Abstract
The reproducibility and agreement of the estimates of stroke volume (SV), cardiac output (CO) and total peripheral resistance (TPR) by transthoracic impedance cardiography (ZCG) and M-mode echocardiography (ECHO) were analyzed before and after the placebo-controlled administration of ascending doses of isosorbide dinitrate and nicorandil in 12 healthy subjects. There was no biostatistical agreement between the two methods in estimating cardiovascular function either before or after dosing (ZCG estimated substantially larger SV, CO and lower TPR). But, ZCG and ECHO estimated about similar overall treatment related changes (across treatments and periods) and reached substantially better agreement when the values were expressed as ratio of the baseline before dosing. Such improvement did not occur when the data were expressed as arithmetic difference from baseline. In spite of the improvement of agreement by expressing the data as ratio of baseline, the coefficients of reproducibility between the two methods (circa 25% of baseline) remained too large to judge the methods interchangeable.
Full text
PDF




Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Aust P. E., Belz G. G., Belz G., Koch W. Comparison of impedance cardiography and echocardiography for measurement of stroke volume. Eur J Clin Pharmacol. 1982;23(6):475–477. doi: 10.1007/BF00637491. [DOI] [PubMed] [Google Scholar]
- Belz G. G., Matthews J., Heinrich J., Wagner G. Controlled comparison of the pharmacodynamic effects of nicorandil (SG-75) and isosorbide dinitrate in man. Eur J Clin Pharmacol. 1984;26(6):681–685. doi: 10.1007/BF00541925. [DOI] [PubMed] [Google Scholar]
- Bland J. M., Altman D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Feb 8;1(8476):307–310. [PubMed] [Google Scholar]
- Chinn S. Statistics in respiratory medicine. 2. Repeatability and method comparison. Thorax. 1991 Jun;46(6):454–456. doi: 10.1136/thx.46.6.454. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fleiss J. L., Wallenstein S., Rosenfeld R. Adjusting for baseline measurements in the two-period crossover study: a cautionary note. Control Clin Trials. 1985 Sep;6(3):192–197. doi: 10.1016/0197-2456(85)90002-9. [DOI] [PubMed] [Google Scholar]
- Hayes R. J. Methods for assessing whether change depends on initial value. Stat Med. 1988 Sep;7(9):915–927. doi: 10.1002/sim.4780070903. [DOI] [PubMed] [Google Scholar]
- Kaiser L. Adjusting for baseline: change or percentage change? Stat Med. 1989 Oct;8(10):1183–1190. doi: 10.1002/sim.4780081002. [DOI] [PubMed] [Google Scholar]
- Kubicek W. G., Karnegis J. N., Patterson R. P., Witsoe D. A., Mattson R. H. Development and evaluation of an impedance cardiac output system. Aerosp Med. 1966 Dec;37(12):1208–1212. [PubMed] [Google Scholar]
- Ng H. W., Walley T., Tsao Y., Breckenridge A. M. Comparison and reproducibility of transthoracic bioimpedance and dual beam Doppler ultrasound measurement of cardiac function in healthy volunteers. Br J Clin Pharmacol. 1991 Sep;32(3):275–282. doi: 10.1111/j.1365-2125.1991.tb03899.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Northridge D. B., Findlay I. N., Wilson J., Henderson E., Dargie H. J. Non-invasive determination of cardiac output by Doppler echocardiography and electrical bioimpedance. Br Heart J. 1990 Feb;63(2):93–97. doi: 10.1136/hrt.63.2.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sakai K., Shiraki Y., Nabata H. Cardiovascular effects of a new coronary vasodilator N-(2-hydroxyethyl)nicotinamide nitrate (SG-75): comparison with nitroglycerin and diltiazem. J Cardiovasc Pharmacol. 1981 Jan-Feb;3(1):139–150. doi: 10.1097/00005344-198101000-00012. [DOI] [PubMed] [Google Scholar]
- Silke B. Non-invasive determination of cardiac output by Doppler echocardiography and electrical bioimpedance. Br Heart J. 1990 Nov;64(5):347–348. doi: 10.1136/hrt.64.5.347-b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Teichholz L. E., Kreulen T., Herman M. V., Gorlin R. Problems in echocardiographic volume determinations: echocardiographic-angiographic correlations in the presence of absence of asynergy. Am J Cardiol. 1976 Jan;37(1):7–11. doi: 10.1016/0002-9149(76)90491-4. [DOI] [PubMed] [Google Scholar]
- de Mey C., Belz G. G. Pitfalls and limitations in the use of impedance cardiography. Br Heart J. 1989 Jan;61(1):128–129. doi: 10.1136/hrt.61.1.128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- de Mey C., Enterling D. Assessment of the hemodynamic response to single passive head up tilt by non-invasive methods in normotensive subjects. Methods Find Exp Clin Pharmacol. 1986 Jul;8(7):449–457. [PubMed] [Google Scholar]
- de Mey C., Enterling D. Noninvasive assessment of cardiac performance by impedance cardiography: disagreement between two equations to estimate stroke volume. Aviat Space Environ Med. 1988 Jan;59(1):57–62. [PubMed] [Google Scholar]
- de Mey C., Hansen-Schmidt S., Enterling D. Postprandial haemodynamic changes: a source of bias in cardiovascular research affected by its own methodological bias. Cardiovasc Res. 1988 Oct;22(10):703–707. doi: 10.1093/cvr/22.10.703. [DOI] [PubMed] [Google Scholar]