Abstract
Thrombotic malfunction of a Björk-Shiley aortic valve prosthesis occurred in three patients 6 to 16 months postoperatively. None of the patients had been taking anticoagulants. Although the presentation was acute, prodromal symptoms could be identified retrospectively in two of the patients. Two patients survived thrombectomy. Postoperative anticoagulant therapy is recommended in patients with these prostheses despite factors that may make such therapy riskier in specific patients. Attention to the character of murmurs and of the closure sound of the prosthetic valve must be part of the routine follow-up. In the emergency situation, when delay must be avoided, catheterization and angiography are unnecessary. The operative approach consists of complete thrombectomy without replacement of the valve or any of its components unless there is obvious periprosthetic leak or prosthetic wear.
Full text
PDF


Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Allard J. R., Fraser G., Dobell A. R. Sudden hemolysis indicating prosthetic valve dysfunction. Can J Surg. 1976 May;19(3):272–273. [PubMed] [Google Scholar]
- Ben-Zvi J., Hildner F. J., Chandraratna P. A., Samet P. Thrombosis on Bjork-Shiley aortic valve prosthesis. Clinical, arteriographic, echocardiographic and therapeutic observations in seven cases. Am J Cardiol. 1974 Oct 3;34(5):538–544. doi: 10.1016/0002-9149(74)90124-6. [DOI] [PubMed] [Google Scholar]
- Björk V. O. A new central-flow tilting disc valve prosthesis. One year's clinical experience with 103 patients. J Thorac Cardiovasc Surg. 1970 Sep;60(3):355–374. [PubMed] [Google Scholar]
- Björk V. O., Henze A. Encapsulation of the Björk-Shiley aortic disc valve prosthesis caused by the lack of anticoagulant treatment. Scand J Thorac Cardiovasc Surg. 1973;7(1):17–20. doi: 10.3109/14017437309139160. [DOI] [PubMed] [Google Scholar]
- Björk V. O., Holmgren A., Olin C., Ovenfors C. O. Clinical and haemodynamic results of aortic valve replacement with the Björk-Shiley tilting disc valve prosthesis. Scand J Thorac Cardiovasc Surg. 1971;5(3):177–191. [PubMed] [Google Scholar]
- Björk V. O. The pyrolytic carbon occluder for the Björk-Shiley tilting disc valve prosthesis. Scand J Thorac Cardiovasc Surg. 1972;6(2):109–113. doi: 10.3109/14017437209134787. [DOI] [PubMed] [Google Scholar]
- Bozer A. Y., Karamehmetoglu A. Thrombosis encountered with Björk-Shiley prosthesis. J Cardiovasc Surg (Torino) 1972 Mar-Apr;13(2):141–143. [PubMed] [Google Scholar]
- Byrd C. L., Yahr W. Z., Greenberg J. J. Long-term results of "simple" thrombectomy for thrombosed Björk-Shiley aortic valve prostheses. Ann Thorac Surg. 1975 Sep;20(3):265–273. doi: 10.1016/s0003-4975(10)64218-x. [DOI] [PubMed] [Google Scholar]
- Cokkinos D. V., Voridis E., Bakoulas G., Theodossiou A., Skalkeas G. D. Thrombosis of two high-flow prosthetic valves. J Thorac Cardiovasc Surg. 1971 Dec;62(6):947–949. [PubMed] [Google Scholar]
- Fernandez J., Morse D., Maranhao V., Gooch A. S. Results of use of the pyrolytic carbon tilting disc Bjork-Shiley aortic prosthesis. Chest. 1974 Jun;65(6):640–645. doi: 10.1378/chest.65.6.640. [DOI] [PubMed] [Google Scholar]
- Gray L. A., Jr, Fulton R. L., Srivastava T. N., Flowers N. C. Surgical treatment of thrombosed Björk-Shiley aortic valve prosthesis. J Thorac Cardiovasc Surg. 1976 Mar;71(3):429–432. [PubMed] [Google Scholar]
- Messmer B. J., Hallman G. L., Liotta D., Martin C., Cooley D. A. Aortic valve replacement: new techniques, hydrodynamics, and clinical results. Surgery. 1970 Dec;68(6):1026–1037. [PubMed] [Google Scholar]