Abstract
Data are presented which show the potential for release of viable microorganisms into the atmosphere from high-vacuum steam sterilizers during the evacuation cycle preceding application of steam under pressure. Bacillus subtilis var. niger spores, Serratia marcescens cells, and T1 coliphage disseminated into the sterilizer chamber as small particles from liquid suspensions, and dried spores of B. subtilis var. niger distributed on bulk discard materials were recovered from the atmosphere around pipes venting steam from the steam ejectors used to create chamber vacuum. Evaluation of the hazard involved is discussed, and the design, fabrication, and installation of a valved filter system for preventing release of viable microorganisms are presented. The filtration system utilized an F-700 water-resistant filter and was shown to eliminate the release of viable airborne microorganisms from a high-vacuum sterilizer. A method is presented for determining size requirements for an atmospheric vent filter in relation to the volume of a sterilizer.
Full text
PDF







Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- BARBEITO M. S., ALG R. L., WEDUM A. G. Infectious bacterial aerosol from dropped Petri dish cultures. Am J Med Technol. 1961 Nov-Dec;27:318–322. [PubMed] [Google Scholar]
- Barbeito M. S., Gremillion G. G. Micriobiological safety evaluation of an industrial refuse incinerator. Appl Microbiol. 1968 Feb;16(2):291–295. doi: 10.1128/am.16.2.291-295.1968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barbeito M. S., Taylor L. A., Seiders R. W. Microbiological evaluation of a large-volume air incinerator. Appl Microbiol. 1968 Mar;16(3):490–495. doi: 10.1128/am.16.3.490-495.1968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- KRUSE R. H. Potential aerogenic laboratory hazards of Cocidioides immitis. Am J Clin Pathol. 1962 Feb;37:150–158. doi: 10.1093/ajcp/37.2.150. [DOI] [PubMed] [Google Scholar]
- MILLER W. S., SCHERFF R. A., PIEPOLI C. R., IDOINE L. S. Physical tracers for bacterial aerosols. Appl Microbiol. 1961 May;9:248–251. doi: 10.1128/am.9.3.248-251.1961. [DOI] [PMC free article] [PubMed] [Google Scholar]
- PHILLIPS G. B., HANEL E., Jr Control of mold contaminants on solid media by the use of actidione. J Bacteriol. 1950 Jul;60(1):104–105. doi: 10.1128/jb.60.1.104-105.1950. [DOI] [PMC free article] [PubMed] [Google Scholar]
- REITMAN M., WEDUM A. G. Microbiological safety. Public Health Rep. 1956 Jul;71(7):659–665. [PMC free article] [PubMed] [Google Scholar]
- Taylor L. A., Barbeito M. S., Gremillion G. G. Paraformaldehyde for surface sterilization and detoxification. Appl Microbiol. 1969 Apr;17(4):614–618. doi: 10.1128/am.17.4.614-618.1969. [DOI] [PMC free article] [PubMed] [Google Scholar]
- WEDUM A. G. LABORATORY SAFETY IN RESEARCH WITH INFECTIOUS AEROSOLS. Public Health Rep. 1964 Jul;79:619–633. [PMC free article] [PubMed] [Google Scholar]