Abstract
One hundred and twenty five measurements of arterial oxygen saturation (Stcao2) obtained with a transcutaneous pulse oximeter were compared with direct arterial oxygen tension (Pao2) in 13 preterm infants with predominantly fetal haemoglobin. Stcao2 of 86-92% corresponded to Pao2 of 5-13 kPa. Stcao2 above 92%, however, was sometimes associated with Pao2 above 13 kPa.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Deckardt R., Steward D. J. Noninvasive arterial hemoglobin oxygen saturation versus transcutaneous oxygen tension monitoring in the preterm infant. Crit Care Med. 1984 Nov;12(11):935–939. doi: 10.1097/00003246-198411000-00002. [DOI] [PubMed] [Google Scholar]
- GARBY L., SJOLIN S., VUILLE J. C. Studies on erythro-kinetics in infancy. II. The relative rate of synthesis of haemoglobin F and haemoglobin A during the first months of life. Acta Paediatr. 1962 May;51:245–254. doi: 10.1111/j.1651-2227.1962.tb06538.x. [DOI] [PubMed] [Google Scholar]
- Kollmeyer K. R., Tsang R. C. Complications of umbilical oxygen electrodes. J Pediatr. 1974 Jun;84(6):894–897. doi: 10.1016/s0022-3476(74)80777-8. [DOI] [PubMed] [Google Scholar]
- Ryan C. A., Barrington K. J., Vaughan D., Finer N. N. Directly measured arterial oxygen saturation in the newborn infant. J Pediatr. 1986 Sep;109(3):526–529. doi: 10.1016/s0022-3476(86)80137-8. [DOI] [PubMed] [Google Scholar]
- Wilkinson A. R., Phibbs R. H., Gregory G. A. Improved accuracy of continuous measurement of arterial oxygen tension in sick newborn infants. Criteria for reading and recalibrating the electrode. Arch Dis Child. 1979 Apr;54(4):307–310. doi: 10.1136/adc.54.4.307. [DOI] [PMC free article] [PubMed] [Google Scholar]