Simple Summary
Cone-beam computed tomography produces three-dimensional images that help dentists measure jawbone before implant placement. We reviewed eight experimental studies of linear measurements and retained two additional reports as background on image artifacts and gray values. Accuracy against a separate reference measurement was distinguished from agreement between scan protocols. Several studies reported small average differences when settings were changed on the same scanner, but the information needed to account reliably for repeated measurements within specimens was insufficient for the planned combined analysis. The revised review therefore presents study-level findings rather than a pooled equivalence claim. Preserved jaws with surrounding tissue sometimes showed larger errors than simpler models, although these comparisons cannot separate the effects of tissue from other study differences. None of the included studies involved living patients. Selected reduced-exposure protocols may preserve average measurements in laboratory conditions, but reliable visualization of critical anatomy and clinical validation remain essential.
Keywords: cone beam computed tomography, dental implants, radiation dosage, alveolar bone, systematic review
Abstract
Background: This review examined whether within-device exposure-setting or nominal voxel-size changes in cone-beam computed tomography (CBCT) affect linear alveolar bone measurements for implant planning, separating accuracy from agreement. Methods: Database-specific search strategies and retrieval counts are reported for PubMed/MEDLINE, Ovid Embase, Scopus, and Web of Science Core Collection. Study selection is presented in the updated PRISMA flow diagram. The reported International Prospective Register of Systematic Reviews (PROSPERO) identifier is CRD420261361509. Studies were grouped by outcome, protocol contrast, and experimental model. The Quality Assessment Tool for In Vitro Studies (QUIN) appraisal was performed by one reviewer. A revision-stage source and analysis-unit audit led to withdrawal of the planned inverse-variance meta-analyses and their ±0.5 mm equivalence claims; structured study-level synthesis was used instead. Results: Eight studies addressed linear measurements; two additional reports were retained only as methodological context. Seven linear-measurement studies compared protocols within a device, and one compared different devices. All were experimental. In one five-skull study, mean absolute errors were 0.50, 0.46, and 0.51 mm at 40, 20, and 7 s, respectively. A cadaver study reported mean absolute errors of approximately 1.10–1.20 mm. Small mean between-protocol differences sometimes coexisted with impaired anatomical visibility. Repeated observations were not equivalent to independent specimens, and several required sampling variances could not be reconstructed. Conclusions: Selected within-device protocol changes may preserve average measurements under controlled experimental conditions. The evidence does not establish statistical equivalence, interchangeability of protocol classes, or clinical safety near critical anatomy. Validation in living patients is needed before broad implementation.
1. Introduction
Cone-beam computed tomography (CBCT) is an established adjunct to clinical examination and two-dimensional radiography for dental implant planning [1]. Cross-sectional images help assess alveolar width, bone height above the mandibular canal or below the maxillary sinus, and proximity to critical anatomy [2]. Incorrect boundary identification can contribute to neurovascular injury or cortical perforation [3]. An error of approximately 1 mm is commonly used as a practical measurement-acceptability threshold in implant-imaging research [4,5,6,7]. It is not a universal threshold for every anatomical site or a prescribed surgical clearance.
The original analysis selected a stricter ±0.5 mm equivalence margin for the mean difference between protocols, with precedent in the sample-size calculation of Kaaber et al. [8]. This concerns a change in measurement performance, whereas a 1 mm acceptability threshold concerns error relative to a reference measurement. Neither value bounds individual errors or replaces a clinically justified safety allowance. Nominal voxel size describes reconstructed sampling, not effective spatial resolution; effective resolution also depends on the system, reconstruction, noise, and object contrast [9,10]. A universal 0.5 mm spatial-resolution limit therefore cannot justify equivalence.
Radiation exposure varies among CBCT devices and protocols [11,12]. Optimization principles include as low as reasonably achievable (ALARA), as low as diagnostically acceptable (ALADA), and as low as diagnostically acceptable, indication-oriented and patient-specific (ALADAIP) [13,14]. Professional recommendations support task-specific optimization [2,15,16]. Tube current, exposure duration, projection number, and rotation arc are exposure-related settings. Nominal voxel size is a separate sampling setting that may co-vary with exposure in manufacturer presets; increasing voxel size does not by itself establish a dose reduction [10,17].
Previous reviews have addressed linear-measurement accuracy, low-dose protocols across dental indications, and low-dose imaging across stages of implant therapy [17,18,19,20]. The present review offers an incremental methodological focus: it separates accuracy against an independent reference from between-protocol agreement, distinguishes within-device from between-device comparisons, and describes outcomes by experimental model. It does not replace those broader reviews.
Within-device comparison reduces confounding from scanner hardware and software but does not isolate a single parameter when a preset changes several settings together. Agreement between two CBCT protocols cannot establish accuracy because both may share bias [8]. Experimental models also differ in attenuation, scatter, and metal-related artifacts [21,22]. Between-study differences in error therefore cannot establish the independent effect of specimen condition.
The aim was to examine whether selected within-device exposure-setting reductions or nominal voxel-size increases affect linear alveolar bone measurements, keeping accuracy and agreement separate, and to describe differences in reported error across experimental settings.
2. Materials and Methods
2.1. Study Protocol and Registration
The review was reported using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses Supplementary File S1 (PRISMA) 2020 framework [23]. The registration identifier supplied with the manuscript is International Prospective Register of Systematic Reviews (PROSPERO) CRD420261361509. Registration timing and the dated historical amendment record were not available for verification; prospective registration is not claimed. The search strategies and study-selection flow are reported in Appendix B and Figure 1.
Figure 1.

Updated study-selection flow, adapted from Supplementary File S1 PRISMA 2020 [23] under CC BY 4.0. The diagram presents the database retrieval counts, screening and full-text dispositions, and the eight linear-measurement studies and two contextual reports. Database-specific search strategies are provided in Appendix B.
The following changes were made during manuscript revision, rather than described as prespecified: source-to-estimate auditing (Table A1); correction of study citations and outcome labels; separation of two non-linear contextual reports from eligible linear-measurement studies; replacement of the original inverse-variance meta-analyses with structured study-level synthesis; and removal of equivalence and unverified sensitivity-analysis claims. These changes are reported here but have not been represented as updates already entered in PROSPERO.
2.2. Eligibility Criteria
Eligible studies reported linear alveolar bone measurements in millimeters and compared two or more acquisition conditions within a CBCT unit, compared CBCT measurements with an independent reference, or both. Patients, cadavers, dry human bone, animal bone, and phantoms were eligible. Specimen type and the number of independent experimental units were recorded. Studies lacking a reconstructible sampling variance remained eligible for descriptive synthesis; insufficient variance information was not treated as evidence that their findings were absent.
For accuracy, an independent reference was a measurement obtained separately from the CBCT examination being evaluated, such as caliper measurement of sectioned specimens, histological measurement, micro-computed tomography, multidetector computed tomography (MDCT), or direct surgical measurement. MDCT is independent in acquisition but is not automatically an error-free reference: suitability depends on validation for the same linear task, sufficient resolution, and co-registration. Another CBCT protocol was classified as a comparator for agreement, not an independent accuracy standard. No MDCT-referenced linear outcome contributed to the present synthesis; the MDCT comparator in Pauwels et al. concerned gray values only [24].
Reviews were used to identify candidate reports, not as primary data sources. Corrections were linked to their primary reports rather than counted as separate studies; the exposure-setting correction to Kaaber et al. was incorporated [8]. Reports restricted to gray values, artifact magnitude, subjective image quality, or surgical placement deviation without an eligible linear measurement were excluded from the linear-measurement synthesis. Pauwels et al. and Shokri et al., previously counted among ten included studies, are retained solely as contextual reports [24,25]. Absence of an independent reference alone does not exclude an otherwise eligible within-device agreement study.
Outcomes were kept separate: accuracy denotes linear measurement error against an independent reference; agreement denotes a signed difference between CBCT protocol measurements. A study could inform both domains when it reported both outcomes, but they were not combined. Gray-value and artifact outcomes were described only as context. Mean absolute error, signed bias, and the difference between marginal means were not treated as interchangeable statistics.
2.3. Information Sources and Search
Appendix B presents the database-specific search strategies and retrieval counts for PubMed/MEDLINE, Ovid Embase, Scopus, and Web of Science Core Collection. PubMed was searched on 28 October 2026 and Ovid Embase on 29 October 2026.
The repeat strategies combine CBCT, implant or alveolar anatomy, and linear-measurement terms. PubMed includes a MeSH term; the database-specific fields are reproduced in Table A2. Additional accuracy, agreement, reliability, and related performance terms are required in Embase, Scopus, and Web of Science, but not in PubMed. PubMed is limited to publications dated 1 January 2000 through 28 October 2026; the other expressions specify calendar years 2000–2026.
The eligibility criteria distinguish accuracy against an independent reference from agreement between CBCT acquisition protocols. An otherwise eligible agreement study does not require an independent reference standard. Figure 1 presents the study-selection flow, and Table A3 summarizes the full-text dispositions recorded in that flow.
2.4. Data Extraction
The original manuscript reports extraction by three authors using a piloted form. During revision, author names, years, titles, and digital object identifiers (DOIs) were reconciled across the text, tables, and bibliography. The audit distinguished a directly reported statistic from a calculation using published summaries and recorded its source location and unit of analysis. Appendix A reports the disposition of all seven original meta-analysis inputs.
Substantive corrections included distinguishing Al-Ekrish’s 55 designated sites from 83 matched height/width outcomes in five skulls; recognizing Ganguly’s 28 table rows as repeated location/measurement summaries rather than 28 independent specimens; and distinguishing Kaaber’s 42 sites from 26 mandibles [6,8,26]. The latter study’s high-resolution versus low-dose contrast was not its most extreme VISO comparison because an ultra-low-dose preset was also tested. Unverifiable paired standard errors (SEs) were removed, not relabeled as specimen-adjusted estimates.
The number of studies retained reflects the evidence identified under the eligibility criteria; a conventional participant-level sample-size calculation is not applicable to this review’s study count. In the primary studies, specimen counts must be distinguished from sites, images, and repeated examiner readings, which do not increase the number of independent specimens.
2.5. Risk of Bias
The original item-level appraisal used the Quality Assessment Tool for In Vitro Studies (QUIN) [27] and was performed by one reviewer; independent duplicate appraisal was not undertaken. No intra-reviewer or inter-reviewer reproducibility estimate is available for the QUIN appraisal. Reliability data from the primary studies concern image measurements and do not establish reproducibility of the review’s ratings. All included studies were experimental, so clinical-study tools and diagnostic-classification checklists were not used for their linear outcomes. The original recorded scores and item-level ratings are retained with their provenance in Figure 2, but the blanket conclusion that every study was at low risk of bias is withdrawn. The interpretation emphasizes sample-size planning, blinding, sampling, and independence of observations. A high checklist score is not a certainty-of-evidence assessment and does not resolve a unit-of-analysis error in the synthesis.
Figure 2.

Original item-level QUIN appraisal, with the original dot-matrix design retained [6,7,8,24,25,26,28,29,30,31]. These single-reviewer judgments have not been independently reassessed. Pauwels and Shokri are contextual reports, not included in linear-measurement studies. Green, adequately specified (2); amber, inadequately specified (1); red, not specified (0); gray, not applicable. Reporting of a statistical method does not establish adequate handling of specimen dependence. No overall low-risk classification is inferred from the grid.
2.6. Synthesis and Handling of Dependent Measurements
The planned accuracy estimand was the within-device difference in mean absolute error (MAE): error under the higher-exposure/finer-voxel protocol minus error under the reduced-exposure/larger-voxel protocol. Negative values would favor the former. The agreement estimand was the signed difference between the measurements themselves in the same direction. A common unit of millimeters does not make these outcomes interchangeable or imply that exposure duration, tube current, and voxel size have equivalent effects.
The original extreme-contrast rule was intended to avoid selecting favorable results after inspecting outcomes and to examine the widest tested change, not to represent the protocol normally used in practice. The revision reports the available protocol-specific findings rather than selecting one unsupported pooled input per study. Bone height was not uniformly separable from width in the sources; combined outcomes are labeled accordingly. A larger reduction is not assumed to be a clinically acceptable default.
Two levels of dependence were distinguished: pairing of protocol measurements at the same location and correlation among locations or repeated observer readings within a specimen. Imputing the paired correlation between protocols does not correct clustering within skulls or mandibles [32]. Where only marginal standard deviations (SDs) were reported, the variance of a paired contrast required the corresponding covariance. Combining observer-, plane-, or device-specific estimates from shared specimens additionally required their joint covariance or specimen-linked data.
A valid specimen-based reanalysis could not be reconstructed for the original set of studies. Substituting the number of specimens for the number of sites while retaining a site-level SD would not reproduce the SD of specimen-level paired means. The original inverse-variance pooling, its confidence and prediction intervals, heterogeneity statistics, and the assertion of robustness in a specimen-based sensitivity analysis were therefore withdrawn. The review does not present a synthetic replacement for unavailable specimen-level results.
The revised synthesis follows a structured approach consistent with Synthesis Without Meta-analysis (SWiM) reporting [33]. Studies are grouped by accuracy versus agreement, exposure-related versus voxel-size changes, and experimental model. Verified means, SDs, and protocol comparisons are tabulated without weighting by site counts. Within-study contrasts and between-study setting comparisons are described separately. No study-counting rule based on statistical significance is used to establish a common effect. The ±0.5 mm margin is retained to explain the planned question, not to declare equivalence from point estimates or descriptive ranges. Neither a valid pooled effect nor a statistical test of equivalence is reported.
3. Results
3.1. Study Selection
The PRISMA flow diagram (Figure 1) reports 3361 records: 1815 from PubMed, 539 from Embase, 604 from Scopus, and 403 from Web of Science. It records 1819 duplicate removals, 1542 records screened, and 1527 records excluded at screening. Fifteen reports were sought for retrieval; three were not retrieved and 12 were assessed for eligibility. Two full-text reports were excluded, with the recorded reasons of no reference standard (n = 1) and non-English full text (n = 1). Eight linear-measurement studies and two contextual reports are presented separately in the review.
3.2. Study Characteristics
Eight studies addressed linear measurements: seven included within-device comparisons [6,7,8,26,28,29,30], and one compared cortical measurements across devices [31]. Five within-device studies used caliper references, although not all supplied a usable mean-absolute-error contrast; two provided agreement without an independent reference. Pauwels et al. and Shokri et al. were retained only as context [24,25]. The experimental models included phantoms, dry bone with or without soft-tissue simulation, preserved human specimens, and bovine bone. None of the included linear-measurement studies involved living patients. Table 1 identifies the actual reference and the repeated-observation structure.
Table 1.
Study characteristics and outcome classification. Eight linear-measurement studies and two contextual reports are shown separately.
| Study | Sample | Unit and Parameter Varied | Reference Standard | Domain |
|---|---|---|---|---|
| Al-Ekrish [26] | 5 skulls; 55 sites; 83 matched outcomes | Iluma; 40/20/7 s; 602/302/117 projections; 120 kV, 3.8 mA; 0.29 mm | Digital caliper | Accuracy |
| Waltrick et al. [7] | 12 hemimandibles; 108 images; 648 readings | i-CAT; nominal voxels 0.2/0.3/0.4 mm | Digital caliper | Accuracy |
| Vasconcelos et al. [29] | 8 mandibles; 8 landmark categories; 3 readers | Kodak 9000; 2–15 mA; 60 kV, 10.8 s; 0.2 mm | Digital caliper | Reference-based differences |
| Ganguly et al. [6] | 4 heads; 28 tabulated summaries; repeated readers | i-CAT; FOV 16 cm × 13 cm; 20 s; 0.3/0.2 mm; kV/mA not reported | Digital caliper | Accuracy |
| Liljeholm et al. [30] | 8 skulls; 160 paired observations | ProMax 3D; standard and 3 reduced presets; exposure/voxel co-vary | None independent | Agreement |
| de Castro et al. [28] | 5 polyurethane mandibles; repeated readings | PaX-i3D; 24 s/720 vs. 15 s/450 projections; 50 kV, 4 mA; 0.2 mm | Digital caliper | Accuracy/agreement |
| Kaaber et al. [8] | 26 mandibles; 42 sites; 3 readers; 2 height planes | X1/AXEOS/VISO; HR/SR/LD; VISO also ULD; corrected exposures [8] | None independent | Agreement |
| Razavi et al. [31] | 2 bovine ribs; 10 implants; 10 readers | i-CAT NG vs. Accuitomo; scanner and settings co-vary | Histology | Between-device |
| Pauwels et al. [24] | PMMA phantom; 13 devices; 30 protocols | Multiple devices and acquisition settings | MDCT gray values | Context only |
| Shokri et al. [25] | 27 bovine bone blocks | Cranex 3D; FOV/voxel co-vary; 4/10 mA | None independent | Context only |
FOV, field of view; PMMA, polymethyl methacrylate; HR, high-resolution preset; SR, standard-resolution preset; LD, low-dose preset; ULD, ultra-low-dose preset. AXEOS, VISO, and i-CAT are CBCT scanner trade names; NG, Next Generation. Preset names do not establish measured effective spatial resolution. Full observation structures and limitations are given in the text and Appendix A.
Acquisition methods varied across studies in scanner, exposure settings, and nominal voxel size (Table 1). Where a preset changed several settings together, the comparison describes their combined performance and cannot isolate a single acquisition parameter. These comparisons do not establish a common acquisition protocol across studies.
3.3. Risk of Bias
Recorded QUIN percentages ranged from 75% to 100%, but these single-reviewer scores should not be interpreted as uniformly low bias or high confidence. Recurring limitations included missing prospective sample-size calculations, unclear or absent blinding in several reports, and few independent specimens relative to the number of observations. Kaaber et al. reported sample-size planning and blinded, randomized image presentation, but its repeated sites and readers still required appropriate covariance handling for any combined estimate [8]. Vasconcelos et al. reported three assessors blinded to exposure parameters; this is retained rather than replaced by a blanket claim of absent blinding [29]. These study-specific procedures do not establish uniform examiner calibration or common image-evaluation criteria across studies; such standardization is not documented in the available review material. Table 2 distinguishes the recorded checklist results from the analysis-unit concerns identified during revision.
Table 2.
Original single-reviewer QUIN scores and limitations relevant to interpretation. Scores are retained as recorded, not as newly duplicated assessments or overall certainty ratings.
| Study | Score | Percentage | Principal Limitation |
|---|---|---|---|
| Kaaber et al. [8] | 24/24 | 100% | Sample-size planning and blinding reported; 42 sites nested within 26 mandibles |
| Waltrick et al. [7] | 23/24 | 95.8% | Protocol blinding not described in original appraisal; repeated readings are not independent |
| Liljeholm et al. [30] | 22/24 | 91.7% | No sample-size calculation in original appraisal; 160 observations from 8 skulls |
| Shokri et al. [25] | 20/24 | 83.3% | No sample-size calculation or blinding in original appraisal; contextual report |
| Vasconcelos et al. [29] | 20/24 | 83.3% | Sample-size planning and randomization limitations; assessors blinded to exposure |
| Pauwels et al. [24] | 18/22 | 81.8% | No sample-size calculation or blinding in original appraisal; contextual report |
| Razavi et al. [31] | 19/24 | 79.2% | Sample-size/blinding limitations in original appraisal; 2 ribs; between-device confounding |
| Ganguly et al. [6] | 18/24 | 75.0% | Post hoc power; sampling, randomization and blinding limitations; 4 heads |
| Al-Ekrish [26] | 18/24 | 75.0% | No prospective sample-size calculation, randomization or blinding in original appraisal; 5 skulls |
| de Castro et al. [28] | 18/24 | 75.0% | Single examiner; no blinding in original appraisal; 5 phantom mandibles |
3.4. Accuracy Findings and the Analysis-Unit Audit
The four studies originally entered in the accuracy meta-analysis investigated three parameter categories: exposure duration with projection number, tube current, and nominal voxel size in two studies [6,7,26,29]. In Al-Ekrish’s study, the mean absolute error was 0.50 mm (SD 0.47) at 40 s, 0.46 mm (SD 0.39) at 20 s, and 0.51 mm (SD 0.46) at 7 s [26]. The 40-minus-7 s difference in rounded means is −0.01 mm. These error summaries concern 83 matched outcomes, not 55 independent specimens.
Waltrick et al. reported an overall measurement error of 0.23 mm (SD 0.20) across the tested voxel sizes, with all observed errors below 1 mm [7]. The review’s previous −0.060 mm contrast and 0.034 mm SE were not retained because their source-cell derivation could not be verified. Ganguly et al. reported overall mean absolute errors of approximately 1.10–1.20 mm and no detected protocol difference in their reported tests [6]. The published aggregate rows do not permit recovery of a four-head paired variance. These are study-specific observations, not equivalence tests.
Vasconcelos et al. reported regional differences between image and caliper measurements for seven current settings [29]. An equal-weight average of the eight published category means is 0.58375 mm at 2 mA and 0.52375 mm at 15 mA, yielding −0.060 mm. This reproduces the arithmetic point contrast but neither establishes its labeling as mean absolute error nor supplies a valid specimen-based SE. The study also found a regional exception in the second-molar measurements and poorer image quality at the lowest current settings. Table 3 and Appendix A distinguish these findings from the withdrawn aggregate inference.
Table 3.
Within-device contrasts and independent units in the four original accuracy inputs. Only supported descriptive arithmetic is retained; no pooled effect is calculated.
| Study | Contrast | Difference (mm) | SE | Independent n | Weight |
|---|---|---|---|---|---|
| Al-Ekrish [26] | 40 s vs. 7 s exposure | −0.010 | — | 5 skulls | — |
| Waltrick et al. [7] | 0.2 vs. 0.4 mm voxel | NR | — | 12 hemimandibles | — |
| Vasconcelos et al. [29] | 15 vs. 2 mA | −0.060 † | — | 8 mandibles | — |
| Ganguly et al. [6] | 0.2 vs. 0.3 mm voxel | NR | — | 4 heads | — |
| Synthesis | Not pooled | — | — | — | — |
NR, original contrast not reverified; —, no SE or inverse-variance weight retained. Al-Ekrish’s −0.010 mm is the difference in rounded mean absolute errors. † Vasconcelos’s −0.060 mm is an equal-category mean of regional image–bone differences, not a validated MAE contrast. The outcome types are not pooled. Source means, SDs, and variance limitations are detailed above and in Appendix A.
The previously stated specimen-based and leave-one-out robustness results are not retained. A reproducible specimen-level sensitivity result was not available, and leave-one-study-out calculations using the same deficient SEs cannot validate them. Table 4 provides the supported agreement findings without a pooled interval; Appendix A records the numerical checks that motivated this change.
Table 4.
Within-device agreement findings, kept separate from accuracy. * Kaaber values are descriptive averages of 18 shared-sample height cells, not independent study estimates. NR, paired contrast not re-verified; —, no specimen-adjusted SE retained.
| Study | Comparison | Difference (mm) | Independent n | SE |
|---|---|---|---|---|
| Liljeholm et al. [30] | Standard minus 3 reduced presets | −0.0181; −0.0038; +0.0600 | 8 skulls | — |
| de Castro et al. [28] | 24 s/720 vs. 15 s/450 projections | NR | 5 mandibles | — |
| Kaaber et al. [8] | HR–LD, all units | +0.040 * | 26 mandibles | — |
| Kaaber et al. [8] | HR–LD, X1/AXEOS; HR–ULD, VISO | −0.00278 * | Same 26 mandibles | — |
| Synthesis | Not pooled | — | — | — |
3.5. Agreement Between Acquisition Protocols
Liljeholm et al. reported standard-minus-reduced protocol differences of −0.0181, −0.0038, and 0.0600 mm for three reduced presets, using 160 paired observations [30]. For the most reduced preset, 0.0600/0.423 reconstructs an observation-level SE of approximately 0.1418 mm from the reported paired t statistic. It does not adjust for the eight skulls. De Castro et al. reported no significant height or thickness difference between 24 s/720-projection and 15 s/450-projection protocols in five polyurethane mandibles [28]. Its unavailable paired variance was not replaced with an assumed specimen-level estimate.
In Kaaber et al., the equal-weight mean of the 18 published HR-minus-LD height contrasts is 0.040 mm [8]. Replacing the six VISO cells with their more extreme HR-minus-ULD contrast changes gives a descriptive mean of −0.00278 mm. Neither average is a validated study-level effect with a recoverable joint variance. Table 4 and Appendix A record the extraction-rule check and the observer-, device-, and plane-specific values. The 2025 correction concerns exposure settings, not these measurement-contrast cells [8].
3.6. Measurement Accuracy and Experimental Setting
Reported errors were larger in some preserved human specimens than in simpler dry-bone models. However, the previously stated monotonic gradient combined with unlike statistics is not retained. Verified examples are Al-Ekrish’s mean absolute errors of 0.46–0.51 mm with simulated soft tissue and Ganguly’s approximately 1.10–1.20 mm in embalmed heads [6,26]. Waltrick’s overall reported error was 0.23 mm [7]. The phantom values previously labeled 0.16–0.22 mm mean absolute error were not retained because they were not verified as means of individual absolute errors. Vasconcelos’s regional differences are labeled as reported rather than converted to mean absolute error. Table 5 preserves these distinctions.
Table 5.
Experimental setting and reported measurement error. MAE, mean absolute error; SD, standard deviation. Razavi’s percentages correspond to 3/6/9 mm measurement levels. Unlike error metrics are not pooled or interpreted as a causal specimen-condition gradient.
| Experimental Setting | Study | Error/Outcome |
|---|---|---|
| Polyurethane mandibles | de Castro et al. [28] | MAE not verified |
| Dry hemimandibles | Waltrick et al. [7] | Overall error 0.23 mm (SD 0.20) |
| Dry skulls with simulated tissue | Al-Ekrish [26] | MAE 0.46–0.51 mm |
| Dry mandibles in water | Vasconcelos et al. [29] | Regional image–bone differences |
| Embalmed heads with tissue | Ganguly et al. [6] | MAE ≈ 1.10–1.20 mm |
| Bovine ribs adjacent to implants | Razavi et al. [31] | Relative error: i-CAT 68/28/18%; Accuitomo 23/5/6% |
Tissue attenuation, scatter, beam hardening, and metal can impair boundary visibility, but specimen condition is confounded with scanner, anatomy, preservation, reference measurement, and observation methods across these studies [6,21,22]. Thus, the findings make boundary visibility an important clinical consideration without establishing that it has a larger independent effect than acquisition settings. Cadaver results should not be treated as a calibrated estimate of patient error; embalming itself may affect visibility [6].
3.7. Between-Device and Contextual Reports
Razavi et al. evaluated cortical thickness adjacent to implants using histological sections, but the scanner and several protocol settings differed together [31]. Mean relative errors at 3, 6, and 9 mm were 68%, 28%, and 18% for i-CAT and 23%, 5%, and 6% for Accuitomo. These device-specific findings illustrate difficulty measuring the thin peri-implant cortex; they do not isolate the effects of dose, voxel size, or metal. Pauwels et al. studied gray values against MDCT numbers, and Shokri et al. studied metal artifacts [24,25]. Their outcomes do not meet the linear-measurement criterion. Table 6 explains each report’s role; lack of an independent reference is not, by itself, a reason to exclude agreement data.
Table 6.
Reasons for retaining one between-device study and two contextual reports outside the within-device linear synthesis.
| Report | Outcome/Comparator | Reason Outside Within-Device Linear Synthesis |
|---|---|---|
| Razavi et al. [31] | Cortical thickness; histological sections | Eligible linear study, but different scanners and several co-varying settings; narrative only. |
| Pauwels et al. [24] | Gray values; MDCT numbers | No eligible linear measurement outcome; methodological context only. |
| Shokri et al. [25] | Metal-artifact magnitude; CBCT comparisons | No eligible linear measurement outcome; context only. Lack of an independent reference alone would not exclude a linear agreement outcome. |
4. Discussion
Selected within-device changes were associated with small reported average measurement differences in experimental models. The revised source audit does not support retaining the original pooled equivalence claims. This is not evidence that reduced-exposure protocols are necessarily inaccurate; it means that the available aggregate reporting cannot establish the precision of a common protocol effect. Exposure duration, tube current, nominal voxel size, and manufacturer presets remain distinct interventions, and conclusions are restricted to the specific tested protocols.
The review’s contribution relative to earlier syntheses is methodological rather than a definitive new pooled estimate: accuracy is separated from agreement, between-device changes are not interpreted as isolated parameter effects, and the analysis-unit problem is made explicit [18,19,20]. The original I2 of 0% meant that statistical heterogeneity was not detected using the submitted inputs; it did not demonstrate identical effects. With four or three studies, tests for heterogeneity have limited power, and their interpretation additionally depends on correctly specified sampling variances [34]. Since those variances were not established, heterogeneity statistics are no longer used to support the conclusion.
The 1 mm practical measurement-error convention and the stricter planned ±0.5 mm between-protocol margin must not be interchanged. Even equivalence in average error would allow both protocols to have substantial absolute error. Close to the inferior alveolar canal, sinus floor, or thin buccal plate, anatomical visibility and the consequences of individual error matter more than whether a mean lies near zero. Neither margin is permission to plan an implant with that clearance from a critical structure [2,3,15].
Image-quality assessment cannot be dismissed merely because average linear measurements change little. Vasconcelos et al. cautioned against the lowest current settings, and Liljeholm et al. found limitations of the most reduced preset for anatomical and bone-quality assessment [29,30]. These observations qualify the choice of an extreme comparison: it tests a boundary of a device’s settings, not an automatically acceptable clinical protocol. Optimization should preserve the information needed for the entire planning task, including the visibility of structures from which measurements are taken.
Intra-examiner repeatability concerns repeated readings by the same examiner, whereas inter-examiner reproducibility concerns readings by different examiners. Both are distinct from the agreement between acquisition protocols and from the reproducibility of the review’s QUIN appraisal. Reliability coefficients do not exclude important individual discrepancies. Kaaber et al. reported height intraclass correlation coefficients (ICCs) above 0.98, while observer differences ranged from −11.29 to 4.24 mm in one pairing [8]. The 15.53 mm value is the full range, not the largest single discrepancy. A high ICC can coexist with problematic absolute agreement and does not establish accuracy against an independent reference. Error distributions, limits of agreement where appropriate, and the proportion crossing a clinically justified threshold are therefore important alongside mean differences. The present evidence does not isolate field of view (FOV) as an independent parameter; restriction of FOV remains a radiation-protection principle rather than a result of this synthesis [12,15]. Linear-measurement findings should not be generalized to diagnostic tasks requiring different levels of detail [10,35,36].
5. Limitations
The available evidence is limited by small numbers of independent experimental specimens, incomplete paired or cluster-adjusted variance reporting, and non-interchangeable outcome summaries. Withdrawing unsupported pooling prevents false precision but does not recover the missing information or establish equivalence. The original QUIN appraisal was performed by one reviewer, and no independent duplicate appraisal is claimed. Some source-table inputs could not be reverified and were omitted rather than imputed. None of the included studies involved living patients; preservation, tissue simulation, lack of movement, and restoration-related artifacts limit direct clinical extrapolation. Between-model comparisons remain confounded, and no independent effect of specimen condition is estimated. These limitations preclude broad recommendations for all low-dose presets, voxel sizes, scanners, or future reconstruction methods.
6. Conclusions
Some tested within-device changes were associated with small average measurement differences under experimental conditions. The revised evidence does not establish statistical equivalence, interchangeable effects across parameter classes, or a universal clinically acceptable protocol. Accuracy against an independent reference remains distinct from between-protocol agreement. Boundary visibility and individual error require attention, especially near critical anatomy. Clinical validation in living patients is needed before broad implementation in routine implant planning.
Declaration of Generative AI and AI-assisted Technologies in the Writing Process.
During manuscript preparation and revision, generative AI assistance was used for drafting, language editing, source-checking support, descriptive audit calculations, and document preparation. Such assistance is not an independent human reviewer and does not constitute author approval or independent duplicate appraisal. Responsibility for verification, interpretation, and the final submitted content remains with the authors.
Acknowledgments
During the preparation of this manuscript/study, the authors used a generative AI assistant (AI) (Claude, Opus 5; Anthropic) for the purposes of drafting and language editing. The authors have reviewed and edited the output and take full responsibility for the content of this publication. Also, the authors would like to thank the Faculty of Dentistry at ADEMA University School and the University of Barcelona for their support in carrying out this work.
Abbreviations
ALADA, as low as diagnostically acceptable; ALADAIP, as low as diagnostically acceptable, indication-oriented and patient-specific; ALARA, as low as reasonably achievable; CBCT, cone-beam computed tomography; CS, cross-sectional; DOI, digital object identifier; FOV, field of view; HR, high-resolution preset; ICC, intraclass correlation coefficient; LD, low-dose preset; MAE, mean absolute error; MDCT, multidetector computed tomography; PMMA, polymethyl methacrylate; PRISMA, Preferred Reporting Items for Systematic Reviews and Meta-Analyses; PROSPERO, International Prospective Register of Systematic Reviews; QUIN, Quality Assessment Tool for In Vitro Studies; SAG, sagittal; SD, standard deviation; SE, standard error; SR, standard-resolution preset; SWiM, Synthesis Without Meta-analysis; ULD, ultra-low-dose preset.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/tomography12110162/s1, Supplementary File S1, PRISMA 2020 reporting checklist [23].
Appendix A. Source-to-Estimate Audit
This audit distinguishes primary-study findings, transparent arithmetic from published summaries, and estimates removed from the synthesis. It is not a claim that unavailable source tables or specimen-level datasets have been recovered. The table covers every study previously assigned an inverse-variance weight. DOI, source location, analysis unit, and the fate of the corresponding estimate are recorded.
Table A1.
Audit of the seven original meta-analysis inputs. Reference numbers identify the primary reports; all linear values are in millimeters.
| Primary Source/Location | Original Review Input |
Source Check | Decision |
|---|---|---|---|
| Al-Ekrish [26]; DOI 10.1016/j.sdentj.2012.05.001; Section 3.2, p. 132; Table 1 | Δ −0.010; SE 0.063; n = 55 | 83 matched height/width outcomes at 55 sites in 5 skulls; means/SDs support the rounded difference, not a cluster-adjusted SE. | Keep published means/SDs; remove SE, weight and pooled contribution. |
| Waltrick et al. [7]; DOI 10.1902/jop.2012.110524; primary abstract, Methods/Results | Δ −0.060; SE 0.034; n = 36 | 12 hemimandibles; 108 images; 648 readings; overall error 0.23 (0.20). Contrast-specific original source cells unavailable for reverification. | Keep verified primary summary; remove unverified protocol contrast/SE. |
| Vasconcelos et al. [29]; DOI 10.11607/jomi.3524; Table 2, p. 1367; Methods, pp. 1365–1366 | Δ −0.060; SE 0.062; n = 80 | Equal average of eight region means reproduces −0.060; source definition and specimen covariance do not validate an MAE paired SE. | Show arithmetic explicitly; no pooled input or MAE relabeling. |
| Ganguly et al. [6]; DOI 10.5624/isd.2016.46.2.93; Table 1, pp. 97–98; Methods, pp. 94–95; Discussion | Δ +0.037; SE 0.091; n = 28 | Four heads; table rows summarize repeated locations/measurements. No specimen-linked paired error distribution is available. | Use reported study summaries; remove row-based effect/SE as an inferential input. |
| Liljeholm et al. [30]; PMID 28944361; Table 3, p. 357 | Δ +0.060; SE ≈ 0.142; n = 160 | Paired t = 0.423 on 159 df reproduces observation-level SE; eight skulls are the specimen units. | Retain all three mean contrasts; do not treat SE as skull-adjusted. |
| de Castro et al. [28]; DOI 10.11607/jomi.8773; primary abstract, Methods/Results | Δ +0.060; SE ≈ 0.346 | Five phantoms; protocol settings and reported tests verified. The unavailable detailed mean/variance table cannot support a newly certified paired SE. | Keep verified descriptive result; omit precise contrast, SE, and previously asserted MAE range. |
| Kaaber et al. [8]; DOI 10.1111/clr.14326; Tables 1, 3 and 5; correction [8] | Δ +0.040; SE ≈ 0.205; denominator 42 | 42 sites in 26 mandibles. HR–LD average differs from the largest-contrast rule. Covariance and complete-case sample counts across cells are not jointly available. | Retain disaggregated contrasts and extraction-rule arithmetic; remove aggregate SE and weight. |
Reproducible Descriptive Checks
Al-Ekrish: 0.50 − 0.51 = −0.010. With an assumed paired correlation r = 0.5, √(0.472 + 0.462 − 2r × 0.47 × 0.46)/√55 = 0.06271; changing 55 to the reported 83 matched outcomes gives 0.05105. Neither calculation handles five-skull clustering. These are demonstrations of the original assumption, not alternative valid SEs [26,32].
Vasconcelos: the eight Table 2 means at 2 mA are 0.68, 0.79, 0.47, 0.57, 0.43, 0.70, 0.41, 0.62 (arithmetic mean 0.58375). At 15 mA, they are 0.76, 0.68, 0.38, 0.44, 0.48, 0.54, 0.36, 0.55 (mean 0.52375). Their difference is −0.060. These are averages of landmark-category summaries, not a recovered specimen-level paired distribution; the source labels the entries as differences between image and bone measurements [29].
Liljeholm: the most reduced comparison is reported as mean difference 0.0600, t = 0.423, and df = 159. The reconstructed paired-observation SE is |0.0600/0.423| = 0.14184. The two less reduced comparisons, −0.0181 and −0.0038, are also retained; the strongest reduction is not substituted for normal clinical practice [30].
Kaaber: read each nine-cell height row in Table 5 as observer 1 (X1, AXEOS, VISO), observer 2 (X1, AXEOS, VISO), observer 3 (X1, AXEOS, VISO). HR–LD cross-sectional values are −0.21, −0.28, 0.21, −0.14, 0.01, 0.24, −0.55, −0.14, 0.20; sagittal values are 0.00, 0.12, 0.26, −0.02, −0.09, 0.38, 0.12, 0.13, 0.48. Their mean is 0.040. Substituting VISO HR–ULD values of 0.10, 0.09, 0.33 (cross-sectional) and 0.06, 0.18, 0.24 (sagittal), obtained by reversing the source’s ULD–HR signs, yields −0.00278. No SE is assigned to either average because the shared-sample covariance is unavailable [8].
Appendix B. Database Search Strategies and Study Selection
This appendix presents the four database-specific search strategies, their retrieval counts, and the study-selection results shown in the updated PRISMA diagram.
Appendix B.1. Database-Specific Search Strategies
Table A2.
Database-specific search strategies and retrieval counts before deduplication. Ovid searches 1–3 are run in sequence.
| Database and Search Details | Exact Search Strategy |
|---|---|
| PubMed/MEDLINE PubMed interface 28 October 2026 1815 records |
(“Cone-Beam Computed Tomography”[Mesh] OR CBCT[tiab] OR “cone beam”[tiab]) AND (implant*[tiab] OR alveolar[tiab] OR edentul*[tiab] OR ridge*[tiab]) AND (“linear measurement*”[tiab] OR “bone measurement*”[tiab] OR “bone height”[tiab] OR “bone width”[tiab] OR “bone thickness”[tiab] OR “cortical thickness”[tiab] OR “ridge dimension*”[tiab]) AND (“2000/01/01”[dp]: “2026/10/28”[dp]) |
| Embase Ovid Advanced Search Embase Classic+Embase 1947 to 2026 Week 43 29 October 2026 539 final records Searches 1 and 2: 2025 each |
Search 1 ((cbct or cone beam) and (implant* or alveolar or edentul* or ridge*) and (linear measurement* or bone measurement* or bone height or bone width or bone thickness or cortical thickness or ridge dimension*)).ti, ab, kf. Search 2 limit 1 to yr = “2000–2026” Search 3 2 and (accura* or agreement or reliab* or reproducib* or repeatab* or precis* or valid* or error* or bias).ti, ab, kf. |
| Web of Science Core Collection Advanced Search 403 records |
TS = (CBCT OR “cone beam”) AND TS = (implant* OR alveolar OR edentul* OR ridge*) AND TS = (“linear measurement*” OR “bone measurement*” OR “bone height” OR “bone width” OR “bone thickness” OR “cortical thickness” OR “ridge dimension*”) AND TS = (accura* OR agreement OR reliab* OR reproducib* OR repeatab* OR precis* OR valid* OR error* OR bias) AND PY = (2000–2026) |
| Scopus Advanced Search 604 records |
TITLE-ABS-KEY(CBCT OR “cone beam”) AND TITLE-ABS-KEY(implant* OR alveolar OR edentul* OR ridge*) AND TITLE-ABS-KEY(“linear measurement*” OR “bone measurement*” OR “bone height” OR “bone width” OR “bone thickness” OR “cortical thickness” OR “ridge dimension*”) AND TITLE-ABS-KEY(accura* OR agreement OR reliab* OR reproducib* OR repeatab* OR precis* OR valid* OR error* OR bias) AND PUBYEAR AFT 1999 AND PUBYEAR BEF 2027 |
PubMed uses the publication-date range 1 January 2000 through 28 October 2026; the other expressions specify calendar years 2000–2026. The Embase, Scopus, and Web of Science strategies include an additional block of accuracy, agreement, reliability, and related performance terms. The search expressions contain no language, human-only, or full-text-availability filters.
Appendix B.2. Study-Selection Flow
The updated PRISMA diagram reports 3361 records identified across the four databases, 1819 duplicate records removed, 1542 records screened, and 1527 records excluded during screening. Fifteen reports were sought for retrieval, three were not retrieved, and 12 were assessed for eligibility. Two reports were excluded after full-text assessment. The review presents eight linear-measurement studies and two contextual reports separately.
Appendix B.3. Full-Text Dispositions
Table A3.
Full-text exclusions and reports not retrieved, as recorded in the PRISMA flow diagram.
| Disposition | Recorded Reason | Reports (n) |
|---|---|---|
| Full-text exclusion | No reference standard | 1 |
| Full-text exclusion | Non-English full text | 1 |
| Report not retrieved | Not retrieved | 3 |
For the synthesis, accuracy denotes measurement error against an independent reference, whereas agreement denotes differences between CBCT protocol measurements. An otherwise eligible agreement study does not require an independent reference. The descriptive accuracy and agreement findings are presented separately in Table 3 and Table 4.
Author Contributions
Conceptualization, P.R.-S. and A.S.E.; methodology, A.A.L.G. and A.S.E.; data collection and analysis, A.A.L.G. and J.L.-L.; data curation and formal analysis and conceptualization, M.P.-D.; validation and formal analysis, M.B.-H.; investigation, A.S.E.; writing—original draft preparation, A.S.E., J.L.-L. and P.R.-S.; writing—review and editing, A.S.E., P.R.-S. and J.L.-L.; writing—review editing and formal analysis and methodology H.A.A.; visualization, A.S.E.; supervision, P.R.-S. and J.L.-L. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
Study-level descriptive results, database-specific search strategies, retrieval counts, and the PRISMA study-selection flow are included in the article and appendices.
Conflicts of Interest
The authors declare no conflicts of interest.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Worthington P., Rubenstein J., Hatcher D.C. The role of cone-beam computed tomography in the planning and placement of implants. J. Am. Dent. Assoc. 2010;141:19S–24S. doi: 10.14219/jada.archive.2010.0358. [DOI] [PubMed] [Google Scholar]
- 2.Tyndall D.A., Price J.B., Tetradis S., Ganz S.D., Hildebolt C., Scarfe W.C., American Academy of Oral and Maxillofacial Radiology Position statement of the American Academy of Oral and Maxillofacial Radiology on selection criteria for the use of radiology in dental implantology with emphasis on cone beam computed tomography. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 2012;113:817–826. doi: 10.1016/j.oooo.2012.03.005. [DOI] [PubMed] [Google Scholar]
- 3.Jacobs R., Quirynen M., Bornstein M.M. Neurovascular disturbances after implant surgery. Periodontology 2000. 2014;66:188–202. doi: 10.1111/prd.12050. [DOI] [PubMed] [Google Scholar]
- 4.Klinge B., Petersson A., Maly P. Location of the mandibular canal: Comparison of macroscopic findings, conventional radiography, and computed tomography. Int. J. Oral Maxillofac. Implants. 1989;4:327–332. [Google Scholar]
- 5.Wyatt C.C., Pharoah M.J. Imaging techniques and image interpretation for dental implant treatment. Int. J. Prosthodont. 1998;11:442–452. [PubMed] [Google Scholar]
- 6.Ganguly R., Ramesh A., Pagni S. The accuracy of linear measurements of maxillary and mandibular edentulous sites in cone-beam computed tomography images with different fields of view and voxel sizes under simulated clinical conditions. Imaging Sci. Dent. 2016;46:93–101. doi: 10.5624/isd.2016.46.2.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Waltrick K.B., Nunes de Abreu Junior M.J., Corrêa M., Zastrow M.D., Dutra V.D. Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: An in vitro study. J. Periodontol. 2013;84:68–77. doi: 10.1902/jop.2012.110524. [DOI] [PubMed] [Google Scholar]
- 8.Kaaber L., Matzen L.H., Spin-Neto R., Schropp L. Low-dose, standard, and high-resolution cone beam computed tomography for alveolar bone measurements related to implant planning: An ex vivo study in human specimens. Clin. Oral Implants Res. 2024;35:1394–1405. doi: 10.1111/clr.14326. Correction in Clin. Oral Implants Res. 2025, 36, 785. https://doi.org/10.1111/clr.14434 . [DOI] [PubMed] [Google Scholar]
- 9.Brüllmann D., Schulze R.K.W. Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today? Dentomaxillofac. Radiol. 2015;44:20140204. doi: 10.1259/dmfr.20140204. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Pauwels R., Araki K., Siewerdsen J.H., Thongvigitmanee S.S. Technical aspects of dental CBCT: State of the art. Dentomaxillofac. Radiol. 2015;44:20140224. doi: 10.1259/dmfr.20140224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Loubele M., Bogaerts R., Van Dijck E., Pauwels R., Vanheusden S., Suetens P., Marchal G., Sanderink G., Jacobs R. Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur. J. Radiol. 2009;71:461–468. doi: 10.1016/j.ejrad.2008.06.002. [DOI] [PubMed] [Google Scholar]
- 12.Ludlow J.B., Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2008;106:106–114. doi: 10.1016/j.tripleo.2008.03.018. [DOI] [PubMed] [Google Scholar]
- 13.Jaju P.P., Jaju S.P. Cone-beam computed tomography: Time to move from ALARA to ALADA. Imaging Sci. Dent. 2015;45:263–265. doi: 10.5624/isd.2015.45.4.263. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Oenning A.C., Jacobs R., Salmon B., DIMITRA Research Group ALADAIP, beyond ALARA and towards personalized optimization for paediatric cone-beam, C.T. Int. J. Paediatr. Dent. 2021;31:676–678. doi: 10.1111/ipd.12797. [DOI] [PubMed] [Google Scholar]
- 15.Harris D., Horner K., Gröndahl K., Jacobs R., Helmrot E., Benic G.I., Bornstein M.M., Dawood A., Quirynen M.E.A.O. guidelines for the use of diagnostic imaging in implant dentistry 2011. A consensus workshop organized by the European Association for Osseointegration at the Medical University of Warsaw. Clin. Oral Implants Res. 2012;23:1243–1253. doi: 10.1111/j.1600-0501.2012.02441.x. [DOI] [PubMed] [Google Scholar]
- 16.Horner K., O’Malley L., Taylor K., Glenny A.-M. Guidelines for clinical use of CBCT: A review. Dentomaxillofac. Radiol. 2015;44:20140225. doi: 10.1259/dmfr.20140225. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Yeung A.W.K., Jacobs R., Bornstein M.M. Novel low-dose protocols using cone beam computed tomography in dental medicine: A review focusing on indications, limitations, and future possibilities. Clin. Oral Investig. 2019;23:2573–2581. doi: 10.1007/s00784-019-02907-y. [DOI] [PubMed] [Google Scholar]
- 18.Carneiro A.L.E., Reis I.N.R., Bitencourt F.V., Salgado D.M.R.A., Costa C., Spin-Neto R. Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: A systematic review and meta-analysis. Dentomaxillofac. Radiol. 2024;53:207–221. doi: 10.1093/dmfr/twae007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Kaaber L., Matzen L.H., Schropp L., Spin-Neto R. Low-dose CBCT protocols in implant dentistry: A systematic review. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 2024;138:427–439. doi: 10.1016/j.oooo.2024.03.013. [DOI] [PubMed] [Google Scholar]
- 20.Fokas G., Vaughn V.M., Scarfe W.C., Bornstein M.M. Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review. Clin. Oral Implants Res. 2018;29:393–415. doi: 10.1111/clr.13142. [DOI] [PubMed] [Google Scholar]
- 21.Schulze R., Heil U., Gross D., Bruellmann D.D., Dranischnikow E., Schwanecke U., Schoemer E. Artefacts in CBCT: A review. Dentomaxillofac. Radiol. 2011;40:265–273. doi: 10.1259/dmfr/30642039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Katsumata A., Hirukawa A., Okumura S., Naitoh M., Fujishita M., Ariji E., Langlais R.P. Effects of image artifacts on gray-value density in limited-volume cone-beam computerized tomography. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2007;104:829–836. doi: 10.1016/j.tripleo.2006.12.005. [DOI] [PubMed] [Google Scholar]
- 23.Page M.J., McKenzie J.E., Bossuyt P.M., Boutron I., Hoffmann T.C., Mulrow C.D., Shamseer L., Tetzlaff J.M., Akl E.A., Brennan S.E., et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. Int. J. Surg. 2021;88:105906. doi: 10.1016/j.ijsu.2021.105906. [DOI] [PubMed] [Google Scholar]
- 24.Pauwels R., Nackaerts O., Bellaiche N., Stamatakis H., Tsiklakis K., Walker A., Bosmans H., Bogaerts R., Jacobs R., Horner K., et al. Variability of dental cone beam CT grey values for density estimations. Br. J. Radiol. 2013;86:20120135. doi: 10.1259/bjr.20120135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Shokri A., Jamalpour M.R., Khavid A., Mohseni Z., Sadeghi M. Effect of exposure parameters of cone beam computed tomography on metal artifact reduction around the dental implants in various bone densities. BMC Med. Imaging. 2019;19:34. doi: 10.1186/s12880-019-0334-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Al-Ekrish A.A. Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls. Saudi Dent. J. 2012;24:127–134. doi: 10.1016/j.sdentj.2012.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Sheth V.H., Shah N.P., Jain R., Bhanushali N., Bhatnagar V. Development and validation of a risk-of-bias tool for assessing in vitro studies conducted in dentistry: The QUIN. J. Prosthet. Dent. 2024;131:1038–1042. doi: 10.1016/j.prosdent.2022.05.019. [DOI] [PubMed] [Google Scholar]
- 28.de Castro H.S., Kehrwald R., Matheus R.A., Gomes A.F., Queiroz P.M. Influence of Low-Dose Protocols of CBCT on Dental Implant Planning. Int. J. Oral Maxillofac. Implants. 2021;36:307–312. doi: 10.11607/jomi.8773. [DOI] [PubMed] [Google Scholar]
- 29.Vasconcelos T.V., Neves F.S., Queiroz de Freitas D., Campos P.S.F., Watanabe P.C.A. Influence of the milliamperage settings on cone beam computed tomography imaging for implant planning. Int. J. Oral Maxillofac. Implants. 2014;29:1364–1368. doi: 10.11607/jomi.3524. [DOI] [PubMed] [Google Scholar]
- 30.Liljeholm R., Kadesjö N., Benchimol D., Hellén-Halme K., Shi X.Q. Cone-beam computed tomography with ultra-low dose protocols for pre-implant radiographic assessment: An in vitro study. Eur. J. Oral Implantol. 2017;10:351–359. [PubMed] [Google Scholar]
- 31.Razavi T., Palmer R.M., Davies J., Wilson R., Palmer P.J. Accuracy of measuring the cortical bone thickness adjacent to dental implants using cone beam computed tomography. Clin. Oral Implants Res. 2010;21:718–725. doi: 10.1111/j.1600-0501.2009.01905.x. [DOI] [PubMed] [Google Scholar]
- 32.Higgins J.P.T., Eldridge S., Li T. Chapter 23: Including variants on randomized trials. In: Higgins J.P.T., Thomas J., Chandler J., Cumpston M., Li T., Page M.J., Welch V.A., editors. Cochrane Handbook for Systematic Reviews of Interventions, Version 6.5. Cochrane; London, UK: 2024. [(accessed on 12 September 2026)]. Sections 23.1 and 23.2. Available online: https://www.cochrane.org/authors/handbooks-and-manuals/handbook/current/chapter-23. [Google Scholar]
- 33.Campbell M., McKenzie J.E., Sowden A., Katikireddi S.V., Brennan S.E., Ellis S., Hartmann-Boyce J., Ryan R., Shepperd S., Thomas J., et al. Synthesis without meta-analysis (SWiM) in systematic reviews: Reporting guideline. BMJ. 2020;368:l6890. doi: 10.1136/bmj.l6890. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Higgins J.P.T., Thompson S.G., Deeks J.J., Altman D.G. Measuring inconsistency in meta-analyses. BMJ. 2003;327:557–560. doi: 10.1136/bmj.327.7414.557. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Tyndall D.A., Rathore S. Cone-beam CT diagnostic applications: Caries, periodontal bone assessment, and endodontic applications. Dent. Clin. N. Am. 2008;52:825–841, vii. doi: 10.1016/j.cden.2008.05.002. [DOI] [PubMed] [Google Scholar]
- 36.Hidalgo Rivas J.A., Horner K., Thiruvenkatachari B., Davies J., Theodorakou C. Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children. Br. J. Radiol. 2015;88:20150559. doi: 10.1259/bjr.20150559. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Study-level descriptive results, database-specific search strategies, retrieval counts, and the PRISMA study-selection flow are included in the article and appendices.