Abstract
Prosthesis‐patient mismatch (PPM) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size. Its main haemodynamic consequence is to generate higher than expected gradients through normally functioning prosthetic valves. This review updates the present knowledge about the impact of PPM on clinical outcomes. PPM is common (20–70% of aortic valve replacements) and has been shown to be associated with worse haemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival. Moreover, as opposed to most other risk factors, PPM can largely be prevented by using a prospective strategy at the time of operation.
Keywords: heart valve disease, heart valve prosthesis, haemodynamic function, Doppler echocardiography, aortic stenosis
Prosthesis‐patient mismatch (PPM) was first described in 1978 by Rahimtoola as follows: “Mismatch can be considered to be present when the effective prosthetic valve area, after insertion into the patient, is less than that of a normal human valve”.1 Inherent in this concept is that a smaller than expected effective orifice area (EOA) in relation to the patient's body surface area (BSA) will result in higher transvalvar gradients. This is best exemplified by the hydraulic equation TPG = Q2/[k × EOA2], which shows that the transvalvar pressure gradient (TPG) is directly related to the square of transvalvar flow (Q) and inversely related to the square of the valve EOA; k is a constant. Hence, the EOA must be proportionate to flow requirement for gradients to remain low. At rest, transvalvar flow is largely related to cardiac output, which in turn is determined by BSA. PPM thus occurs when the EOA of the prosthetic valve is too small in relation to a patient's body size.1,2,3,4 The immediate consequence is the persistence of abnormally high TPGs.
For instance, assuming a normal cardiac index of 3 l/min/m2, the implantation of a prosthesis with an EOA of 1.3 cm2 in a patient with a BSA of 1.5 m2 will theoretically result in a mean TPG of about 13 mm Hg. The mean TPG will theoretically be 35 mm Hg if the same prosthesis is implanted in a patient with a BSA of 2.5 m2 (table 15). Moreover, the difference in TPGs between these two patients would be even more important during exercise, given that gradients are a square function of flow.
Table 1 Theoretical comparison of mean transvalvar pressure gradient in five hypothetical patients receiving the same prosthetic valve but having different body surface areas.
| Patient number | |||||
|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | |
| Body surface area (m2) | 1.5 | 1.75 | 2.0 | 2.25 | 2.5 |
| Cardiac output (l/min) | 4.5 | 5.25 | 6.0 | 6.75 | 7.5 |
| Valve EOA (cm2) | 1.3 | 1.3 | 1.3 | 1.3 | 1.3 |
| Mean pressure gradient (mm Hg) | 13 | 17 | 22 | 28 | 35 |
For this simulation, mean pressure gradient was calculated assuming a cardiac index of 3 l/min/m2, a heart rate of 65 beats/min, and a systolic ejection time of 300 ms.
EOA, effective orifice area.
Reproduced and modified from Pibarot and Dumesnil5 with permission of Remedica Publishing.
PARAMETER USED TO DEFINE MISMATCH
Consistent with these physiological considerations, the parameter used to characterise PPM is the indexed EOA—that is, the EOA of the prosthesis divided by the patient's BSA. The indexed EOA is in fact the only parameter that has been found to consistently correlate with postoperative gradients. Figure 1 shows that the relation between the TPG and the indexed EOA is curvilinear and that gradients increase exponentially when the indexed EOA is ⩽ 0.8 to 0.9 cm2/m2.6 On the basis of this relation, an indexed EOA ⩽ 0.85 cm2/m2 is generally regarded as the threshold for PPM in the aortic position with values between 0.65–0.85 cm2/m2 being classified as moderate PPM and those < 0.65 cm2/m2 as severe PPM.2,3,6,7,8 Such categorisation is important because the impact of PPM on clinical outcomes increases with severity. Depending on studies, the reported prevalence of moderate PPM varies between 20–70%, whereas that of severe PPM is between 2% and 11%.6,7,8,9,10,11,12,13,14,15,16,17,18,19,20
Figure 1 Correlation between mean transvalvar gradient and indexed effective orifice area in patients with a stented bioprosthesis (n = 51; solid circles), a stentless bioprosthesis (n = 194; open circles), an aortic homograft (n = 55; triangles), and a pulmonary autograft (n = 96; squares). Several points overlap. Reproduced from Pibarot and Dumesnil6 with permission of the American College of Cardiology.
It should be noted that some authors have attempted to characterise PPM by using the internal geometric area (IGA) of the prosthesis rather than the EOA because IGA is more reproducible. The IGA is a static manufacturing specification based on the ex vivo measurement of the diameter of the prosthesis. The criteria used for its measurement unfortunately differ from one type of prosthesis to another so that, for instance, the IGA grossly overestimates the EOA but to a much larger extent in the case of a bioprosthesis than in the case of a mechanical prosthesis (fig 2).18,19 Hence, the relation between IGA and EOA varies extensively depending on the type and size of prosthesis and it has been shown that the indexed IGA cannot be used to predict postoperative gradients (fig 3A).14,19,20,21 This observation is further corroborated by the recent data of Koch et al21 that, as compared with homografts, pericardial valves have similar values for indexed IGA but more or less two‐fold values for peak and mean gradients. The same authors also found no relation between the indexed IGA and functional improvement after an aortic valve replacement (AVR). As well, most studies that have used the indexed IGA have failed to find any significant relation between this parameter and adverse clinical outcomes (table 2).21,22,23,24,25 This should, however, come as no surprise since, as mentioned, the indexed IGA does not bear any relation whatsoever to postoperative haemodynamic function.14,19,20,21 In contrast, the indexed EOA has consistently been shown to correlate with postoperative gradients (fig 1, fig 3B), as well as being highly predictive of adverse outcomes (table 2).2,5,6,7,8,9,10,11,13,14,15,16,17,18,19,20,26,27,28,29,30
Figure 2 View of a bioprosthesis and a bileaflet mechanical valve with the leaflets in a fully open position. The area highlighted in pink is the effective orifice area (EOA). Reproduced and modified from Pibarot and Dumesnil5 with permission of Remedica Publishing.
Figure 3 Correlation between postoperative mean gradient and parameters calculated at the time of operation: (A) indexed internal geometric area and (B) projected indexed EOA. Open circles represent stentless bioprostheses; closed circles represent stented bioprostheses. Reproduced from Pibarot et al20 with permission of the Society of Thoracic Surgeons.
Table 2 Summary of the clinical impact of prosthesis‐patient mismatch (PPM).
| Outcome variable | Parameter* | Clinical impact of PPM | Reference |
|---|---|---|---|
| LV mass and function | |||
| LV mass regression | EOA‐I | Decreased 4.5% with PPM v 23% without PPM (p = 0.0001) | 29 |
| EOA‐I | Decreased at 2 years from baseline Δ−47 g with PPM v Δ−77 g without PPM (p = 0.002) | 15 | |
| IGA‐I | No influence on LV mass regression | 22 | |
| LV systolic function (cardiac index) | EOA‐I | Decreased at 5 years from baseline Δ−0.54 (l/min/m2) with PPM v Δ−0.17 (l/min/m2) without PPM (p = 0.04) | 8 |
| Improvement of NYHA classification | EOA‐I | Functional class improved +1.5 with PPM v +1.9 without PPM (p = 0.009) | 8 |
| IGA‐I | No influence on functional recovery | 21 | |
| Freedom from CHF | EOA‐I | 60% increase in the risk of CHF | 26 |
| EOA‐I | 71% with severe PPM v 86% with moderate PPM and 87% without PPM 8 years postoperatively (p = 0.02) | 14 | |
| Freedom from late cardiac events | EOA‐I | 56% with severe PPM v 80% with moderate PPM and 94% without PPM 15 years postoperatively (p = 0.03) | 11 |
| EOA‐I | 75% with PPM v 87% without PPM 5 years postoperatively (p = 0.005) | 19 | |
| Early mortality | EOA‐I | 7.9% with PPM v 4.6% without PPM (p = 0.03) | 10 |
| EOA‐I | 26% with severe PPM v 6% with moderate PPM and 3% without PPM (p<0.001) | 13 | |
| IGA‐I | No influence on early mortality | 23 | |
| IGA‐I | No influence on early mortality | 24 | |
| IGA‐I | 1–2% increase in early mortality (p = 0.002) | 25 | |
| Freedom from late mortality | EOA‐I | 75% with PPM v 84% without PPM 12 years postoperatively (p = 0.0004) | 10 |
| EOA‐I | 41% with severe PPM v 65% with moderate PPM and 74% without PPM 8 years postoperatively (p = 0.002) | 14 | |
| EOA‐I | 82% with PPM v 93% without PPM 5 years postoperatively (p = 0.01) | 19 | |
| IGA‐I | No influence on late mortality | 24 | |
| IGA‐I | No influence on late mortality | 25 | |
| High risk patients | |||
| Early mortality (LVEF <35%) | NA | 47% with small prosthesis v 15% with larger prosthesis (p = 0.03) | 30 |
| Early mortality (LVEF <40%) | EOA‐I | 67% with severe PPM v 16% with moderate PPM and 7% without PPM (p<0.001) | 13 |
*Parameter used to define PPM.
CHF, congestive heart failure; EOA‐I, indexed effective orifice area; IGA‐I, indexed internal geometric orifice area; LV, left ventricular; LVEF, left ventricular ejection fraction; NA, not available; NYHA, New York Heart Association.
In view of the above, the following points need to be emphasised. Firstly, it is not the size (labelled size or IGA) of the prosthesis that matters but rather its EOA and in whom you implant it. Secondly, the only parameter yet shown to be valid to define PPM is the indexed EOA. Thirdly, the indexed IGA and labelled valve size cannot be used to identify PPM or to characterise its severity. To avoid any confusion about the interpretation of results of the different studies, the terminology used to describe these phenomena also apparently must be consistent and without ambiguity. Hence, “indexed orifice area” should not be used without specifying whether it is the indexed IGA or the indexed EOA. Also, given that the term PPM stems from a haemodynamic concept, its use should be reserved for data relating to haemodynamic function (that is, indexed EOAs and gradients), whereas the results of studies based on the analysis of the IGA or labelled valve size would be more appropriately described in terms of patient‐prosthesis size but without using the term mismatch, thus avoiding much confusion. The use of adequate terminology has important clinical implications, since, as mentioned, patient‐prosthesis size has little relevance to adverse clinical outcomes, whereas PPM can be viewed as a major risk factor in this regard.6,7,8,10,11,13,14,15,19,26,27,28,29
IMPACT OF PPM ON CLINICAL OUTCOMES
Table 2 summarises the impact of PPM on clinical outcomes. As indicated above, the studies based on the indexed IGA are not really descriptive of PPM. They are given in the interest of completeness and to dispel any confusion but will not be described in any further detail.
Left ventricular mass and function and adverse cardiac events
In a study of 1103 patients with a porcine bioprosthetic valve, Del Rizzo et al29 found a strong and independent relation between the indexed EOA and the extent of left ventricular (LV) mass regression after AVR. In a smaller series, Tasca et al15 also found that the normalisation of LV mass is negatively and independently influenced by PPM. Whereas some authors have found that the persistence of PPM results in less regression of LV hypertrophy, others have reported that patients with PPM or small prostheses exhibited significant reductions in LV mass and, on this basis, concluded that PPM was not an important issue.12,22,31,32 Interestingly, Tasca et al27 found that patients with PPM nonetheless exhibit LV mass regression after operation but that the extent of such regression varies considerably from one patient to another and can be largely related to the extent of valve EOA increase after operation. These findings remind us of the following important pathophysiological concepts.27,33 Firstly, even in the presence of PPM, surgery normally results in improved haemodynamic function, the extent of which can be quite important. Secondly, a more optimal result can be expected if PPM is completely avoided. Thirdly, in analysing the results of AVR, it is important to remember that the relation between gradients and the indexed EOA is curvilinear and that the implications for a given patient will be directly related to his or her original and final positions on the indexed EOA‐gradient curve (fig 1).
Besides LV mass, other factors including the coronary flow reserve (CFR) may influence the normalisation of LV function and the regression of symptoms after AVR. It is well known that patients with aortic stenosis may experience angina pectoris and present ECG signs suggestive of myocardial ischaemia, even if they have angiographically normal epicardial coronary arteries.34,35 Angina is associated with a major increase in the risk of sudden death in these patients and is relieved immediately after AVR, whereas LV hypertrophy may regress over the next several months to years.34,36,37 The reduction in CFR is the key factor responsible for myocardial ischaemia in patients with aortic stenosis and this may contribute to the development of LV dysfunction, symptoms, and adverse outcomes.38 In a study of 20 patients with aortic stenosis with angiographically normal coronary arteries, Rajappan et al38 found that the severity of impairment of CFR measured by positron emission tomography was related to the severity of valve stenosis (valve EOA, gradient, LV systolic pressure) and diastolic perfusion time rather than to LV mass. In a subsequent study, the same team reported that changes in CFR after AVR were not directly related to regression of LV mass but rather were dependent on the change in valve EOA achieved with AVR.39 These findings suggest that PPM may compromise the postoperative normalisation of CFR and thereby have a detrimental impact on LV function after AVR.
Consistently, we have found that PPM is associated with a significant reduction in cardiac index during the postoperative course, the greatest deterioration being seen in patients with the most severe PPM.8 Moreover, the incidence of congestive heart failure was significantly higher in patients with PPM. These findings are consistent with those of Milano et al,11 who reported that patients with severe PPM had many more late cardiac events (most of them being congestive heart failure) after AVR (fig 4) and that PPM was an independent predictor of these late cardiac events. More recently, Ruel et al26 analysed the factors associated with persistent or recurrent heart failure in 1563 patients having undergone AVR and found that PPM defined as an indexed EOA ⩽ 0.80 cm2/m2 was an independent risk factor associated with a 60% increase in the risk of heart failure. Other recent studies also further confirmed that PPM is a strong risk factor for the occurrence of cardiac events after AVR (table 2).14,19
Figure 4 Freedom from late cardiac events in patients with non‐significant (indexed EOA (EOAi) >0.9 cm2/m2; squares), moderate (EOAi >0.6 cm2/m2 and ⩽0.9 cm2/m2; solid circles), or severe (EOAi ⩽0.6 cm2/m2; open circles) mismatch. Reproduced from Milano et al11 with permission of the Society of Thoracic Surgeons.
Early mortality
The impact of PPM on early mortality may be particularly important given that the left ventricle is more vulnerable during the early postoperative period and that it may thus be more sensitive to the increased haemodynamic burden imposed by PPM. In this regard, several studies reported that early mortality is significantly increased in patients with PPM (table 2).10,12,13,30 In a series of 2154 patients, Rao et al10 found that 30 day mortality was significantly higher (7.9% v 4.6%, p = 0.03) in patients with PPM. However, PPM was found to be an independent predictor only of late mortality, not of early mortality. We recently showed that PPM had a profound impact on early mortality in a series of 1265 consecutive patients undergoing AVR.13 In‐hospital mortality was 4.6% in this series and moderate PPM had a risk ratio of 2.1 (95% confidence interval (CI) 1.2 to 3.7), whereas severe PPM had a risk ratio of 11.4 (95% CI 4.4 to 29.5). Moreover, the adverse impact of PPM was much more pronounced in patients having an impaired LV ejection fraction (⩽ 40%). As fig 5 shows, the mortality risk was relatively low (mortality 2–5%) in the case of patients with preserved LV function who had non‐significant or moderate PPM. On the other hand, the mortality risk was dramatically increased (mortality 67%) in the patients having poor LV function and, concomitantly, severe PPM. In addition, even moderate PPM had a highly detrimental impact in the context of depressed LV function (mortality 16%). These findings are consistent with the concept that a failing ventricle is much more sensitive than a normal ventricle to an increase in afterload. In light of these results, avoidance of potential PPM should become a particularly mandatory consideration for patients with LV dysfunction. From the standpoint of pathophysiology, it would also make sense to consider that these high risk patients have a decreased ventricular reserve and are thus more vulnerable to the different degrees of PPM, particularly in the critical perioperative period.
Figure 5 Relative risk ratio for short term mortality according to valve prosthesis‐patient mismatch and preoperative left ventricular ejection fraction (LVEF). Numbers above the bars indicate the relative risk ratio for mortality compared with the group with non‐significant mismatch and normal LVEF. Reproduced from Blais et al13 with permission from the American Heart Association.
Late mortality
Several previous studies of relatively small numbers of patients found no negative impact of PPM on mid‐term (up to eight years) mortality.8,11,12 However, in a study of 2516 patients who underwent AVR with a stented bioprosthetic valve, Rao et al10 reported that freedom from valve related death at 12 years was significantly (p = 0.004) lower in patients with an indexed EOA ⩽ 0.75 cm2/m2 (75.5%) compared with those with a larger indexed EOA (84.2%). Moreover, the patient's age and indexed valve EOA were found to be independent predictors of valve related death. In a recent study by Tasca et al19 of 315 patients with pure aortic stenosis, five year overall survival and cardiac event‐free survival were significantly (p ⩽ 0.01) lower in patients with PPM (mean (SD) 82 (3)%, and 75 (4)%, respectively) than in those without PPM (93 (3)% and 87 (4%)). In another recent study by Mohty‐Echahidi et al14 of 388 patients with a 19 or 21 mm Standard St Jude valve, eight year survival (41 (8)%) of patients with severe PPM was significantly less than that of patients with moderate PPM (65 (5)%, p = 0.026) or no PPM (74 (5%), p = 0.002). On multivariate analysis after adjustment for other predictors of outcome, severe PPM was associated with higher mortality than was no PPM.
When analysed collectively, these previous studies suggest that the greatest impact of PPM on survival is in the early postoperative period when the left ventricle is most vulnerable. They also suggest that PPM has a significant impact on late mortality. A possible explanation for this late effect may be that patients with PPM undergoing long term valve degeneration or development of pannus have less EOA “reserve” and will therefore develop severe valve stenosis more rapidly than will the patients without PPM undergoing the same processes. As well, given that their LV is subjected to an increased haemodynamic burden at the outset, patients with PPM may not be able to tolerate as well or as long the additional workload associated with other concomitant diseases such as hypertension, coronary artery disease, and mitral insufficiency.
Bleeding complications
Vincentelli et al28 reported that abnormalities of von Willebrand factor and associated bleeding complications are common in patients with severe aortic stenosis. They also showed that von Willebrand abnormalities are directly related to the TPG and the stenosis induced shear stress. Interestingly, these abnormalities were generally improved by AVR in patients with no PPM but, on the other hand, they persisted in those with PPM. Further studies are now needed to determine whether the persistence of high shear stress caused by PPM is associated with an increased occurrence of bleeding complications after AVR.
PREVENTION OF MISMATCH
As opposed to most other risk factors associated with adverse clinical outcomes, PPM is modifiable and can be largely avoided by using a simple strategy at the time of operation.6,20 Our original description of this strategy was as follows. Step 1: calculate the patient's BSA from his or her weight and height. Step 2: multiply BSA by 0.85 cm2/m2, the result being the minimum EOA that the prosthesis to be implanted should have to avoid PPM. For example, if the patient's BSA is 1.80 m2, then 1.80 × 0.85 = 1.53 cm2 is the minimum EOA to avoid PPM. Step 3: choose the prosthesis in light of the result obtained in step 2 and the reference values for the different types and sizes of prosthesis (table 36,40,41,42). Hence, the EOA of the prosthesis to be implanted in the example chosen would have to be > 1.53 cm2 to completely avoid PPM and if, for example, the surgeon had intended to implant a Carpentier‐Edwards Perimount prosthesis, it would have had to be a size ⩾ 23. Fortunately, most prosthesis manufacturers have now made this exercise easier by providing charts that give the projected indexed EOAs for the patient's BSA and prosthesis sizes (fig 6). With regard to reference values for EOA and indexed EOA, three caveats are worth reiterating. Firstly, the values should be derived from in vivo rather than in vitro data, since in vitro data are usually too optimistic, particularly in the case of stentless valves.43 Secondly, values derived from geometric measurements (for example, internal diameters or geometric areas) are inadequate, since they do not predict postoperative TPGs (figs 2 and 3).20 Thirdly, when using these charts (or table 3), it is important to remember that there are often important discrepancies between the sizes of the different types of prostheses and the size for a given patient's annulus; the labelled size that fits may vary from one type of prosthesis to another.
Table 3 Normal reference values of EOAs* for prosthetic valves.
| Valve type | Prosthetic valve size (mm) | Reference | |||||
|---|---|---|---|---|---|---|---|
| 19 | 21 | 23 | 25 | 27 | 29 | ||
| Stented bioprosthetic valves | |||||||
| Medtronic Mosaic | 1.20 | 1.22 | 1.38 | 1.65 | 1.80 | 2.00 | 6 |
| Hancock II | NA | 1.18 | 1.33 | 1.46 | 1.55 | 1.60 | 6 |
| Carpentier‐Edwards Perimount | 1.10 | 1.30 | 1.50 | 1.80 | 1.80 | NA | 6 |
| Stentless bioprosthetic valves | |||||||
| Medtronic Freestyle | 1.15 | 1.35 | 1.48 | 2.00 | 2.32 | NA | 6 |
| St Jude Medical Toronto SPV | – | 1.30 | 1.50 | 1.70 | 2.00 | 2.50 | 6 |
| Prima Edwards | 0.80 | 1.10 | 1.50 | 1.80 | 2.30 | 2.80 | 6 |
| Mechanical valves | |||||||
| Medtronic‐Hall | 1.19 | 1.34 | NA | NA | NA | NA | 6 |
| St Jude Medical Standard | 1.04 | 1.38 | 1.52 | 2.08 | 2.65 | 3.23 | 6 |
| St Jude Medical Regent | 1.60 | 2.00 | 2.20 | 2.50 | 3.60 | 4.40 | 40 |
| MCRI On‐X | 1.50 | 1.70 | 2.00 | 2.40 | 3.20 | 3.20 | 41 |
| Carbomedics | 1.00 | 1.54 | 1.63 | 1.98 | 2.41 | 2.63 | 6 |
| Sorin Bicarbon | NA | 1.66 | 1.96 | NA | NA | NA | 42 |
*Expressed as mean values available in the literature.
Figure 6 Example of a chart provided by the manufacturers for the risk assessment of prosthesis‐patient mismatch. The chart gives the projected indexed effective orifice area (EOAi) for each level of patient's body surface area (BSA; left side) and size (top of chart) of a given model of prosthesis (hypothetical model in this example). Green cells indicate that the projected EOAi is > 0.85 cm2/m2, yellow cells indicate borderline values, and red cells indicate a risk of mismatch.
If PPM is projected with the type of prosthesis that was originally intended to be implanted, the following options can be considered:
Option 1: implant another type of prosthesis with a larger EOA such as a stentless bioprosthesis, a new generation mechanical prosthesis, or an aortic homograft (see table 3).
Option 2: enlarge the aortic root to accommodate a larger prosthesis of the same type.
Option 3: accept PPM in light of other clinical conditions.
The alternative options to avoid PPM (options 1 and 2) should of course be considered in light of the patient's clinical condition and overall risk to benefit ratio. If, for example, the projected indexed EOA is 0.75–0.80 cm2/m2 in a patient who is sedentary and has good LV function and it is evaluated that option 1 or 2 would significantly increase the operative risk, accepting this level of PPM may be the best option for this particular patient. If, on the other hand, severe PPM is projected or if the patient has poor LV function, the risks associated with PPM would be much higher and would probably outweigh any additional risks considered to be associated with options 1 and 2. Whatever the option chosen, we advocate performing this simple exercise in all cases, since it allows an enlightened clinical decision about the overall risks confronting the patient.
Castro et al44 used this prospective strategy for 657 consecutive patients whereby the aortic root was enlarged (option 2) whenever the indexed EOA was projected to be < 0.85 cm2/m2. As a result, the overall incidence of PPM in this series was only 2.5% instead of the 17% that would have been expected had the prospective strategy not been used. Moreover, operative mortality was not increased as a result of the root enlargement that was performed by a novel technique consisting of the insertion of a patch made of Hemasheild (Boston Scientific, Natick, Massachusetts, USA). These results show that the prospective strategy to avoid PPM is feasible and can be applied with success. More recent studies that have used an independent sizer to measure the actual diameter of the annulus also show that most stented bioprosthesis will result in an unacceptably high incidence of PPM in patients with an annulus < 22 mm (45% of the series) and are further evidence that the prospective strategy should be mandatory in all cases.16,17
IMPLICATIONS FOR THE INTERPRETATION OF HIGH POSTOPERATIVE GRADIENTS
A simple algorithm may be used to assess abnormally high gradients in the postoperative period (fig 745). The first step is to compare the EOA measured by Doppler echocardiography with the normal reference value of EOA (see table 3) for the type and size of prosthesis that has been implanted in the patient. If the measured EOA is much lower than the normal reference value and, moreover, has decreased over time during follow up, the diagnosis of intrinsic dysfunction should be raised. Causes of such dysfunction include the presence of leaflet calcific degeneration, thrombosis, pannus, or endocarditis. Occasionally, an abnormally high velocity jet corresponding to a localised gradient may be recorded by continuous wave Doppler interrogation through the smaller central slit‐like orifice of a bileaflet mechanical prosthesis. If, on the other hand, the EOA is comparable with its normal reference value, the next step is to calculate the indexed EOA. An indexed EOA < 0.85 cm2/m2 indicates the presence of PPM. And, lastly, it is also important to keep in mind that both phenomena—that is, PPM and intrinsic dysfunction—may coexist.
Figure 7 Algorithm used for evaluating abnormally high transvalvar pressure gradients after aortic valve replacement. LVOT, left ventricular outflow tract; PPM, prosthesis‐patient mismatch. Reproduced and modified from Dumesnil and Pibarot45 with permission from Marcel Dekker Inc.
PPM IN THE MITRAL POSITION
Rahimtoola and Murphy46 were the first to describe the case of a patient with PPM in the mitral position. In subsequent studies, Dumesnil et al3,4 then showed that the indexed EOA of mitral prostheses should ideally not be less than 1.2–1.3 cm2/m2 to avoid abnormally high residual TPGs. In a recent series, we found that PPM defined as an indexed EOA ⩽ 1.2 cm2/m2 is common (71%) after mitral valve replacement and that it is associated with persisting pulmonary hypertension.47 Indeed, the prevalence of pulmonary hypertension decreased from 69% before operation to 19% after operation in patients with no PPM, whereas it remained unchanged in those with PPM (69% before v 68% after replacement). Consistently, Masuda et al48 found that the maximum transprosthetic flow velocity is a strong determinant of the pulmonary capillary wedge pressure in children with mitral prostheses. In this series, the indexed mitral valve area did not correlate with pulmonary pressure; however, it should be pointed out that the valve area was derived from the IGA and not from the EOA. These results show that mitral PPM is not a rare occurrence and definitely warrants further documentation.
The prevention of PPM in the mitral position is a particularly demanding challenge because, as opposed to the aortic position, for the mitral position there is no alternative technique allowing implantation of a larger prosthesis. The preventive strategy should therefore be focused on the implantation of the prosthesis having the largest EOA for a given size. This observation also underlines the need for the development of better performing mitral prostheses and provides further motivation for repairing rather than replacing the valve whenever possible.
CONCLUSION
PPM is a common and modifiable risk factor leading to worse haemodynamic function, less regression of LV hypertrophy, more cardiac events, and lower survival. The projected indexed EOA should be systematically calculated at the time of the operation to estimate the risk of PPM and, if PPM is anticipated, alternative options should be considered in light of the patient's overall clinical condition and risk to benefit ratio.
ACKNOWLEDGEMENTS
This work was supported in part by Canadian Institutes of Health Research grants (MOP‐10929, MOP‐57745, and MOP‐67123). Dr Pibarot holds the Canada Research Chair in Valvular Heart Disease, Canadian Institutes of Health Research, Ottawa, Canada.
Abbreviations
AVR - aortic valve replacement
BSA - body surface area
CFR - coronary flow reserve
EOA - effective orifice area
IGA - internal geometric area
LV - left ventricular
PPM - prosthesis‐patient mismatch
TPG - transvalvar pressure gradient
Footnotes
Drs Pibarot and Dumesnil have received financial support from Medtronic, St Jude Medical, and Edwards Life Science to conduct some of the studies presented in this article. They are also consultants and are on the speakers' bureaus of St Jude Medical and Medtronic.
References
- 1.Rahimtoola S H. The problem of valve prosthesis‐patient mismatch. Circulation 19785820–24. [DOI] [PubMed] [Google Scholar]
- 2.Dumesnil J G, Honos G N, Lemieux M.et al Validation and applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiography. J Am Coll Cardiol 199016637–643. [DOI] [PubMed] [Google Scholar]
- 3.Dumesnil J G, Yoganathan A P. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg 19926S34–S38. [PubMed] [Google Scholar]
- 4.Dumesnil J G, Honos G N, Lemieux M.et al Validation and applications of mitral prosthetic valvular areas calculated by Doppler echocardiography. Am J Cardiol 1990651443–1448. [DOI] [PubMed] [Google Scholar]
- 5.Pibarot P, Dumesnil J G. Patient‐prosthesis mismatch and the predictive use of indexed effective orifice area: is it relevant? Card Surg Today 2003143–51. [Google Scholar]
- 6.Pibarot P, Dumesnil J G. Hemodynamic and clinical impact of prosthesis‐patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol 2000361131–1141. [DOI] [PubMed] [Google Scholar]
- 7.Pibarot P, Honos G N, Durand L G.et al The effect of patient‐prosthesis mismatch on aortic bioprosthetic valve hemodynamic performance and patient clinical status. Can J Cardiol 199612379–387. [PubMed] [Google Scholar]
- 8.Pibarot P, Dumesnil J G, Lemieux M.et al Impact of prosthesis‐patient mismatch on hemodynamic and symptomatic status, morbidity, and mortality after aortic valve replacement with a bioprosthetic heart valve. J Heart Valve Dis 19987211–218. [PubMed] [Google Scholar]
- 9.Yun K L, Jamieson W R E, Khonsari S.et al Prosthesis‐patient mismatch: hemodynamic comparison of stented and stentless aortic valves. Semin Thorac Cardiovasc Surg 19991198–102. [PubMed] [Google Scholar]
- 10.Rao V, Jamieson W R E, Ivanov J.et al Prosthesis‐patient mismatch affects survival following aortic valve replacement. Circulation 2000102III5–III9. [DOI] [PubMed] [Google Scholar]
- 11.Milano A D, De Carlo M, Mecozzi G.et al Clinical outcome in patients with 19‐mm and 21‐mm St. Jude aortic prostheses: comparison at long‐term follow‐up, Ann Thorac Surg 20027337–43. [DOI] [PubMed] [Google Scholar]
- 12.Hanayama N, Christakis G T, Mallidi H R.et al Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thorac Surg 2002731822–1829. [DOI] [PubMed] [Google Scholar]
- 13.Blais C, Dumesnil J G, Baillot R.et al Impact of prosthesis‐patient mismatch on short‐term mortality after aortic valve replacement. Circulation 2003108983–988. [DOI] [PubMed] [Google Scholar]
- 14.Mohty‐Echahidi D, Malouf J F, Girard S E.et al The impact of prosthesis‐patient mismatch on long‐term survival in patients with small St Jude Medical mechanical prostheses in the aortic position. Circulation 2006113420–426. [DOI] [PubMed] [Google Scholar]
- 15.Tasca G, Brunelli F, Cirillo M.et al Impact of valve prosthesis‐patient mismatch on left ventricular mass regression following aortic valve replacement. Ann Thorac Surg 200579505–510. [DOI] [PubMed] [Google Scholar]
- 16.Botzenhardt F, Eichinger W B, Guenzinger R.et al Hemodynamic performance and incidence of patient‐prosthesis mismatch of the complete supraannular perimount magna bioprosthesis in the aortic position. Thorac Cardiovasc Surg 200553226–230. [DOI] [PubMed] [Google Scholar]
- 17.Botzenhardt F, Eichinger W B, Bleiziffer S.et al Hemodynamic comparison of bioprostheses for complete supra‐annular position in patients with small aortic annulus. J Am Coll Cardiol 2005452054–2060. [DOI] [PubMed] [Google Scholar]
- 18.Muneretto C, Bisleri G, Negri A.et al The concept of patient‐prosthesis mismatch. J Heart Valve Dis 200413S59–S62. [PubMed] [Google Scholar]
- 19.Tasca G, Mhagna Z, Perotti S.et al Impact of prosthesis‐patient mismatch on cardiac events and mortality after aortic valve replacement in patients with pure aortic stenosis. Circulation 2006113570–576. [DOI] [PubMed] [Google Scholar]
- 20.Pibarot P, Dumesnil J G, Cartier P C.et al Patient‐prosthesis mismatch can be predicted at the time of operation. Ann Thorac Surg 200171S265–S268. [DOI] [PubMed] [Google Scholar]
- 21.Koch C G, Khandwala F, Estafanous F G.et al Impact of prosthesis‐patient size on functional recovery after aortic valve replacement. Circulation 20051113221–3229. [DOI] [PubMed] [Google Scholar]
- 22.Knez I, Rienmüller R, Maier R.et al Left ventricular architecture after valve replacement due to critical aortic stenosis: an approach to dis‐/qualify the myth of valve prosthesis‐patient mismatch? Eur J Cardiothorac Surg 200119797–805. [DOI] [PubMed] [Google Scholar]
- 23.Fernandez J, Chen C, Laub G W.et al Predictive value of prosthetic valve area index for early and late clinical results after valve replacement with the St Jude Medical valve prosthesis. Circulation 199694II109–II12. [PubMed] [Google Scholar]
- 24.Medalion B, Blackstone E H, Lytle B W.et al Aortic valve replacement: is valve size important? J Thorac Cardiovasc Surg 2000119963–974. [DOI] [PubMed] [Google Scholar]
- 25.Blackstone E H, Cosgrove D M, Jamieson W R.et al Prosthesis size and long‐term survival after aortic valve replacement. J Thorac Cardiovasc Surg 2003126783–793. [DOI] [PubMed] [Google Scholar]
- 26.Ruel M, Rubens F D, Masters R G.et al Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves. J Thorac Cardiovasc Surg 2004127149–159. [DOI] [PubMed] [Google Scholar]
- 27.Tasca G, Brunelli F, Cirillo M.et al Impact of the improvement of valve area achieved with aortic valve replacement on the regression of left ventricular hypertrophy in patients with pure aortic stenosis. Ann Thorac Surg 2005791291–1296. [DOI] [PubMed] [Google Scholar]
- 28.Vincentelli A, Susen S, Le Tourneau T.et al Acquired von Willebrand syndrome in aortic stenosis. N Engl J Med 2003349343–349. [DOI] [PubMed] [Google Scholar]
- 29.Del Rizzo D F, Abdoh A, Cartier P.et al Factors affecting left ventricular mass regression after aortic valve replacement with stentless valves. Semin Thorac Cardiovasc Surg 199911114–120. [PubMed] [Google Scholar]
- 30.Connolly H M, Oh J K, Schaff H V.et al Severe aortic stenosis with low transvalvular gradient and severe left ventricular dysfunction: result of aortic valve replacement in 52 patients. Circulation 20001011940–1946. [DOI] [PubMed] [Google Scholar]
- 31.Freed D H, Tam J W, Moon M C.et al Nineteen‐millimeter prosthetic aortic valves allow normalization of left ventricular mass in elderly women. Ann Thorac Surg 2002742022–2025. [DOI] [PubMed] [Google Scholar]
- 32.Tasca G, Brunelli F, Cirillo M.et al Mass regression in aortic stenosis after valve replacement with small size pericardial bioprosthesis. Ann Thorac Surg 2003761107–1113. [DOI] [PubMed] [Google Scholar]
- 33.Dumesnil J G. Invited commentary. Ann Thorac Surg 2005791296 [Google Scholar]
- 34.Gould K L, Carabello B A. Why angina in aortic stenosis with normal coronary arteriograms? Circulation 20031073121–3123. [DOI] [PubMed] [Google Scholar]
- 35.Marcus M L, Doty D B, Hiratzka L F.et al Decreased coronary reserve: a mechanism for angina pectoris in patients with aortic stenosis and normal coronary arteries. N Engl J Med 19823071362–1366. [DOI] [PubMed] [Google Scholar]
- 36.Lester S J, Heilbron B, Gin K.et al The natural history and rate of progression of aortic stenosis. Chest 19981131109–1114. [DOI] [PubMed] [Google Scholar]
- 37.Monrad E S, Hess O M, Murakami T.et al Time course of regression of left ventricular hypertrophy after aortic valve replacement. Circulation 1988771345–1355. [DOI] [PubMed] [Google Scholar]
- 38.Rajappan K, Rimoldi O, Camici P G.et al Factors influencing coronary microcirculatory function in patients with aortic stenosis after aortic valve replacement. Circulation 2002106II640 [Google Scholar]
- 39.Rajappan K, Rimoldi O E, Camici P G.et al Functional changes in coronary microcirculation after valve replacement in patients with aortic stenosis. Circulation 20031073170–3175. [DOI] [PubMed] [Google Scholar]
- 40.Bach D S, Sakwa M P, Goldbach M.et al Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve. Ann Thorac Surg 2002742003–2009. [DOI] [PubMed] [Google Scholar]
- 41.Chambers J, Ely J L. Early postoperative echocardiographic hemodynamic performance of the On‐X prosthetic heart valve: a multicenter study. J Heart Valve Dis 19987569–573. [PubMed] [Google Scholar]
- 42.De Carlo M, Milano A D, Nardi C.et al Serial Doppler echocardiographic evaluation of small‐sized sorin bicarbon prostheses. J Thorac Cardiovasc Surg 2003126337–343. [DOI] [PubMed] [Google Scholar]
- 43.Dumesnil J G, LeBlanc M H, Cartier P.et al Distinctive hemodynamic features of the Freestyle aortic bioprosthesis as compared to stented bioprosthesis. Ann Thorac Surg 199866S130–S133. [DOI] [PubMed] [Google Scholar]
- 44.Castro L J, Arcidi J M J, Fisher A L.et al Routine enlargement of the small aortic root: a preventive strategy to minimize mismatch. Ann Thorac Surg 20027431–36. [DOI] [PubMed] [Google Scholar]
- 45.Dumesnil J G, Pibarot P. Perioperative evaluation of aortic valve surgery by Doppler‐echocardiography. In: Denault A, Couture P, eds. Transesophageal echocardiography multimedia manual: a perioperative transdisciplinary approach. New York: Marcel Dekker, 2005361–379.
- 46.Rahimtoola S H, Murphy E. Valve prosthesis‐patient mismatch: a long‐term sequela. Br Heart J 198145331–335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Li M, Dumesnil J G, Mathieu P.et al Impact of valve prosthesis‐patient mismatch on pulmonary arterial pressure after mitral valve replacement. J Am Coll Cardiol 2005451034–1040. [DOI] [PubMed] [Google Scholar]
- 48.Masuda M, Kado H, Tatewaki H.et al Late results after mitral valve replacement with bileaflet mechanical prosthesis in children: evaluation of prosthesis‐patient mismatch. Ann Thorac Surg 200477913–917. [DOI] [PubMed] [Google Scholar]