Comparison of three manual ventilation devices using an intubated mannequin

S Hussey, C Ryan, B Murphy
PMCID: PMC1721775  PMID: 15499138

Abstract

Objective: To compare three devices for manual neonatal ventilation.

Design: Participants performed a two minute period of ventilation using a self inflating device, an anaesthesia bag with attached manometer, and a Neopuff device. An intubated neonatal mannequin, approximating a 1 kg infant with functional lungs, was used for the study. Target ventilation variables included a rate of 40 breaths per minute, peak inspiratory pressure (PIP) of 20 cm H2O, and positive end expiratory pressure (PEEP) of 4 cm H2O. The circuit was attached to a laptop computer for data recording.

Results: Thirty five participants were enrolled, including consultant neonatologists, paediatricians, and anaesthetists, paediatric and anaesthetic registrars, and neonatal nurses. The maximum PIP recorded using the self inflating bag, anaesthetic bag, and Neopuff device were 75.9, 35.5, and 22.4 cm H2O respectively. There were significant differences between the devices for mean PIP (30.7, 18.1, and 20.1 cm H2O), mean PEEP (0.2, 2.8, and 4.4 cm H2O), mean airway pressure (7.6, 8.5, and 10.9 cm H2O), % total breaths ⩽ 21 cm H2O PIP (39%, 92%, and 98%), and % total breaths ⩾ 30 cm H2O PIP (45%, 0, and 0). There was no difference between doctors and allied health professionals for the variables examined.

Conclusion: The anaesthetic bag with manometer and Neopuff device both facilitate accurate and reproducible manual ventilation. Self inflating devices without modifications are not as consistent by comparison and should incorporate a manometer and a PEEP device, particularly when used for resuscitation of very low birthweight infants.

Full Text

The Full Text of this article is available as a PDF (60.7 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Brion L. P., Goyal M., Suresh B. R. Sudden deterioration of intubated newborn: four steps to the differential diagnosis and initial management. J Perinatol. 1992 Sep;12(3):281–293. [PubMed] [Google Scholar]
  2. Carbine D. N., Finer N. N., Knodel E., Rich W. Video recording as a means of evaluating neonatal resuscitation performance. Pediatrics. 2000 Oct;106(4):654–658. doi: 10.1542/peds.106.4.654. [DOI] [PubMed] [Google Scholar]
  3. Finer N. N., Barrington K. J., Al-Fadley F., Peters K. L. Limitations of self-inflating resuscitators. Pediatrics. 1986 Mar;77(3):417–420. [PubMed] [Google Scholar]
  4. Finer N. N., Rich W., Craft A., Henderson C. Comparison of methods of bag and mask ventilation for neonatal resuscitation. Resuscitation. 2001 Jun;49(3):299–305. doi: 10.1016/s0300-9572(01)00311-2. [DOI] [PubMed] [Google Scholar]
  5. Gregory G. A., Kitterman J. A., Phibbs R. H., Tooley W. H., Hamilton W. K. Treatment of the idiopathic respiratory-distress syndrome with continuous positive airway pressure. N Engl J Med. 1971 Jun 17;284(24):1333–1340. doi: 10.1056/NEJM197106172842401. [DOI] [PubMed] [Google Scholar]
  6. Hernandez L. A., Peevy K. J., Moise A. A., Parker J. C. Chest wall restriction limits high airway pressure-induced lung injury in young rabbits. J Appl Physiol (1985) 1989 May;66(5):2364–2368. doi: 10.1152/jappl.1989.66.5.2364. [DOI] [PubMed] [Google Scholar]
  7. Kain Z. N., Berde C. B., Benjamin P. K., Thompson J. E. Performance of pediatric resuscitation bags assessed with an infant lung simulator. Anesth Analg. 1993 Aug;77(2):261–264. doi: 10.1213/00000539-199308000-00009. [DOI] [PubMed] [Google Scholar]
  8. Kanter R. K. Evaluation of mask-bag ventilation in resuscitation of infants. Am J Dis Child. 1987 Jul;141(7):761–763. doi: 10.1001/archpedi.1987.04460070063025. [DOI] [PubMed] [Google Scholar]
  9. Karsdon J., Stijnen T., Berger H. M. The effect of a manometer on the mean airway pressure during hand ventilation, an in vitro study. Eur J Pediatr. 1989 Apr;148(6):574–576. doi: 10.1007/BF00441562. [DOI] [PubMed] [Google Scholar]
  10. Lefkowitz William. Oxygen and resuscitation: beyond the myth. Pediatrics. 2002 Mar;109(3):517–519. doi: 10.1542/peds.109.3.517. [DOI] [PubMed] [Google Scholar]
  11. Lindner W., Vossbeck S., Hummler H., Pohlandt F. Delivery room management of extremely low birth weight infants: spontaneous breathing or intubation? Pediatrics. 1999 May;103(5 Pt 1):961–967. doi: 10.1542/peds.103.5.961. [DOI] [PubMed] [Google Scholar]
  12. Martell R. J., Soder C. M. Laerdal infant resuscitators are unreliable as free-flow oxygen delivery devices. Am J Perinatol. 1997 Jul;14(6):347–351. doi: 10.1055/s-2007-994158. [DOI] [PubMed] [Google Scholar]
  13. Mondolfi A. A., Grenier B. M., Thompson J. E., Bachur R. G. Comparison of self-inflating bags with anesthesia bags for bag-mask ventilation in the pediatric emergency department. Pediatr Emerg Care. 1997 Oct;13(5):312–316. doi: 10.1097/00006565-199710000-00003. [DOI] [PubMed] [Google Scholar]
  14. Muscedere J. G., Mullen J. B., Gan K., Slutsky A. S. Tidal ventilation at low airway pressures can augment lung injury. Am J Respir Crit Care Med. 1994 May;149(5):1327–1334. doi: 10.1164/ajrccm.149.5.8173774. [DOI] [PubMed] [Google Scholar]
  15. O'Donnell C. P. F., Davis P. G., Morley C. J. Neonatal resuscitation: review of ventilation equipment and survey of practice in Australia and New Zealand. J Paediatr Child Health. 2004 Apr;40(4):208–212. doi: 10.1111/j.1440-1754.2004.00339.x. [DOI] [PubMed] [Google Scholar]
  16. Palme-Kilander C. Methods of resuscitation in low-Apgar-score newborn infants--a national survey. Acta Paediatr. 1992 Oct;81(10):739–744. doi: 10.1111/j.1651-2227.1992.tb12094.x. [DOI] [PubMed] [Google Scholar]
  17. Perlman Jeffrey M. Resuscitation--air versus 100% oxygen. Pediatrics. 2002 Feb;109(2):347–349. doi: 10.1542/peds.109.2.347. [DOI] [PubMed] [Google Scholar]
  18. Spears R. S., Jr, Yeh A., Fisher D. M., Zwass M. S. The "educated hand". Can anesthesiologists assess changes in neonatal pulmonary compliance manually? Anesthesiology. 1991 Oct;75(4):693–696. doi: 10.1097/00000542-199110000-00022. [DOI] [PubMed] [Google Scholar]
  19. Stewart A. R., Finer N. N., Peters K. L. Effects of alterations of inspiratory and expiratory pressures and inspiratory/expiratory ratios on mean airway pressure, blood gases, and intracranial pressure. Pediatrics. 1981 Apr;67(4):474–481. [PubMed] [Google Scholar]
  20. Upton C. J., Milner A. D. Endotracheal resuscitation of neonates using a rebreathing bag. Arch Dis Child. 1991 Jan;66(1 Spec No):39–42. doi: 10.1136/adc.66.1_spec_no.39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Weinberger Barry, Laskin Debra L., Heck Diane E., Laskin Jeffrey D. Oxygen toxicity in premature infants. Toxicol Appl Pharmacol. 2002 May 15;181(1):60–67. doi: 10.1006/taap.2002.9387. [DOI] [PubMed] [Google Scholar]
  22. Whyte S. D., Sinha A. K., Wyllie J. P. Neonatal resuscitation--a practical assessment. Resuscitation. 1999 Jan;40(1):21–25. doi: 10.1016/s0300-9572(98)00143-9. [DOI] [PubMed] [Google Scholar]
  23. Zmora E., Merritt T. A. Control of peak inspiratory pressure during manual ventilation. A controlled study. Am J Dis Child. 1982 Jan;136(1):46–48. doi: 10.1001/archpedi.1982.03970370048012. [DOI] [PubMed] [Google Scholar]

Articles from Archives of Disease in Childhood Fetal and Neonatal Edition are provided here courtesy of BMJ Publishing Group

RESOURCES