Abstract
Background:
The auditory brainstem response (ABR) is an essential tool for assessing the hearing sensitivity of newborns and other patients who are unable to provide behavioral responses. During a typical ABR session, clinicians rely on extensive training and expertise to rapidly identify the hearing threshold of the patient by adaptively optimizing the acoustic stimuli presented to the patient. Nevertheless, ABR sessions are notoriously long, which may lead to an incomplete diagnosis or necessitate additional appointments, which may impose a significant burden on the caregiver. Furthermore, clinical best practice requires clinicians to identify ABRs based on visual inspection of the recordings, which may introduce errors that lead to inaccurate diagnoses.
Purpose:
The long-term goal of this study was to develop a clinical decision support system (CDSS) to enable faster and objective ABR threshold estimation. This study sought to validate an algorithm that suggests optimal stimuli based on accruing data during an ABR session.
Research Design:
Two estimates of ABR thresholds at 500, 1000, 2000, and 4000 Hz were measured. The first estimate followed the algorithm and the second estimate followed clinical best practice. Behavioral audiograms were also measured.
Study Sample:
Fifty adult ears with typical hearing and hearing loss up to 80 dB HL.
Data Collection and Analysis:
ABRs were recorded using the Interacoustics Eclipse platform in a sound-attenuating booth. The two ABR thresholds were compared via the mean absolute difference and mean bias. The intraclass correlation coefficient (ICC) between the two ABR thresholds and behavioral thresholds were computed.
Results:
We found that 72 percent of the ears had a mean absolute difference of approximately 10 dB or less and 96 percent had a mean absolute difference of approximately 20 dB or less. The ICC for the algorithm and clinician threshold estimates was 0.78, indicating good agreement.
Conclusions:
This algorithm can estimate ABR thresholds with clinically acceptable accuracy for a majority of ears. Additional development and validation are necessary before the algorithm is clinically viable.
Clinical Relevance Statement:
A CDSS that suggests optimal stimuli for rapid and objective ABR threshold estimation would permit shorter clinical sessions, allow clinicians to spend more time on other aspects of clinical care, and expand access to rural areas that do not have dedicated audiology clinics.
Keywords: audiogram, auditory brainstem response, Bayesian adaptive algorithm, clinical decision support system, Gaussian process model, threshold
INTRODUCTION
The auditory brainstem response (ABR) is a physiological assessment of hearing sensitivity and a fundamental component of diagnostic hearing-health care for newborns and other populations who are unable to respond to behavioral audiometry. However, the procedure requires a high level of clinical expertise and is time-intensive. The goal of the current experiment was to validate a time-efficient ABR threshold estimation algorithm among a population of adults with typical hearing and those with hearing loss.
The ABR is evoked by the presentation of an acoustic stimulus and can be measured noninvasively through scalp electrodes (Davidson et al., 1990). Both wide-band (e.g., click) and narrow-band (e.g., tone-burst, narrow-band chirp) stimuli may be presented at different levels. Typically, a measured ABR comprises a time-domain waveform that is the culmination of hundreds to thousands of time-synchronous averages that exhibit characteristic peaks and troughs associated with specific neural generators. These features (e.g., wave V) typically decrease in amplitude when the stimulus level is decreased, such that an ABR threshold may be defined as the lowest stimulus level for which a particular wave feature may be observed above the baseline noise. For narrow-band stimuli, the ABR threshold can be measured across frequencies, and this threshold contour may be used as a proxy for behavioral audiograms during diagnosis and prescription in hearing loss interventions (Hall, 2007; Neely et al., 1988); however, discrepancies between ABR and behavioral audiograms do exist (Hall, 2007).
The ABR is a critical tool for diagnosing infant patients who have not passed their newborn hearing screening. In a typical session, the caregiver and infant are in a sound-attenuating booth. The clinician seeks to estimate the ABR threshold at multiple frequencies generally including 500, 1000, 2000, and 4000 Hz. However, ABR testing is notoriously difficult because of the very weak target signal. During ABR testing, an averaged waveform usually requires approximately 2 to 3 minutes to resolve, and any excessive movement or vocalization of the infant may require testing to be paused or ended. A typical session measuring eight ABR thresholds (four frequencies and two ears) can require approximately 60–90 minutes (American Speech-Language-Hearing Association, 2023). Frequently, the clinician may not be able to provide a complete diagnosis during a single session; therefore, additional appointments are necessary to avoid risking an incomplete diagnosis. These scenarios impose a significant burden on caregivers and are a barrier to accessing essential hearing-health care.
Many approaches have been devised to decrease the time required to measure the ABR threshold. Most of these have sought to minimize the amount of time required to resolve an averaged waveform for one stimulus condition (frequency and level). One common method is to define a statistical metric, such as the single or multiple point variance (Elberling & Don, 1984) that tracks the variance of one or more points in the time domain waveform during synchronous averaging. This provides a confidence interval for the presence of an ABR and can be used to adaptively determine the minimum number of averages required per stimulus condition. Other methods involve presenting multiple stimuli simultaneously, for example, by interleaving frequencies (Buran et al., 2020) or presenting stimuli to both ears simultaneously (Polonenko & Maddox, 2019). These techniques, which seek to minimize the time required to resolve one or more averaged waveforms, can reduce the measurement time for a single waveform by a factor of two or three. Many other approaches to reduce testing time have been proposed in the literature (Sininger et al., 2020).
Another approach is to minimize the total number of stimuli required to estimate the threshold. In clinical practice, standard ABR threshold estimation protocols continue to require highly trained clinicians for accurate and efficient diagnoses. To expedite the session, experienced clinicians typically determine their stimulus choice adaptively based on the waveform characteristics (amplitude and latency) of the data as it is collected, often referred to as sequential testing. For example, if a robust response is observed for a given stimulus level, then the clinician may choose to decrease the subsequent stimulus level by a larger interval than that if the response had been weak. In practice, this saves time by reducing the number of stimuli presented at suprathreshold levels and instead prioritizes presenting stimuli at or near the threshold. However, discerning an observable waveform near the ABR threshold is challenging, and within and between clinician estimates are known to have high variability, thus introducing potential subjectivity and bias (Vidler & Parker, 2004).
Numerous approaches for objective ABR detection have been developed over the years. Some are statistical approaches based on the F-ratio or other signal-to-noise estimates, such as the aforementioned single point variance method (Elberling & Don, 1984). In a previous study, three objective measures based on correlation, variance ratio, and multiple pre–post z-tests were evaluated; however, they were not as sensitive as expert visual inspection (Arnold, 1985). More recently, cross-correlation of neighboring levels was found to be a good method for objective ABR threshold estimation (Suthakar & Liberman, 2019). However, this technique requires that ABRs are measured for adjacent levels (e.g., every 5 or 10 dB), which makes it best adapted for laboratories that measure ABRs in animal models following a predetermined list of levels. Additional approaches that compare ABR waveforms against an idealized waveform template, with the ABR threshold defined by a statistical measure such as cross-covariance, have been developed (Tanaka et al., 2023). However, many of these objective threshold estimations were developed as post hoc tools and perform better when more data are available, whereas clinical constraints favor reducing the amount of data necessary for determining ABR thresholds.
Gaussian process (GP) models and Bayesian adaptive procedures are a promising approach to time-efficient and objective estimation of physiological responses such as ABRs (Pillow & Park, 2016). In recent years, several such procedures used a GP model to derive interim ABR thresholds paired with a Bayesian adaptive procedure to optimize subsequent testing (Chesnaye et al., 2023, 2025; Petersen & Shen, 2024).
The GP models are probabilistic and do not make a priori assumptions about waveform morphologies as do parametric-based or template-based models. Instead, GP models are fitted on the basis of the covariance between data points throughout the stimulus space, for example, time, frequency, and level in the case of ABRs. Because of the assumed covariance, a sparse sampling of the stimulus space may be sufficient to provide a faithful representation of the ABR amplitude across the entire space. When the GP model has been fitted, the desired characteristics of the system (e.g., ABR threshold) may be derived from model predictions, thereby reducing the total amount of required data for a given experiment. To further increase efficiency, GP models are often paired with a Bayesian adaptive sampling strategy that uses model predictions to inform optimal subsequent stimulus selection for maximum information gain as data continue to be collected.
In a recent study, a GP model and adaptive sampling strategy were used to estimate the ABR threshold for a single frequency at one time (Chesnaye et al., 2023). In this configuration, the GP model was fitted to the amplitudes of the averaged ABR waveforms collected thus far at various levels for the same test frequency. An optimal subsequent stimulus was chosen to minimize the uncertainty along an interim threshold estimate. This approach was shown to reduce the testing time by a factor of two in a simulation experiment. The authors then developed a second model based on the previous study by exploiting the covariance between level, frequency, and wave V amplitude (Chesnaye et al., 2025). This model was validated among a cohort of 31 adult participants with and without hearing loss; the GP-predicted ABR threshold was compared against human raters following clinical best practice and behavioral audiograms. The results demonstrated good agreement (6.5 dB mean absolute difference) between GP and behavioral audiograms, again reducing testing time by a factor of two.
Concurrently, the authors of the present article developed a similar ABR threshold estimation algorithm (Petersen & Shen, 2024). Like that reported by Chesnaye et al. (2025), this algorithm pairs a GP model and Bayesian adaptive sampling strategy to leverage covariance across frequency and level. Unlike that reported by Chesnaye et al. (2025), the GP model is fitted to, and subsequently predicts, the time-domain ABR waveform voltage trace rather than the wave V amplitude growth function. This algorithm was previously validated on datasets comprising four animal species in a simulated experiment and demonstrated the robustness of the GP model framework for a diverse set of waveform morphologies.
In the current study, the algorithm described by Petersen & Shen (2024) that was designed for autonomous threshold assessment in laboratory animals has been configured specifically for clinical ABR assessments in humans. In its current implementation, the algorithm operates as a clinical decision support system (CDSS). The CDSS paradigm has been used across various medical disciplines since the 1980s and leverage data to enhance objectivity in the clinician’s decision-making strategy (Sutton et al., 2020). Often, a CDSS will use machine learning based on phenomenological or data-driven models coupled with a patient’s history and medical condition to provide probable diagnoses or statistically optimal intervention. In this paradigm, the CDSS does not replace the clinician; instead, it contributes objective data to be used in conjunction with clinical expertise. In the current CDSS-based ABR threshold estimation protocol, the adaptively chosen stimuli conditions are suggested to the clinician based on the accrued data and model predictions. The long-term goal is to combine clinician expertise with machine learning approaches to improve the time-efficiency and objectivity of ABR testing. In the current experiment, the baseline performance of the ABR threshold estimation algorithm with minimal interference by the clinician was assessed.
The algorithm was validated among a cohort of adults with typical hearing and hearing loss. Adult participants were chosen because they are able to respond to behavioral testing of audiograms and match the participant population of a similar recent experiment (Chesnaye et al., 2025). In the current experiment, ABR thresholds were first estimated following the stimuli adaptively chosen by the algorithm. In this scenario, the clinician was tasked with monitoring the quality of the data, for example, by ensuring sufficient waveform averages and pausing data collection during excess noise events. Afterwards, the clinician followed clinical best practice to obtain ABR thresholds. It was hypothesized that the algorithm-estimated and clinician-estimated ABR thresholds would agree within a clinically acceptable tolerance, and that the algorithm-derived ABR thresholds would correlate with behavioral thresholds.
METHODS
Bayesian Adaptive Algorithm
Clinician–Algorithm Interaction
In the current experiment, the ABR threshold estimation CDSS protocol was seamlessly integrated into a typical ABR measurement setup. The clinician used a dedicated commercially available ABR acquisition system to present acoustic stimuli to participants and recorded their ABRs inside a sound-attenuating booth. Independently, a second computer located outside the booth was continuously running the ABR threshold algorithm (custom program written in MATLAB). After each averaged ABR was obtained using the commercial software, the clinician exported the newly acquired waveform to a folder saved to a server accessible to both computers using the export functionality of the research module. The MATLAB program running on the second computer continuously monitored the shared folder for a new waveform; when one was saved there, the MATLAB program read it in, and the GP model was refitted with all the existing data, including the most recent waveform. After fitting the GP model, a suggested subsequent stimulus was displayed on a screen visible to the clinician through the booth window, who then presented the new stimulus to the participant. This CDSS configuration was chosen because it requires no modifications to, or computational resources from, the existing clinical ABR system. Furthermore, this protocol allowed the clinician to monitor the data quality in real-time and pause data collection during excess noise events to ensure sufficient synchronous averaging of the ABR waveforms.
Description of the Algorithm
The algorithm paired a GP model with a Bayesian adaptive sampling strategy to iteratively estimate the ABR threshold (Figure 1). For this study, the following vocabulary was defined when describing the algorithm:
Figure 1.

Schematic of the adaptive algorithm. The loop shown between “display stimulus” and “choose subsequent stimulus?” comprises a single iteration. One run of the algorithm is complete upon arrival at “final threshold estimate” at the bottom.
Stimulus: a given frequency and level pair
Waveform: the time-synchronously averaged time-domain voltage trace following the presentation of a stimulus that may or may not contain archetypal ABR characteristics such as wave V
Iteration: the presentation of a stimulus, acquisition of the averaged waveform, fitting of the GP model with all waveforms collected thus far, and either displaying the adaptively determined subsequent stimulus or terminating the procedure and determination of the final ABR threshold estimates
Run of the algorithm: conducting the full set of iterations from the first stimulus condition until the termination criteria have been achieved.
Holistically, the algorithm operated by fitting a model to the averaged ABR waveforms measured in response to a small number of stimuli conditions (frequency and level), and it used the model to predict the ABR waveforms across the full stimulus space covering all relevant frequencies and levels. Estimated ABR thresholds were then derived from the model predictions, which may occur at stimuli conditions that have not yet been tested. Additional ABR waveforms were measured at stimuli conditions chosen to reduce the model uncertainty and, thus, improve the ABR threshold estimates.
GP Model
A GP model is a nonparametric statistical model that accounts for covariance between data points across the stimulus space. Kernels are simple mathematical functions used to encode the covariance relation between variables for a specific dimension and may be chosen to impose assumptions about how data points in the stimulus space may covary. The kernels are defined by hyperparameters, or the kernel function constants, which can be fitted to any existing data, typically comprising a subset of the entire stimulus space. A fitted GP model (e.g., with hyperparameters that were fitted based on existing data) may be used to predict responses (e.g., ABR waveforms) for novel stimuli that are not included in the training set.
The GP model in the current article is trained on the ABR amplitudes (in µV) as a function of time, frequency, and level. The fitted model can predict the ABR waveform for a given pair of level and frequency. The time and frequency dimensions were both modeled using the squared exponential kernel. The level dimension was modeled by a composite kernel structured as the sum of the linear and squared exponential kernels. The latter was chosen to encode amplitude growth with level while providing enough flexibility to capture floor and ceiling effects; see Petersen & Shen (2024) for additional details. The GPML toolbox (Rasmussen & Nickisch, 2010) was used for all GP modeling in the current article.
Adaptive Algorithm
The algorithm began with an initialization stage during which waveforms were collected for a small number of preselected stimuli. After the initialization stage was complete, the adaptive section followed three steps: (1) the GP model was fitted to all averaged waveforms collected so far and used to predict waveforms across the stimulus space; (2) an interim ABR threshold contour across frequencies was determined; and (3) either a subsequent stimulus was chosen or the interim ABR threshold became the final ABR threshold estimate depending on the termination criteria.
Initial Stimuli.
The stimuli used in the initialization stage were chosen to provide the GP model with archetypal versions of the individual’s ABRs; therefore, they typically include examples across the frequency range of interest and at expected suprathreshold levels. The primary objective of this step was to provide sufficient data to ensure that the first fitted GP model captured global characteristics of the dataset while minimizing time spent collecting ABRs following predetermined list of stimuli and instead prioritized time spent on the adaptive portion of the algorithm.
Threshold Estimate.
The ABR threshold was defined as the lowest level that exceeds a minimum tolerance for the peak-to-peak amplitude of the predicted ABR waveform within a predefined time window following the onset of the stimulus. The fixed tolerance was defined as the maximum peak-to-peak value, which is the maximum voltage value minus the minimum voltage value in the time window (Chesnaye et al., 2025). Generally, a lower tolerance leads to a lower ABR threshold estimate and is one of the parameters that must be determined based on the characteristics of the ABR acquisition system and the desired confidence in detecting a response to the stimulus (Petersen & Shen, 2024).
Choice of Subsequent Stimulus.
Each subsequent stimulus was chosen to be the frequency and level along the ABR threshold contour for which the GP model prediction had the highest variance.
Hyperparameter Initialization.
Each time the GP model was fitted, a set of hyperparameter initialization values was included in the optimization function (Pillow & Park, 2016). These values may be chosen based on expected covariance between datapoints across the stimulus space. In this algorithm, the initialization values were chosen from a random distribution defined by upper and lower bounds. Each hyperparameter initialization value was chosen from a unique distribution, and a new value was drawn each time the GP model was fitted. The upper and lower bounds from which the values were chosen were determined by Petersen & Shen (2024) using animal ABRs and independently validated on adult human ABRs in pilot studies. The final optimized hyperparameter values were generally stable across a wide range of initialization values, stimuli, and recording parameters (Petersen & Shen, 2024).
Experimental Procedure
Participants
The current experiment was approved by the Institutional Review Board at the University of Washington (IRB ID STUDY00015828). Thirty-one adults (16 female and 15 male participants) from 18 to 94 years of age were recruited for the current experiment. Older adults (n = 25; average, 72 years; minimum, 56 years; maximum, 94 years) in this experiment were recruited from a roster developed during a different study in the same laboratory and selected to include varying degrees of hearing loss (Kursun et al., 2025). Younger adults (n = 6; average, 21 years; minimum, 18 years; maximum, 23 years) were recruited from the general public for the purpose of including participants with typical hearing.
Nineteen of the participants completed testing for both ears, whereas one ear was tested for the remaining participants, resulting in a total of 50 ears tested. Each participant completed signed consent, otoscopy, tympanometry, and standard air and bone conduction testing from 250 to 8000 Hz (Figure 2). Air conduction thresholds were tested using ER3A insert phones with electroacoustic properties similar to those used for the ABR measurements. Six of the participants had typical hearing (<20 dB HL at all frequencies), and 25 participants had mild to profound hearing loss (Clark, 1981). The hearing loss group included 23 participants with sensorineural hearing loss and two participants with mixed hearing loss. Three participants had asymmetric hearing loss (American Academy of Otolaryngology–Head and Neck Surgery Foundation, 2021) with a maximum pure tone average asymmetry of 38.6 dB.
Figure 2.

Behavioral audiograms of all participants shown in gray (air conduction only). The mean and standard deviation are depicted in black. Hearing impairment ranged from typical hearing to profound hearing loss.
ABR Acquisition System
The experiment used the commercially available Interacoustics Eclipse EP25 ABR platform. Stimuli were presented via a RADIOEAR IP30 insert earphone, and responses were measured using Ambu Neuroline 720 surface electrodes. Narrow-band CE-chirp stimulus with rarefaction polarity was used for all four frequencies and presented at rates of 21.7 chirps per second for 500 and 1000 Hz and 27.7 chirps per second for 2000 and 4000 Hz (Elberling & Don, 2010). Possible stimulus levels used in this study ranged from 20 to 90 dB nHL in increments of 5 dB. The entire stimulus space included 60 independent frequency/level combinations that may be chosen by the algorithm at each iteration.
Algorithm Configuration
Initial Stimuli.
The initial stimuli used in this experiment were 80 dB nHL at 500, 1000, 2000, and 4000 Hz, with an additional 2000 Hz stimulus at 50 dB nHL. Pilot testing revealed that including high levels for all frequencies in the stimulus space, in addition to one or several mid-levels, provided a sufficient set of initial data for fitting the GP model.
Threshold Estimate.
The time window used to search for an ABR was 5–15 ms after stimulus onset, which is likely to include wave V across a large range of possible latencies (Ormundo et al., 2024). A peak-to-peak tolerance of 0.15 µV was established during pilot testing to minimize the deviation between the algorithm and expert clinician threshold estimates.
Termination Criteria.
The algorithm was terminated after 15 total waveforms were collected, and the GP model was fitted for the final time. These 15 waveforms included those from the five predetermined initial stimuli plus 10 adaptively chosen stimuli. These can be measured at any of the four frequencies and 15 levels, for a total of 60 total possible conditions. Simulated experiments and pilot studies indicated that 15 waveforms were sufficient to reduce the average deviation between the algorithm-estimated and clinician-estimated thresholds <10 dB for the majority of ears.
Hyperparameter Initialization Values.
The initialization ranges for the GP model hyperparameters are shown in Table 1. These values were chosen based on simulation and pilot experiments.
Table 1.
Hyperparameter Initialization
| Time | Level | Frequency | ||||
|---|---|---|---|---|---|---|
| Sq. exp. | Lin. | Sq. exp. | Sq. exp. | |||
| Log(ℓ) | σ | Log(ℓ) | Log(ℓ) | σ | Log(ℓ) | σ |
| −10, −8 | −5, −1 | 2, 4 | 2, 4 | 1, 2 | −2, 0 | 1, 3 |
Hyperparameters and ranges from which each hyperparameter was randomly drawn to initialize the Gaussian process (GP) model fitting procedure. The squared exponential (Sq. exp.) is defined by k(xp, xq) = σ2 exp(-|xp - xq|2/(2ℓ2)) and the linear kernel is defined by k(xp, xq) = (xp xq)/ℓ2, thus relating the covariance between arbitrary data points xp and xq (Petersen & Shen, 2024).
Additional Data Processing.
The waveforms were first processed by the Eclipse ABR platform. Artifact rejection was set at 39.6 µV and the recorded waveforms were processed with Bayesian weighting after applying high-pass and low-pass filters with cutoff frequencies at 33 and 5000 Hz, respectively. In all cases, final waveforms included a minimum of 1,000 averages.
Next, the time-domain waveforms were preprocessed in preparation for use in fitting the GP model. First, the traces were scaled to units of µV. Next, they were downsampled from 15000 to 6000 Hz, which is a sampling rate that is sufficiently high to capture the significant ABR characteristics while reducing computational load required to fit the GP model. In addition to windowing the ABR threshold detection search range from 5 to 15 ms, frequency-specific and level-specific normative latencies were used to time-shift the ABR waveforms to better time-align wave V across levels and frequencies (Neely et al., 1988; Rasetshwane et al., 2013) (Equation 3 and Table 3 for the 4-ms stimuli option). Approximate time alignment of wave V using normative values improved the GP model predictions in a simulated experiment. Reverse time-shifting is applied to the model-predicted waveforms to recuperate the original latencies.
ABR Measurement Protocol
Disposable electrodes were mounted on each participant’s head in a two-channel (Fz, A1, A2, Fpz) montage. Electrode sites were cleaned with electrode prep pads and NuPrep electrode gel when applicable. Inter-electrode impedances were <3 kΩ throughout the experiment. During the ABR measurement session, participants were reclined in a comfortable chair located in a dim sound-attenuating booth. The clinician (B.C.) was co-located in the booth to monitor for excess movement. Waveform averaging was determined online by the clinician following clinical experience.
For each participant, the clinician first followed the algorithm-suggested stimuli, yielding a final algorithm-estimated ABR threshold following the 15th iteration. After the algorithm run was complete, the clinician, who was blinded to the algorithm-estimated threshold, adhered to best-practice guidelines for ABR threshold estimation and obtained additional waveforms as necessary. Best-practice guidelines included focused attention to wave V (most likely highest peak and lowest trough relationships) within an estimated range of 1–12 ms, with repetition of waveforms to confirm the response (British Society of Audiology, 2025). This protocol yielded two ABR threshold estimates: the algorithm estimate using a fixed number of 15 stimuli and the clinician estimate that required a variable number of additional waveforms. The minimum algorithm-estimated and clinician-estimated estimated thresholds were fixed at 20 dB nHL, corresponding to typical hearing.
The algorithm and clinician estimates were first obtained for the left ear. The procedure was repeated for the right ear when time allowed, typically depending on the participant’s comfort and level of background noise. The order of threshold estimates (algorithm estimate followed by the clinician estimate) allowed the clinician to use waveforms collected during the algorithm portion to inform their clinical threshold estimate, thereby reducing total testing time for a given ear. This allowed for testing both ears on many of the participants in a single session.
Although the protocol was configured as a CSDD, the current experiment was designed to validate the algorithm performance. The clinician’s responsibility during the algorithm-driven portion of the data collection was limited to ensure quality control of the averaged waveforms.
RESULTS
Comparison of ABR Thresholds
The performance of the algorithm was evaluated by comparing the algorithm-estimated ABR threshold to the clinician-estimated ABR threshold. Because of the monaural configuration of the algorithm, all analyses were completed for each ear and independent of the other when both ears were tested for a single participant. The algorithm-estimated and clinician-estimated ABR thresholds as well as the behavioral thresholds are shown in Figure 3.
Figure 3.

Auditory brainstem response (ABR) thresholds for each participant estimated by the algorithm (dashed black), estimated by a clinical audiologist (solid blue), and estimated by the behavioral audiogram (dotted red). Both ABR thresholds are shown in units of dB nHL and the behavioral audiogram threshold is shown in units of dB HL. In-panel annotations provide the mean absolute difference between the clinician-estimated and algorithm-estimated ABR thresholds averaged over frequency.
Two general metrics were defined: the bias and the absolute difference. The bias is the frequency-by-frequency difference between the clinician-estimated and algorithm-estimated ABR thresholds. The absolute difference is the magnitude of the bias. The absolute difference may be interpreted as a global measure of agreement between the two estimates, while the bias may point to systematic overestimation or underestimation by the algorithm. In general, these values may be reported for individual frequencies, averaged over frequency for a given ear, averaged over all ears for a given frequency, or averaged over frequency and ears, depending on the desired granularity of the data. Absolute differences ≤10 dB may be considered a clinically acceptable estimate in many cases (Beahan et al., 2012). Absolute differences of 10–20 dB, although not optimal, are within the range of inter-rater estimates of ABR thresholds in many cases (Vidler & Parker, 2004).
In addition to absolute difference and bias, the algorithm-estimated and clinician-estimated thresholds were compared using the intraclass correlation coefficient (ICC) calculated for all frequencies and all ears. The ICC was calculated using a two-way, mixed-effects, absolute agreement model and denoted as ICC(A,k) (McGraw & Wong, 1996). The ICCs with values of <0.5, 0.5–0.75, 0.75–0.9, and >0.9 indicated poor, moderate, good, and excellent reliability, respectively (Koo & Li, 2016).
The absolute difference and bias, averaged over all frequencies and ears, were 7.5 dB and 1.9 dB, respectively, indicating good agreement (e.g., <10 dB) between the clinician-estimated and algorithm-estimated thresholds when the dataset was considered in its entirety. The absolute difference and bias for each frequency (0.5, 1.0, 2.0, and 4.0 kHz), averaged over all ears, were 6.2, 7.2, 7.5, and 8.9 dB, respectively, and 0.0, 0.0, 2.3, and 5.1 dB, respectively.
Of the 50 ears, 36 showed mean absolute differences across frequency ≤10 dB, and 48 showed differences ≤20 dB. This demonstrated that the clinician-estimated and algorithm-estimated thresholds absolute differences were within clinically acceptable agreement for 72 percent of the ears and within clinically observed inter-rater differences for 96 percent of the ears. Similarly, when bias was averaged over frequencies, 37 and 48 ears had absolute values ≤10 dB and ≤20 dB, respectively, demonstrating no significant bias between the clinician-estimated and algorithm-estimated threshold estimates.
The distribution of absolute difference and bias averaged over frequency are shown in Figure 4. The absolute difference distribution was skewed toward 0 dB, with decreasing occurrences of larger absolute difference values. The bias distribution was centered at approximately 0 dB, with relatively symmetric positive and negative tails. This indicated that the algorithm did not have a significant systematic overestimation or underestimation of the ABR threshold.
Figure 4.

Histograms of the mean differences between the clinician-estimated and algorithm-estimated thresholds averaged over frequency for each ear. The mean absolute difference and mean bias difference are depicted in the top and bottom panels, respectively.
The absolute difference and bias were also assessed on a per-frequency basis. For 500, 1000, 2000, and 4000 Hz, 43, 38, 40, and 37 of the ears, respectively, had an absolute difference ≤10 dB, and 47, 48, 46, and 42 of the ears, respectively, had an absolute difference ≤20 dB or less. The bias values for 500, 1000, 2000, and 4000 Hz averaged over ears were 0, 0, 2.3, and 5.1 dB, respectively, indicating that the algorithm underestimated clinician thresholds slightly at 2000 and 4000 Hz.
The algorithm-estimated and clinician-estimated thresholds had an ICC of 0.78, indicating good agreement (95 percent confidence interval, 0.70–0.83).
In addition to a group-level analysis, the mean absolute difference and bias were separately calculated for the typical hearing group and hearing impairment groups. For this analysis, typical hearing was defined as behavioral thresholds <20 dB HL at 500, 1000, 2000, and 4000 Hz. The mean absolute difference and bias values for the typical hearing and hearing impairment groups were 3.9, −3.6, 8.5, and 3.4, dB, respectively. This indicated that the algorithm-estimated and clinician-estimated ABR thresholds matched better for the typical hearing group.
Convergence of the Algorithm-Estimated ABR Threshold with the Clinician-Estimated ABR Threshold
The adaptive algorithm was designed to improve the ABR threshold estimate as additional waveforms were added to the data used by the GP model. To examine this convergence, the absolute difference and bias were evaluated as a function of algorithm iteration (Figure 5). Note that no predictions are shown for n < 5 because the GP model was not fitted until after the five initialization stimuli had been collected.
Figure 5.

Convergence of the algorithm-estimated auditory brainstem response (ABR) threshold and the clinician-estimated ABR threshold after each iteration of the algorithm (i.e., number of waveforms used to train the model). The top panel shows the absolute difference (black) and bias (blue) between estimates averaged across all participants and frequencies. Error bars denote standard deviation from the mean. The bottom panel shows the absolute difference and bias between estimates for each frequency individually. Solid, dashed, dashed-dotted, and dotted lines denote 0.5, 1.0, 2.0, and 4.0 kHz, respectively.
Figure 5 shows the absolute difference and bias averaged over all frequencies and ears, representing the aggregate convergence of ABR threshold estimates for the dataset as a whole (50 ears with 4 frequencies per ear, yielding 200 data points per iteration). As expected, the agreement between the algorithm-estimated and clinician-estimated thresholds improved as more data were supplied to the GP model. The mean absolute difference was <10 dB by the time the ninth waveform was included. Similarly, the mean bias began with a large magnitude that trended toward zero as more data were collected and stayed within ± 2 dB after the eighth waveform was included. The absence of a significant bias indicated that the ABR threshold tolerance value was well-suited for this dataset when averaged across frequencies.
Figure 5 also shows frequency-specific absolute difference and bias. The absolute difference and bias indicated that agreement between the clinician-estimated and algorithm-estimated thresholds improved as more data were included. However, by approximately the tenth waveform, there was a systematic underestimation of thresholds at higher frequencies that led to increased absolute differences. This indicated that the same peak-to-peak tolerance may not be suitable to define the ABR threshold across different frequency bands.
Correlation Between Behavioral and Algorithm-Estimated ABR Thresholds
The algorithm-estimated and clinician-estimated ABR thresholds may also be compared to the behavioral thresholds. However, it is known that behavioral thresholds are typically lower than ABR thresholds, which may be modified using different correction factors based on the measurement hardware, protocol, age, and hearing loss configuration (Bagatto et al., 2024). The ABR thresholds were compared with behavioral thresholds without applying correction factors because of the different heterogeneous configurations of hearing loss in the participant cohort.
The behavioral thresholds and clinician-estimated ABR thresholds had an ICC of 0.85 (95 percent confidence interval, 0.68–0.92), which indicated good reliability of the clinician-estimated ABR threshold estimates. The behavioral thresholds and algorithm-estimated ABR thresholds had an ICC of 0.63 (95 percent confidence interval, 0.5–0.73), indicating moderate reliability.
Further, the per-frequency bias values at 500, 1000, 2000, and 4000 Hz averaged over all ears were −10.0, −8.3, 0, and 2.5 dB, respectively. This bias indicated that, especially at 500 and 1000 Hz, the algorithm predicted thresholds approximately 10 dB higher than those measured behaviorally. This systematic difference between behavioral and ABR thresholds was expected and the values were consistent with commonly applied correction factors found in the literature (Bagatto et al., 2024; Ferm et al., 2013; Ferm & Lightfoot, 2015). Agreement between the bias reported here and normative correction factors does not account for pathology-specific discrepancies between behavioral and ABR thresholds expected for certain configurations of hearing loss, including severity (McCreery et al., 2015). Furthermore, there is currently no consensus regarding the correct application of correction factors when normalizing ABR thresholds to match behavioral thresholds (Norrix & Velenovsky, 2017).
DISCUSSION
The adaptive algorithm estimated ABR thresholds within a 10-dB mean difference compared with those estimated by an experienced clinician for 36 of 50 ears while using approximately one-half of the stimulus conditions required in a typical clinical setting.
Agreement with Previous Studies
A similar algorithm that pairs a GP model with an adaptive sampling strategy for efficient and objective ABR threshold detection has been recently validated (Chesnaye et al., 2025). Similar to the algorithm described in the current study, that algorithm was designed for sequential testing in which interim threshold estimates are generated during data collection rather than during post hoc analyses. It was validated among a cohort of adults with typical hearing and hearing loss with a mean absolute difference of 6.5 dB compared with a mean absolute difference of 7.5 dB in the current study.
There were several differences between our study and that by Chesnaye et al. (2025). Chesnaye et al. (2025) reported a time reduction of approximately 50 percent compared with that associated with visual examiners. However, their time-savings metric was determined by the total number of preprocessed and artifact-free epochs used to estimate threshold per frequency. In contrast, the algorithm in the current study was run until to a fixed number (15) of stimuli and a variable number of epochs per averaged waveform was determined in real-time by the audiologist. Therefore, a direct comparison of time-savings for the two algorithms was not possible.
An additional difference between the algorithm reported by Chesnaye et al. (2025) and the algorithm in the current study is that the GP model in their study was designed to model the wave V amplitude growth function with stimulus level, whereas the GP model in the current study modeled the time-domain voltage trace. This time-domain trace implicitly included the wave V amplitude growth function in addition to other waveform morphology characteristics such as wave V latency.
Algorithm Limitations
The algorithm presented in the current study exhibited some limitations that must be addressed. One such limitation was the peak-to-peak tolerance used to define the presence or absence of a response in the time-domain waveform. The peak-to-peak amplitude was defined by the maximum and minimum voltages in the 5- to 15-ms window after stimulus onset, which is often, but not always, dominated by wave V. This definition was occasionally insufficiently sensitive for the identification of wave V peaks with small amplitudes, which are more easily identified by experienced clinicians in conjunction with other waveform characteristics, such as latency trends across stimulus levels. Additionally, the peak-to-peak tolerance was sometimes exceeded by the early portions of non-ABR generators, such as the postauricular muscle reflex or the middle latency response. In these cases, the algorithm may interpret the non-ABR waveform features as a response, artificially lowering the algorithm threshold estimate. To mitigate this risk, it may be best to use earlobe electrode placement to avoid postauricular muscle responses altogether. However, a more sophisticated definition of the presence of an ABR, rather than peak-to-peak tolerance, may be more robust against wave characteristics caused by non-ABR generators.
Similarly, the algorithm was validated using a narrow set of stimulus and data processing configurations. Specifically, the CE-chirp narrow-band stimuli used in this experiment are known to produce larger wave V amplitude responses in some cases because of increased synchronization of neural responses (Ferm et al., 2013). Other stimuli types may require additional modifications of the peak-to-peak tolerance used to define the presence or absence of an ABR. Additionally, data processing schemes such as the averaging method (unweighted, Bayesian, Kalman-weighted, etc.) may impact the resulting averaged waveforms, which may alter the ABR threshold estimated by the algorithm. These implementation details remain to be tested. However, provided that rigorous artifact rejection is applied during averaging, specific averaging choices should not have a significant impact on the results; instead, they should be accounted for during future fine-tuning of the algorithm.
An additional limitation observed in this experiment was the occasional overfitting of the GP model to the measured data. This typically occurred when few of the averaged waveforms used to train the GP model contained a significant ABR, and the model was instead fitted to waveforms dominated by noise. In these cases, the GP predictions matched the measured data very closely and predicted weak responses (i.e., low-voltage amplitudes) for stimuli conditions that have not been measured, resulting in a nonmonotonic growth function with the stimulus level. However, this does not always negatively impact the threshold estimate, which was defined as the lowest level that exceeds a peak-to-peak tolerance. Dips in predicted amplitudes for higher levels are irrelevant when the peak-to-peak tolerance has been exceeded at lower levels. Nevertheless, enforcing a monotonic, or nearly monotonic, growth function with the stimulus level may improve the validity of the GP model.
Finally, the algorithm created stimulus suggestions solely based on the averaged waveforms collected previously in the session and did not account for participant characteristics such as age, sex, self-reported hearing difficulties, or experience with hearing aids. Experienced clinicians may also account for individual patient profiles while conducting ABRs that go beyond waveform characteristics, which may inform their interpretation of waveform features and stimulus sampling strategies. Additional considerations include objective measures such as immittance and otoacoustic emissions, as well as patient factors such as known developmental differences, neurological considerations, and family history.
Study Limitations
A limitation of the current study was the reliance on one clinician-estimated threshold against which the algorithm-estimated threshold was validated. The subjective nature of ABR threshold estimation is a known issue that can lead to large inter-rater variability. Ideally, ABR thresholds estimated via visual inspection should be corroborated by more than one clinician to improve accuracy and objectivity when determining an ABR threshold against which to compare the algorithm estimate. However, the strong correlation between the behavioral and clinician-estimated ABR thresholds (ICC = 0.85) in the current study demonstrated that the ground truth ABR estimates for a majority of ears were reasonable.
A second limitation of the current study was that the algorithm is intended for infants but was validated using a cohort of adults. However, validation of experimental algorithms using adult populations is a common step prior to validation using clinically vulnerable populations such as infants with hearing loss. Experiments involving ABRs for infants are inherently more challenging and introduce a greater burden on caregivers and participants whose medical care may involve significant complications. Therefore, initial validation experiments using adult cohorts is a necessary precursor to those including infant populations. An additional benefit was that adults can provide reliable behavioral thresholds that may be used in corroboration with clinician-estimated ABR thresholds to validate the algorithm.
An additional limitation of the study was that there was no direct time-savings estimate available to benchmark the algorithm against the clinician-estimated ABR thresholds. Ideally, the total number of waveforms used by the clinician to estimate ABR threshold should be compared against the 15 waveforms used by the algorithm. However, during this experiment, the clinician, in some instances, collected waveforms in excess of what was required for a clinical diagnosis; these were intended to be used in post hoc analyses outside the scope of the current article. Therefore, the comparison between the total waveforms collected and the 15 used by the algorithm was not relevant. However, the 15 waveforms used by the algorithm may be compared against records from a large clinical database in which threshold estimates at 500, 1000, 2000, and 4000 Hz required a median of approximately 32 waveforms (approximately double the number used by the algorithm).
Future Directions
Although the current study demonstrated some promising results, this ABR threshold estimation algorithm is not ready for deployment in a clinical setting. Necessary improvements must be made from an engineering perspective. Additional improvements could be made, such as refinement of the adaptive sampling strategy for more efficient ABR threshold estimation, selective weighting of ABR waveforms to favor high signal-to-noise measurements, and improved objectivity of the final threshold estimate via more robust statistical metrics.
Additional work must be performed to facilitate the practical implementation of this algorithm with clinical systems. In its current configuration, the algorithm is run on a standalone installation of the MATLAB programming environment. Although the computational load is relatively low (that is, the algorithm can be run on a typical laptop), integrating the workflow in a seamless manner is essential for clinical adoption. This may involve translating the algorithm to run as an executable file on a free programming platform such as Python or building the computation directly into commercial ABR software through an industry collaboration.
Following these engineering-oriented modifications, the algorithm must be validated among the intended population of infants with hearing loss. This will require close collaboration with clinicians who work with this vulnerable population to ensure the safety and comfort of their patients.
Although the results of the current study demonstrate that the algorithm-estimated and clinician-estimated ABR thresholds matched reasonably well, the clinician estimates exhibited stronger correlation with behavioral thresholds. This suggests that the algorithm may not be suitable for replacing clinicians; instead, it may be useful as a supporting tool during clinical care. A promising model for clinical integration is the CDSS paradigm. In this scenario, the algorithm may visualize the waveforms predicted by the GP model across the stimulus space and suggest statistically optimized stimulus conditions. However, the clinician retains the authority of stimulus selection and may deviate from the algorithm suggestions based on their clinical experience, requirements of the patient, and possible extenuating circumstances of the environment (e.g., hospital room, neonatal intensive care unit, etc.). Validation of the CDSS paradigm using comparisons of ABR threshold estimates obtained by a clinician alone and those obtained by a clinician in collaboration with the algorithm will be an important step toward clinical integration.
Acknowledgments
The authors would like to thank the participants of this study, as well as Iris M. Choe and Nicholas R. McArthur for their assistance with scheduling participants and data collection.
Abbreviations:
- ABR
auditory brainstem response
- CDSS
clinical decision support system
- GP
Gaussian process
- ICC
intraclass correlation coefficient
- Ω
ohm
Footnotes
Any mention of a product, service, or procedure in the Journal of the American Academy of Audiology does not constitute an endorsement of the product, service, or procedure by the American Academy of Audiology.
REFERENCES
- American Academy of Otolaryngology–Head and Neck Surgery Foundation. (2021). American academy of otolaryngology-head and neck surgery position statement: Red flags-warning of ear disease. www.entnet.org/resource/position-statement-red-flags-warning-of-ear-disease/.
- American Speech-Language-Hearing Association. (2023). Identifying and managing hearing loss in school-age children. www.asha.org/siteassets/ais/ais-identifying-and-managing-hearing-loss-in-school-age-children.pdf?srsltid=AfmBOoqV59sEUDa2HF6SJdJMpXuNUPMp-YNR-9XFyTXX2gf5pFt0Rlq5.
- Arnold, S. A. (1985). Objective versus visual detection of the auditory Brain-stem response. Ear and Hearing, 6(3), 144–150. 10.1097/00003446-198505000-00004. [DOI] [PubMed] [Google Scholar]
- Bagatto, M., El-Naji, R., Purcell, D., Easwar, V., Pigeon, M., … Scollie, S. (2024). Correction factor evaluation and between-system comparison of behavioral threshold predictions from auditory brainstem response measures in infants. American Journal of Audiology, 33(4), 1104–1117. 10.1044/2024_AJA-24-00018. [DOI] [PubMed] [Google Scholar]
- Beahan, N., Kei, J., Driscoll, C., Charles, B., & Khan, A. (2012). High-frequency pure-tone audiometry in children: a test-retest reliability study relative to ototoxic criteria. Ear and Hearing, 33(1), 104–111. 10.1097/AUD.0b013e318228a77d. [DOI] [PubMed] [Google Scholar]
- British Society of Audiology. (2025). Auditory brainstem response (ABR) Testing in Newborns. [Online]. Available from www.thebsa.org.uk/wp-content/uploads/2025/06/OD100_Auditory-Brainstem-Response-ABR-Testing-in-Newborns-July-2025-.pdf. [accessed January 2026].
- Buran, B. N., Elkins, S., Kempton, J. B., Porsov, E. V., Brigande, J. V., & David, S. V. (2020). Optimizing auditory brainstem response acquisition using interleaved frequencies. Journal of the Association for Research in Otolaryngology: JARO, 21(3), 225–242. 10.1007/s10162-020-00754-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Clark, J. G. (1981). Uses and abuses of hearing loss classification. ASHA, 23(7), 493–500. [PubMed] [Google Scholar]
- Chesnaye, M., Simpson, D., Schlittenlacher, J., & Bell, S. (2023). Gaussian processes for hearing threshold estimation using auditory brainstem responses. IEEE Transactions on Bio-Medical Engineering, 71(3), 803–819. 10.1109/TBME.2023.3318729. [DOI] [PubMed] [Google Scholar]
- Chesnaye, M. A., Simpson, D. M., Schlittenlacher, J., Laugesen, S., & Bell, S. L. (2025). Audiogram estimation performance using auditory evoked potentials and gaussian processes. Ear and Hearing, 46(1), 230–241. 10.1097/AUD.0000000000001570. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Davidson, S. A., Wall, L. G., & Goodman, C. M. (1990). Preliminary Studies on the Use of an ABR Amplitude Projection Procedure for Hearing Aid Selection. https://journals.lww.com/ear-hearing/fulltext/1990/10000/preliminary_studies_on_the_use_of_an_abr_amplitude.3.aspx. [DOI] [PubMed]
- Elberling, C., & Don, M. (1984). Quality estimation of averaged auditory brainstem responses. Scandinavian Audiology, 13(3), 187–197. 10.3109/01050398409043059. [DOI] [PubMed] [Google Scholar]
- Elberling, C., & Don, M. (2010). A direct approach for the design of chirp stimuli used for the recording of auditory brainstem responses. The Journal of the Acoustical Society of America, 128(5), 2955–2964. 10.1121/1.3489111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ferm, I., & Lightfoot, G. (2015). Further comparisons of ABR response amplitudes, test time, and estimation of hearing threshold using frequency-specific chirp and tone pip stimuli in newborns: Findings at 0.5 and 2 kHz. International Journal of Audiology, 54(10), 745–750. 10.3109/14992027.2015.1058978. [DOI] [PubMed] [Google Scholar]
- Ferm, I., Lightfoot, G., & Stevens, J. (2013). Comparison of ABR response amplitude, test time, and estimation of hearing threshold using frequency specific chirp and tone pip stimuli in newborns. International Journal of Audiology, 52(6), 419–423. 10.3109/14992027.2013.769280. [DOI] [PubMed] [Google Scholar]
- Hall, J. W. (2007). New handbook of auditory evoked responses. Boston, MA, Pearson. [Google Scholar]
- Koo, T. K., & Li, M. Y. (2016). A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Journal of Chiropractic Medicine, 15(2), 155–163. 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kursun, B., Shola, C., Cunio, I. E., Langley, L., & Shen, Y. (2025). Variability of preference-based adjustments on hearing aid frequency–gain response. Journal of Speech, Language, and Hearing Research: JSLHR, 68(4), 2006–2025. 10.1044/2024_JSLHR-24-00215. [DOI] [PubMed] [Google Scholar]
- McCreery, R. W., Kaminski, J., Beauchaine, K., Lenzen, N., Simms, K., & Gorga, M. P. (2015). The impact of degree of hearing loss on auditory brainstem response predictions of behavioral thresholds. Ear & Hearing, 36(3), 309–319. 10.1097/AUD.0000000000000120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McGraw, K. O., & Wong, S. P. (1996). Forming inferences about some intraclass correlation coefficients. Psychological Methods, 1(1), 30–46. 10.1037/1082-989X.1.1.30. [DOI] [Google Scholar]
- Neely, S. T., Norton, S. J., Gorga, M. P., & Jesteadt, W. (1988). Latency of auditory Brain-stem responses and otoacoustic emissions using tone-burst stimuli. The Journal of the Acoustical Society of America, 83(2), 652–656. 10.1121/1.396542. [DOI] [PubMed] [Google Scholar]
- Norrix, L. W., & Velenovsky, D. (2017). Unraveling the mystery of auditory brainstem response corrections: the need for universal standards. Journal of the American Academy of Audiology, 28(10), 950–960. 10.3766/jaaa.16112. [DOI] [PubMed] [Google Scholar]
- Ormundo, D. D. S., Fávero, M. L., & Lewis, D. R. (2024). Audiogram estimation by auditory brainstem response with NB CE-Chirp LS stimulus in normal hearing infants. International Archives of Otorhinolaryngology, 28(2), e294–e300. 10.1055/s-0043-1776727. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Petersen, E. A., & Shen, Y. (2024). Multispecies initial numerical validation of an efficient algorithm prototype for auditory brainstem response hearing threshold estimation. The Journal of the Acoustical Society of America, 156(3), 1674–1687. 10.1121/10.0028537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pillow, J., & Park, M. (2016). Adaptive Bayesian methods for closed-loop neurophysiology, p 3–18. Closed Loop Neuroscience. Amsterdam, the Netherlands, Elsevier. 10.1016/B978-0-12-802452-2.00001-9. [DOI] [Google Scholar]
- Polonenko, M. J., & Maddox, R. K. (2019). The parallel auditory brainstem response. Trends in Hearing, 23, 2331216519871395. 10.1177/2331216519871395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rasetshwane, D. M., Argenyi, M., Neely, S. T., Kopun, J. G., & Gorga, M. P. (2013). Latency of tone-burst-evoked auditory Brain stem responses and otoacoustic emissions: Level, frequency, and rise-time effects. The Journal of the Acoustical Society of America, 133(5), 2803–2817. 10.1121/1.4798666. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rasmussen, C., & Nickisch, H. (2010). Gaussian processes for machine learning (GPML) toolbox. Journal of Machine Learning Research, 11, 3011–3015. [Google Scholar]
- Sininger, Y. S., Hunter, L. L., Roush, P. A., Windmill, S., Hayes, D., & Uhler, K. U. (2020). Protocol for rapid, accurate, electrophysiologic, auditory assessment of infants and toddlers. Journal of the American Academy of Audiology, 31(6), 455–468. 10.3766/jaaa.19046. [DOI] [PubMed] [Google Scholar]
- Suthakar, K., & Liberman, M. C. (2019). A simple algorithm for objective threshold determination of auditory brainstem responses. Hearing Research, 381, 107782. 10.1016/j.heares.2019.107782. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sutton, R. T., Pincock, D., Baumgart, D. C., Sadowski, D. C., Fedorak, R. N., & Kroeker, K. I. (2020). An overview of clinical decision support systems: Benefits, risks, and strategies for success. NPJ Digital Medicine, 3, 17. 10.1038/s41746-020-0221-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tanaka, K., Ohara, S., Matsuzaka, T., Matsugaki, A., Ishimoto, T., … Nakano, T. (2023). Quantitative threshold determination of auditory brainstem responses in mouse models. International Journal of Molecular Sciences, 24(14), 11393. 10.3390/ijms241411393. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vidler, M., & Parker, D. (2004). Auditory brainstem response threshold estimation: subjective threshold estimation by experienced clinicians in a computer simulation of the clinical test. International Journal of Audiology, 43(7), 417–429. 10.1080/14992020400050053. [DOI] [PubMed] [Google Scholar]