Background: Adequate ventilation is the key to successful neonatal resuscitation. Positive pressure ventilation (PPV) is initiated with manual ventilation devices via face masks. These devices may be used with a manometer to measure airway pressures delivered. The expiratory tidal volume measured at the mask (VTE(mask)) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation.
Aim: To assess the effect of viewing a manometer on the peak inspiratory pressures used, the volume delivered, and leakage from the face mask during PPV with two manual ventilation devices in a model of neonatal resuscitation.
Methods: Participants gave PPV to a modified resuscitation mannequin using a Laerdal infant resuscitator and a Neopuff infant resuscitator at specified pressures ensuring adequate chest wall excursion. Each participant gave PPV to the mannequin with each device twice, viewing the manometer on one occasion and unable to see the manometer on the other. Data from participants were averaged for each device used with the manometer and without the manometer separately.
Results: A total of 7767 inflations delivered by the 18 participants were recorded and analysed. Peak inspiratory pressures delivered were lower with the Laerdal device. There were no differences in leakage from the face mask or volumes delivered. Whether or not the manometer was visible made no difference to any measured variable.
Conclusions: Viewing a manometer during PPV in this model of neonatal resuscitation does not affect the airway pressure or tidal volumes delivered or the degree of leakage from the face mask.
Full Text
The Full Text of this article is available as a PDF (67.5 KB).
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Howells Rachel, Madar John. Newborn resuscitation training--which manikin. Resuscitation. 2002 Aug;54(2):175–181. doi: 10.1016/s0300-9572(02)00111-9. [DOI] [PubMed] [Google Scholar]
- O'Donnell C. P. F., Davis P. G., Morley C. J. Neonatal resuscitation: review of ventilation equipment and survey of practice in Australia and New Zealand. J Paediatr Child Health. 2004 Apr;40(4):208–212. doi: 10.1111/j.1440-1754.2004.00339.x. [DOI] [PubMed] [Google Scholar]
- O'Donnell C. P. F., Davis P. G., Morley C. J. Positive pressure ventilation at neonatal resuscitation: review of equipment and international survey of practice. Acta Paediatr. 2004 May;93(5):583–588. doi: 10.1111/j.1651-2227.2004.tb02981.x. [DOI] [PubMed] [Google Scholar]
- Palme-Kilander C. Methods of resuscitation in low-Apgar-score newborn infants--a national survey. Acta Paediatr. 1992 Oct;81(10):739–744. doi: 10.1111/j.1651-2227.1992.tb12094.x. [DOI] [PubMed] [Google Scholar]
- Phillips B., Zideman D., Wyllie J., Richmond S., van Reempts P., European Resuscitation Council European Resuscitation Council Guidelines 2000 for Newly Born Life Support. A statement from the Paediatric Life Support Working Group and approved by the Executive Committee of the European Resuscitation Council. Resuscitation. 2001 Mar;48(3):235–239. doi: 10.1016/s0300-9572(00)00382-8. [DOI] [PubMed] [Google Scholar]
- Saugstad O. D. Practical aspects of resuscitating asphyxiated newborn infants. Eur J Pediatr. 1998 Jan;157 (Suppl 1):S11–S15. doi: 10.1007/pl00014284. [DOI] [PubMed] [Google Scholar]
- Tracy M., Downe L., Holberton J. How safe is intermittent positive pressure ventilation in preterm babies ventilated from delivery to newborn intensive care unit? Arch Dis Child Fetal Neonatal Ed. 2004 Jan;89(1):F84–F87. doi: 10.1136/fn.89.1.F84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zmora E., Merritt T. A. Control of peak inspiratory pressure during manual ventilation. A controlled study. Am J Dis Child. 1982 Jan;136(1):46–48. doi: 10.1001/archpedi.1982.03970370048012. [DOI] [PubMed] [Google Scholar]