Abstract
The mean current required for ventricular defibrillation was measured and found to be 0.35 +/- SE 0.03 A/kg body weight, which is about one-third of that predicted from animal experiments. There was no apparent correlation between the current required and body weight (r = -0.007 +/- SE 0.213). There is no evidence of need for defibrillators storing more than 400 J.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Anderson G. J., Suelzer J. The efficacy of trapezoidal wave forms for ventricular defibrillation. Chest. 1976 Aug;70(2):298–300. doi: 10.1378/chest.70.2.298. [DOI] [PubMed] [Google Scholar]
- BROOKS D. K., FELDMAN S. A. Metabolic acidosis. A new approach to 'neostigmine resistant curarisation'. Anaesthesia. 1962 Apr;17:161–169. doi: 10.1111/j.1365-2044.1962.tb13447.x. [DOI] [PubMed] [Google Scholar]
- DeSilva R. A., Lown B. Energy requirement for defibrillation of a markedly overweight patient. Circulation. 1978 Apr;57(4):827–830. doi: 10.1161/01.cir.57.4.827. [DOI] [PubMed] [Google Scholar]
- EDMARK K. W. SIMULTANEOUS VOLTAGE AND CURRENT WAVEFORMS GENERATED DURING INTERNAL AND EXTERNAL DIRECT CURRENT PULSE DEFIBRILLATION. Surg Forum. 1963;14:262–264. [PubMed] [Google Scholar]
- Geddes L. A., Grubbs S. S., Wilcox P. G., Tacker W. A., Jr The thoracic windows for electrical ventricular defibrillation current. Am Heart J. 1977 Jul;94(1):67–72. doi: 10.1016/s0002-8703(77)80345-1. [DOI] [PubMed] [Google Scholar]
- Gold J. H., Schuder J. C., Stoeckle H., Granberg T. A., Hamdani S. Z., Rychlewski J. M. Comparison of transthoracic square-wave defibrillation experience in the dog and calf. Med Instrum. 1978 Jan-Feb;12(1):20–23. [PubMed] [Google Scholar]
- LOWN B., AMARASINGHAM R., NEUMAN J. New method for terminating cardiac arrhythmias. Use of synchronized capacitor discharge. JAMA. 1962 Nov 3;182:548–555. [PubMed] [Google Scholar]
- Lown B., Crampton R. S., DeSilva R. A., Gascho J. Health-manpower needs. N Engl J Med. 1978 Jun 1;298(22):1252–1253. doi: 10.1056/NEJM197806012982210. [DOI] [PubMed] [Google Scholar]
- Tacker W. A., Jr, Galioto F. M., Jr, Giuliani E., Geddes L. A., McNamara D. G. Energy dosage for human trans-chest electrical ventricular defibrillation. N Engl J Med. 1974 Jan 24;290(4):214–215. doi: 10.1056/NEJM197401242900408. [DOI] [PubMed] [Google Scholar]
- Tacker W. A., Jr, Geddes L. A., Cabler P. S., Moore A. G. Electrical threshold for defibrillation of canine ventricles following myocardial infarction. Am Heart J. 1974 Oct;88(4):476–481. doi: 10.1016/0002-8703(74)90208-7. [DOI] [PubMed] [Google Scholar]
- Tacker W. A., Jr, Guinn G. A., Geddes L. A., Bourland J. D., Korompai F. L., Rubio P. A. The electrical dose for direct ventricular defibrillation in man. J Thorac Cardiovasc Surg. 1978 Feb;75(2):224–226. [PubMed] [Google Scholar]