). In addition, the use of Cas12a can reduce the steps required for RNA processing and transcription (Zheng et al., 2026). However, current CRISPR/Cas12a-based detection methods generally require pre-amplification of the target nucleic acid to achieve diagnostic sensitivity, which necessitates the use of isothermal amplification. Combining isothermal amplification with CRISPR/Cas systems may facilitate the rapid detection and analysis of nucleic acids (Akimbekova et al., 2026). Although CRISPR/Cas-based methods have been developed to detect H. pylori, the vacAs1 gene has not yet been targeted (Liu Y. et al., 2026; Xu et al., 2026).
Electrochemical biosensors have been recognized as promising analytical technologies due to the unique advantages, including ease of operation, low cost, fast response and high sensitivity. Several electrochemical biosensors based on different modified electrodes have been developed for H. pylori detection (Kumar et al., 2025; Liu X. et al., 2026; Rafique et al., 2024). Despite these successes, most of these strategies either require complex fabrication of electrochemically responsive interfaces or suffer from potential cross-reactivity. These shortcomings may lead to trade-offs in sensitivity, specificity and user-friendliness. Previous studies have shown that nucleic acid detection systems based on CRISPR/Cas12a achieve specific detection through complementary base pairing between crRNA and target sequence, enabling accurate nucleic acid targeting. For example, Yang et al. developed an electrochemical DNA sensor based on CRISPR/Cas12a for the detection of hepatitis B virus (HBV) DNA, which exhibited high sensitivity and selectivity (Yang et al., 2026). From this perspective, combining ssDNA with CRISPR technology to construct electrochemical DNA sensors is expected to provide improved biosensing performance for the precise quantification of H. pylori.
In this study, we constructed a CRISPR/Cas12a-based electrochemical biosensor combined with RAA for the detection of VacAs1 gene of H. pylori (Figure 1). First, AuNPs were electrodeposited on the electrode surface, followed by immobilization of thiolated single-stranded DNA (SH-ssDNA) vis Au-S covalent bonding. crRNA was designed to specifically recognize target DNA through complementarity with the target sequence and the protospacer adjacent motif (PAM). Upon PAM recognition, Cas12a activated its indiscriminate trans-cleavage activity toward ssDNA, thereby cleaving non-target ssDNA. In the presence of target DNA amplified by RAA, the trans-cleavage ability of Cas12a was activated, leading to cleavage of the surface-immobilized ssDNA from the electrode. Electrochemical signals were recorded by differential pulse voltammetry (DPV) before and after the introduction of the target nucleic acid. By combining RAA with CRISPR/Cas12a, this electrochemical biosensor provided a novel ultrasensitive and highly specific platform for future pathogen detection.
FIGURE 1.

Principle of the CRISPR-based electrochemical biosensor. (A) First, genomic DNA was extracted, followed by RAA amplification and Cas12a-mediated trans-cleavage. (B) The electrochemical biosensor was used to detect RAA products. AuNPs were electrodeposited on the electrode, and the ssDNA reporter was then immobilized on the modified electrode. 6-mercapto hexanol (MCH) was used to block the unreacted active sites. In the presence of the target, the trans-cleavage activity of Cas12a was activated, and the SH-ssDNA reporter was cleaved from the electrode, resulting in a high electrochemical current. In the absence of the target, Cas12a/crRNA did not trigger trans-cleavage activity of the SH-ssDNA reporter, resulting in a low electrochemical currents.
2. Materials and methods
2.1. Materials
EnGen®Lba Cas12a and NEBuffer™ 2.1 were purchased from New England Biolabs (Beijing) Co., Ltd. (Beijing, China). RAA basic kit was purchased from Jiangsu Qitian Gene Technology Co., Ltd. (Jiangsu, China). Tri - (2-carboxyethyl) phosphine hydrochloride (TCEP) and 6-mercaptohexyl alcohol (MCH) were purchased from Wuhan Servicebio Technology Co., Ltd. (Wuhan, China). Artificial saliva (pH 6.8) was obtained from Shanghai Yuanye Bio-Technology Co., Ltd. (Shanghai, China). The Rapid Bacterial Genomic DNA Isolation Kit, Mag-MK Soil & Stool genomic DNA extraction kit, and Ezup Saliva/Urine DNA Extraction Kit were purchased from Sangon Biotech (Shanghai) Co., Ltd. (Shanghai, China). All electrochemical measurements were performed on a CHI 660E Electrochemical Workstation (Shanghai Chenhua Instrument Co., LTD., China).
2.2. Design of primers and crRNA for vacAs1 gene
Primers for amplification of the vacAs1 gene of H. pylori were designed using the NCBI Prime-BLAST online tool. crRNA was designed using the CRISPOR online tool. The primers, crRNA, ssDNA-FQ and SH-ssDNA (listed in Supplementary Table S1) were synthesized by Sangon Biotech (Shanghai) Co., Ltd. (Shanghai, China).
2.3. Bacterial culture and genomic DNA extraction
A total of seven bacteria strains were used in this study; including H. pylori (ATCC 43504), H. pylori type II (vacA-negative) and five clinical isolates of non-H. pylori strains (Pseudomonas aeruginosa, Escherichia coli, Enterococcus faecalis, Acinetobacter baumannii and Staphylococcus aureus), all of which were kept in our laboratory. The culture conditions and CFU quantitative method for these bacteria were described in the Supporting Information. Subsequently, the bacteria were scraped off and suspended in phosphate-buffered saline (PBS). According to the manufacturer’s instructions, 1 mL of each bacterial culture suspension was used for DNA extraction using the Rapid Bacterial Genomic DNA Isolation Kit. The purity and concentration of extracted genomic DNA were measured with an ultramicrospectrophotometer (UltraM-C200, Shanghai, China). The DNA was then stored at −20 °C for subsequent RAA tests.
2.4. RAA amplification
The H. pylori vacAs1 gene sequence retrieved from NCBI was selected as the target of RAA amplification. The NCBI Primer-BLAST online tool was used to design primers for the RAA reaction based on the H. pylori vacAs1 gene. Both the forward and reverse primers were 30 nucleotides (nt) in length, and the expected amplicon size was approximately 200 bp (listed in Supplementary Table S1). The RAA reaction was performed according to the instructions of the RAA amplification kit, with minor modifications. Briefly, the RAA precipitate was re-suspended in a provided buffer. Each reaction consisted of 25 μL of RAA buffer solution, 10 μM forward and reverse primers, 2 μL of target DNA template, and 20 mM magnesium acetate (MgAc), supplemented with sterile water to a final volume of 50 μL. The mixture was then amplified at 37 °C for 40 min. It should be noted that the DNA concentrations reported in this study referred to the final concentration in the 50 μL RAA reaction mixture. The detailed DNA dilution procedure was provided in the Supporting Information.
2.5. Preparation of the SH-ssDNA modified gold electrode
Before modification, the gold electrode was successively polished with 0.3 μm and 0.05 μm aluminum oxide, ultrasonicated in anhydrous ethanol and distilled water for 5 min, and finally dried with N2. A layer of AuNPs was electrodeposited on electrode surface in 2 mL of 1% HAuCl4 solution at −0.2 V for 40 s. Meanwhile, the SH-ssDNA probe was treated with 10 µM TCEP and incubated in the dark at 37 °C for 2 h to reduce the S-S bond. The SH-ssDNA probe was then diluted to 1 µM with 10 mM Tris buffer containing 100 mM NaCl, 10 mM MgCl2, and 10 mM EDTA. The diluted SH-ssDNA was added to the surface of the modified electrode and incubated in the dark at 37 °C for 5 h. The electrode was then immersed in 2 mM MCH solution and stored overnight at 4 °C to block the unreacted active sites. After each modification step, the electrode was washed with 10 mM Tris buffer to remove non-absorbed residual molecules. The modified electrode was either used immediately for subsequent test or stored in 10 mM Tris buffer (containing 100 mM NaCl) at 4 °C for short-term storage.
2.6. In vitro digestion of Cas12a/crRNA
According to the target sequence, crRNA was designed using Cas-Designer (http://www.rgenome.net/cas-designer/) (listed in Supplementary Table S1). Following the method of Li et al. (2021) with slight modifications, electrochemical analysis of the trans-cleavage reporter was carried out. The Cas12a/crRNA cleavage reaction system consisted of 200 nM EnGen LbaCas12a, 200 nM crRNA and 1×NEBuffer 2.1. After the RAA reaction, 5 μL of the amplification product was mixed with 15 μL Cas12a/crRNA lysate to form Cas12a/crRNA-target DNA complex. The complex was then added to the surface of the SH-ssDNA-modified electrode, incubated at 37 °C for 4 min, and washed with 10 mM Tris buffer. Finally, electrochemical detection was performed. To verify the RAA-CRISPR/Cas12a system, a 200 nM ssDNA-FQ probe was added to the Cas12a/crRNA lysate and analyzed on a real-time fluorescent quantitative PCR instrument at 37 °C for 45 min, with the fluorescence signal measured every minute.
2.7. Electrochemical measurements
Differential pulse voltammetry (DPV) was used to evaluate the response of the electrochemical sensor. The DPV parameters were as follows: voltage range was −0.1∼ 0.5 V, amplitude was 0.05 V, pulse width was 0.05 s, and sampling width was 0.0167 s. DPV measurements were performed before and after the addition of Cas12a/crRNA-target DNA complex, and the change of peak current was analyzed. The change in peak current was calculated as △I=I0-I, where I0 was the peak current of the modified electrode before the reaction, and I was the peak current after the reaction. Cyclic voltammetry (CV) and electrochemical impedance technique (EIS) were used to characterize the electrode surface during assembly. The CV parameters were set as follows: scanning speed was 50 mV/s, and scanning potential range was −0.2 ∼ 0.6 V. EIS measurements were performed over a frequency range of 0.1 Hz–100 KHz with an amplitude of 5 mV. All electrochemical measurements were performed in PBS (0.1 M, pH 7.0) containing 10 mM K3 [Fe(CN)6]/K4 [Fe(CN)6].
2.8. Spiked sample analysis
This study was conducted in accordance with the Helsinki Declaration and relevant clinical trial laws and regulations of China. The collection of samples has obtained the ethical approval document (Approval number: 1598-01). The stool samples used in this study were residual samples from routine stool testing of patients who had been confirmed negative for H. pylori infection by the urea breath test. H. pylori strain ATCC 43504 was cultured to the logarithmic growth stage and serially diluted 10-fold with sterile saline. For each 1 g stool sample or 1 mL artificial saliva, 8 mL of sterile saline and 1 mL of bacteria solution at different concentrations were added sequentially. After thorough mixing, genomic DNA was extracted for CRISPR-based electrochemical analysis. qPCR was used to verify the accuracy of the developed method, and the qPCR procedure was described in the Supporting Information. In parallel, the initial concentration of inoculated bacteria solution was determined by plate counting.
To evaluate the sensitivity and specifcity of the developed sensor, experiments were conducted in triplicate, and the results were presented as mean ± standard deviation (SD). Statistical analysis and graph visualization were performed using GraphPad Prism software.
3. Results and discussion
3.1. Verification of the CRISPR-based electrochemical biosensor for H. pylori detection
To verify the feasibility of the CRISPR-based electrochemical biosensor for nucleic acid detection, the vacA gene (HPATCC43504_00804) of H. pylori was selected as the target sequence, and the corresponding crRNA was designed based on this sequence. Genomic DNA was subjected to RAA reaction to generate abundant target dsDNA. Cas12a, guided by crRNA, specifically recognized and cleave the target dsDNA, thereby activating its trans-cleavage activity. Since Cas12a tended to cleave ssDNA, ssDNA-FQ probes were added to the Cas12a/crRNA digestion solution to verify the feasibility of the CRISPR/Cas12a system. As shown in Figure 2A, the target dsDNA was successfully amplified by RAA, with a length of about 200 bp. The amplified product was then subjected to Cas12a/crRNA digestion. When both the Cas12a/crRNA complex and the target dsDNA were present, the fluorescence intensity increased rapidly. In contrast, in the absence of Cas12a, crRNA, or target dsDNA, the fluorescence signal nearly unchanged (Figure 2B). These results demonstrated that the RAA-CRISPR/Cas12a system could be used for H. pylori detection and that Cas12a could be activated to cleave non-specific ssDNA indiscriminately.
FIGURE 2.

Feasibility analysis of CRISPR-based electrochemical biosensor for H. pylori detection. (A) RAA amplification of the vacAs1 gene of H. pylori. Lanes 1-6 and lane 7 corresponded to the RAA amplification of the gene and the negative control, respectively; lane M was the GN100 bp DNA ladder marker. (B) Real-time fluorescence analysis of non-specific ssDNA cleavage based on the CRISPR/Cas12a system. (C) CV curves characterizing the assembly and reaction process of the CRISPR-based electrochemical biosensor. (a) Bare gold electrode, (b) AuNPs/gold electrode, (c) ssDNA/AuNPs/gold electrode, (d) MCH/AuNPs/gold electrode, (e) Cas12a/crRNA-target DNA/MCH/AuNPs/gold electrode. (D) EIS curves characterizing the assembly and reaction process of the CRISPR-based electrochemical biosensor. (a) Bare gold electrode, (b) AuNPs/gold electrode, (c) ssDNA/AuNPs/gold electrode, (d) MCH/AuNPs/gold electrode, (e) Cas12a/crRNA-target DNA/MCH/AuNPs/gold electrode.
Subsequently, the CRISPR/Cas12a system was applied to the electrochemical biosensor to investigate its trans-cleavage activity on modified electrodes. The Cas12a/crRNA-target DNA complex was added to the SH-ssDNA-modified electrode for the trans-cleavage reaction. The interfacial properties of the gold electrode at different stages were characterized by CV and EIS, and the current signals before and after the reaction were determined by DPV. As shown in Figure 2C, a well-defined [Fe(CN)6]3-/4 oxidation-reduction peak pair (curve a) was observed on the bare gold electrode. After AuNPs were electrodeposited on the electrode, the peak current was further enhanced due to the large specific surface area and good conductivity of the AuNPs, which increased the effective area of the electrode and accelerated the electron transfer rate (curve b). When SH-ssDNA was immobilized on the gold electrode, the negatively charged SH-ssDNA repelled [Fe(CN)6]3-/4 in the electrolyte through electrostatic repulsion, hindering electron transfer and thereby reducing the current signal (curve c). After the electrode surface was blocked with MCH, the electron transfer was further hindered due to the insulating properties of MCH, leading to a further decrease in the current signal (curve d). When the Cas12a/crRNA-target DNA complex was added to the modified electrode, abundant immobilized SH-ssDNA was cleaved by activated Cas12a, reducing the negative charge on the electrode surface and weakening the electrostatic repulsion between [Fe(CN)6]3-/4- and the SH-ssDNA, which resulted in an enhanced current signal in curve e compared with curve d. The results were further validated by EIS, and the results shown in Figure 2D were consistent with those obtained by CV. Furthermore, SEM was performed to examine the morphological changes of AuNPs. As shown in Supplementary Figure S1, the AuNPs assembled on the electrode surface by electrodeposition exhibited a uniformly distributed particle morphology, with an average particle size of approximately 80 nm. In summary, the CRISPR-based electrochemical detection method could be applied to the detection of H. pylori.
3.2. Optimization the experimental conditions
Since AuNPs used as electrode modifiers can enhance the electrochemical signal, the CV responses of AuNPs-modified electrodes prepared with different deposition times were investigated in 10 mM [Fe(CN)6]3-/4- phosphate buffer solution to optimize the deposition time of AuNPs. As shown in Figure 3A, the response current increased with deposition time and reached the maximum at 40 s. When the deposition time exceeded 40 s, the response current decreased, owing to excessive deposition of AuNPs on the electrode surface, which formed a thick and dense barrier layer that hindered electron transfer. Therefore, 40 s was selected as the optimal electrochemical deposition time.
FIGURE 3.

Optimization of experimental conditions. (A) Effect of AuNPs electrodeposition time on the detection signal. (B) Effect of ssDNA concentration on the detection signal. (C) Effect of Cas12a/crRNA concentration on trans-cleavage activity. (D) Effect of Cas12a trans-cleavage time on the detection signal. Error bars represent mean ± SD, where n = 3 replicates.
Next, the concentration of ssDNA on the working electrode was optimized to achieve efficient surface chemistry-based trans-cleavage. As shown in Figure 3B, as the ssDNA concentration increased from 0.25 μM to 1 μM, the change in peak current (ΔI) increased accordingly. However, when the concentration exceeded 1 μM, ΔI decreased. Therefore, 1 μM was selected as the optimal ssDNA concentration.
Trans-cleavage activity was key to signal transduction, and the Cas12a/crRNA concentration directly affected trans-cleavage efficiency. Therefore, the Cas12a/crRNA concentrations were optimized. As shown in Figure 3C, ΔI was the largest when both Cas12a and crRNA concentrations were 200 nM. Therefore, 200 nM Cas12a and 200 nM crRNA were selected as the optimal concentrations for subsequent detection.
Finally, the trans-cleavage reaction time was investigated. As shown in Figure 3D, ΔI increased continuously with time up to 4 min and then plateaued. Therefore, 4 min was selected as the CRSPR reaction time for subsequent experiments.
3.3. Sensitivity evaluation
The sensitivity of the CRISPR-based electrochemical sensors was evaluated using genomic DNA of H. pylori at different concentrations (5.4 × 10−13-5.4 × 10−19 M). As shown in Figure 4A, with increasing genomic DNA concentration, more Cas12a/crRNA complexes were activated, leading to enhanced trans-cleavage of SH-ssDNA on the electrode surface. Consequently, the current signal increased, and the change in peak current (ΔI) showed a good linear relationship with the logarithm of genomic DNA concentration. The linear regression equation was y = 18.57x+375.8 (R2 = 0.9812), where y was the change in peak current before and after the reaction, and x was the logarithm of genomic DNA concentration (Figure 4B). The limit of detection (LOD) was derived to be 9.5 × 10−20 M based on the 3SB/m criterion (details were provided in the Supporting Information).
FIGURE 4.

(A) DPV curves of the CRISPR-based electrochemical sensors for the detection of genomic DNA at different concentrations; (B) Linear relationship between the change in peak current (ΔI) and the logarithm of the genomic DNA concentration; (C) DPV curves of the CRISPR-based electrochemical sensors for detecting of different bacterial concentrations; (D) Linear relationship between ΔI and the logarithm of the bacterial concentration. Control refered to the replacement of the template with H2O. Error bars represented mean ± SD, where n = 3 replicates.
To further clarify the relationship between the reported DNA concentration and the actual number of target molecules in each reaction, the genome copies per 50 μL RAA reaction were calculated based on the H. pylori genome size of approximately 1.65 Mb (corresponding to ∼1.81 fg per genome copy). At the lowest experimentally tested concentration of 5.4 × 10−19 M, the corresponding genome copies per 50 μL reaction was approximately 16.3. The LOD was calculated as 9.5 × 10−20 M (S/N = 3), corresponding to approximately 2.9 genome copies per 50 μL reaction.
To the best of our knowledge, no CRISPR/Cas12a-based electrochemical sensing platform for H. pylori has been reported. Therefore, we compared our method with previously reported H. pylori detection methods. As can be seen from Supplementary Table S2, the CRISPR-based electrochemical method developed in this study showed improved performance in terms of detection range and LOD compared with previously reports (Yin et al., 2025; Wang et al., 2023). This improvement was attributed to the signal amplification effect of AuNPs and the trans-cleavage activity of Cas12a.
The CRISPR-based electrochemical sensor was further applied to the detection of H. pylori in pure culture medium. As the bacterial concentration increased, the current signal gradually increased (Figure 4C). A linear relationship was observed between the change in peak current and the logarithm of the bacterial concentration. The linear detection range was 101-107 CFU/mL, and the linear regression equation was y = 20.29x-1.143 (R2 = 0.9775), where y was the change in peak current and x was the logarithm of the bacterial concentration (Figure 4D). Calculated by the above method, the LOD was 5 CFU/mL (details were provided in the Supporting Information).
3.4. Selectivity, stability and reproducibility of the CRISPR-based electrochemical sensor
To evaluate the specificity of the sensor against H. pylori, Pseudomonas aeruginosa, Escherichia coli, Enterococcus faecalis, Acinetobacter baumannii, Staphylococcus aureus and H. pylori type II were selected as non-target bacteria. The results were shown in Figure 5A. The sensor exhibited a significant electrochemical response to H. pylori (at a bacterial concentration of 103 CFU/mL). Although the concentrations of the six non-target bacteria were ten times that of the target (at a bacterial concentration of 104 CFU/mL), the electrochemical response of the sensor to the non-target bacteria was not significantly different from the background signal. Moreover, when H. pylori was mixed with these non-target bacteria, the change in peak current (ΔI) was almost the same as that observed with the target alone, indicating that the detection of H. pylori was not affected by the non-target bacteria. Therefore, the sensor showed good selectivity for the detection of the vavAs1 gene of H. pylori.
FIGURE 5.

(A) Specificity of the sensor towards H. pylori using Pseudomonas aeruginosa, Escherichia coli, Enterococcus faecalis, Acinetobacter baumannii, Staphylococcus aureus and H.pylori Type II as non-target bacteria. (B) Electrochemical signal changes of the sensor stored at 4 °C for 20 days, with the signal recorded every 5 days. (C) ΔI values obtained from five independently prepared sensors for the detecting of the same concentration of H. pylori. Error bars represented mean ± SD, where n = 3 replicates.
Stability is an important parameter of sensor performance and can affect its practical applicability. The sensor was stored in 10 mM Tris buffer (containing 100 mM NaCl) at 4 °C, and the signal was recorded every 5 days. The results were shown in Figure 5B. After 20 days, the change of the electrochemical signal remained at 87.2% of the initial signal, indicating that the developed sensor had excellent storage stability.
Finally, five independently prepared sensors were used to detect the same concentration of H. pylori (at a bacterial concentration of 103 CFU/mL), as shown in Figure 5C. The obtained ΔI values were similar, with a relative standard deviation (RSD) of 3.51%, indicating that the reproducibility of the sensor was within an acceptable range.
3.5. Spike-recovery experiment
To evaluate the potential of the sensor for practical application, spike-recovery experiments were performed using H. pylori-negative stool samples and artificial saliva spiked with cultured H. pylori. Different spike levels of bacterial solutions (1 × 102, 1 × 104, and 1 × 106 CFU/mL) were added to stool and artificial saliva samples, and genomic DNA was extracted for CRISPR-based electrochemical detection. As shown in Table 1, the recovery rates of H. pylori in stool samples ranged from 94.74% to 103.32%, with an RSD of 2.86%–5.78%; the recovery rates in artificial saliva samples ranged from 94.66% to 98.93%, with an RSD of 3.64%–5.29%. In addition, the detection results obtained by the sensor were in good agreement with those obtained by qPCR (Supplementary Table S2). These results indicated that the prepared sensor could be used for the detection of the vacAs1 gene of H. pylori in spiked stool and artificial saliva samples with good recovery.
TABLE 1.
Recovery of H. pylori in spiked stool and artificial saliva samples (n = 3 replicates). Data were presented as mean ± SD. RSD, relative standard deviation.
| Samples | Spiked Hp(CFU/mL) | Detected Hp(CFU/mL) | Recovery rate (%) | RSD (%) |
|---|---|---|---|---|
| Stool | 0 | Negative | - | - |
| 1 × 102 | 0.94 × 102 | 94.74 | 2.86 | |
| 1 × 104 | 1.03 × 104 | 103.32 | 5.76 | |
| 1 × 106 | 0.97 × 106 | 96.99 | 5.78 | |
| Artificial saliva | 0 | Negative | - | - |
| 1 × 102 | 0.95 × 102 | 94.69 | 5.29 | |
| 1 × 104 | 0.98 × 104 | 98.03 | 4.91 | |
| 1 × 106 | 0.95 × 106 | 94.66 | 3.64 |
4. Conclusion
In this study, we developed a CRISPR-based electrochemical biosensor for the ultrasensitive and highly specific detection of the vacAs1 gene of H. pylori. Compared with previous Cas12a-based signal amplification strategies, the CRISPR-based electrochemical sensor platform not only took full advantage of the specific RNA-guided recognition ability of Cas12a to achieve high specificity, but also converted target recognition into a detectable electrochemical signal, thereby enhancing sensitivity. Under optimized conditions, the detection limit of the platform for H. pylori genomic DNA was 9.5 × 10−20 M. Moreover, as a reliable platform, the CRISPR-based electrochemical biosensor can be adapted to other Cas systems through rational crRNA design, enabling the detection of various pathogenic bacteria or nucleic acids. Therefore, the CRISPR-based electrochemical sensors held promise for a wide range of nucleic acid biosensing applications, facilitating the monitoring of clinical pathogens and the development of point-of-care diagnostic systems.
Funding Statement
The author(s) declared that financial support was received for this work and/or its publication. We acknowledge the support funded by the Key Scientific Research project of Henan Higher Education Institutions (Grant No.: 25A320042), the 2025 Special Project for Research on Traditional Chinese Medicine in Henan Province (Jointly Supported by Henan Province Hospital of TCM) (Grant No.: 2025LHZX2067, 2025LHZ2012 and 2025LHZ2068) and the Henan Provincial Science and Technology Research Project (Grant No.: 252102310101).
Footnotes
Edited by: Biswajit Saha, Indian Institute of Technology Gandhinagar, India
Reviewed by: Nafiseh Fahimi-Kashani, Isfahan University of Technology, Iran
Qiaoqiao CI, Virginia Tech, United States
Data availability statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.
Ethics statement
The studies involving humans were approved by Henan Provincial Hospital of Traditional Chinese Medicine (The Second Affiliated Hospital of Henan University of Chinese Medicine) Ethics Committees. The studies were conducted in accordance with the local legislation and institutional requirements. The human samples used in this study were acquired from a by-product of routine care or industry. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements.
Author contributions
XY: Conceptualization, Writing – original draft. RZ: Formal Analysis, Writing – review and editing. LW: Data curation, Writing – review and editing. SL: Data curation, Writing – review and editing. YS: Methodology, Writing – review and editing. XL: Methodology, Writing – review and editing. CJ: Resources, Writing – review and editing. ZL: Resources, Writing – review and editing. YG: Supervision, Writing – review and editing. YL: Funding acquisition, Writing – review and editing.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fbioe.2026.1999211/full#supplementary-material
References
- Akimbekova A., Shaizadinova A., Amanzholova M., Bulashev A., Abeldenov S. (2026). Integrated RPA-CRISPR/Cas12a technology for rapid detection of Salmonella enterica . Diagn. (Basel) 16 (9), 1371. 10.3390/diagnostics16091371 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Arif M., He M., Samreen, Hays H., Wang L., Zhang X., et al. (2026). Targeted catalytic eradication of antibiotic-resistant Helicobacter pylori by a pH-activated Fe-doped nanozyme. J. Colloid Interface Sci. 718, 140494. 10.1016/j.jcis.2026.140494 [DOI] [PubMed] [Google Scholar]
- Chen H., Wang Y., Wang Z., Su W., Shao Y., Zhang G., et al. (2026). Helicobacter pylori activates histone lactylation to promote gastric cancer progression and immune evasion through the HAS2/c-MYC/PD-L1 axis. Gut Microbes 18 (1), 2661416. 10.1080/19490976.2026.2661416 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fang C., Li P., Jia D., Li Y., Zhang S., Zhang R., et al. (2026). Monoclonal stool antigen test for diagnosing Helicobacter pylori in chronic atrophic gastritis: a prospective primary care study. Front. Cell. And Infect. Microbiol. 16, 1811628. 10.3389/fcimb.2026.1811628 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gao Z., Liu G. (2026). Advancing point-of-care testing for Helicobacter pylori toward CRISPR-Cas-enabled diagnostics. ACS Sens. 11 (4), 2923–2939. 10.1021/acssensors.6c00089 [DOI] [PubMed] [Google Scholar]
- Huang T., Wu T., Xiong X., Zhao X., Zhang J., Wang C., et al. (2026). Direct/enrichment dual-mode lateral flow immunoassay enabled by a composite magnetic-fluorescent nanolabel for ultrasensitive detection of Helicobacter pylori . Mikrochim. Acta 193 (6), 373. 10.1007/s00604-026-08014-6 [DOI] [PubMed] [Google Scholar]
- Kumar R. R., Wang C. Y., Bharti A. M., Hu W. C., Zhang Y. C., Yu L. S., et al. (2025). Hierarchical graphene/Au/polyaniline nanostructured electrode for dual-modality electrochemical LAMP biosensing of Helicobacter pylori . Anal. Chem. 97 (45), 25148–25157. 10.1021/acs.analchem.5c04200 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li F., Ye Q., Chen M., Zhou B., Zhang J., Pang R., et al. (2021). An ultrasensitive CRISPR/Cas12a based electrochemical biosensor for Listeria monocytogenes detection. Biosens. and Bioelectron. 179, 113073. 10.1016/j.bios.2021.113073 [DOI] [PubMed] [Google Scholar]
- Liu C., Ji C. (2026). Screening of putative inhibitors for cytotoxin VacA from Helicobacter pylori . Sci. Rep. 16 (1), 25949. 10.1038/s41598-026-54825-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu Y., Chen S., Zhang C., Xue H., Abubakar M. U., Yin S., et al. (2026a). Asy-RPA/PCR combined with One-crRNA-CRISPR/Cas12a for simultaneous detection of multiple Clarithromycin resistance mutations in Helicobacter pylori . Nanomedicine 74, 102942. 10.1016/j.nano.2026.102942 [DOI] [PubMed] [Google Scholar]
- Liu X., Lin X., Liu Y., Yin S., Zhuoma Z., Zhang Y., et al. (2026b). Dual-mode microfluidic biosensor based on catalytic hairpin assembly synergized with bismuth-based MOF-loaded platinum nanozyme for sensitive detection of Helicobacter pylori DNA. Talanta 298 (Pt B), 129019. 10.1016/j.talanta.2025.129019 [DOI] [PubMed] [Google Scholar]
- Ma L., Liu Y., Ma Y., Wang J., Chen Y., Deng F., et al. (2026). Multiplexed Thermus thermophilus Argonaute-triggered tri-color fluorescent palette biosensing for rapid detection and genotyping of Helicobacter pylori . Chem. Eng. J. 536, 175785. 10.1016/j.cej.2026.175785 [DOI] [Google Scholar]
- Mbuthia O., Mehmood A., Mahjabeen I., Kayani M. A., Obiero J., Matsumoto T., et al. (2026). Integrated virulence-resistance profiling of Helicobacter pylori reveals context-dependent pathogenic signatures in gastric cancer. Front. Cell Infect. Microbiol. 16, 1812371. 10.3389/fcimb.2026.1812371 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Okoye C. D., Caso G. C., Cover T. L. (2026). Helicobacter pylori geographic variation: a window into global differences in gastric cancer incidence. Microbiol. Mol. Biol. Rev. 90 (3), e0020424. 10.1128/mmbr.00204-24 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rafique S., Akram R., Nasir R., Naz N., Shafique A., Bashir S., et al. (2024). An aptamer and hydroxyapatite-silver-zinc oxide-based novel electrochemical sensor for ultrasensitive H. pylori detection. Microchim. Acta 191 (12), 751. 10.1007/s00604-024-06834-y [DOI] [PubMed] [Google Scholar]
- Singla A., Jain U., Chauhan N. (2026). Bio-interfaces-based detection of gastric carcinogenic bacterium Helicobacter pylori: enabling early and high precision diagnosis. Expert Rev. Mol. Diagnostics 26 (4), 325–343. 10.1080/14737159.2026.2665802 [DOI] [PubMed] [Google Scholar]
- Skrebinska S., Megraud F., Daugule I., Santare D., Isajevs S., Liepniece-Karele I., et al. (2022). Who could be blamed in the case of discrepant histology and serology results for Helicobacter pylori detection? Diagn. (Basel) 12 (1), 133. 10.3390/diagnostics12010133 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Son Y. S., Kwon Y. H., Lee M. S., Kwon O., Jeong Y. J., Mun S. J., et al. (2025). Helicobacter pylori VacA-induced mitochondrial damage in the gastric pit cells of the antrum and therapeutic rescue. Biomaterials 314, 122842. 10.1016/j.biomaterials.2024.122842 [DOI] [PubMed] [Google Scholar]
- Song Q., Wu H., Ma Z., Huang T., Zhu X., Zhang Z., et al. (2026a). The functional and catalytic landscape of urease reveals a conserved target against Helicobacter pylori . Gut Microbes 18 (1), 2653575. 10.1080/19490976.2026.2653575 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Song X., Tang X., Tay A., Benghezal M., Marshall B. J., Tang H., et al. (2026b). Helicobacter pylori vacuolating cytotoxin A: structure, biological functions, genetic polymorphisms, and therapeutic perspectives. Biomolecules 16 (7), 1068. 10.3390/biom16071068 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang B., Gan Q., Tong Y., Qiao Y., Han M., Zhang R., et al. (2023). A visual diagnostic detection of Helicobacter pylori and the gastric carcinoma-related virulence genes (cagA and vacA) by a fluorescent loop-mediated isothermal amplification (LAMP). Talanta 256, 124260. 10.1016/j.talanta.2023.124260 [DOI] [PubMed] [Google Scholar]
- Wang P., Zhao J., Lv B., Zhang X., Li M., Chen S., et al. (2026). Genomic insights into antibiotic resistance, virulence traits and phylogenetic lineages of 141 clinical Helicobacter pylori isolates from Eastern China. Front. Cell Infect. Microbiol. 15, 1726881. 10.3389/fcimb.2025.1726881 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Xu L., Jin J., Lyu W., Liang X., Wang Q., Zhang J., et al. (2026). Point-of-care profiling of H. pylori virulence and antibiotic resistance from endoscopic biopsies using an integrated restriction enzyme-CRISPR microfluidic platform. Biosens. and Bioelectron. 299, 118438. 10.1016/j.bios.2026.118438 [DOI] [PubMed] [Google Scholar]
- Yang M., Hu J., Zhang Z., Huang L., Yu Y., Qiu D., et al. (2026). Homogeneous electrochemical sensor for sensitive detection of HBV DNA based on magnetic separation and CRISPR/Cas12a protein trans-cleavage. Talanta 305, 129676. 10.1016/j.talanta.2026.129676 [DOI] [PubMed] [Google Scholar]
- Yin S., Liu Y., Yang X., Lubanga N., Tai P., Xiong M., et al. (2025). Rapid visual detection of Helicobacter pylori and vacA subtypes by Dual-Target RAA-LFD assay. Clin. Chim. Acta 564, 119927. 10.1016/j.cca.2024.119927 [DOI] [PubMed] [Google Scholar]
- Zan J., Guo W., Tan W., Lan J., Li T., Huang Y., et al. (2026). High-affinity monoclonal antibody pairs-based time-resolved fluorescent microsphere-lateral flow and electrochemical immunosensing platforms for rapid and quantitative detection of Helicobacter pylori in saliva. Talanta 308, 129920. 10.1016/j.talanta.2026.129920 [DOI] [PubMed] [Google Scholar]
- Zheng J., Wen Z., Li Y. (2026). Rapid detection of Enterococcus faecalis using RPA-CRISPR/Cas12a-assisted technology. Microbiol. Spectrum. 14 (7), e0016826. 10.1128/spectrum.00168-26 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhu Y., Xu W., Li Y. (2026). Dual mechanisms of crRNA 3'-extension-mediated Cas12a attenuation enable programmable one-pot CRISPR diagnostics. Anal. Chemistry. 98 (21), 15830–15842. 10.1021/acs.analchem.6c01964 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.