Abstract
Pulse oximeters are being widely used for non-invasive, simultaneous assessment of haemoglobin oxygen saturation. They are reliable, accurate, relatively inexpensive and portable. Pulse oximeters are often used for estimating heart rate at rest and during exercise. However, at present the data available to validate their use as heart rate monitors are not sufficient. We evaluated the accuracy of two oximeters (Radiometer, ear and finger probe; Ohmeda 3700, ear probe) in monitoring heart rate during incremental exercise by comparing the pulse oximeters with simultaneous ECG readings. Data were collected on eight men (713 heart rate readings) during graded cycle ergometer and treadmill exercise to volitional fatigue. Analysis by linear regression revealed that general oximeter readings significantly correlated with those of ECG (r = 0.91, P less than 0.0001). However, comparison of heart rate at each level of work showed that oximeter readings significantly (P less than 0.05) under-estimated rates above 155 beats/min. These results indicate that the use of pulse oximeters as heart rate monitors during strenuous exercise is questionable. This inaccuracy may well originate from the instability of the probes, sweating, other artefacts during exercise, and measurement of different components in the cardiovascular cycle.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Bruce R. A. Exercise testing for evaluation of ventricular function. N Engl J Med. 1977 Mar 24;296(12):671–675. doi: 10.1056/NEJM197703242961206. [DOI] [PubMed] [Google Scholar]
- Jennis M. S., Peabody J. L. Pulse oximetry: an alternative method for the assessment of oxygenation in newborn infants. Pediatrics. 1987 Apr;79(4):524–528. [PubMed] [Google Scholar]
- Ludwig H. Heart- or respiration-rate calculator. Med Biol Eng Comput. 1977 Nov;15(6):700–702. doi: 10.1007/BF02457932. [DOI] [PubMed] [Google Scholar]
- Maxwell L. G., Harris A. P., Sendak M. J., Donham R. T. Monitoring the resuscitation of preterm infants in the delivery room using pulse oximetry. Clin Pediatr (Phila) 1987 Jan;26(1):18–20. doi: 10.1177/000992288702600102. [DOI] [PubMed] [Google Scholar]
- Ramanathan R., Durand M., Larrazabal C. Pulse oximetry in very low birth weight infants with acute and chronic lung disease. Pediatrics. 1987 Apr;79(4):612–617. [PubMed] [Google Scholar]
- Solimano A. J., Smyth J. A., Mann T. K., Albersheim S. G., Lockitch G. Pulse oximetry advantages in infants with bronchopulmonary dysplasia. Pediatrics. 1986 Nov;78(5):844–849. [PubMed] [Google Scholar]