AIMS—To process video slit lamp biomicroscopic fundus image sequences in order to generate wide field, high quality fundus image montages which might be suitable for photodocumentation.
METHODS—Slit lamp biomicroscopic fundus examination was performed on human volunteers with a contact or non-contact lens. A stock, charge coupled device camera permitted image capture and storage of the image sequence at 30 frames per second. Acquisition time was approximately 30 seconds. Individual slit lamp biomicroscope fundus image frames were aligned and blended with custom developed software.
RESULTS—The developed algorithms allowed for highly accurate alignment and blending of partially overlapping slit lamp biomicroscopic fundus images to generate a seamless, high quality, wide field montage.
CONCLUSIONS—Video image acquisition and processing algorithms allow for mosaicking and enhancement of slit lamp biomicroscopic fundus images. The improved quality and wide field of view may confer suitability for inexpensive, real time photodocumentation of disc and macular abnormalities.



Full Text

The Full Text of this article is available as a PDF (98.7 KB).

Figure 1  .

Figure 1  

(A) Unprocessed slit lamp biomicroscopic fundus images of an eye with presumed ocular histoplasmosis syndrome following laser treatment for juxtafoveal choroidal neovascularisation. (B) Slit lamp fundus biomicroscopic images following automated extraction of the illuminated portion of the fundus and elimination of specular reflections.

Figure 2  .

Figure 2  

(A) Unblended mosaic of slit lamp fundus biomicroscope images for the eye depicted in Figure 1. (B) Final blended mosaic of slit lamp derived fundus biomicroscopic images.

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Berger J. W., Patel T. R., Shin D. S., Piltz J. R., Stone R. A. Computerized stereochronoscopy and alternation flicker to detect optic nerve head contour change. Ophthalmology. 2000 Jul;107(7):1316–1320. doi: 10.1016/s0161-6420(00)00157-3. [DOI] [PubMed] [Google Scholar]
  2. Berger J. W., Shin D. S. Computer-vision-enabled augmented reality fundus biomicroscopy. Ophthalmology. 1999 Oct;106(10):1935–1941. doi: 10.1016/S0161-6420(99)90404-9. [DOI] [PubMed] [Google Scholar]
  3. Lee M. S., Shin D. S., Berger J. W. Grading, image analysis, and stereopsis of digitally compressed fundus images. Retina. 2000;20(3):275–281. [PubMed] [Google Scholar]
  4. Madjarov B. D., Berger J. W. Automated, real time extraction of fundus images from slit lamp fundus biomicroscope video image sequences. Br J Ophthalmol. 2000 Jun;84(6):645–647. doi: 10.1136/bjo.84.6.645. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Rivero M. E., Bartsch D. U., Otto T., Freeman W. R. Automated scanning laser ophthalmoscope image montages of retinal diseases. Ophthalmology. 1999 Dec;106(12):2296–2300. doi: 10.1016/S0161-6420(99)90529-8. [DOI] [PubMed] [Google Scholar]
  6. Shin D. S., Javornik N. B., Berger J. W. Computer-assisted, interactive fundus image processing for macular drusen quantitation. Ophthalmology. 1999 Jun;106(6):1119–1125. doi: 10.1016/S0161-6420(99)90257-9. [DOI] [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group

RESOURCES