Skip to main content
Emergency Medicine Journal : EMJ logoLink to Emergency Medicine Journal : EMJ
. 2003 Nov;20(6):524–525. doi: 10.1136/emj.20.6.524

Standard pulse oximeters can be used to monitor respiratory rate

P Leonard 1, T Beattie 1, P Addison 1, J Watson 1
PMCID: PMC1726217  PMID: 14623838

Abstract

Background: One of the most important limitations of standard pulse oximeters is the inability to detect changes in respiratory rate until oxygenation is affected. This study sought to determine if analysis of the plethysmogram by wavelet transforms would enable the determination of changes in respiratory rate at an earlier stage.

Methods: Ten healthy adult volunteers were monitored, breathing at baseline and predetermined respiratory rates, using a standard pulse oximeter. Photo-plethysmograms captured in an attached lap top computer were then analysed using wavelet transforms.

Results: Determination of baseline respiratory rate and subsequent changes including apnoea were easily identified.

Comment: Wavelet transforms permit the accurate determination of respiratory rate by a standard pulse oximeter.

Full Text

The Full Text of this article is available as a PDF (133.3 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Addison Paul S., Watson James N., Clegg Gareth R., Steen Petter A., Robertson Colin E. Finding coordinated atrial activity during ventricular fibrillation using wavelet decomposition. IEEE Eng Med Biol Mag. 2002 Jan-Feb;21(1):58-61, 65. doi: 10.1109/51.993194. [DOI] [PubMed] [Google Scholar]
  2. Bracic M., Stefanovska A. Wavelet-based analysis of human blood-flow dynamics. Bull Math Biol. 1998 Sep;60(5):919–935. doi: 10.1006/bulm.1998.0047. [DOI] [PubMed] [Google Scholar]
  3. Vegfors M., Lindberg L. G., Pettersson H., Oberg P. A. Presentation and evaluation of a new optical sensor for respiratory rate monitoring. Int J Clin Monit Comput. 1994 Aug;11(3):151–156. doi: 10.1007/BF01132363. [DOI] [PubMed] [Google Scholar]
  4. de Trafford J., Lafferty K. What does photoplethysmography measure? Med Biol Eng Comput. 1984 Sep;22(5):479–480. doi: 10.1007/BF02447713. [DOI] [PubMed] [Google Scholar]