Abstract
Background: Endothelial examination of organ culture stored corneas is usually done manually and on several mosaic zones. Some banks use an image analyser that takes account of only one zone. This method is restricted by image quality, and may be inaccurate if endothelial cell density (ECD) within the mosaic is not homogeneous. The authors have developed an analyser that has tools for automatic error detection and correction, and can measure ECD and perform morphometry on multiple zones of three images of the endothelial mosaic.
Methods: 60 human corneas were divided into two equal groups: group 1 with homogeneous mosaics, group 2 with heterogeneous ones. Three standard microscopy video images of the endothelium, graded by quality, were analysed either in isolation (so called mono-image analysis) or simultaneously (so called tri-image analysis), with 50 or 300 endothelial cells (ECs) counted. The automated analysis was compared with the manual analysis, which concerned 10 non-adjacent zones and about 300 cells. For each analysis method, failures and durations were studied according to image quality.
Results: All corneas were able to undergo analysis, in about 2 or 7.5 minutes for 50 and 300 ECs respectively. The tri-image analysis did not increase analysis time and never failed, even with mediocre images. The tri-image analysis of 300 ECs was always most highly correlated with the manual count, particularly in the heterogeneous cornea group (r=0.94, p<0.001) and prevented serious count errors.
Conclusions: This analyser allows reliable and rapid analysis of ECD, even for heterogeneous endothelia mosaics and mediocre images.
Keywords: endothelium, cornea, organ culture, automatic analysis, multi-image analysis
Cornea storage techniques, whether short or long term, require analysis of endothelial cell density (ECD). In most eye banks, cornea delivery for grafting is authorised only if ECD is greater than 2000 cells/mm2 after storage. This cell count is done either by specular microscopy during short term storage at +4°C or by standard microscopy during organ culture,1 the most common technique in Europe.2 In Europe, counting is most often done manually. Study of the 1999 register3 of French cornea banks (Table 1) which lists the data of 2425 corneas stored in organ culture at +31°C, showed a substantial difference in ECDs between banks. This was 32.6% (mean 2634 (SD 231), median 2675, range 2140–3030) at the start of storage and 30.3% (mean 2430, SD 219, median 2360, range 1986–2848) at delivery. It is important to note that all ECDs had been evaluated manually, except in bank 4, which had a semiautomated analysis system that required a technician to digitise the cell contours visible on the screen. Such differences in ECDs were not justified by differences in mean donor age (64.0 years, SD 6.2, range 49–71.5), mean death procurement time (13.7 hours, SD 7.2, range 8–29.4), or mean storage time (17.6 days, SD 2.2, range 14–21). It was legitimate to think that these large ECD differences between banks were due to manual count errors and/or errors in calibration, as may occur in any measuring system.4,5
Table 1.
Cornea bank | Corneas delivered (number) | Donor age (years) | Death-procurement time (hours) | ECD at start of organ culture (cells/mm2) | Organ culture time (days) | ECD at delivery (cells/mm2) |
1 | 126 | 62.0 | 9.25 | 2140 | 15.4 | 1986 |
2 | 102 | 67.2 | 14.2 | 2441 | 15.5 | 2131 |
3 | 31 | 54.0 | NA | 2316 | 19 | 2190 |
4 | 288 | 71.5 | 29.4 | nd | 18.5 | 2218 |
5 | 85 | 58.0 | 14 | NA | 16 | 2300 |
6 | 211 | 66.0 | NA | 2390 | 16 | 2308 |
7 | 73 | 68.0 | 11 | 2660 | 18.5 | 2340 |
8 | 64 | 64.0 | 11 | NA | 16 | 2496 |
9 | 91 | 50.0 | 8 | 2680 | 16 | 2540 |
10 | 93 | 49.4 | 12.4 | 2748 | 14.5 | 2561 |
11 | 152 | NA | 12 | 2675 | 16 | 2567 |
12 | 267 | 61.0 | 8 | 2800 | 17.3 | 2600 |
13 | 107 | 70.7 | 13.4 | 2738 | 21 | 2643 |
14 | 49 | 64.0 | 9.75 | 2806 | 17.3 | 2710 |
15 | 133 | 67.5 | 11.7 | 2645 | 14 | 2745 |
16 | 142 | 64.4 | 8.3 | 3030 | 18 | 2848 |
17 | 411 | NA | NA | NA | NA | NA |
**Only bank 4 used a computer aided analysis tool based on digitisation by tracing with the mouse the perimeter of 100 cells visible on the screen, an objective but laborious and time consuming method. The ECDs of our bank (number 2) and of the 15 others were measured manually during this period. Since our bank has been using our prototype analyser, our mean ECDs have increased a mean 4% (data not presented) and now approach those of bank 4. ECD = endothelial cell density, NA = data not available, nd = analysis not done
We believe the use of an effective, automated cell counting system is indispensable. Existing systems are of only limited use, especially because of difficulties when the image of the endothelial mosaic under analysis is mediocre. This is often the case with standard microscopy, and requires either numerous contour touch ups or a simple marking of the presumed cell centres. Moreover, whereas the technician manually observes and analyses several non-adjacent zones of the mosaic (generally 5–10 zones, depending on the bank), and a large number of cells, current analysers only take account of one endothelial zone with a limited number of cells, approximately 50 to 100. The operator's choice of a single zone of analysis may thus be responsible for a selection bias, especially if ECD within the cornea is not uniform. We present a prototype analyser with automatic error detection and correction tools, which is able to measure ECD simultaneously on several endothelial images. Its purpose is to be effective even when images are medium or only adequate, and to limit measurement error when mosaics are heterogeneous.
MATERIAL AND METHODS
Manual counts
These were done after making visible the intercellular spaces by incubation for 4 minutes with 0.9% sodium chloride, through a graduated reticle (reference 519966, Leica, Wetzlar, Germany) placed in the ocular of a Leica Laborlux S standard direct optical microscope (Leica). At the moment when the cell contours were most visible, the technician counted the ECs, contained in one microsquare of the reticle, excluding those touching two adjacent borders. He conducted an overall screening of the mosaic of the central 8 mm. This consisted of 10 counts on 10 non-adjacent zones of each cornea, avoiding the folds, which cause overcounting because of parallax error. He then multiplied the mean of his 10 measurements by a corrective factor derived from calibrating the microscope with a micrometric slide (Merkeurolab, Strasbourg, France) in order to obtain the ECD in cells/mm2. All counts were done by the same experienced technician (over 3000 counts).
Image quality and mosaic homogeneity
At the same time, when the intercellular spaces seemed well dilated (neither too much nor too little), he took three video images (Sony SC 75 CE mono CCD) of wide field (width 1000 μm, height 750 μm) of three zones, chosen at random but non-adjacent, of the paracentral endothelium. These images were saved in a tagged image format file (TIFF). Each image was examined blind and quality graded (Table 2) “adequate,” “medium,” or “good” (Figs 1A, B, C). In addition, each cornea was graded “homogeneous” if the ECDs between the three images and within each image appeared similar, or “heterogeneous” otherwise. We analysed 60 corneas, divided into two equal groups: group 1 of homogeneous corneas (30 corneas, 3 × 30 images) and group 2 of heterogeneous corneas (30 corneas, 3 × 30 images). The corneas all came from phakic eyes, and had undergone mean storage of 14 days (SD 4, range 7–31) at +31°C in Inosol (Opsia, Toulouse, France). They were first counted manually, then with our analyser.
Table 2.
Image quality | Score | Criteria |
Good | 2 | Excellent view of cell contours |
Low background noise | ||
Cells visible on over ⅔ of image area | ||
Medium | 1 | Good view of cell contours |
Moderate background noise | ||
Cells visible on ⅓ to ⅔ of image area | ||
Adequate | 0 | Poor view of cell contours |
High background noise | ||
Cells visible on less than ⅓ of image area |
For the analyses in tri-image mode, overall image quality was graded “good” if the three images obtained scores of 2/2/2 or 2/2/1, “medium” if the scores were 2/1/1, 2/1/0, or 1/1/1, and “adequate” if the scores were 1/1/0, 1/0/0, or 0/0/0
Analyser and automatic analysis procedures
Our prototype analyser comprised an IBM compatible PC, 800 MHZ Pentium, 256 Mb RAM, equipped with a video frame grabber (Data translation DT 3155, Malboro, MA, USA) installed on the same microscope as used for the manual counts. Calibration was done with the same micrometric slide. The three mosaic images were viewed side by side on a 22 inch monitor. The analysis program, written in C++, was based on a generalist image analysis software, ImageTool (University of Texas Health Science Center, San Antonio, TX, USA, available at ftp://maxrad6.uthscsa) and specifically adapted to human corneal endothelium. In particular, it took account of the frequently poor quality of images, with the following functions: selection on one image of several zones of interest according to free forms; automatic elimination of surface artefacts smaller than 50 μm2; automatic detection of probable errors in contour recognition (EC of area greater than 1500 μm2 or whose longest length was more than twice the shortest width that could correspond to poorly separated ECs). The automated analysis went as follows. Each selected zone on each image was successively subject to thresholding, which transformed the starting image, formed of 256 grey levels, into a binary image. The analyser then automatically eliminated the artefacts, and with another colour underlined EC contours that were too wide or too long (Fig 1D). If necessary, the technician then did touch ups by tracing and/or erasing contour segments. These touch ups were made easier by a zoom lens, by the possibility to return to the original image, and by the real time display of the number of ECs taken into account. The final analysis report was then printed and archived in hypertext markup language (HTML) for remote transmission to the surgeon. The report included all parameters of counting (ECD, number of ECs counted) and of morphometry (Fig 2).
For each cornea two analyses methods were compared, with precisely 50 or 300 ECs being taken into account in each method: the “mono-image” method, which analysed one or more zones of a single image (Fig 2A), and the “tri-image” method, which simultaneously analysed one or more zones of three images of the mosaic (Fig 2B). In total, eight automatic analyses (three mono-image 50, three mono-image 300, one tri-image 50, one tri-image 300) were done for each cornea and compared with the manual count. The percentage of severe discrepancy in ECD, as defined by a deviation of plus or minus 20% between the manual and automated counts ((manual ECD − automated ECD)/manual ECD) was compared in each group according to the count method. The tri-image 300 method reproduced most faithfully the strategy of global screening of the mosaic done by the technician during the manual count. The technician timed each analysis (from opening the image files to final printout of the report). The duration and percentage failure of the analysis methods were compared according to the quality of 180 images. The morphometric data, though supplied by the analyser, were not considered for this study and will be the subject of a subsequent study.
Statistics
Comparison of the ECDs and durations of each method was done with Student's t test for paired samples. The correlations between the different methods of automated and manual analysis were studied (Pearson's r coefficient) and tested against the null hypothesis of no correlation (r=0). In all the statistical tests, p<0.05 was deemed significant.
RESULTS
Image appearance, analysis failure, and duration (Table 3)
Table 3.
“Adequate” image | “Medium” image | “Good” image | All grades | |||||||||
Analysis method | Time | Failures | No | Time | Failures | No | Time | Failures | No | Time | Failures | No |
Mono-image 50 | 174 | 0% | 17 | 110 | 0% | 79 | 79 | 0% | 84 | 101 | 0% | 180 |
(98, 69–387) | (86, 43–600) | (27, 38–168) | (72, 38–600) | |||||||||
Tri-image 50 | 141 | 0% | 11 | 124 | 0% | 23 | 96 | 0% | 26 | 115 | 0% | 60 |
(49, 95–260) | (49, 59–243) | (26, 62–141) | (44, 59–260) | |||||||||
Mono-image 300 | 698 | 59% | 17 | 547 | 8% | 79 | 356 | 0% | 84 | 456 | 9% | 180 |
(287, 405–1230) | (234, 240–1200) | (126, 145–781) | (216, 145–1230) | |||||||||
Tri-image 300 | 577 | 0% | 11 | 486 | 0% | 23 | 367 | 0% | 26 | 451 | 0% | 60 |
(147, 399–767) | (110, 258–689) | (140, 138–693) | (152, 138–767) |
Mean time required was just under 2 minutes to count 50 cells and 7.5 minutes to count 300. With rare exceptions, automated analysis was far shorter than a manual count, which took 10–15 minutes. It should be noted that video image capture time allows the cornea to be returned very quickly to its storage medium. In the case of good images, analysis times were significantly shortened, falling to about 1.5 minutes for 50 cells (p<0.001) and to 6 minutes for 300 cells (p<0.001). Analysis time in tri-image mode was not longer than in mono-image mode, whichever the cell count. On the contrary, analysis time was substantially shorter for adequate images (not significant) because of multiple zone selection or a properly visible mosaic, which reduced the need for touch ups. Note finally that the only failed analyses were in mono-image 300 mode (p=0.014) and for medium or adequate images. Mean durations are expressed in seconds (SD, range).
The 60 corneas could be analysed by at least one of the eight automated methods. Of a total of 480 theoretical analyses, 464 (97%) could be done. Analysis failures were encountered only in mono-image 300 mode. Images were judged “adequate,” “medium,” or “good” in respectively 10% (n=17), 44% (n=79), and 46% (n=84) of cases. When images were of good quality, especially when 50 cells were counted, touch ups were not necessary. Analysis of the “adequate” images required more touch ups and lasted longer. The mean duration, (not including the time to capture the three images, which took about 1 minute), was less than 2 minutes for 50 cells and 7.5 minutes for 300 cells. The tri-image method was no slower than the mono-image method. On the contrary, it tended to be faster in the case of only adequate images (difference not significant). Manual analyses lasted about 10–15 minutes, comprising 5–7 minutes to count and a similar time to write the report.
Comparison of analysis methods and manual count
In manual mode, the technician counted a mean of 339 cells (SD 56, range 192–476). The ECDs provided by the automated analysis methods, in group 1 of homogeneous corneas and group 2 of heterogeneous corneas, are presented respectively in Tables 4 and 5. The manual count always slightly underestimated ECD, by mean values of 3.8% (group 1) and 7.1% (group 2) in comparison with all the automated analyses. In group 1, all the automated analysis methods were strongly correlated with the manual count (r between 0.78 and 0.93, p<0.001) (Fig 3). But in group 2 (Fig 4), the strongest correlations with the manual count were in tri-image 300 mode (r = 0.94, p <0.001), and the weakest were in mono-image 50 mode (r = 0.54, p <0.001) and then mono-image 300 mode (r = 0.59, p <0.001).
Table 4.
Analysis method | No | Mean ECD (cells/mm2(SD, range) | Mean ECD (cells/mm2) discrepancy with manual count (p) | Pearson's r* |
Mono 50 (image 1) | 30 | 2473 (389, 1773–3816) | 139 (p=0.003) | 0.81 |
Mono 50 (image 2) | 30 | 2400 (357, 1805–3460) | 66 (p=0.123) | 0.78 |
Mono 50 (image 3) | 30 | 2381 (329, 1912–3344) | 47 (p=0.172) | 0.83 |
Mono 300 (image 1) | 28† | 2480 (363, 1782–3533) | 128 (p=0.001) | 0.88 |
Mono 300 (image 2) | 28† | 2390 (324, 1814–3367) | 76 (p=0.040) | 0.82 |
Mono 300 (image 3) | 29† | 2398 (313, 1897–3174) | 63 (p=0.032) | 0.88 |
Tri image 50 | 30 | 2447 (363, 1865–3636) | 113 (p=0.002) | 0.90 |
Tri image 300 | 30 | 2422 (312, 1862–3448) | 87 (p=0.001) | 0.93 |
Manual count | 30 | 2334 (245, 1781–2990) | – | – |
*all p values <0.001.
†Missing data: 300 cells uncountable due to adequate or medium image quality. ‡‡The manual count underestimated ECD by a mean 3.8% (mean: 90 cells/mm2, range 47–139) in comparison with the automated analyses. Correlations between automated-analysis methods and manual count were good (r = 0.78–0.88) in mono-image and excellent (r = 0.90–0.93) in tri-image.
ECD = endothelial cell density, SD = standard deviation, Mono 50 = analysis of one image and 50 cells, Mono 300 = analysis of one image and 300 cells, Tri-image 50 = analysis of three images simultaneously and 50 cells, tri-image 300 = analysis of three images simultaneously and 300 cells.
Table 5.
Analysis method | No | Mean ECD (SD, range) Cells/mm2 | Mean ECD discrepancy with manual count (cells/mm2, p) | Pearson's r* |
Mono 50 (image 1) | 30 | 2139 (581, 921–3134) | 66 (p=0.35) | 0.76 |
Mono 50 (image 2) | 30 | 2220 (455, 1464–3184) | 147 (p=0.064) | 0.54 |
Mono 50 (image 3) | 30 | 2273 (633, 937–3754) | 200 (p=0.007) | 0.82 |
Mono 300 (image 1) | 26† | 2265 (452, 1259–2923) | 155 (p=0.007) | 0.81 |
Mono 300 (image 2) | 28† | 2296 (514, 1555–3484) | 229 (p=0.009) | 0.59 |
Mono 300 (image 3) | 25† | 2241 (526, 933–3246) | 183 (p=0.012) | 0.77 |
Tri image 50 | 30 | 2159 (442, 1410–2932) | 86 (p=0.019) | 0.91 |
Tri image 300 | 30 | 2204 (412, 1545–2994) | 131 (p=0.001) | 0.94 |
Manual count | – | 2073 (420, 1248–3094) | – | – |
†Missing data: 300 cells uncountable due to adequate or medium image quality. ‡As with the group 1 corneas, the manual count underestimated ECD by a mean 7.1% (mean: 148 cells/mm2, range 66–229) in comparison with the automated analyses. The best correlations with manual analysis were obtained in tri-image mode with a 300 cells count (r = 0.94), then in tri-image 50 mode (r = 0.91). However, the correlations were weaker in mono-image mode and even poor in some modes (r = 0.54 and r = 0.59).
ECD = endothelial cell density, SD = standard deviation, Mono 50 = analysis of one image and 50 cells, Mono 300 = analysis of one image and 300 cells, tri-image 50 = analysis of three images simultaneously and 50 cells, tri-image 300 = analysis of three images simultaneously and 300 cells.
*all p values <0.001.
Serious discrepancies in ECD measurement (of 20% and more) between manual and automatic methods were observed for heterogeneous corneas in 31.1% and 29.1% of cases in mono-image 50 or 300 mode respectively, but less often in tri-image 50 or 300 modes (respectively 3.3% of cases, p<0.006, and 6.7% of cases, p<0.01). In group 1 such serious discrepancies were far rarer (only three out of 240 analyses) irrespective of the analysis method and the number of cells counted.
DISCUSSION
We developed a prototype automated analyser which, using three standard microscopy video images, allows rapid and effective measurement of ECD and morphometry of organ culture stored corneas, even with adequate or medium quality images and heterogeneous mosaics. It reduces the human bias that exists with other analysers, which only analyse one mosaic zone chosen by the technician.
The use of computerised image analysis tools reduces interindividual and intraindividual variability by lowering the risk of error, and saves time compared to manual counting.6–8 Such tools were originally developed for in vivo analysis of specular microscopy images9–17 and then for endothelial examination during short term storage at +4°C.18,19 The contours of ECs viewed in specular mode are non modified and relatively well contrasted. The moderate descemetic folds and stromal oedema seen in short storage allow generation of easy to analyse images. However, some authors stress that the accuracy of automated specular image analysers remains linked to image quality13,15,16,20 and to the size of the visible field.10
Automated analysis of the mosaic in optical microscopy, the most common technique in European banks that use organ culture, is more difficult given the often poorer image quality.6,21 EC contours can only be viewed after osmotically induced dilation of the intercellular spaces, typically by a solution of 0.9% NaCl.22 This dilation varies from one cornea to another and between zones of the same cornea.6 It must not be excessive, because analysis of too retracted cells is very difficult. The use of 0.9% NaCl for about 4 minutes helped us obtain medium (44%) and good (46%) images without recourse to other osmotic agents liable to enhance viewing of the cell contours, such as 1.8% sucrose.23 Furthermore, the technician must capture the images when the intercellular spaces are well dilated, and adjust the focus, light, and the microscope condenser to improve contour definition and contrast and the percentage of visible cells, and to limit background noise. Thresholding, which is done independently on each zone of each image, must also receive special attention since it reduces the need for touch ups. When images were of good quality, very limited and sometimes no touch ups were needed. Prolonging organ culture increases the number of folds23 and the stromal oedema,24 and can also degrade images. Barisani, with his mono-image analyser, had 44% failure beyond 21 days' organ culture.6 This was not the case in our study, because even in the case of prolonged storage (up to 31 days in our series), analysis was always possible in tri-image mode, even with a high cell count (300). The possibility to choose several zones on the same image according to the free forms, avoiding zones where the mosaic was not visible, and to repeat the operation on the three images, prevented failure even with mediocre images. Lastly, direct acquisition of the video images, without an intermediate photographic stage, limited information loss and eliminated the risk of errors linked to image magnification,16,18 because the entire optical and computing chain was directly calibrated.
All automated analysers currently used in optical and specular microscopy select a single parallelepipedic zone for analysis. This has two drawbacks: (1) this form does not always allow elimination from analysis of zones where the image is worst, as stated above, or of zones of folds responsible for ECD overestimation due to parallax error; (2) since there is only one zone, which is often limited in size, this reduces accordingly the number of cells counted and the accuracy of the ECD measurement. Hirst points out that such an analysis is only representative of the whole endothelium in cases where the mosaic is particularly homogeneous.11,14 Development of wide field specular microscopy, enabling 2000 to 3000 cells to be viewed,25 allowed better mosaic sampling and analysis of a very large number of cells (over 400).12 However, the technology of the time limited its use because of excessive analysis time, close to 30 minutes, and was only suited to in vivo specular microscopy. With our analyser, selection of several zones on three wide field optical microscopy images makes it theoretically possible to analyse between 4000 and 6000 cells on a corneal area of 2.25 mm2, constituting a sample of over 4% of the central 8 mm area (50.26 mm2). This is the widest field analysed in the literature. We have thus shown that analysis of a single image of 50 or even 300 cells could be faulted, especially if the mosaic is heterogeneous. This was not the case with our original analysis method, which considers three images simultaneously.
Corneas with a heterogeneous endothelium are not rare. Observation of thousands of consecutive images of endothelial mosaics led us to note that about one cornea in five may be deemed heterogeneous (data not shown). Some zones, where polymegethism and pleomorphism are greatest, may have a lower ECD than others where the mosaic is more uniform. This was shown by cartographic studies evaluating the spacial variations of the ECD in corneal pathologies such as keratoconus.26 This has also been noted in scarring procedures (especially cataract surgery)27 which are generally excluded from procurement. However, ageing is a factor in heterogeneity28–30 that banks increasingly face, and particularly in Europe, where donor age is high.31,32 In our series, in tri-image 300 mode the 30 heterogeneous corneas had a mean ECD of 2204 cells/mm2 (SD 412, range 1545–2994) and 19 of these (63%) were grafted. In this group, mono-image analysis gave substantial discrepancies in ECD measurement, more than 20% (greater or less) than the manual count, for nearly a third of the corneas. These could thus have been wrongly eliminated or delivered. The possibility, in tri-image 300 mode, to analyse simultaneously several non-adjacent zones on several images and a large number of cells limits technician bias in selecting a single mosaic zone: it reproduces his strategy of overall screening of multiple mosaic zones and takes account of 300 cells as when he counts manually. This analysis method is most strongly correlated with manual analysis, particularly for heterogeneous corneas (r = 0.94), but also for homogeneous ones (r =0.93).
Analysis time is necessarily lengthened by counting of a large number of cells. While the number of ECs counted increases the validity of ECD measurement through improved sampling, it also increases the risk of touch up related error, the laboriousness of the operation, and the cost of analysis. Most authors thus deemed counting of 75 to 100 ECs, in specular33–35 and optical35,36 microscopy, a good compromise, assuming the endothelial mosaic is relatively homogeneous. Moreover, the tools of that time could typically analyse only a relatively restricted field of the mosaic, and computers were slower and did not offer automated error detection or correction. Time is saved by programming these features, in particular the elimination of surfaces smaller than 50 μm2 corresponding to debris, nuclei, or contour artefacts. Likewise, automatic detection of probable contour recognition error enables touch ups to be guided, thus making them more rapid and accurate, an appreciable factor in morphometric validity. Our mean analysis time was just under 7.5 minutes for 300 cells. We felt it unreasonable to design a process exceeding 10 minutes, the duration of a manual analysis, in order to make the analyser routinely available to the technician. The choice of three images, rather than four, five or even more, was partly imposed by this time constraint. The tri-image mode does not increase analysis time; and in the case of mediocre images substantially shortens it by allowing selection of multiple zones of interest where the mosaic is clearly visible. Furthermore, these error detection and correction tools enhance system user friendliness and make learning to use it easier, a skill acquired after a few counts. Lastly, two major benefits should be emphasised: (1) because analysis can be deferred, capture of the three images takes about 1 minute, which considerably reduces the time spent by the corneas out of the storage medium; (2) image archiving (in TIFF format) and results transmission via a computer network (thanks to the HTML format) save considerable time.
CONCLUSIONS
Since January 2001our analyser has been routinely used in our bank, and analysis in tri-image mode of 300 cells is now our gold standard for cornea delivery. This analysis method has markedly increased our cornea delivery rate, but also allowed more stringent selection, especially for heterogeneous mosaics. This is important for our clinical research, where accurate ECD data at delivery are required for postoperative study of diminishing ECDs. More recently, our laboratory's acquisition of a second analyser has proved particularly useful in the experimental evaluation of new storage media, where slight ECD and morphometric variations must be highlighted. Provision of this system to Europe's ophthalmological community may represent a step towards the standardisation and improvement of endothelial examination within and between cornea banks.
Acknowledgments
We thank the AIRE association (which supports local medical research) for funding (2000–01 research programme)
Part of this paper was presented at the 14th annual meeting of the European Eye Bank Association, held on 18–19 January 2002 in Paris, France. It won the best oral presentation award.
Proprietary interest: none.
REFERENCES
- 1.Delbosc B, Borderie V. Methods of preservation of the human corneas. J Fr Ophtalmol 1997;20:221–40. [PubMed] [Google Scholar]
- 2.Maas-Reijs J, Pels E, Tullo AB. Eye banking in Europe 1991–1995. Acta Ophthalmol Scand 1997;75:541–3. [DOI] [PubMed] [Google Scholar]
- 3.French directory of corneal storage centres. 7th ed. 1999.
- 4.Ohno K, Nelson LR, McLaren JW, et al. Comparison of recording systems and analysis methods in specular microscopy. Cornea 1999;18:416–23. [DOI] [PubMed] [Google Scholar]
- 5.Benetz BA, Diaconu E, Bowlin SJ, et al. Comparison of corneal endothelial image analysis by Konan SP8000 noncontact and Bio-Optics Bambi systems. Cornea 1999;18:67–72. [PubMed] [Google Scholar]
- 6.Barisani-Asenbauer T, Baumgartner I, Grabner G, et al. Automated digital image analysis of organ culture preserved donor corneas. Ophthalmic Res 1993;25:94–9. [DOI] [PubMed] [Google Scholar]
- 7.Seitz B, Muller EE, Langenbucher A, et al. Reproducibility and validity of a new automatic method of specular microscopy analysis of corneal endothelium. Ophthalmologe 1997;94:127–35. [DOI] [PubMed] [Google Scholar]
- 8.Reinhard T, Spelsberg H, Holzwarth D, et al. Endothelial evaluation of corneal transplants by digital imaging. Klin Monatsbl Augenheilkd 1999;214:407–11. [DOI] [PubMed] [Google Scholar]
- 9.Rao GN, Waldron WR, Aquavella JV. Morphology of graft endothelium and donor age. Br J Ophthalmol 1980;64:523–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Lester JM, McFarland JL, Bursell SE, et al. Automated morphometric analysis of corneal endothelial cells. Invest Ophthalmol Vis Sci 1981;20:407–10. [PubMed] [Google Scholar]
- 11.Hirst LW, Auer C, Abbey H, et al. Quantitative analysis of wide-field endothelial specular photomicrographs. Am J Ophthalmol 1984;97:488–95. [DOI] [PubMed] [Google Scholar]
- 12.Hirst LW, Sterner RE, Grant DG. Automated analysis of wide-field specular photomicrographs. Cornea 1984;3:83–7. [PubMed] [Google Scholar]
- 13.Nishi O, Hanasaki K. Automated morphometry of corneal endothelial cell: use of video camera and video tape recorder. Br J Ophthalmol 1988;72:68–73. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Hirst LW, Yamauchi K, Enger C, et al. Quantitative analysis of wide-field specular microscopy. II. Precision of sampling from the central corneal endothelium. Invest Ophthalmol Vis Sci 1989;30:1972–9. [PubMed] [Google Scholar]
- 15.Nishi O, Hanasaki K. Automated determination of polygonality of corneal endothelial cells. Cornea 1989;8:54–7. [PubMed] [Google Scholar]
- 16.Siertsema JV, Landesz M, van den Brom H, et al. Automated video image morphometry of the corneal endothelium. Doc Ophthalmol 1993;85:35–44. [DOI] [PubMed] [Google Scholar]
- 17.Landesz M, Siertsema JV, Van Rij G. Comparative study of three semiautomated specular microscopes. J Cataract Refract Surg 1995;21:409–16. [DOI] [PubMed] [Google Scholar]
- 18.Hartmann C, Koditz W. Automated morphometric endothelial analysis combined with video specular microscopy. Cornea 1984;3:155–67. [PubMed] [Google Scholar]
- 19.Matsuda M, Yee RW, Glasser DB, et al. Specular microscopic evaluation of donor corneal endothelium. Arch Ophthalmol 1986;104:259–62. [DOI] [PubMed] [Google Scholar]
- 20.Fabian E, Mertz M, Koditz W. Morphometry of the endothelium using automated television image analysis. Klin Monatsbl Augenheilkd 1983;182:218–23. [DOI] [PubMed] [Google Scholar]
- 21.Delbosc B, Cantenot L, Fellmann D, et al. In vitro semi-automated morphometry of preserved human corneal endothelium. J Fr Ophtalmol 1986;9:445–54. [PubMed] [Google Scholar]
- 22.Sperling S. Evaluation of the endothelium of human donor corneas by induced dilation of intercellular spaces and trypan blue. Graefes Arch Clin Exp Ophthalmol 1986;224:428–34. [DOI] [PubMed] [Google Scholar]
- 23.Pels E, Schuchard Y. Organ-culture preservation of human corneas. Doc Ophthalmol 1983;56:147–53. [DOI] [PubMed] [Google Scholar]
- 24.Muller LJ, Pels E, Vrensen GF. The effects of organ-culture on the density of keratocytes and collagen fibers in human corneas. Cornea 2001;20:86–95. [DOI] [PubMed] [Google Scholar]
- 25.Koester CJ, Roberts CW, Donn A, et al. Wide field specular microscopy. Clinical and research applications. Ophthalmology 1980;87:849–60. [DOI] [PubMed] [Google Scholar]
- 26.Corkidi G, Marquez J, Garcia-Ruiz M, et al. Cartographic system for spatial distribution analysis of corneal endothelial cells. Med Biol Eng Comput 1994;32:421–6. [DOI] [PubMed] [Google Scholar]
- 27.Mishima S. Clinical investigations on the corneal endothelium-XXXVIII Edward Jackson Memorial Lecture. Am J Ophthalmol 1982;93:1–29. [DOI] [PubMed] [Google Scholar]
- 28.Yee RW, Matsuda M, Schultz RO, et al. Changes in the normal corneal endothelial cellular pattern as a function of age. Curr Eye Res 1985;4:671–8. [DOI] [PubMed] [Google Scholar]
- 29.Sherrard ES, Novakovic P, Speedwell L. Age-related changes of the corneal endothelium and stroma as seen in vivo by specular microscopy. Eye 1987;1:197–203. [DOI] [PubMed] [Google Scholar]
- 30.Bourne WM, Nelson LR, Hodge DO. Central corneal endothelial cell changes over a ten-year period. Invest Ophthalmol Vis Sci 1997;38:779–82. [PubMed] [Google Scholar]
- 31.Borderie VM, Scheer S, Touzeau O, et al. Donor organ cultured corneal tissue selection before penetrating keratoplasty. Br J Ophthalmol 1998;82:382–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Gain P, Thuret G, Chiquet C, et al. Cornea procurement from very old donors: post organ culture cornea outcome, and recipient graft outcome. Br J Ophthalmol 2002;86:404–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Sugar A. Clinical specular microscopy. Surv Ophthalmol 1979;24:21–32. [DOI] [PubMed] [Google Scholar]
- 34.Waring GOd, Bourne WM, Edelhauser HF, et al. The corneal endothelium. Normal and pathologic structure and function. Ophthalmology 1982;89:531–90. [PubMed] [Google Scholar]
- 35.Geroski DH, Edelhauser HF. Morphometric analysis of the corneal endothelium. Specular microscopy vs alizarin red staining. Invest Ophthalmol Vis Sci 1989;30:254–9. [PubMed] [Google Scholar]
- 36.Doughty MJ, Fonn D, Trang Nguyen K. Assessment of the reliability of calculations of the coefficient of variation for normal and polymegethous human corneal endothelium. Optom Vis Sci 1993;70:759–70. [DOI] [PubMed] [Google Scholar]