Skip to main content
British Medical Journal logoLink to British Medical Journal
. 1970 Jul 18;3(5715):135–137. doi: 10.1136/bmj.3.5715.135

Bacteriological Studies of the Modified Kiil Dialyser

D M Jones, Barbara M Tobin, G R Harlow, A J Ralston
PMCID: PMC1702287  PMID: 5431083

Abstract

Bacterial proliferation in dialysis fluid during haemodialysis may be associated with rigors, hypotension, and bacteraemia. Investigations carried out in a period in which rigors were particularly common showed the source of bacteria to be parts of the gasket system of the Kiil dialyser, areas that are inaccessible to disinfectants.

Full text

PDF

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Curtis J. R., Wing A. J., Coleman J. C. Bacillus cereus bacteraemia. A complication of intermittent haemodialysis. Lancet. 1967 Jan 21;1(7482):136–138. doi: 10.1016/s0140-6736(67)91036-7. [DOI] [PubMed] [Google Scholar]
  2. KIDD E. E. BACTERIAL CONTAMINATION OF DIALYSING FLUID OF ARTIFICIAL KIDNEY. Br Med J. 1964 Apr 4;1(5387):880–882. doi: 10.1136/bmj.1.5387.880. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from British Medical Journal are provided here courtesy of BMJ Publishing Group

RESOURCES