Abstract
Background and purpose
We aimed to characterize temporal trends in implant selection, estimate associations between patient, surgeon, and institutional characteristics and implant choice, and quantify surgeon- and institution-related variation at first-time aseptic revision knee arthroplasty (rTKA).
Methods
This retrospective registry-based study used National Joint Registry data from England, Wales, Northern Ireland, the Isle of Man, Guernsey, and Jersey. Elective aseptic revisions of primary total knee replacements undertaken between January 1, 2009 and September 30, 2023 were included. Outcomes were metaphyseal fixation device (MFD) use, hinged-construct use, absence of intramedullary stem fixation, and mobile-bearing use. Temporal trends were described. Multilevel regression models estimated associations between patient, surgeon, and institutional characteristics and implant selection, as well as surgeon- and institution-related variation.
Results
The study comprised 12,950 patients (12,950 knees), with procedures undertaken by 1,250 surgeons across 402 institutions. The use of MFDs increased from 8.5% in 2009 to 37% in 2023. High-volume surgeons had greater odds of MFD use (odd ratio [OR] 1.69, 95% confidence interval [CI] 1.27–2.23; P < 0.001) and lower odds of revision without intramedullary stem fixation (OR 0.58, CI 0.48–0.72; P < 0.001). Compared with aseptic loosening, instability was strongly associated with hinge use (OR 5.96, CI 5.07–7.00; P < 0.001). Measured fixed effects explained little variation in MFD and mobile-bearing use, with substantial variation observed between surgeons and institutions.
Conclusion
Implant selection changed substantially over time and varied with patient factors and surgeon/institutional context. Measured factors explained few of the examined implant decisions, highlighting substantial practice variation and the need for clearer evidence-based guidance to support implant choices in this setting.
Implant survival following revision knee arthroplasty (rTKA) decreases with each successive revision procedure [1]. rTKA requires the surgeon to convert a failed and often mechanically compromised knee into a stable, well-fixed, and functional reconstruction. Surgical decision-making requires selection of an appropriate level of constraint, consideration of intramedullary stems and metaphyseal fixation devices (MFDs), and, in some implant systems, a choice between fixed- and mobile-bearing designs [2].
Evidence to support implant selection at rTKA is lacking. In 2020, the James Lind Alliance (JLA) Priority Setting Partnership (PSP) on “the problematic knee replacement” highlighted implant selection as one of the “Top 10” research priorities within revision knee surgery [3]. More recently, the British Association for Surgery of the Knee (BASK) issued A Guide to Good Practice for Revision Knee Replacement; yet due to limited evidence on outcomes for different revision implants, it was unable to make strong recommendations for practice in this domain [4].
Implant selection is particularly challenging in younger patients, men, and those undergoing early first-time rTKA following primary arthroplasty, as these groups are at an increased risk of re-revision [5]. Registry data suggests implant choice may influence subsequent outcomes, with hinged prostheses and revision procedures performed without intramedullary stem fixation associated with inferior implant survival [6-8]. Reconstructive practice is evolving, with registry data showing increasing use of MFDs at first-time aseptic rTKA in Australia and the United States [6,9].
We aimed primarily to characterize temporal trends in implant selection at first-time aseptic rTKA, and secondarily to estimate associations between patient, surgeon, and institutional characteristics and implant choice, and to quantify surgeon- and institution-related variation in implant selection.
Methods
Study design
This observational registry-based study utilized routinely collected data from the National Joint Registry for England, Wales, Northern Ireland, the Isle of Man, Guernsey, and Jersey (NJR). Reporting followed the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement [10].
Sample
Linkage to primary total knee arthroplasty (pTKA) ensured inclusion of first revisions only. Procedures performed between January 1, 2009 and September 30, 2023 in which 1 or both tibial and femoral components were revised were eligible. For patients with bilateral rTKA procedures, only the first procedure was included (or the procedure with the lowest njrrevisionindexno if simultaneous) to fulfil the assumption of independence of observations. Procedures entailing isolated revision of the patellar component or isolated polyethylene exchange were excluded. Revisions of unicompartmental knee replacements and those undertaken for infection, fracture, progressive arthritis, or “other” were excluded. Where multiple revision diagnoses were recorded, indication was assigned to one of 10 groups using the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) hierarchical system [11].
Mobile-bearing use was examined in a prespecified subgroup of condylar-constrained revisions in which bearing design could be classified as fixed or mobile.
Implant selection variables
The aspects of implant selection examined were constraint, metaphyseal fixation, intramedullary stem fixation, and bearing design. Constraint was determined from the bearing and/or femoral component and classified as unconstrained, posterior-stabilized (PS), condylar-constrained (CCK), or hinge, according to NJR definitions [12]. Metaphyseal fixation was classified as sleeves, cones, or neither at the construct level. Intramedullary stem characteristics included use, design, and length. Stem length was dichotomized as ≤ 75 mm (short) or > 75 mm (long), consistent with previous literature [2,13]. Constructs were considered long-stemmed if any stem exceeded 75 mm and short-stemmed if all stems were ≤ 75 mm. Bearing design was classified as fixed or mobile from registry descriptions. Manufacturer was identified from bearing, femoral component, or tibial tray details, as available.
Statistics
A statistical analysis plan was published a priori and is publicly available [14]. Analyses were conducted using Stata version 18.0 (StataCorp LLC, College Station, TX, USA).
Descriptive analyses and temporal trends
Categorical variables were summarized using counts and proportions, and continuous variables using means with standard deviations (SD) and medians with interquartile ranges (IQR). Temporal trends in constraint, stem use and design, MFD use, and bearing design were presented graphically and quantitatively using annual counts and proportions.
Multilevel regression analyses
Binary mixed-effects logistic regression models addressed the secondary study objectives. Separate models examined 4 binary outcomes: MFD use, hinged construct use, absence of intramedullary stem fixation, and mobile-bearing use. An odds ratio (OR) > 1 therefore indicated greater odds of the specified outcome.
Patient-, surgeon-, and institution-level variables were selected a priori based on previous literature and clinical plausibility. These comprised age, sex, American Society of Anesthesiologists (ASA) grade, survival time of the primary implant, revision indication, mean annual surgeon revision volume, mean annual unit revision volume, and funding provider. ASA grades 3 and 4 were combined (ASA 3–4). Mean annual surgeon and unit revision volumes were calculated from a separate NJR dataset of 112,587 unlinked rTKA procedures using a previously described method [15]. Surgeon volume was dichotomized as ≥ 9 vs < 9 cases annually, based on evidence identifying ≥ 9 cases as an outcome-related threshold for aseptic rTKA [16]. Fixed-effect estimates were mutually adjusted and reported as adjusted OR) with 95% confidence intervals (CIs).
Random intercepts were specified for institution and surgeon nested within institution. This approach was taken because implant selection may reflect both individual surgical practice and institution-specific factors, such as local procurement policies and implant availability.
Sensitivity analyses
Models were repeated with additional adjustment for implant manufacturer to assess the extent to which manufacturer accounted for surgeon- and institution-level variation. Manufacturer adjustment was not undertaken for mobile-bearing use because of substantial collinearity between bearing design and manufacturer. All models were also repeated among surgeons operating at a single institution to assess the influence of cross-institutional practice on variance estimates.
Variance partitioning and explanatory power
Residual clustering was summarized using the intraclass correlation coefficient (ICC). Higher-level variance was defined as the combined variance attributable to the institution and surgeon-within-institution random effects. The relative institutional and surgeon-within-institution contributions were calculated as each variance component divided by this combined higher-level variance. Marginal R² represented the variation explained by the fixed effects, whereas conditional R² represented the variation explained by the full model, incorporating both fixed and random effects.
Model diagnostics and missing data
Model adequacy was assessed through convergence and the magnitude, precision, and stability of parameter estimates. Linearity of continuous predictors on the logit scale was assessed using LOWESS-based diagnostics. Multicollinearity was evaluated using variance inflation factors, all of which were < 5. There was no missing data for variables entering each model; differences in sample size reflected prespecified eligibility for the relevant implant-selection analysis.
Ethics, registration, funding, and disclosures
The study was approved by the NJR research committee (reference RSC2024/13) and the Healthcare Quality Improvement Partnership Data Access Request Group (reference HQIP479). Patient consent for data collection was obtained by the NJR. Ethical approval was not required, as stipulated by the Health Research Authority (HRA) guidance [17]. BT is jointly funded by the National Joint Registry, Orthopaedic Research UK, and the Royal College of Surgeons of England through a 2-year research fellowship. Complete disclosure of interest forms according to ICMJE are available on the article page, doi: 10.2340/17453674.2026.47125
Results
Study population
Data was available for 40,796 first-time linked revisions of pTKA. After exclusions, the principal cohort comprised 12,950 patients (12,950 knees), with procedures undertaken by 1,250 surgeons across 402 institutions. The prespecified analysis of mobile-bearing use included 6,842 procedures, undertaken by 925 surgeons across 354 institutions (Figure 1).
Figure 1.

Cohort flowchart. CCK = condylar-constrained knee; DAIR = debridement and implant retention; MFD = metaphyseal fixation device; OA = osteoarthritis; PE = polyethylene.
Mean age was 69 years (SD 9.3), 7,367 (57%) patients were female, and most were ASA grade 2 (8,548; 66%) or 3 (3,594; 28%) (Table 1). Revision occurred a mean 5.4 years (SD 4.0) after primary arthroplasty, most commonly for aseptic loosening (6,750; 52%), malalignment (2,120; 16%), or instability (2,032; 16%). Most procedures were undertaken by high-volume surgeons (7,743; 60%) and were NHS funded (11,902; 92%). Full-component revision accounted for 10,930 (84%) procedures. Condylar-constrained implants were used in 6,979 (54%) revisions and hinged implants in 1,785 (14%). Overall, 1,760 (14%) procedures were performed without intramedullary stems and 3,096 (24%) used an MFD. The condylar-constrained subgroup had broadly similar characteristics, although full-component revision, stem use, and MFD use were more common (Table 1).
Table 1.
Cohort characteristics. Values are count (%) or as specified
| Characteristic | Primary analysis n = 12,950 | CCK subgroup n = 6,842 |
|---|---|---|
| Age | ||
| mean (SD) | 69 (9.3) | 69 (8.9) |
| median (IQR) | 69 (13.0) | 69 (12.3) |
| Sex | ||
| Female | 7,367 (57) | 3,758 (55) |
| Male | 5,583 (43) | 3,084 (45) |
| ASA grade | ||
| 1 | 711 (5.5) | 338 (4.9) |
| 2 | 8,548 (66) | 4,469 (65) |
| 3 | 3,594 (28) | 1,984 (29) |
| 4 | 97 (0.7) | 51 (0.7) |
| Years since primary | ||
| mean (SD) | 5.4 (4.0) | 5.8 (4.1) |
| median (IQR) | 4.3 (5.4) | 4.8 (5.6) |
| Indication | ||
| Aseptic loosening | 6,750 (52) | 3,801 (56) |
| Malalignment | 2,120 (16) | 1,155 (17) |
| Instability | 2,032 (16) | 953 (14) |
| Stiffness | 906 (7.0) | 417 (6.1) |
| Unexplained pain | 697 (5.4) | 293 (4.3) |
| Component wear | 445 (3.4) | 223 (3.3) |
| Mean annual surgeon volume | ||
| < 9 | 5,207 (40) | 2,558 (37) |
| ≥ 9 | 7,743 (60) | 4,284 (63) |
| Mean annual unit volume | ||
| mean (SD) | 43.9 (39.5) | 43.5 (38.3) |
| median (IQR) | 29.3 (39.6) | 30.4 (38.6) |
| Funding | ||
| National Health Service | 11,902 (92) | 6,355 (93) |
| Independent | 1,048 (8.1) | 487 (7.1) |
| Manufacturer | ||
| DePuy | 4,728 (37) | 2,719 (40) |
| Zimmer Biomet | 4,010 (31) | 1,851 (27) |
| Stryker | 1,835 (14.) | 1,249 (18) |
| Smith & Nephew | 1,560 (12) | 992 (15) |
| Other | 817 (6.3) | 31 (0.5) |
| Revision type | ||
| Full component | 10,930 (84) | 6,582 (96) |
| Tibia only | 1,315 (10) | 116 (1.7) |
| Femur only | 705 (5.4) | 144 (2.1) |
| Constraint | ||
| Unconstrained | 1,251 (9.7) | – |
| Posterior-stabilized | 2,935 (23) | – |
| Condylar-constrained | 6,979 (54) | 6,842 (100) |
| Hinge | 1,785 (14) | – |
| Intramedullary stem use | ||
| No stems | 1,760 (14) | 207 (3.0) |
| Long cementless | 5,134 (40) | 2,972 (43) |
| Long cemented | 1,859 (14) | 1,289 (19) |
| Short cementless | 2,288 (18) | 1,405 (21) |
| Short cemented | 1,358 (11) | 558 (8.2) |
| Combination of stem designs | 551 (4.3) | 411 (6.0) |
| Metaphyseal fixation device | ||
| None | 9,853 (76) | 4,600 (67) |
| Sleeves | 2,714 (21) | 1,983 (29) |
| Cones | 382 (2.9) | 259 (3.8) |
| Cone/sleeve combination | 1 (0.0) | 0 (0.0) |
| Bearing design | ||
| Fixed | – | 4,453 (65) |
| Mobile | – | 2,389 (35) |
ASA = American Society of Anesthesiologists; CCK = condylar-constrained knee.
Temporal trends in implant selection
Temporal trends are shown in Figure 2 and Table S1 (see Supplementary data). Use of intramedullary stems increased over time, with procedures performed without stems decreasing from 28% in 2009 to 6.9% in 2023. Use of MFDs increased from 8.5% to 37%. Long cementless stems remained the most common design but declined from 53% to 32%, while sleeve use increased from 8.2% to 28%. Use of short cemented stems increased from 3.4% to 13%, alongside an increase in cone use from 0.2% to 8.3%.
Figure 2.

Temporal changes in implant selection (2009–2023). Analysis on bearing design comprises only condylar-constrained subgroup defined in Methods and Figure 1 (n = 6,842). Quantitative summary of data provided in Supplementary material.
Use of condylar-constrained implants increased from 35% to 66%, while unconstrained and posterior-stabilized implants declined from 21% to 4.3% and 35% to 12%, respectively. Hinged implant use increased from 8.7% to 17%. Within the condylar-constrained subgroup, mobile-bearing use increased from 21% to 45% (see Figure 2; Table S1).
Factors associated with implant selection
MFD use was greater among high-volume surgeons (OR 1.69, CI 1.27–2.23; P < 0.001), patients with ASA grade ≥ 3 (OR 1.35, CI 1.02–1.78; P = 0.04), and with increasing primary implant survival (OR 1.09 per year, CI 1.07–1.10; P < 0.001) (Table 2). Compared with aseptic loosening, MFD use was lower for all other indications, particularly component wear (OR 0.35, CI 0.24–0.50; P < 0.001) and unexplained pain (OR 0.36, CI 0.27–0.49; P < 0.001).
Table 2.
Results of the binary multilevel logistic regression models. Values are odds ratio with 95% confidence interval and P value or as specified
| Factor | MFD use | Hinge use | No stem fixation | Mobile-bearing use a |
|---|---|---|---|---|
| Patient and procedural factors | ||||
| Age, per year | 1.00 (0.99–1.01) 0.9 | 1.04 (1.03–1.05) < 0.001 | 0.99 (0.98–0.99) < 0.001 | 1.00 (0.99–1.01) 0.9 |
| Primary implant survival, per year | 1.09 (1.07–1.10) < 0.001 | 1.03 (1.02–1.05) < 0.001 | 0.95 (0.93–0.97) < 0.001 | 1.07 (1.04–1.11) < 0.001 |
| Sex | ||||
| Female | Reference | Reference | Reference | Reference |
| Male | 1.09 (0.97–1.22) 0.2 | 0.62 (0.55–0.70) < 0.001 | 1.16 (1.03–1.30) 0.02 | 1.08 (0.87–1.32) 0.5 |
| ASA grade | ||||
| 1 | Reference | Reference | Reference | Reference |
| 2 | 1.10 (0.85–1.43) 0.5 | 1.00 (0.74–1.36) 0.9 | 0.76 (0.60–0.96) 0.02 | 0.85 (0.53–1.37) 0.5 |
| 3–4 | 1.35 (1.02–1.78) 0.04 | 1.60 (1.16–2.20) 0.004 | 0.51 (0.30–0.66) < 0.001 | 1.23 (0.74–2.04) 0.4 |
| Revision indication | ||||
| Aseptic loosening | Reference | Reference | Reference | Reference |
| Malalignment | 0.69 (0.58–0.82) < 0.001 | 1.27 (1.05–1.54) 0.01 | 1.11 (0.93–1.34) 0.3 | 0.73 (0.54–1.00) 0.046 |
| Component wear | 0.35 (0.24–0.50) < 0.001 | 1.92 (1.42–2.59) < 0.001 | 2.97 (2.20–4.00) < 0.001 | 0.41 (0.22–0.77) 0.006 |
| Instability | 0.65 (0.54–0.77) < 0.001 | 5.96 (5.07–7.00) < 0.001 | 1.17 (0.97–1.40) 0.1 | 0.66 (0.48–0.92) 0.01 |
| Stiffness | 0.62 (0.49–0.78) < 0.001 | 1.79 (1.39–2.32) < 0.001 | 2.23 (1.80–2.76) < 0.001 | 0.80 (0.51–1.26) 0.4 |
| Unexplained pain | 0.36 (0.27–0.49) < 0.001 | 0.62 (0.43–0.91) 0.01 | 3.98 (3.18–4.98) < 0.001 | 0.76 (0.45–1.27) 0.3 |
| Surgeon and institutional factors | ||||
| High-volume surgeon (≥ 9 revisions/year) | 1.69 (1.27–2.23) < 0.001 | 1.18 (0.96–1.45) 0.1 | 0.58 (0.48–0.72) < 0.001 | 0.72 (0.40–1.32) 0.3 |
| Mean unit volume, revision/year | 1.00 (0.99–1.01) 0.5 | 1.01 (1.00–1.01) 0.007 | 1.00 (0.99–1.01) 0.9 | 1.01 (0.99–1.02) 0.5 |
| Funding | ||||
| NHS | Reference | Reference | Reference | Reference |
| Independent | 1.19 (0.83–1.71) 0.3 | 1.39 (1.04–1.880) 0.03 | 0.88 (0.67–1.17) 0.4 | 1.28 (0.59–2.78); 0.5 |
| Variance partitioning and explanatory power | ||||
| Institutional share of higher-level variance, % | 54 | 32 | 22 | 63 |
| Surgeon-within-institution share of higher-level variance, % | 45 | 68 | 78 | 37 |
| Overall ICC | 0.656 | 0.339 | 0.353 | 0.897 |
| Marginal R² | 0.036 | 0.145 | 0.068 | 0.007 |
| Conditional R² | 0.668 | 0.435 | 0.397 | 0.898 |
Model comprises only the subgroup of condylar-constrained revisions described in the Methods and Figure 1 (n = 6,842).
Use of hinged implants increased with age (OR 1.04 per year, CI 1.03–1.05; P < 0.001) and was lower in males (OR 0.62, CI 0.55–0.70; P < 0.001). Revision indication showed the strongest association, with instability associated with substantially greater hinge use than aseptic loosening (OR 5.96, CI 5.07–7.00; P < 0.001).
Absence of intramedullary stem fixation was less common among high-volume surgeons (OR 0.58, CI 0.48–0.72; P < 0.001), older patients, those with ASA grade ≥ 3 (OR 0.51, CI 0.39–0.66; P < 0.001), and with longer primary implant survival (OR 0.95 per year, CI 0.93–97; P < 0.001). Compared with aseptic loosening, stemless revision was more common for component wear, stiffness, and unexplained pain.
Associations with mobile-bearing use were generally small or imprecise. Longer primary implant survival was associated with greater use (OR 1.07 per year, CI 1.04–1.11; P < 0.001), while use was lower for component wear, instability, and malalignment than for aseptic loosening.
Implant selection varied substantially by manufacturer, particularly for MFDs, which were used predominantly with DePuy Synthes (Warsaw, IN, USA) implants, while differences by revision indication were most pronounced for constraint level and the use of intramedullary stems (Tables S2 and S3 see Supplementary data).
Surgeon- and institution-related variation in implant selection
Clustering of implant selection by surgeon and institution was greatest for mobile-bearing use (ICC 0.897) and MFD use (ICC 0.656), compared with absence of stem fixation (ICC 0.353) and hinge use (ICC 0.339). Measured fixed effects explained little of the variation in mobile-bearing use (marginal R² 0.007), MFD use (0.036), and absence of stem fixation (0.068), but more for hinge use (0.145).
For mobile-bearing use, 63% of higher-level variance occurred between institutions and 37% between surgeons within institutions. For MFD use, the corresponding proportions were 54% and 46%. Hinge use and absence of stem fixation varied predominantly between surgeons within the same institution, accounting for 68% and 78% of higher-level variance, respectively.
In practical terms, mobile-bearing and MFD use were therefore particularly associated with where, and by whom, the revision was performed. After adjustment for measured patient and clinical factors, variation in hinge use and stem fixation was concentrated predominantly between individual surgeons practising within the same institution.
Sensitivity analyses
After adjustment for manufacturer, manufacturer was strongly associated with MFD and hinge use but had little influence on absence of stem fixation (Tables S4–S6, see Supplementary data). For MFD use, marginal R² increased from 0.036 to 0.450, ICC fell from 0.656 to 0.497, and the institutional share of higher-level variance decreased from 54% to 31%. For hinge use, marginal R² increased from 0.145 to 0.360 while ICC remained similar (0.339 vs 0.365). Analyses restricted to surgeons operating at a single institution were broadly consistent with the primary analyses.
In practical terms, differences in implant manufacturer accounted for a substantial proportion of institutional variation in MFD use. Manufacturer was also associated with hinge use, but substantial variation between surgeons and institutions remained after adjustment. Similar results among surgeons working at a single institution suggest that the overall findings were not driven by surgeons practising across multiple hospitals.
Discussion
In this large registry-based study of first-time elective aseptic rTKA, implant selection changed substantially over time and varied markedly between surgeons and institutions. Measured patient and clinical characteristics explained relatively little of this variation. Mobile-bearing use varied predominantly between institutions, MFD use at both institutional and surgeon-within-institution levels, and hinge and stem use predominantly between surgeons within the same institution.
Temporal trends in implant selection at rTKA are poorly described. Use of MFDs has increased substantially in the United States, from 7.5% of first-time aseptic rTKA procedures in 2008 to 43% in 2022, with similar trends reported by the AOANJRR, where use increased from 3.6% of revisions for aseptic loosening in 2003 to 43% in 2024 [6,9]. International data describing temporal trends in constraint selection remains limited, although CCK is currently the most frequently used constraint level in Switzerland and Norway, while PS implants remain most prevalent at first-time aseptic rTKA in Australia [6,18,19]. Our findings also suggest changes in fixation strategy; however, the observed reduction in long cementless stems should not be interpreted as an equivalent reduction in diaphyseal fixation, as a construct comprising a short cementless stem and sleeve may still achieve metaphyseal–diaphyseal fixation, for example [20]. Metaphyseal cones are frequently used with short cemented stems, which may partly explain the parallel increase in use of these components over the study period [21].
A recent multi-center study identified a diagnosis of aseptic loosening, increased lead time from primary arthroplasty to first revision, and ASA grade ≥ 3 to be associated with Anderson Orthopaedic Research Institute (AORI) type ≥ 2B bone loss at first-time aseptic rTKA [22]. These factors were also associated with greater MFD use in the present study. However, they explained relatively little of the observed variation, which remained substantial at both institutional and surgeon-within-institution levels. Sensitivity analyses suggested that manufacturer accounted for a large proportion of this variation, highlighting the potential influence of wider industry and market factors on implant selection. Given the increasing use and potential cost implications of MFDs, further research is needed to understand this variation and support more consistent, evidence-based use at first-time aseptic rTKA.
In contrast to MFD use, institutional variation in hinge selection was less marked, and the ICC changed little after inclusion of implant manufacturer in the model despite a substantial increase in marginal R². This statistical pattern is consistent with manufacturer choice occurring later in the decision pathway, for example where surgeons change implant systems to access a preferred hinge design or one procured by the institution. Revision indication was strongly associated with hinge use, most notably for instability. Data from the AOANJRR similarly shows that hinged implants account for 33% of first-time revisions performed for instability [5]. Higher-level variation in hinge selection occurred predominantly between surgeons working within the same institution. This may reflect unmeasured differences in case-mix between surgeons or differing individual surgical thresholds for utilizing hinged prostheses.
Younger patients, men, and those undergoing early first-time rTKA following primary arthroplasty are known to be at an increased risk of re-revision [5]. Our study highlights that these groups were more often managed without intramedullary stem fixation. Absence of intramedullary stem fixation is known to be associated with an increased risk of re-revision [6,7]. Our findings highlight this strategy was more common among low-volume surgeons; however, substantial variation also occurred between surgeons practicing within the same institution after adjusting for surgeon volume. This suggests that individual operative practice plays an important role in stem selection and, in the context of previous evidence, supports multidisciplinary team discussion when a stemless strategy is being considered, particularly in patients already at elevated risk of further revision surgery.
Our study also demonstrates a substantial shift in bearing design among condylar-constrained revisions, with increasing use of mobile-bearing constructs over time. Bearing selection showed marked institutional variation, most likely explained by local procurement arrangements. High-volume surgeons likely performed more complex cases. However, the associations between surgeon volume and some aspects of implant selection remained evident after mutual adjustment for the measured patient and clinical characteristics. First-time aseptic rTKA procedures undertaken by surgeons performing ≥ 9 revisions annually have been shown to be associated with a significantly lower re-revision risk (OR 0.77 CI 0.62–0.95, P = 0.02) [16]. Whilst the reasons for this are likely multifactorial, our study provides insight into 1 potential mechanism to explain these findings. Future studies might examine whether the observed differences in implant selection between low- and high-volume surgeons are associated with subsequent re-revision risk.
Strengths
This study is strengthened by its large sample size but has several limitations. As with any registry-based study, some misclassification may have arisen from incorrect or incomplete data.
Limitations
The NJR does not record several factors that may influence implant selection, including bone loss, ligament competence, and intraoperative findings, so some surgeon- or institution-level variation may reflect unmeasured case complexity. Nevertheless, the magnitude of variation, particularly for MFD and mobile-bearing use, suggests that higher-level practice patterns remain important. Surgeons were modelled as nested within institutions, although some operated across more than 1 institution; however, restricting analyses to single-institution surgeons did not materially alter the conclusions. The NJR also does not routinely record the tibial or femoral distribution of stem use or the fixation technique used for implanted stems. Finally, given the large sample size, regression analyses were highly powered, such that small differences may have reached statistical significance despite limited clinical importance.
Conclusion
We showed that implant selection at first-time elective aseptic rTKA changed substantially over time, with increasing use of MFDs, mobile-bearing designs, and higher-constraint constructs. Measured patient and clinical characteristics explained relatively little of the examined implant decisions, with substantial variation occurring at surgeon and institutional levels.
In perspective, further research is needed to examine the factors that influence surgeons’ implant-selection decisions and to inform more consistent, evidence-based practice in this setting.
Supplementary data
Tables S1–S6 are available as supplementary data on the article page, doi: 10.2340/17453674.2026.47125
Supplementary Material
Funding Statement
BT is jointly funded by the National Joint Registry, Orthopaedic Research UK, and the Royal College of Surgeons of England through a 2-year research fellowship.
BT: conceptualization, methodology, formal analysis, investigation, data curation, writing—original draft, writing—review and editing, visualization, project administration, funding acquisition. CH, DD, PB: conceptualization, methodology, writing—review and editing, supervision.
The authors would like to thank the patients and staff of all the hospitals who have contributed data to the National Joint Registry. They are grateful to the Healthcare Quality Improvement Partnership (HQIP), the NJR Research Committee and staff at the NJR for facilitating this work. The authors have conformed to the NJR’s standard protocol for data access and publication. The views expressed represent those of the authors and do not necessarily reflect those of the National Joint Registry Steering Committee, Research Sub-committee or the Healthcare Quality Improvement Partnership (HQIP), who do not vouch for how the information is presented.
Handling co-editors: Keijo Mäkelä and Robin Christensen
Acta thanks Scott Crawford and Simon Kovač for help with peer review of this manuscript.
References
- 1.Hald J T, Fenstad A M, El-Galaly A, Odgaard A, Furnes O N. Similar incidence and implant survival of multiply revised knee arthroplasties in Norway and Denmark: a register study of the Danish and Norwegian Knee Arthroplasty registries from 1998–2021. Acta Orthop 2026; 97: 110-16. doi: 10.2340/17453674.2026.45411. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Zhang J, Li E, Zhang Y. Prostheses option in revision total knee arthroplasty, from the bench to the bedside: (1) basic science and principles. EFORT Open Rev 2022; 7: 174-87. doi: 10.1530/EOR-21-0089. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.James Lind Alliance/National Institute for Health and Care Research . Problematic knee replacement Top 10 Priorities. 2020. Available from: https://www.jla.nihr.ac.uk/priority-setting-partnerships/Revision-knee-replacement
- 4.BASK Revision Knee Working Group . A guide to good practice for revision knee replacement. BASK; 2023. Available from: https://bask.ac.uk/wp-content/uploads/2025/07/Revision_Knee_Guide_to_Good_Practice.pdf [Google Scholar]
- 5.Deere K, Whitehouse M R, Kunutsor S K, Sayers A, Price A J, Mason J, et al. How long do revised and multiply revised knee replacements last? A retrospective observational study of the National Joint Registry. Lancet Rheumatol 2021; 3: e438-e446. doi: 10.1016/S2665-9913(21)00079-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Lewis P L, Gill D R, McAuliffe M J, Stoney J D, Vertullo C J, Wall C J, et al. Hip, knee and shoulder arthroplasty: 2025 annual report, Australian Orthopaedic Association National Joint Replacement Registry. Adelaide: AOA; 2025. doi: 10.25310/MXFR3061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Hoskins W, Gusho C, McAuliffe M, Wall C, Xu Q, Vince K G. The AAHKS Surgical Techniques and Technologies Award: what is the ideal technique and implant in aseptic revision total knee arthroplasty? An analysis of second revision and component loosening rates from the Australian Orthopaedic Association National Joint Replacement Registry. J Arthroplasty 2026; 41: S11-S20.e9. doi: 10.1016/j.arth.2026.05.046. [DOI] [PubMed] [Google Scholar]
- 8.Tyas B, Hewitt C, Deehan D, Baker P. Condylar-constrained prostheses are associated with favourable implant survival following first-time aseptic revision knee arthroplasty: an analysis from the National Joint Registry. Knee Surg Sports Traumatol Arthrosc 2026; 1-17. doi: 10.1002/ksa.70548. [DOI] [PubMed] [Google Scholar]
- 9.Hinman A D, Chang R N, Kelly M P, Fasig B H, Paxton E W, Royse K E. Impact of metaphyseal sleeves and porous cones on risk reduction for revision total knee arthroplasty: aseptic re-revision in a United States integrated health care system. J Arthroplasty 2025; 40: 2375-80.e1. doi: 10.1016/j.arth.2025.02.050. [DOI] [PubMed] [Google Scholar]
- 10.von Elm E, Altman D G, Egger M, Pocock S J, Gøtzsche P C, Vandenbroucke J P. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Int J Surg 2014; 12: 1495-9. doi: 10.1016/j.ijsu.2014.07.013 [DOI] [Google Scholar]
- 11.Australian Orthopaedic Association National Joint Replacement Registry . Hip, knee & shoulder arthroplasty: 2010 annual report. Adelaide: AOA; 2010. Available from: https://aoanjrr.sahmri.com/annual-reports-2010 [Google Scholar]
- 12.National Joint Registry . National Joint Registry 22nd Annual Report 2025. 2025. Available from: https://reports.njrcentre.org.uk/ [PubMed]
- 13.Abdelatif E A, Abu Mukh A, Elsaid A N S, Youssef A O, Foissey C, Servien E, et al. Inferior outcome of stand-alone short versus long tibial stem in revision total knee arthroplasty: a retrospective comparative study with minimum 2 year follow-up. SICOT J 2025; 11: 3. doi: 10.1051/sicotj/2024054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Tyas B, Hewitt C, Deehan D, Baker P. Statistical analysis plan for investigating implant selection and patient outcomes in elective aseptic revision knee arthroplasty using linked NJR–HES–CivReg–PROMs data. Open Science Framework; 2025. doi: 10.17605/OSF.IO/X7S58. [DOI] [Google Scholar]
- 15.Valsamis E M, Collins G S, Pinedo-Villanueva R, Whitehouse M R, Rangan A, Sayers A, et al. Association between surgeon volume and patient outcomes after elective shoulder replacement surgery using data from the National Joint Registry and Hospital Episode Statistics for England: population based cohort study. BMJ 2023; 381: e075355. doi: 10.1136/bmj-2023-075355. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Matthews A H, Gray W K, Evans J P, Evans J T, Lamb S E, Porteous A, et al. Higher surgeon volume reduces early failure in first time revision of non-infected total knee arthroplasty: an analysis using data from the United Kingdom National Joint Registry. Knee Surg Sports Traumatol Arthrosc 2025; 33: 3286-97. doi: 10.1002/ksa.12690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Health Research Authority . Do I need NHS ethics approval? 2025. Available from: https://www.hra-decisiontools.org.uk/ethics/
- 18.Swiss National Hip & Knee Joint Registry . Annual report of the SIRIS Registry Hip & Knee, 2012–2024. 2025. Available from: https://www.siris-implant.ch/de/Downloads&category=16
- 19.Norwegian Arthroplasty Register . Norwegian National Network for Arthroplasty and Hip Fractures. 2025. Available from: https://www.helse-bergen.no/nasjonalt-kvalitets-og-kompetansenettverk-for-leddproteser-og-hoftebrudd/arsrapporter/#annual-reports
- 20.Nadorf J, Kinkel S, Gantz S, Jakubowitz E, Kretzer J P. Tibial revision knee arthroplasty with metaphyseal sleeves: the effect of stems on implant fixation and bone flexibility. PLoS One 2017; 12: e0177285. doi: 10.1371/journal.pone.0177285. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Behery O A, Shing E Z, Yu Z, Springer B D, Fehring T K, Otero J E. Survivorship and radiographic evaluation of metaphyseal cones with short cemented stems in revision total knee arthroplasty. J Arthroplasty 2022; 37: 330-5. doi: 10.1016/j.arth.2021.10.027. [DOI] [PubMed] [Google Scholar]
- 22.Tyas B, Hewitt C, Deehan D, Walker R, Kottam L, Baker P, et al. Major bone loss at elective aseptic revision of a primary total knee replacement is uncommon, but intraoperative defects are frequently underestimated pre-operatively. Knee 2026; 62: 104499. doi: 10.1016/j.knee.2026.104499. [DOI] [PubMed] [Google Scholar]
Associated Data