. 1999 Feb;81(2):206–211. doi: 10.1136/hrt.81.2.206

In vitro validation of the luminal measurement of a novel catheter based moulding technique

S Eccleshall 1, P Jordan 1, N Buller 1, J Townend 1
PMCID: PMC1728921  PMID: 9922361

Abstract

Objective—To investigate a modified angioplasty balloon catheter, which uses a novel balloon polymer to produce luminal moulds.
Design—The catheter was tested in polyurethane phantoms of diameter 1.5 to 4.0 mm. Inflations were to 1.4 atmospheres for 20 seconds at 37°C. The moulds were viewed by reinflating the balloon to 0.34 atmospheres and quantified using macrophotography and caliper measurement.
Results—Evidence of systematic error was found with lumen diameters ⩽ 2.0 mm, accuracy being 0.32 to 0.80 mm and precision 0.23 to 0.24 mm. However, between 2.5 and 4.0 mm the accuracy of measuring luminal diameters was 0.01 mm, the precision 0.06 mm, and the absolute mean error 0.05 mm. The results for percentage diameter stenosis were −1.15%, 0.86%, and 1.21%, and for lesion length they were 0.20, 0.60, and 0.41 mm, respectively.
Conclusions—The prototype catheter shows a high degree of accuracy and precision in phantoms of diameter 2.5 to 4.0 mm—the range within which most interventional work is performed. Further work on this technique as a method of direct three dimensional moulding of the coronary artery lumen is warranted.

 Keywords: moulds; luminal moulding; balloon catheter; coronary artery

Full Text

The Full Text of this article is available as a PDF (125.5 KB).

Figure 1  .

Figure 1  

The catheter: the balloon (a) with central guidewire (b) attaches proximally to the outer tube (c) and distally to the inner tube (d). The outer tube is in continuity with the manifold (e) producing a lumen between the balloon and inflation port (f). The inner tube passes through the manifold proximally at an O-ring (g), which allows it to move freely during inflation and balloon shortening.

Figure 2  .

Figure 2  

(A) Diagram illustrating an assembled complex phantom. The discs of Tufset (a) are mounted on metal rods (b) to produce a stenotic lumen (c). (B) Photograph of a mould of a complex phantom. The balloon is mounted on a mandrel adjacent to the calibration cylinder. Diameters are measured as shown (solid arrows). Lesion length is measured between the indentations (dotted arrow).

Figure 3  .

Figure 3  

Modified Bland-Altman plots. (A) The difference between the balloon and phantom for simple lumen diameters in the range 1.5 to 4.0 mm. (B) The difference between the balloon and phantom diameters for complex models in the range 2.5 to 4.0 mm. (C) The difference between the balloon and phantom for changes in luminal diameter ⩽ 1.0 mm. The dotted lines represent the mean signed difference ± 2 SD.    

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Alfonso F., Macaya C., Goicolea J., Hernandez R., Bañuelos C., Iñiguez A., Zamorano J., Zarco P. Angiographic changes (Dotter effect) produced by intravascular ultrasound imaging before coronary angioplasty. Am Heart J. 1994 Aug;128(2):244–251. doi: 10.1016/0002-8703(94)90475-8. [DOI] [PubMed] [Google Scholar]
  2. Bland J. M., Altman D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Feb 8;1(8476):307–310. [PubMed] [Google Scholar]
  3. Coy K. M., Maurer G., Siegel R. J. Intravascular ultrasound imaging: a current perspective. J Am Coll Cardiol. 1991 Dec;18(7):1811–1823. doi: 10.1016/0735-1097(91)90525-e. [DOI] [PubMed] [Google Scholar]
  4. Di Mario C., von Birgelen C., Prati F., Soni B., Li W., Bruining N., de Jaegere P. P., de Feyter P. J., Serruys P. W., Roelandt J. R. Three dimensional reconstruction of cross sectional intracoronary ultrasound: clinical or research tool? Br Heart J. 1995 May;73(5 Suppl 2):26–32. doi: 10.1136/hrt.73.5_suppl_2.26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Gurley J. C., Nissen S. E., Booth D. C., DeMaria A. N. Influence of operator- and patient-dependent variables on the suitability of automated quantitative coronary arteriography for routine clinical use. J Am Coll Cardiol. 1992 May;19(6):1237–1243. doi: 10.1016/0735-1097(92)90330-p. [DOI] [PubMed] [Google Scholar]
  6. Haase J., van der Linden M. M., Di Mario C., van der Giessen W. J., Foley D. P., Serruys P. W. Can the same edge-detection algorithm be applied to on-line and off-line analysis systems? Validation of a new cinefilm-based geometric coronary measurement software. Am Heart J. 1993 Aug;126(2):312–321. doi: 10.1016/0002-8703(93)91045-g. [DOI] [PubMed] [Google Scholar]
  7. Keane D., Haase J., Slager C. J., Montauban van Swijndregt E., Lehmann K. G., Ozaki Y., di Mario C., Kirkeeide R., Serruys P. W. Comparative validation of quantitative coronary angiography systems. Results and implications from a multicenter study using a standardized approach. Circulation. 1995 Apr 15;91(8):2174–2183. doi: 10.1161/01.cir.91.8.2174. [DOI] [PubMed] [Google Scholar]
  8. Kearney P. P., Starkey I. R., Sutherland G. R. Intracoronary ultrasound: current state of the art. Br Heart J. 1995 May;73(5 Suppl 2):16–25. doi: 10.1136/hrt.73.5_suppl_2.16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Kimura B. J., Bhargava V., DeMaria A. N. Value and limitations of intravascular ultrasound imaging in characterizing coronary atherosclerotic plaque. Am Heart J. 1995 Aug;130(2):386–396. doi: 10.1016/0002-8703(95)90457-3. [DOI] [PubMed] [Google Scholar]
  10. Nishimura R. A., Edwards W. D., Warnes C. A., Reeder G. S., Holmes D. R., Jr, Tajik A. J., Yock P. G. Intravascular ultrasound imaging: in vitro validation and pathologic correlation. J Am Coll Cardiol. 1990 Jul;16(1):145–154. doi: 10.1016/0735-1097(90)90472-2. [DOI] [PubMed] [Google Scholar]
  11. Ozaki Y., Keane D., Herrman J. P., Foley D., Haase J., den Boer A., di Mario C., Serruys P. W. Coronary arteriography for quantitative analysis: experimental and clinical comparison of cinefilm and video recordings. Am Heart J. 1995 Mar;129(3):471–475. doi: 10.1016/0002-8703(95)90270-8. [DOI] [PubMed] [Google Scholar]
  12. Reiber J. H., Serruys P. W., Kooijman C. J., Wijns W., Slager C. J., Gerbrands J. J., Schuurbiers J. C., den Boer A., Hugenholtz P. G. Assessment of short-, medium-, and long-term variations in arterial dimensions from computer-assisted quantitation of coronary cineangiograms. Circulation. 1985 Feb;71(2):280–288. doi: 10.1161/01.cir.71.2.280. [DOI] [PubMed] [Google Scholar]
  13. Roelandt J. R., di Mario C., Pandian N. G., Wenguang L., Keane D., Slager C. J., de Feyter P. J., Serruys P. W. Three-dimensional reconstruction of intracoronary ultrasound images. Rationale, approaches, problems, and directions. Circulation. 1994 Aug;90(2):1044–1055. doi: 10.1161/01.cir.90.2.1044. [DOI] [PubMed] [Google Scholar]
  14. Tabbara M., Kopchok G., White R. A. In vitro and in vivo evaluation of intraluminal ultrasound in normal and atherosclerotic arteries. Am J Surg. 1990 Dec;160(6):556–560. doi: 10.1016/s0002-9610(05)80745-7. [DOI] [PubMed] [Google Scholar]
  15. Uehata A., Matsuguchi T., Bittl J. A., Orav J., Meredith I. T., Anderson T. J., Selwyn A. P., Ganz P., Yeung A. C. Accuracy of electronic digital calipers compared with quantitative angiography in measuring coronary arterial diameter. Circulation. 1993 Oct;88(4 Pt 1):1724–1729. doi: 10.1161/01.cir.88.4.1724. [DOI] [PubMed] [Google Scholar]
  16. Von Birgelen C., Di Mario C., Reimers B., Prati F., Bruining N., Gil R., Serruys P. W., Roelandt J. R. Three-dimensional intracoronary ultrasound imaging. Methodology and clinical relevance for the assessment of coronary arteries and bypass grafts. J Cardiovasc Surg (Torino) 1996 Apr;37(2):129–139. [PubMed] [Google Scholar]
  17. Wenguang L., Gussenhoven W. J., Zhong Y., The S. H., Di Mario C., Madretsma S., van Egmond F., de Feyter P., Pieterman H., van Urk H. Validation of quantitative analysis of intravascular ultrasound images. Int J Card Imaging. 1991;6(3-4):247–253. doi: 10.1007/BF01797856. [DOI] [PubMed] [Google Scholar]

Articles from Heart are provided here courtesy of BMJ Publishing Group

RESOURCES