Convolutional neural network metal artifact reduction in ultra-high resolution photon counting detector CT of middle- and inner-ear implants

Kevin J Treb 1,2, Maryam Sadeghian 1,3, Wenchao Cao 1,4, Paul J Farnsworth 1, John I Lane 1, John C Benson 1, Christopher P Favazza 1, Andrea Ferrero 1, Andrew D Missert 1, Matthew L Carlson 5, Shuai Leng 1,3,*
PMCID: PMC13783213  NIHMSID: NIHMS2220476  PMID: 42875916

Abstract

Background:

Metallic inner- and middle-ear implants with submillimeter structures requiring ultra-high resolution (UHR) imaging invariably result in metal artifacts at CT, which can obscure adjacent microstructures and subtle pathologic findings. Current metal artifact reduction techniques can be less effective for small metal objects or metal in heterogeneous background anatomy, and can sometimes introduce new artifacts or noise.

Purpose:

To develop and evaluate an image-domain convolutional neural network to reduce metal artifacts near inner- and middle-ear implants in UHR photon-counting-detector (PCD)-CT.

Methods:

For network training, three stapes prosthesis models containing titanium and platinum were randomly oriented within a head-sized phantom and imaged on a PCD-CT with a UHR protocol. Metal and metal+artifact images were isolated by subtraction of the phantom background followed by CT number thresholding. Isolated artifact signal was enhanced via multiplication with a randomized factor ranging from one to a pre-determined artifact amplification scalar hyperparameter. Paired training inputs (with artifacts) and labels (without artifacts) were created by adding metal-free patient temporal bone images with metal+artifact and metal images, respectively. Combinations of 9 patient CT volumes and 24 metal CT volumes were used to create 216 CT volumes for training and validation with an 80%/20% training/validation split. The network was retrained with different amplification scalars (5, 10, 20, 30, 40) and with either mean-squared-error or mean-absolute-error loss functions to determine the optimal combination of these hyperparameters. The network was tested on 7 CT volumes from patients with three different implant types: stapes prostheses, cochlear implants, and a transverse sinus stent. Network performance was quantified through the artifact index, noise magnitude, CT number preservation of metal and bone, and dimensions of metal measured as the full-width at half-maximum of CT number line profiles.

Results:

Networks trained with mean-squared-error demonstrated a trade-off between metal artifact reduction and preservation of bone CT number as the artifact amplification scalar changed. The trade-off was not observed when mean-absolute error was used. Using the network with hyperparameters that best preserved bone and reduced artifacts simultaneously (mean-absolute-error, amplification scalar 20), the artifact index was improved by 83.2% (p < 0.001) across all implant types, and image noise was reduced by 46.8% (p < 0.001). Artifact reduction was significantly better for cases with stapes prostheses than the other implant types (p < 0.004), although artifact index was still greatly improved for all implant types (91.1% for the stapes prostheses, 81.3% for the CI, and 72.9% for the transverse sinus stent). Metal implant dimensions were not significantly different between the network input and output images (p = 0.67).

Conclusion:

An image-domain convolutional neural network with task-specific hyperparameters reduced artifact severity and image noise in UHR PCD-CT of different inner- and middle-ear implants without loss of fine image detail.

Keywords: photon counting detector CT, convolutional neural networks, deep learning, metal artifact reduction, ultra-high resolution CT

Introduction

CT is commonly used for post-operative visualization of middle- and inner-ear implants and adjacent microanatomy due to its relatively good spatial resolution, rapid scan time, and wide availability.1-5 More recently, ultra-high resolution (UHR) photon-counting-detector CT (PCD-CT) demonstrated superior spatial resolution compared to conventional energy-integrating-detector CT (EID-CT) for improved evaluation of middle-ear prostheses.6 However, most otologic implants contain metallic alloys, and therefore result in significant metal artifacts at CT. The severity of metal artifacts can be similar between PCD-CT and EID-CT for neuroimaging,7 and can negatively impact the visualization of both metallic implant substructures as well as anatomical microstructures near the metal. Therefore, methods to reduce metal artifacts in UHR PCD-CT are desired for middle- and inner-ear implant imaging.

In general, hardware-based data correction and software-based data replacement techniques for metal artifact reduction (MAR) have been established.8,9 Hardware-based MAR techniques generally attempt to increase the effective x-ray energy used to create the CT image (e.g., virtual monoenergetic imaging, high-energy-threshold images) in order to reduce the beam-hardening and scatter portions of metal artifacts.10-13 With multi-energy PCD-CT, detector-based energy thresholding and spectral post-processing can be used to reconstruct images preferentially from higher-energy x-rays, while down weighting or eliminating the contribution of lower-energy x-rays. However, virtual monoenergetic images above approximately 70–80 keV incur a noise penalty due to suboptimal utilization of lower-energy photons.10 This can be especially troublesome with UHR image acquisition and high-resolution reconstructions, where image noise is increased as a result of maximally leveraging spatial resolution. Additionally, higher-energy images do not address non-linear partial volume effects, which contribute to metal artifacts.8 Lastly, spectral acquisition and post-processing is not currently available on commercial PCD-CT with the sharpest resolution parameters, which are needed to take full advantage of UHR PCD-CT technology.

Software-based MAR techniques are well established and have found wide success in both EID-CT and PCD-CT.13-17 However, commercially-available MAR algorithms have been shown to be less effective for small metal implants or for metal in heterogeneous background anatomy.13,18 For middle- and inner-ear implant imaging, both conditions of very small metal implants and heterogeneous bone/air/soft tissue background are present. Further, MAR algorithms can sometimes lead to loss of spatial resolution or new artifacts.9,15 For inner- and middle-ear imaging, any loss of spatial resolution is unacceptable due to the spatial resolution constraints imposed by the extremely small structures of the temporal bone, ossicles, implants, etc. Lastly, similar to hardware-based MAR techniques, commercially available software corrections are not currently available with the sharpest resolution parameters in UHR PCD-CT.

Due to the lack of available MAR techniques with commercial UHR PCD-CT, in-house MAR software is desirable. Further, an effective solution must reduce artifacts near small implants within heterogeneous anatomy while preserving spatial resolution and not increasing noise. Lastly, for use within a clinical informatics environment where raw projection data may be unavailable, image-domain techniques become advantageous over projection-domain techniques.19

Image-domain machine learning models have demonstrated potential to overcome the limitations of currently-available MAR algorithms,19-28 with supervised methods using paired data performing better in recovering anatomical details than unsupervised methods.25 Cao et al. introduced a supervised deep convolutional neural network (CNN) with an image-domain metal artifact simulation framework to reduce metal artifacts in CT near ablation probes for interventional oncology applications.19 With this approach, spatial resolution was preserved by the CNN, although the imaging task was such that the spatial resolution was chosen for typical abdominal imaging. Additionally, the anatomical background mostly contained soft tissue, although in principle the methodology can be generalized to heterogeneous background structures within the image data. Therefore, the specific image-domain CNN approach of Cao et al. potentially can overcome the limitations of conventional MAR techniques for inner- and middle-ear imaging as long as it can generalize to the variety of shapes, sizes, and materials used in inner- and middle-ear implants.

The purpose of this work was to develop a supervised image-domain CNN to reduce metal artifacts near inner- and middle-ear implants in UHR PCD-CT, and to evaluate its generalizable performance for this task across network hyperparameters and metal implant types.

Methods

Image Data Collection

All image data used in this study were acquired with a commercial PCD-CT scanner (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany) in UHR mode using a clinical temporal bone protocol. Images were acquired at 120 kV, 120x0.2 mm collimation, helical pitch of 0.5, and CARE keV IQ level set to 220, resulting in a typical volume CT dose index (16 cm phantom) of 34 mGy. Images of each ear were reconstructed with 80 mm FOV, 1024x1024 matrix, 0.2/0.1 mm slice thickness/increment, Hr84 reconstruction kernel, and quantum iterative reconstruction strength 3. For comparison, additional reconstructions were performed with the commercial iterative metal artifact reduction (iMAR) provided on the scanner and the sharpest compatible reconstruction kernel (Hr64).

For the network training data, 3 stapes prosthesis models (Grace Medical, Memphis, TN) containing titanium and platinum (MEITI Piston, Grace Piston, Lippy Bucket Handle) that were observed clinically to result in metal artifacts were placed within a semi-anthropomorphic adult head phantom (Model 007TE Adult Head, CIRS INC, Norfolk, VA) containing air, soft tissue- and bone-mimicking materials. For each prosthesis, acquisitions were performed for 8 random prosthesis orientations within an air cavity of the phantom (Figure 1) using the UHR PCD-CT temporal bone protocol. In addition to the phantom data, internal images from 9 patient cases were retrospectively collected to be used with the phantom images for network training. The patient images used in the training data did not have any metallic implants in the head or neck regions.

Figure 1:

Figure 1:

a. Semi-anthropomorphic head phantom used in the training data. b. 3D model of the stapes prostheses used in the training data.

Images from a retrospectively collected internal cohort of 6 independent patient cases and 1 fresh frozen (non-formalin fixed) human cadaver head specimen were used for network fine-tuning and testing. Each case had either a stapes prosthesis (2 cases), a cochlear implant (4 cases), or a transverse dural sinus stent (1 case) implanted. Network fine-tuning was performed on a subset containing one stapes prosthesis and one cochlear implant case: the remaining five cases were reserved for testing. The human cadaver head specimen was obtained through our institute anatomy laboratory and the cadaveric portion of this study was performed with the approval of our institute biospecimen committee.

Training Data Curation

The training data curation methodology is outlined in Figures 2 and 3, and follows the general metal artifact simulation framework described by Cao et al.19 In summary, the metal and artifact signals are isolated from the phantom images, and then superimposed onto the metal-free patient images within the training data to create paired artifact-corrupted and artifact-free images containing both metal and anatomical features.

Figure 2:

Figure 2:

Flowchart outlining the methodology to isolate metal signal and artifact signal from phantom images to be used for model training. The scalar multiplication factor is a hyperparameter that was fine-tuned.

Figure 3:

Figure 3:

Methodology to create paired training input and label images. Metal + enhanced artifact and metal only images are superimposed onto metal-free patient images to create training data with features of both metal implants and anatomy.

The first step of the training data curation (Figure 2) is to isolate the metal+artifact signal from the phantom background. To do this, phantom only images from scans without any metal are subtracted from phantom+metal images containing each metal implant. To further isolate the metal signal from the metal+artifact image, global CT number thresholding is performed with an empirically determined threshold unique to each implant model. This thresholding step also has the advantage of removing the background image noise in addition to removing artifacts. After the metal only and metal+artifact images are created, artifact only images are created by subtracting the metal only images from the metal+artifact images.

In order to enhance the metal artifacts in the training data, the artifact only images were amplified via multiplication with a randomized factor ranging from one up to an empirical artifact amplification scalar. In addition to enhancing the metal artifact and the image noise, the randomization in this step augments the training data and broadens the distribution of metal artifacts and noise seen by the network during training. In this work, the artifact amplification scalar was one of the hyperparameters that was varied during fine-tuning of the network. Different network instances were trained with amplification scalar values of 5, 10, 20, 30, or 40. The paired metal+enhanced artifact and metal only image stacks from each implant and orientation were then randomly spatially translated and flipped in a paired manner for further data augmentation.

For the next step of training data curation (Figure 3), the metal+enhanced artifact and the metal only images were added to a stack of patient images to create images for inputs and labels, respectively, for supervised network training. Images from each of the 3 implant models, each with 8 implant orientations, were added to each stack of images from the 9 patient exams within the training data. This resulted in 216 unique combinations of implant, orientation, and patient.

Convolutional Neural Network Training

From the input and label images, CNN input patches were extracted by randomly selecting a 128x128 pixel ROI. For each slice n in the image volume, input slices n-1, n, and n+1 were paired with output slice n. For each patch, an empirical maximum pixel value threshold was used to determine if the 128x128x1 label patch contained any metal. If the label patch contained metal, then it was used in the CNN training data along with its corresponding 128x128x3 input patch.

The deep CNN model used a simplified U-net architecture with four levels of down-sampling (max pool operation) and up-sampling. Each module sequentially performs convolution, batch normalization, and rectified linear unit activation. The loss function was the second hyperparameter that was varied for network fine-tuning: For each of the amplification scalar hyperparameter values, separate networks were trained with mean absolute error (MAE) or mean squared error (MSE) as the loss function. A total of 10 different networks were trained, covering each combination of different hyperparameters. For each of the 10 different networks, all model parameters were updated during training. Each network was applied to the images reserved for network fine-tuning to determine the optimal hyperparameters. The network was implemented in Python using the TensorFlow library (Google AI, Inc.) and Keras interface.

Quantitative Image-Quality Evaluation

For each case in the network fine-tuning dataset, performance was evaluated using the following quantitative metrics. First, the background standard deviation σbackground of voxel CT number was measured in a uniform region of the brain. The noise reduction from the CNN was calculated as the percent difference between input and output standard deviations.

To quantify the reduction in metal artifacts, the standard deviation σartifact was also measured in ROIs positioned over metal artifact regions in an otherwise uniform tissue background to calculate the artifact index,

Artifact Index=σartifact2−σbackground2.

The artifact index was calculated across all cases, as well as separated by implant type (stapes prosthesis, cochlear implant, transverse sinus stent).

To evaluate preservation of fine image detail, CT number line profiles through each of the metal implants and nearby trabecular bone were measured in transaxial image slices. The full-width at half-maximum (FWHM) and the maximum CT number of the line profiles were calculated to represent the object dimension and visibility in each image. Preservation of small objects were calculated based on their maximum CT numbers CT#max as

%Preservation=1−CT#max,in−CT#max,outCT#max,in,

where the subscripts in and out represent the CNN input and output images, respectively. Mean and standard deviations for all quantitative metrics were calculated across ten non-consecutive image slices, each separated by at least 5 slices.

Statistical Analysis

Statistical significance was evaluated using two-tailed paired-sample t-tests to compare artifact index and noise magnitude between the network input and output images. Two-tailed two-sample t-tests assuming unequal variances were used to compare the percent reduction in artifact index between different implant types. P-values less than 0.05 were deemed statistically significant. If a significant difference was observed in the two-tailed t-test, then a one-tailed t-test was performed and a p-value reported. All statistical analyses were performed on MATLAB version R2023a (MathWorks, Natick, MA, USA).

Results

Network Optimization

All CNN output images had markedly reduced metal artifacts and noise compared to the original images (Table 1). However, preservation of metal and bone structures differed between the networks with different parameters. Networks with MSE loss demonstrated a trade-off between preservation of metal CT number and bone CT number, as well as a trend of improved artifact reduction with worsened bone preservation, suggesting that these networks falsely identified bone structures as artifacts. These trends were not as consistent in the MAE networks. Instead for MAE networks, improved artifact reduction was related to improved bone preservation for amplification scalars of 20 and below. Above 20, both bone and metal signal were lost. None of the networks demonstrated worsening of spatial resolution, although some demonstrated apparent resolution improvements. However, the apparent improvement was caused by loss of true structures in the image, and therefore represents degraded image-quality. Across all models trained with the span of investigated hyperparameters on the same fine-tuning dataset, the level of artifact reduction ranged from 67%-92%, noise reduction ranged from 43%-70%, and preservation of metal and bone ranged from 2.5%-99.7% and 21.4%-75.7%, respectively.

Table 1:

Quantitative metrics for network input and output images in the validation data with different network hyperparameters (mean ± standard error).

Loss
Function
Amplification
Scalar
Noise
Magnitude
(HU)
Artifact
Index (HU)
FWHM
(mm)
Metal
Preservation
(%)
Trabecular
Bone
Preservation
(%)
N/A (Network Input) 76.9 ± 1.0 579.0 ± 114.6 0.32 ± 0.01 100 100
MSE 5 33.0 ± 1.6 64.0 ± 17.1 0.31 ± 0.01 99.7 ± 0.3 21.4 ± 15.9
10 38.0 ± 1.4 46.5 ± 9.5 0.31 ± 0.01 96.5 ± 1.9 16.1 ± 14.7
20 35.3 ± 2.1 180.2 ± 51.8 0.32 ± 0.01 80.5 ± 25.9 64.1 ± 13.5
30 35.8 ± 1.6 190.0 ± 7.0 0.21 ± 0.01 20.8 ± 22.7 75.7 ± 7.7
40 23.2 ± 1.6 122.8 ± 24.8 0.24 ± 0.01 2.5 ± 1.2 64.6 ± 8.8
MAE 5 44.0 ± 1.4 113.9 ± 28.9 0.33 ± 0.01 99.5 ± 0.0 29.9 ± 21.9
10 40.4 ± 1.4 102.6 ± 30.7 0.32 ± 0.01 98.7 ± 0.3 41.8 ± 23.3
20 42.5 ± 1.5 86.6 ± 7.0 0.33 ± 0.01 96.2 ± 3.7 74.7 ± 16.1
30 39.5 ± 2.1 102.0 ± 4.0 0.13 ± 0.01 10.4 ± 13.5 50.3 ± 25.9
40 42.7 ± 1.6 178.2 ± 79.8 0.25 ± 0.01 17.2 ± 16.2 64.0 ± 20.6

MSE = Mean Squared Error, MAE = Mean Absolute Error, FWHM = Full-Width at Half-Maximum, HU=Hounsfield Units.

Across all networks with different hyperparameters, the network with MAE loss and an amplification scalar of 20 resulted in relatively good overall preservation of fine metal and bone structures, while reducing image noise and having the third strongest artifact reduction of all the hyperparameter combinations for the validation subset of the fine-tuning cases. Therefore, these hyperparameters were chosen for the image analysis on the remaining test cases.

Image-Quality Evaluation

For the stapes prosthesis test cases (Figure 4), the CNN removed false metal artifacts in one case and improved resolvability in another case. For the cochlear implant test cases (Figure 5), the CNN output images had improved or similar resolvability of the individual electrodes compared to the original images. For the transverse sinus stent case (Figure 6), artifact reduction was most prominently noted about the metal markers at the end of the stent, and there was overall improved resolution of the stent mesh. All Figures had matched display window width/level between the original and CNN output images.

Figure 4:

Figure 4:

Representative PCD-CT images of stapes prostheses with and without deep convolutional neural network (CNN) processing in axial and sagittal planes. The commercial iMAR image using the sharpest reconstruction kernel available with iMAR (Hr64) on the scanner provides insufficient image-quality. False metallic structures due to artifacts and the resolvability of prosthesis structures are improved by the neural network. Window width/level: 4000/1000 HU.

Figure 5:

Figure 5:

Representative PCD-CT images of cochlear implants both with and without deep convolutional neural network (CNN) processing. Metal artifacts are greatly reduced and resolvability of individual electrodes is improved. Window width/level: 4000/1000 HU.

Figure 6:

Figure 6:

Representative PCD-CT images of a transverse dural sinus stent from a single case both with and without deep convolutional neural network (CNN) processing. Metal artifacts from the markers at the end of the stent and the resolution of the stent mesh are improved by the network. Window width/level: 4000/1000 HU (top); 1750/650 HU (bottom).

Compared to the commercial iMAR images using the sharpest compatible reconstruction kernel (Figure 4), the CNN output image greatly improved visibility of fine image details and resolvability of the metal in the axial plane. The commercial iMAR also falsely converted metal artifacts in air background into a soft-tissue-like background near the stapes prosthesis. In sagittal reformatted planes, the lower-resolution commercial iMAR image spatially distorted the metal implants due to less isotropic resolution properties. Streak artifacts visible in sagittal reformats were also introduced by the commercial iMAR processing. Overall, the commercial iMAR using the sharpest compatible reconstruction provided insufficient image-quality.

For all of the CNN test cases, image noise magnitude was significantly less in the CNN output images than the input images (p < 0.001). The average noise magnitude was 80.7 HU (95% CI: 76.1 HU – 85.3 HU) for the input images and 42.5 HU (95% CI: 41.0 HU – 44.0 HU) for the output images, representing an average noise reduction of 46.8% (95% CI: 45.6% - 48.0%).

The artifact index across the test cases was significantly less in the output images than the inputs (p < 0.001). Across these cases, the average artifact index was 651.1 HU (95% CI: 536.7 HU – 765.6 HU) for the input images and 93.1 HU (95% CI: 74.3 HU – 111.9 HU) for the output images, representing an average artifact reduction of 83.2% (95% CI: 79.0% - 87.3%). When grouped by type of implant, the reduction in artifact index remained statistically significant for each implant type (p < 0.001). The artifact reductions were 91.1% (95% CI: 88.3% - 93.9%) for the stapes prostheses, 81.3% (95% CI: 75.0% - 87.6%) for the CI, and 72.9% (95% CI: 60.3% - 85.5%) for the transverse sinus stent. The artifact reduction was significantly better for the stapes prostheses than for either the cochlear implant (p = 0.004) or the transverse sinus stent (p = 0.005). The artifact reduction was not significantly different between the cochlear implant and the sinus stent (p = 0.16).

Full-width at half-maximum values for the metal implants were not significantly different between the network input and output images (p = 0.67), with an average difference of 0.003 mm (95% CI: −0.011 mm – 0.017 mm). Visual inspection of the line profiles demonstrated similar object widths and CT number intensities (Figure 7). Outside of the metal, the CNN output line profiles demonstrated less severe CT number undershoot and overshoot than that of the input profile due to artifacts.

Figure 7:

Figure 7:

Representative CT number line profiles through the stapes prosthesis in Figure 4 and the dural sinus stent in Figure 6. Line profiles from convolutional neural network (CNN) input and output images demonstrated comparable CT number profiles, with CT number undershoot and overshoot due to artifacts reduced in the output profile.

Discussion

The results of this study indicate that the image-domain metal artifact reduction CNN significantly reduced the severity of metal artifacts near stapes prostheses, cochlear implants, and a transverse dural sinus stent in patient cases despite only being trained on stapes prostheses. Additionally, spatial resolution and small object dimensions were not affected, fine metal and bony structures were mostly preserved, and image noise was improved by the CNN. Together, these image-quality improvements enabled greater visibility and resolvability of the small metallic implants. Further, the UHR CNN MAR images were of superior quality to the commercial iMAR images using the sharpest compatible reconstruction kernel in terms of both artifact reduction and spatial resolution. The results of this study may have considerable clinical importance, as even small differences in prosthesis depth, position, and ossicular coupling inform the need for and risk of surgical intervention.

The strength of CNN artifact reduction was significantly different between the stapes prostheses and the other metal implants. This is likely due to only stapes prostheses being included in the training data. Therefore, the model has not seen any metal artifacts or metallic structures directly from cochlear implants or stents. The ability of the CNN to still significantly reduce metal artifacts near all three types of implants demonstrates its ability to generalize through learning generalizable features universal to metal and metal artifacts from the stapes prostheses within the training data. Further, these results combined with the preservation of surrounding anatomy support the hypothesis that features specific to metal artifacts can be identified in the image domain without relying on additional information from the projection domain.

During training data curation, CT number thresholding is utilized to separate metal signal from artifact signal. During this process, the noise- and artifact-corrupted image background remains in the training input data, but is removed in the training output data by setting the CT number of all pixels not containing metal to the same value. In this manner, the network learns from the data to jointly reduce artifacts and image noise. Since both noise and artifacts are affected by the amplification scalar hyperparameter, one might expect the noise magnitude and artifact index to be related for networks with different amplification scalars. However, this was not observed in this work. Rather, noise magnitude was much more consistent than the artifact index across different networks. This observation may be due to more complex overlapping features between the artifact, metal, and bone signals potentially confounding the reduction in metal artifacts. Further, the amplification scalar represents a maximum value of a randomly sampled uniform distribution, i.e., a value of 20 means that the artifacts of each individual training sample is multiplied by a random number ranging from 1-20. Therefore, one might expect that higher amplification would improve model performance. For example, since the model with amplification scalar of 40 contains amplifications ranging from 1-40, the ranges of 1-5, 1-10, 1-20, and 1-30 are subsets that are still covered by this model. However, the “unified” model with amplification of 40 did not have the optimal performance in this work. Overall, the impact of different hyperparameters (loss function, artifact amplification scalar) on network performance was unintuitive, supporting the need for fine-tuning highly nonlinear models given a specific task.

This work follows the general methodology proposed by Cao et al. for interventional abdominal imaging19 although substantial modifications were made to accommodate the more demanding UHR PCD-CT temporal bone imaging task. First, due to the background tissue heterogeneity present near the temporal bone rather than the abdomen, the training data were acquired with a phantom that contained air, soft tissue, and bone-equivalent materials, rather than a uniform water tank. Although less heterogeneous than a completely anthropomorphic phantom, the semi-anthropomorphic head phantom used in this work for training data acquisition allows the metal artifacts in the training data to contain a broader distribution of features for the CNN to learn. Second, the optimal artifact enhancement hyperparameter in this work was 20 (MAE), which is much larger than the 4 (MSE) used by Cao et al. for abdominal imaging. Although in this work, it was found that the 5-MSE network greatly reduced metal artifacts and preserved metal structures, these hyperparameters resulted in loss of bony structures. For ear imaging, increased artifact amplification might have been necessary to further distinguish the distribution of artifact features from features of the temporal bone in order to retain anatomical features.

Metal artifact reduction near cochlear implants in CT has been performed previously using image-domain deep learning.21 Wang et al. used a conditional generative adversarial network for this task, demonstrating alongside this work that image-domain deep learning is generally a viable solution for MAR in ear implants. The work presented herein differs from that of Wang et al. in that UHR methods were used in this work (0.08 mm x 0.08 mm x 0.20 mm voxels) compared to more typical resolution used by Wang et al. (0.25 mm x 0.25 mm x 0.30 mm voxels). Additionally, the performance metrics are different between these works, as Wang et al. focused on evaluating the impact of artifact reduction on segmentation of anatomical structures. This makes direct comparison between the works difficult, especially since different implant types (cochlear implant, stapes prosthesis, sinus stent) were used herein. In future work, different deep learning techniques such as generative adversarial networks may be investigated for the task of UHR metal artifact reduction to potentially further improve performance.

A limitation of this work is the lack of comparison to a commercial MAR solution with matched resolution to the CNN input images, since commercial MAR algorithms equipped on the PCD-CT system are not currently available for the higher-resolution parameters used in the UHR temporal bone protocol. Further, spectral CT data could not be included in the CNN since multi-energy capabilities of the commercial PCD-CT system are not currently available with the parameters for the UHR protocol. Future upgrades to the commercial scanner might enable these additional avenues to be pursued to further improve the network performance and to provide additional comparisons. Further, since the network training relies on empirical CT number thresholding to separate metal and artifact signals, it is currently unclear how well a network trained only on 120 kV images would generalize to images acquired at different tube potentials. Lastly, a limitation of this work is the absence of paired pre- and post-implantation images from the same patients to provide references for evaluating anatomical preservation. This limitation could be addressed with a prospective study.

Conclusions

An image-domain supervised deep convolutional neural network was developed to reduce metal artifacts in UHR PCD-CT of inner- and middle-ear implants. After hyperparameter fine-tuning, the network preserved bony anatomy and reduced image noise while generalizing to different types of ear implants.

Acknowledgement Statement

This work was supported by an NIH grant (R01DC022228), a Mayo Clinic Radiology Research Internal Grant, and Mayo Clinic Graduate School of Biomedical Sciences. The authors would like to thank Grace Medical for providing the prostheses used in this study. The authors would also like to thank Kevin Kimlinger for his assistance in the manuscript submission.

Footnotes

Conflict of Interest Statement

The authors have no relevant conflicts of interest to disclose.

References

  • 1.Kosling S, Bootz F. CT and MR imaging after middle ear surgery. Eur J Radiol. 2001;40(2):113–118. [DOI] [PubMed] [Google Scholar]
  • 2.Ali HI, Khater NH. Otosclerosis and complications of stapedectomy: CT and MRI correlation. Alexandria Journal of Medicine. 2019;54(3):197–201. [Google Scholar]
  • 3.Tames H, Padula M, Sarpi MO, et al. Postoperative Imaging of the Temporal Bone [published online ahead of print 20210319]. Radiographics. 2021;41(3):858–875. [DOI] [PubMed] [Google Scholar]
  • 4.Altmann S, Abello Mercado MA, Ucar FA, et al. Ultra-High-Resolution CT of the Head and Neck with Deep Learning Reconstruction-Assessment of Image Quality and Radiation Exposure and Intraindividual Comparison with Normal-Resolution CT [published online ahead of print 20230424]. Diagnostics (Basel). 2023;13(9). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Adams JK, Marinelli JP, De Jong R, Spear SA, Erbele ID. Modern Trends in Otologic Surgery and Implications for Residency Training [published online ahead of print 20240122]. Otolaryngol Head Neck Surg. 2024;170(5):1404–1410. [DOI] [PubMed] [Google Scholar]
  • 6.Sadeghian M, Benson JC, Farnsworth PJ, et al. Impact of Spatial Resolution on CT Imaging of Middle Ear Prostheses: Comparison of Photon-Counting and Energy-Integrating Detectors [published online ahead of print 20260504]. AJNR Am J Neuroradiol. 2026;47(5):1341–1348. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Schmitt N, Wucherpfennig L, Rotkopf LT, et al. Metal artifacts and artifact reduction of neurovascular coils in photon-counting detector CT versus energy-integrating detector CT - in vitro comparison of a standard brain imaging protocol [published online ahead of print 20220820]. Eur Radiol. 2023;33(2):803–811. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Katsura M, Sato J, Akahane M, Kunimatsu A, Abe O. Current and Novel Techniques for Metal Artifact Reduction at CT: Practical Guide for Radiologists. Radiographics. 2018;38(2):450–461. [DOI] [PubMed] [Google Scholar]
  • 9.Hsieh SS, Leng S, Rajendran K, Tao S, McCollough CH. Photon Counting CT: Clinical Applications and Future Developments [published online ahead of print 20200828]. IEEE Trans Radiat Plasma Med Sci. 2021;5(4):441–452. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Yu L, Leng S, McCollough CH. Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol. 2012;199(5 Suppl):S9–S15. [DOI] [PubMed] [Google Scholar]
  • 11.Mangold S, Gatidis S, Luz O, et al. Single-source dual-energy computed tomography: use of monoenergetic extrapolation for a reduction of metal artifacts. Invest Radiol. 2014;49(12):788–793. [DOI] [PubMed] [Google Scholar]
  • 12.Zhou W, Bartlett DJ, Diehn FE, et al. Reduction of Metal Artifacts and Improvement in Dose Efficiency Using Photon-Counting Detector Computed Tomography and Tin Filtration. Invest Radiol. 2019;54(4):204–211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lee YH, Park KK, Song HT, Kim S, Suh JS. Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software [published online ahead of print 20120204]. Eur Radiol. 2012;22(6):1331–1340. [DOI] [PubMed] [Google Scholar]
  • 14.Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M. Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys. 2010;37(10):5482–5493. [DOI] [PubMed] [Google Scholar]
  • 15.Lell MM, Meyer E, Kuefner MA, et al. Normalized metal artifact reduction in head and neck computed tomography. Invest Radiol. 2012;47(7):415–421. [DOI] [PubMed] [Google Scholar]
  • 16.Aissa J, Boos J, Schleich C, et al. Metal Artifact Reduction in Computed Tomography After Deep Brain Stimulation Electrode Placement Using Iterative Reconstructions. Invest Radiol. 2017;52(1):18–22. [DOI] [PubMed] [Google Scholar]
  • 17.Byl A, Klein L, Sawall S, Heinze S, Schlemmer HP, Kachelriess M. Photon-counting normalized metal artifact reduction (NMAR) in diagnostic CT [published online ahead of print 20210528]. Med Phys. 2021;48(7):3572–3582. [DOI] [PubMed] [Google Scholar]
  • 18.Huang JY, Kerns JR, Nute JL, et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging [published online ahead of print 20150114]. Phys Med Biol. 2015;60(3):1047–1067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Cao W, Parvinian A, Adamo D, et al. Deep convolutional-neural-network-based metal artifact reduction for CT-guided interventional oncology procedures (MARIO) [published online ahead of print 20240214]. Med Phys. 2024;51(6):4231–4242. [DOI] [PubMed] [Google Scholar]
  • 20.Huang X, Wang J, Tang F, Zhong T, Zhang Y. Metal artifact reduction on cervical CT images by deep residual learning [published online ahead of print 20181127]. Biomed Eng Online. 2018;17(1):175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Wang J, Zhao Y, Noble JH, Dawant BM. Conditional Generative Adversarial Networks for Metal Artifact Reduction in CT Images of the Ear [published online ahead of print 20180926]. Med Image Comput Comput Assist Interv. 2018;11070:3–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Liang K, Zhang L, Yang H, Yang Y, Chen Z, Xing Y. Metal artifact reduction for practical dental computed tomography by improving interpolation-based reconstruction with deep learning. Med Phys. 2019;46(12):e823–e834. [DOI] [PubMed] [Google Scholar]
  • 23.Hyun CM, Bayaraa T, Yun HS, Jang TJ, Park HS, Seo JK. Deep learning method for reducing metal artifacts in dental cone-beam CT using supplementary information from intra-oral scan [published online ahead of print 20220825]. Phys Med Biol. 2022;67(17). [DOI] [PubMed] [Google Scholar]
  • 24.Kim S, Ahn J, Kim B, Kim C, Baek J. Convolutional neural network-based metal and streak artifacts reduction in dental CT images with sparse-view sampling scheme [published online ahead of print 20220810]. Med Phys. 2022;49(9):6253–6277. [DOI] [PubMed] [Google Scholar]
  • 25.Li G, Ji L, You C, et al. MARGANVAC: metal artifact reduction method based on generative adversarial network with variable constraints [published online ahead of print 20231002]. Phys Med Biol. 2023;68(20). [DOI] [PubMed] [Google Scholar]
  • 26.Selles M, Slotman DJ, van Osch JAC, et al. Is AI the way forward for reducing metal artifacts in CT? Development of a generic deep learning-based method and initial evaluation in patients with sacroiliac joint implants [published online ahead of print 20230420]. Eur J Radiol. 2023;163:110844. [DOI] [PubMed] [Google Scholar]
  • 27.Wang H, Xie Q, Zeng D, Ma J, Meng D, Zheng Y. OSCNet: Orientation-Shared Convolutional Network for CT Metal Artifact Learning [published online ahead of print 20240102]. IEEE Trans Med Imaging. 2024;43(1):489–502. [DOI] [PubMed] [Google Scholar]
  • 28.Kleber CEJ, Karius R, Naessens LE, et al. Advancements in supervised deep learning for metal artifact reduction in computed tomography: A systematic review [published online ahead of print 20240907]. Eur J Radiol. 2024;181:111732. [DOI] [PubMed] [Google Scholar]

RESOURCES