Abstract
The irradiance produced by phototherapy units in use on a neonatal unit was measured. The phototherapy tubes were found to decline at a variable rate. Heat shields used to decrease heat and evaporative losses in the very low birthweight infant reduce irradiance and may appreciably limit the effectiveness of phototherapy.
Full text
PDF


Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Baumgart S., Fox W. W., Polin R. A. Physiologic implications of two different heat shields for infants under radiant warmers. J Pediatr. 1982 May;100(5):787–790. doi: 10.1016/s0022-3476(82)80594-5. [DOI] [PubMed] [Google Scholar]
- Brown A. K., Kim M. H., Wu P. Y., Bryla D. A. Efficacy of phototherapy in prevention and management of neonatal hyperbilirubinemia. Pediatrics. 1985 Feb;75(2 Pt 2):393–400. [PubMed] [Google Scholar]
- Marks K. H., Friedman Z., Maisels M. J. A simple device for reducing insensible water loss in low-birth-weight infants. Pediatrics. 1977 Aug;60(2):223–226. [PubMed] [Google Scholar]
- Tan K. L. The pattern of bilirubin response to phototherapy for neonatal hyperbilirubinaemia. Pediatr Res. 1982 Aug;16(8):670–674. doi: 10.1203/00006450-198208000-00018. [DOI] [PubMed] [Google Scholar]