Abstract
Three procedures (radioimmunoassay, fluorescence immunoassay, and enzyme immunoassay) for the determination of gentamicin and tobramycin levels were compared. These systems were evaluated on the basis of accuracy, reproducibility, specificity, and cost. All three systems showed a high degree of accuracy and precision. The fluorescence immunoassay gave significantly lower values for gentamicin and tobramycin at levels below 5 microgram/ml. Of the three systems, the radioimmunoassay was the least expensive, provided that a minimum of 10 analyses were performed daily; single tests required more technical time and were more costly. Costs of single tests by fluorescence immunoassay and enzyme immunoassay systems were similar. The enzyme immunoassay system performed equally well for both gentamicin and tobramycin in giving rapid and accurate results.
Full text
PDF




Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Burd J. F., Wong R. C., Feeney J. E., Carrico R. J., Boguslaski R. C. Homogeneous reactant-labeled fluorescent immunoassay for therapeutic drugs exemplified by gentamicin determination in human serum. Clin Chem. 1977 Aug;23(8):1402–1408. [PubMed] [Google Scholar]
- Lantz C. H., Lawrie D. J., Witebsky F. G., MacLowry J. D. Evaluation of serum gentamicin assay procedures for a clinical microbiology laboratory. J Clin Microbiol. 1980 Oct;12(4):583–589. doi: 10.1128/jcm.12.4.583-589.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mahon W. A., Ezer J., Wilson T. W. Radioimmunoassay for measurement of gentamicin in blood. Antimicrob Agents Chemother. 1973 May;3(5):585–589. doi: 10.1128/aac.3.5.585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Phaneuf D., Francke E., Neu H. C. Rapid, reproducible enzyme immunoassay for gentamicin. J Clin Microbiol. 1980 Mar;11(3):266–269. doi: 10.1128/jcm.11.3.266-269.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sabath L. D., Casey J. I., Ruch P. A., Stumpf L. L., Finland M. Rapid microassay of gentamicin, kanamycin, neomycin, streptomycin, and vancomycin in serum or plasma. J Lab Clin Med. 1971 Sep;78(3):457–463. [PubMed] [Google Scholar]
- Smith A. L., Waitz J. A., Smith D. H., Oden E. M., Emerson B. B. Comparison of enzymatic and microbiological gentamicin assays. Antimicrob Agents Chemother. 1974 Sep;6(3):316–319. doi: 10.1128/aac.6.3.316. [DOI] [PMC free article] [PubMed] [Google Scholar]