Abstract
Methods: The dynamic response of mini-Wright (MW), Vitalograph (V), TruZone (TZ), MultiSpiro (MS) and pneumotachograph (PT) flow meters was tested by delivering two differently shaped flow-time profiles from a computer controlled explosive decompression device fitted with a fast response solenoid valve. These profiles matched population 5th and 95th centiles for rise time from 10% to 90% of PEF and dwell time of flow above 90% PEF. Profiles were delivered five times with identical chamber pressure and solenoid aperture at PEF. Any difference in recorded PEF for the two profiles indicates a poor dynamic response.
Results: The absolute (% of mean) flow differences in l/min for the V, MW, and PT PEF meters were 25 (4.7), 20 (3.9), and 2 (0.3), respectively, at PEF ≈500 l/min, and 25 (10.5), 20 (8.7) and 6 (3.0) at ≈200 l/min. For TZ and MS meters at ≈500 l/min the differences were 228 (36.1) and 257 (39.2), respectively, and at ≈200 l/min they were 51 (23.9) and 1 (0.5). All the meters met ATS accuracy requirements when tested with their waveforms.
Conclusions: An improved method for testing the dynamic response of flow meters detects marked overshoot (underdamping) of TZ and MS responses not identified by the 26 ATS waveforms. This error could cause patient misclassification when using such meters with asthma guidelines.
Full Text
The Full Text of this article is available as a PDF (187.6 KB).
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Hankinson J. L., Reynolds J. S., Das M. K., Viola J. O. Method to produce American Thoracic Society flow-time waveforms using a mechanical pump. Eur Respir J. 1997 Mar;10(3):690–694. [PubMed] [Google Scholar]
- Janata Karin, Holzer Michael, Laggner Anton N., Müllner Marcus. Cardiac troponin T in the severity assessment of patients with pulmonary embolism: cohort study. BMJ. 2003 Feb 8;326(7384):312–313. doi: 10.1136/bmj.326.7384.312. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miles J. F., Tunnicliffe W., Cayton R. M., Ayres J. G., Miller M. R. Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan. Thorax. 1996 Apr;51(4):403–406. doi: 10.1136/thx.51.4.403. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miller M. R., Dickinson S. A., Hitchings D. J. The accuracy of portable peak flow meters. Thorax. 1992 Nov;47(11):904–909. doi: 10.1136/thx.47.11.904. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miller M. R., Jones B., Xu Y., Pedersen O. F., Quanjer P. H. Peak expiratory flow profiles delivered by pump systems. Limitations due to wave action. Am J Respir Crit Care Med. 2000 Jun;161(6):1887–1896. doi: 10.1164/ajrccm.161.6.9805085. [DOI] [PubMed] [Google Scholar]
- Miller M. R., Lloyd J., Bright P. Recording flow in the first second of a maximal forced expiratory manoeuvre: influence of frequency content. Eur Respir J. 2002 Mar;19(3):530–533. doi: 10.1183/09031936.02.00227102. [DOI] [PubMed] [Google Scholar]
- Pedersen O. F., Miller M. R., Sigsgaard T., Tidley M., Harding R. M. Portable peak flow meters: physical characteristics, influence of temperature, altitude, and humidity. Eur Respir J. 1994 May;7(5):991–997. [PubMed] [Google Scholar]
- Pedersen O. F., Rasmussen T. R., Kjaergaard S. K., Miller M. R., Quanjer P. H. Frequency response of variable orifice type peak flow meters: requirements and testing. Eur Respir J. 1995 May;8(5):849–855. [PubMed] [Google Scholar]
- Pedersen O. F., Rasmussen T. R., Omland O., Sigsgaard T., Quanjer P. H., Miller M. R. Peak expiratory flow and the resistance of the mini-wright peak flow meter. Eur Respir J. 1996 Apr;9(4):828–833. doi: 10.1183/09031936.96.09040828. [DOI] [PubMed] [Google Scholar]
- Sly P. D., Cahill P., Willet K., Burton P. Accuracy of mini peak flow meters in indicating changes in lung function in children with asthma. BMJ. 1994 Feb 26;308(6928):572–574. doi: 10.1136/bmj.308.6928.572. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.