Skip to main content
Annals of Surgery logoLink to Annals of Surgery
. 1979 Jan;189(1):68–74. doi: 10.1097/00000658-197901000-00014

Moist bacterial strike-through of surgical materials: confirmatory tests.

H Laufman, J D Siegal, S C Edberg
PMCID: PMC1396949  PMID: 758867

Abstract

New tests consisting of modifications of the inverted Mason jar test confirm our previously reported studies which showed that woven and nonwoven surgical materials vary greatly in their ability to serve as barriers against moist bacterial strike-through. Among the woven materials, only tightly woven Pima cloth or materials treated with Quarpel waterproofing process or with polythene layer lamination was invariably resistant. However, tight-woven Pima cloth, which had been treated with Quarpel became permeable after 100 washing-sterilizing cycles. Of the nonwoven materials, single-layer nonwoven materials tended to unevenly permeable to moist bacterial strike-through. Only the front and sleeves of nonwoven gowns reinforced with polyethelene layer were invariably resistant to moist contamination.

Full text

PDF

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Laufman H., Eudy W. W., Vandernoot A. M., Harris C. A., Liu D. Strike-through of moist contamination by woven and nonwoven surgical materials. Ann Surg. 1975 Jun;181(6):857–862. doi: 10.1097/00000658-197506000-00018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Moylan J. A., Balish E., Chan J. Intraoperative bacterial transmission. Surg Gynecol Obstet. 1975 Nov;141(5):731–733. [PubMed] [Google Scholar]