Skip to main content
Applied Microbiology logoLink to Applied Microbiology
. 1974 Jul;28(1):95–99. doi: 10.1128/am.28.1.95-99.1974

Evaluation of Miniature Test for Bacteriuria Using Dehydrated Media and Nitrite Pads

Craig M Moffat 1, Michael R Britt 1, John P Burke 1
PMCID: PMC186601  PMID: 4844272

Abstract

A new dip-inoculum method for detecting bacteriuria which utilizes dehydrated media pads and a nitrite pad attached to a small plastic strip was evaluated in hospitalized patients. Discrepant interpretations were made by independent observers in 9.3% of the specimens with > 105 colonies per ml. The media pads failed to support growth of yeast and gave variable results with Staphylococcus epidermidis and non-group D streptococci. False-negative culture results commonly occurred if the patients were receiving antibiotics. The nitrite test occasionally remained positive for brief periods after the elimination of bacteriuria by antibiotics. Conditions and drugs (especially phenazopyridine) which discolor urine interfered with reading both the culture and nitrite tests. Although not suitable for hospital use, or for monitoring therapy, the test strip is probably as reliable as the calibrated loop-streak plate culture for office screening.

Full text

PDF
95

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Birch D. F. Dipslide screening of hospitalized patients for bacteriuria. Med J Aust. 1971 May 29;1(22):1176–1178. doi: 10.5694/j.1326-5377.1971.tb88104.x. [DOI] [PubMed] [Google Scholar]
  2. Cohen S. N., Kass E. H. A simple method for quantitative urine culture. N Engl J Med. 1967 Jul 27;277(4):176–180. doi: 10.1056/NEJM196707272770403. [DOI] [PubMed] [Google Scholar]
  3. Craig W. A., Kunin C. M., DeGroot J. Evaluation of new urinary tract infection screening devices. Appl Microbiol. 1973 Aug;26(2):196–201. doi: 10.1128/am.26.2.196-201.1973. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Craig W. A., Kunin C. M. Quantitative urine culture method using a plastic "paddle" containing dual media. Appl Microbiol. 1972 May;23(5):919–922. doi: 10.1128/am.23.5.919-922.1972. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Harper W. E. The value of serial colony counts and the use of dipslides as an aid in screening specimens from catheterized patients. Med J Aust. 1972 Aug 12;2(7):369–371. doi: 10.5694/j.1326-5377.1972.tb47353.x. [DOI] [PubMed] [Google Scholar]
  6. Kunin C. M., Bergeron J. A. A simple quantitative urine culture method using an internally coated plastic pipette. Am J Clin Pathol. 1972 Oct;58(4):371–375. doi: 10.1093/ajcp/58.5.371. [DOI] [PubMed] [Google Scholar]
  7. McAllister T. A., Arneil G. C., Barr W., Kay P. Assessment of plane dipslide quantitation of bacteriuria. Nephron. 1973;11(2):111–122. doi: 10.1159/000180224. [DOI] [PubMed] [Google Scholar]
  8. McAllister T. A. The day of the dipslide. Nephron. 1973;11(2):123–133. doi: 10.1159/000180225. [DOI] [PubMed] [Google Scholar]
  9. Norden C. W., Kass E. H. Bacteriuria of pregnancy--a critical appraisal. Annu Rev Med. 1968;19:431–470. doi: 10.1146/annurev.me.19.020168.002243. [DOI] [PubMed] [Google Scholar]

Articles from Applied Microbiology are provided here courtesy of American Society for Microbiology (ASM)