Abstract
The purpose of this study was to determine the optimum membrane filter structure and characteristics for recovery of coliform organisms. Additionally, other factors such as sterilization method and membrane composition were examined. Fecal coliform growth tests with varied samples indicated that the most critical factor in recovery was surface pore morphology and not other factors previously suspected. Fecal coliform counts showed a dramatic increase, with increasing surface opening sizes. Membrane structures with surface openings large enough to surround the entrapped bacteria are required for optimum growth of fecal coliform organisms. Maximum fecal coliform recoveries are obtained using membranes composed of mixed esters of cellulose exhibiting a surface opening diameter of 2.4 μm and a retention pore size of 0.7 μm.
Full text
PDF






Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Dutka B. J., Jackson M. J., Bell J. B. Comparison of autoclave and ethylene oxide-sterilized membrane filters used in water quality studies. Appl Microbiol. 1974 Sep;28(3):474–480. doi: 10.1128/am.28.3.474-480.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]