Abstract
Artificial cadaveric femoral neck fractures were internally fixed with five different devices and subjected to cyclical loading of 0-1.0 kilonewtons (approximately one body weight) whilst in an anatomical position. Displacement of the proximal fragment was detected by a transducer and charted. Bone strength was assessed by a preliminary control loading phase on the intact bone. Efficiency of each fracture fixator could then be directly compared by the relative movement in each case. Five specimens each were tested with Moore's Pins, Trifin Nail, Garden Screws and a sliding screw-plate (OEC Ltd). By the criteria of the experiment, which put a severe shearing load on the implant, none of these devices reliably bore the representative body weight. An extended barrel-plate, which supported the sliding screw almost up to the fracture line, was then made. This device, employing some of Charnley's concepts, tolerated body weight in four cases out of five.
Full text
PDF





Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Bentley G. Treatment of nondisplaced fractures of the femoral neck. Clin Orthop Relat Res. 1980 Oct;(152):93–101. [PubMed] [Google Scholar]
- CHARNLEY J., BLOCKEY N. J., PURSER D. W. The treatment of displaced fractures of the neck of the femur by compression. J Bone Joint Surg Br. 1957 Feb;39-B(1):45–65. doi: 10.1302/0301-620X.39B1.45. [DOI] [PubMed] [Google Scholar]
- GARDEN R. S. STABILITY AND UNION IN SUBCAPITAL FRACTURES OF THE FEMUR. J Bone Joint Surg Br. 1964 Nov;46:630–647. [PubMed] [Google Scholar]
- Kofoed H. Femoral neck fractures in young adults. Injury. 1982 Sep;14(2):146–150. doi: 10.1016/0020-1383(82)90049-3. [DOI] [PubMed] [Google Scholar]
- Kyle R. F., Wright T. M., Burstein A. H. Biomechanical analysis of the sliding characteristics of compression hip screws. J Bone Joint Surg Am. 1980 Dec;62(8):1308–1314. [PubMed] [Google Scholar]
- Outerbridge R. E. Perosseous venography in the diagnosis of viability in subcapital fractures of the femur. Clin Orthop Relat Res. 1978 Nov-Dec;(137):132–139. [PubMed] [Google Scholar]
- SCHUMPELICK W., JANTZEN P. M. A new principle in the operative treatment of trochanteric fractures of the femur. J Bone Joint Surg Am. 1955 Jul;37-A(4):693–698. [PubMed] [Google Scholar]