Skip to main content
British Medical Journal logoLink to British Medical Journal
. 1971 Aug 28;3(5773):512–516. doi: 10.1136/bmj.3.5773.512

Leakage of Oxygen from Blood and Water Samples Stored in Plastic and Glass Syringes

Peter V Scott, J N Horton, W W Mapleson
PMCID: PMC1800464  PMID: 5565518

Abstract

Theory and experiment showed that samples of blood and water stored in 2-ml and 5-ml syringes made of polypropylene, polystyrene, or S.A.N. co-polymer exchanged oxygen with their surroundings. In the first hour the exchange was due mainly to equilibration with the plastic of the syringe and only in small degree to permeation through the plastic. With high initial tension or with blood of low haemoglobin concentration the exchange can result in errors in Po2 of up to 6% in two minutes and 16% in 30 to 60 minutes. With all-glass syringes the exchange was much slower but, even so, after 24 hours was important in all but a few of 18 interchangeable glass syringes. Therefore unless analysis can be started immediately all-glass syringes are to be preferred, and for prolonged storage even these should be selected.

Full text

PDF
512

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. ASMUSSEN E., NIELSEN M. The disappearance of oxygen from shed blood at body temperature. Scand J Clin Lab Invest. 1961;13:297–300. [PubMed] [Google Scholar]
  2. Adams A. P., Morgan-Hughes J. O., Sykes M. K. pH and blood-gas analysis. Methods of measurement and sources of error using electrode systems. Anaesthesia. 1967 Oct;22(4):575–597. doi: 10.1111/j.1365-2044.1967.tb10156.x. [DOI] [PubMed] [Google Scholar]
  3. CHARLTON G., READ D., READ J. CONTINUOUS INTRA-ARTERIAL PO2 IN NORMAL MAN USING A FLEXIBLE MICROELECTRODE. J Appl Physiol. 1963 Nov;18:1247–1251. doi: 10.1152/jappl.1963.18.6.1247. [DOI] [PubMed] [Google Scholar]
  4. Fletcher G., Barber J. L. Effect of sampling technique on the determination of PaO2, during oxygen breathing. J Appl Physiol. 1966 Mar;21(2):463–468. doi: 10.1152/jappl.1966.21.2.463. [DOI] [PubMed] [Google Scholar]
  5. HEDLEY-WHYTE J., LAVER M. B. O2 SOLUBILITY IN BLOOD AND TEMPERATURE CORRECTION FACTORS FOR PO2. J Appl Physiol. 1964 Sep;19:901–906. doi: 10.1152/jappl.1964.19.5.901. [DOI] [PubMed] [Google Scholar]
  6. Hilty H., Karendal B. Effect of syringe material on oxygen tension in stored blood. Acta Soc Med Ups. 1969;74(3-4):193–205. [PubMed] [Google Scholar]
  7. Kelman G. R. Digital computer subroutine for the conversion of oxygen tension into saturation. J Appl Physiol. 1966 Jul;21(4):1375–1376. doi: 10.1152/jappl.1966.21.4.1375. [DOI] [PubMed] [Google Scholar]
  8. LAVER M. B., SEIFEN A. MEASUREMENT OF BLOOD OXYGEN TENSION IN ANESTHESIA. Anesthesiology. 1965 Jan-Feb;26:73–101. doi: 10.1097/00000542-196501000-00013. [DOI] [PubMed] [Google Scholar]
  9. LeBlanc A. D., Johnson P. C. The handling of xenon-133 in clinical studies. Phys Med Biol. 1971 Jan;16(1):105–109. doi: 10.1088/0031-9155/16/1/310. [DOI] [PubMed] [Google Scholar]
  10. Mapleson W. W., Horton J. N., Ng W. S., Imrie D. D. The response pattern of polarographic oxygen electrodes and its influence on linearity and hysteresis. Med Biol Eng. 1970 Nov;8(6):585–593. doi: 10.1007/BF02478232. [DOI] [PubMed] [Google Scholar]
  11. Severinghaus J. W. Blood gas calculator. J Appl Physiol. 1966 May;21(3):1108–1116. doi: 10.1152/jappl.1966.21.3.1108. [DOI] [PubMed] [Google Scholar]

Articles from British Medical Journal are provided here courtesy of BMJ Publishing Group

RESOURCES