Abstract
Aims
To evaluate the speed of onset of bronchodilation following salbutamol administered via a metered-dose inhaler with a spacer (pMDI + Volumatic) and a dry-powder inhaler (Diskus), as well as the relative potencies of these devices in asthmatic patients with methacholine-induced bronchoconstriction.
Methods
Eighteen patients inhaled methacholine (MCh) until FEV1 decreased by 35% of control. Following administration of placebo, 200 µg salbutamol or 400 µg salbutamol through the pMDI + Volumatic or the Diskus, we calculated the time elapsed from drug administration and the appearance of a 90% increase in post-MCh forced vital capacity (FVC), FEV1 and volume-adjusted mid-expiratory flow (recovery times). The salbutamol doses to be delivered by the two inhalation devices to achieve similar recovery times and the relative potencies of the devices were calculated by using the 2-by-2 Finney parallel regression method.
Results
For all functional variables, recovery times were significantly (P < 0.01) shorter in pMDI + Volumatic than Diskus trials. The salbutamol doses to be delivered by the Diskus to achieve recovery times for FVC, FEV1 and volume-adjusted mid-expiratory flow similar to those obtained with 200 µg salbutamol administered via the pMDI + Volumatic were 558 (95% CI 537, 579) µg, 395 (95% CI 388, 404) µg and 404 (95% CI 393, 415) µg, respectively, and corresponded to relative potencies of 2.79 (95% CI 2.68, 2.90), 1.98 (95% CI 1.94, 2.02), and 2.02 (95% CI 1.96, 2.07).
Conclusions
Administration of salbutamol via the pMDI + Volumatic provides faster reversal of induced bronchoconstriction than via the Diskus. The salbutamol dose targeting the lungs with the pMDI + Volumatic is approximately twice that with the Diskus.
Keywords: asthma, dry-powder inhaler, functional antagonism, spacers, speed of bronchodilation
Introduction
Short-acting β2-adrenoceptor bronchodilators are widely used for the treatment of acute asthma attacks [1]. Compared with oral administration, inhaled bronchodilators provide a more rapid onset of action, greater airway response and reduced risk of systemic side-effects [1]. The pressurized metered-dose inhalers (pMDIs) and dry-powder inhalers (DPIs) are commonly used for administering β2-adrenergic bronchodilators, such as salbutamol. Studies performed in asthmatic patients with spontaneous [2] or induced bronchoconstriction [3] have shown that salbutamol administration via a DPI or a pMDI in conjunction with a spacer provides faster bronchodilation than the pMDI alone. Whether a difference exists in the speed of the bronchodilator response to bronchodilator agents administered via either a DPI or a pMDI with a spacer in asthmatic patients with natural or induced bronchoconstriction remains to be established. Rapid onset of action is an important feature of bronchodilator drugs in patients with asthma, as prompt relief from bronchoconstriction will give reassurance of effect and may be a key factor in patient compliance [1]. Furthermore, because rapidity of action of inhaled bronchodilators depends on the amount of drug delivered to the appropriate receptors in the airways [4], it is important that clinicians know whether different inhalation devices, such as the DPI and the pMDI with a spacer, are bioequivalent, i.e. if they are able to deliver essentially the same drug amount to the relevant receptor.
One of the most commonly used methods for the in vivo assessment of bioequivalence between different formulations (e.g. powdered or aerosolized) of the same bronchodilator drug is based on the measurement of the degree of bronchodilation in patients with spontaneous bronchospasm. With this method, however, the results are often difficult to interpret, especially in patients with different degrees of baseline airway obstruction. To overcome this problem, some authors have evaluated the magnitude of bronchodilation in patients with mild degrees of induced airway narrowing [3, 5], usually corresponding to a 20% decrease in baseline FEV1. In such experimental conditions, however, even low doses of bronchodilators are likely to result in near maximal bronchodilation [6–9], so that differences in the magnitude of responses to different agents or to varied formulations of the same agent may be difficult to detect.
Evidence from in vitro [10] and in vivo [11] studies in animals, and from dose–response studies in humans [12–14], indicate that the dose of β2-adrenoceptor agonists required to overcome induced bronchoconstriction is related not only to the degree of airway narrowing, but also to the type of bronchoconstrictor used (e.g. methacholine or histamine) and the dose required to achieve the desired level of bronchoconstriction [15–17]. With methacholine (MCh), the larger the dose required to obtain a given level of airway narrowing, the larger the dose of β2-adrenoceptor agonist needed to reverse it: a phenomenon called functional antagonism [17]. With this in mind, we have hypothesized that a bioassay based on a strong functional antagonism between MCh and salbutamol might be particularly suitable to evaluate differences in the amount of salbutamol that reaches β-adrenergic receptors in the airways when the drug is administered via a pMDI with a spacer or via a DPI. Therefore, we have used a clinical bioassay based on a moderate [18] degree of MCh-induced bronchoconstriction to compare the speed of the bronchodilator response following salbutamol inhaled via either a DPI (the Diskus) or a pMDI in conjunction with a large-volume spacer (pMDI + Volumatic) in asthmatic patients. In addition, we have calculated the relative potency of the Diskus and the pMDI + Volumatic, i.e. the ratio of the salbutamol doses administered by the two inhalation devices producing an equal response. An attempt has also been made to ascertain which pulmonary function variable is most sensitive to detect differences in the speed of bronchodilation following salbutamol administered via the two inhalation devices. The present investigation is a posthoc analysis performed on data collected in a previous study that investigated the magnitude of the bronchodilator response following salbutamol inhaled via these different inhalation devices in asthmatic patients [19].
Methods
Patients
We studied 18 nonsmoking outpatients (seven females, mean age 29 years, range 18–35 years) with intermittent asthma [1]; their baseline FEV1 value corresponded (mean 95% confidence interval, CI) to 100.4% (93.4, 107.4) of predicted [20]. None of the patients had suffered from recent (<4 weeks) respiratory tract infections and all were steroid naïve and were not receiving regular medications for their asthma. All patients had first been tested for bronchial hyper-responsiveness, and the mean (95% CI) value of provocative MCh concentration causing a 20% fall in FEV1 was 1.64 mg ml−1 (1.27, 2.01). This study was approved by the local ethics committee and individual informed consent was obtained.
Methacholine-induced bronchoconstriction
Patients tidally inhaled for 2 min doubling MCh concentrations (acetyl-β-methylcholine chloride, Lofarma, Milano, Italy) ranging from 0.06 to 16 mg ml−1 through a DeVilbiss no. 646 nebulizer (DeVilbiss Co, Somerset, PA, USA) driven by a constant airflow (8 l min−1) and preceded by 0.9% saline as a control solution. The bronchial response was evaluated by measuring the FEV1 90–120 s after administration of either saline or each MCh concentration by means of a water-sealed spirometer (Collins Survey II, Braintree, MA, USA) connected to an X-Y recorder to obtain volume-time tracings of the forced vital capacity (FVC).
Protocol
This was a double-blind, double-dummy, six-way crossover study, randomized for different treatments and inhalation devices. On each study day, MCh inhalation was preceded by baseline assessments of FVC, FEV1 and mean forced expiratory flow rate measured between 25 and 75% of the FVC (FEF25–75). The latter variable is considered an expression of small airway patency [21]. When FEV1 had fallen by ≈35% of the postsaline (control) value following MCh inhalation, patients inhaled placebo or 200 or 400 µg of salbutamol, either through the Diskus DPI (Diskus 200 µg salbutamol per actuation, GlaxoSmithKline Inc., Ware, Hertfordshire, UK) or the pMDI (HFA134a salbutamol sulphate, 100 µg salbutamol per actuation, Glaxo Operations, Speke, UK) in conjunction with the Volumatic spacer device (Volumatic, Allen & Hanbury’s, Uxbridge, Middlesex, UK), while keeping the inspiratory flow rate as close as possible to 30 l min−1[19]. Values of FVC, FEV1 and FEF25–75 were determined at 3-min intervals for 15 min, with further assessments performed 30, 45 and 60 min after drug administration. Additional details regarding the techniques for salbutamol and placebo administration, as well as the blinding procedures, have already been described [19].
Data analysis
Power calculations indicated that 12 patients were required to provide a greater than 80% chance of detecting a 20% difference between treatments in the speed of onset of bronchodilation (as assessed by FEV1 changes) at the 5% level of significance [3, 19]. The largest of three reproducible (±5% coefficient of variation) FVC, FEV1 and FEF25–75 measurements, the latter obtained from the single best test curve, was used for calculations [18]. Because of the dependence on changes in lung volumes, values of FEF25–75 recorded following placebo and salbutamol administration were corrected for baseline FVC and expressed as isoFEF25–75[3, 22]. As all placebo trials attained overlapping results, the latter were averaged and considered as a whole in all subsequent analyses. Functional changes over time induced by placebo and salbutamol administration were evaluated by linear and nonlinear regression analysis, respectively [19].
Scrutiny of results revealed that, in some trials, the decay in lung function provoked by MCh inhalation could not be fully reversed by salbutamol within the 60-min observation period. For instance, with 200 µg salbutamol the increases in FVC, FEV1 and isoFEF25–75 values turned out to be only slightly greater than 90% of the corresponding control values (Figure 1). Therefore, recovery times were always calculated as the time elapsed from salbutamol administration and the appearance of a 90% increase [23] in post-Mch FVC, FEV1 and isoFEF25–75 values. Recovery times were subsequently compared by repeated measures analysis of variance (anova) followed by the Bonferroni’s test for multiple comparisons. We also noted that placebo administration caused relatively small increases in post-MCh lung function variables; as these increases consistently failed to reach the 90% level (Figure 1), recovery times were not calculated in placebo trials. Conceivably, salbutamol-induced increases in lung function values over time result from summation of the pharmacological effect by salbutamol plus a modest bronchodilator effect by placebo. Therefore, to calculate the net bronchodilator response to salbutamol, the post-MCh FVC, FEV1 and isoFEF25–75 increases induced by placebo administration were subtracted, at each considered time interval, from the corresponding increases observed following salbutamol administration. Recovery times for placebo-corrected changes in lung function variables over time were subsequently calculated with the same procedure as that described above. To evaluate differences in recovery times between FVC, FEV1 and isoFEF25–75, the recovery times calculated in each patient were pooled, and compared by anova followed by the Bonferroni’s test for multiple comparisons. Comparisons between dose–response curves obtained in pMDI + Volumatic and Diskus trials were performed by using the Finney 2-by-2 parallel regression bioassay analysis [24]. In essence, after having demonstrated a significant dose–response relationship and the absence of deviation from parallelism [24], this statistical method constructs a two-point dose–response curve for each of the two inhalation devices studied. Based on these curves, we calculated, for each considered pulmonary function variable, the salbutamol doses to be delivered by the two inhalation devices to achieve equal recovery times, as well as the ratio between these two doses, i.e. the relative potency [24]. In addition, to evaluate which pulmonary function variable was most sensitive in detecting differences in recovery times observed in trials performed with the two different inhalation devices, we calculated the ratio of the root mean squared error (s) to the slope (b) of the previously determined dose–response curves [25, 26]. The lower the s : b ratio below 1, the greater the sensitivity [25, 26]. Because in the present study all slope (b) values of the dose–response curves were negative (i.e. decreases in recovery times with increasing salbutamol doses), the calculated s : b values were more conveniently expressed as absolute values. Comparisons between s : b values for FVC, FEV1 and isoFEF25–75 were performed by repeated measures anova followed by the Bonferroni’s test. Data analyses and representation were performed using statistical and graphic software (Graphpad Prism release 3.01; Graphpad Software, San Diego, CA, USA), and sample size and power calculations were performed using dedicated software (PS Power and Sample Size Calculations, release 2.1.30, Vanderbilt University Medical Center, Nashville, TN, USA). In all instances, a P value less than 0.05 was taken as significant.
Figure 1.
Time course of mean (18 patients) percentage changes in postmethacholine (MCh) forced vital capacity (FVC), forced expiratory volume in 1 s (FEV1) and volume-adjusted forced expiratory flow rate between 25 and 75% of the FVC (isoFEF25–75) following administration of placebo (mean values of two different preparations, X symbols), 200 (open circles) and 400 (filled circles) µg of salbutamol through the pressurized metered dose inhaler with the Volumatic spacer (solid lines) and the Diskus inhaler (dotted lines). In each panel, the thick dotted line marks the level of a 90% increase in postmethacholine value used to calculate the recovery times (see Methods). C, control (premethacholine) conditions
Results
In all trials, MCh inhalation induced comparable FEV1 decreases up to a level that closely approached −35% of the corresponding postsaline value (mean maximum FEV1 fall 35.33%, 95% CI 34.90, 35.77). The mean (95% CI) MCh concentration required to produce a 35% fall in post-saline FEV1 was 4.83 mg ml−1 (3.89, 5.77). Reductions in FEV1 were always accompanied by decreases in FVC (−20.27%, 95% CI −18.37, −22.17) and isoFEF25–75 (−55.33%, 95% CI −52.2, −56.05). Methacholine-induced bronchoconstriction was well tolerated by all patients, and no adverse effects were observed.
Salbutamol administration via either inhalation device caused a rapid rise (within 3–6 min) in lung function variables, followed by further, slower increases up to a plateau level that was achieved within 30 min for FVC, and FEV1, and within 45 min for isoFEF25–75 (Figure 1). Individual regression coefficients for postsalbutamol changes in lung function variables over time ranged from 0.90 to 0.99 for FVC, from 0.88 to 0.99 for FEV1, and from 0.96 to 0.99 for isoFEF25–75 (P < 0.01). Following placebo, all variables displayed a slower, linear increase that consistently failed to achieve the corresponding control value (Figure 1).
For all lung function variables, mean recovery times were dose-dependently shortened by salbutamol inhalation (Figure 2); independently of the salbutamol dose, mean recovery times for each considered variable were significantly (P < 0.01) shorter in pMDI + Volumatic than in Diskus trials (Table 1). As shown in Figure 3, recovery times for FVC were significantly shorter than those for FEV1 and isoFEF25–75 (P < 0.01).
Figure 2.
Parallel regression analysis for forced vital capacity (FVC), forced expiratory volume in 1 s (FEV1) and volume-adjusted forced expiratory flow rate between 25 and 75% of the FVC (isoFEF25–75) recovery times observed in pMDI + Volumatic (filled circles) and Diskus (empty circles) trials. The arrowed dotted lines indicate the calculated salbutamol dose to be delivered by the Diskus to achieve recovery times equal to those observed following 200 µg salbutamol administered via the pMDI + Volumatic
Table 1.
Mean (95% confidence interval) recovery times (in min) for FVC, FEV1 and isoFEF25–75 following administration of 200 and 400 µg salbutamol via the pressurized metered-dose inhaler with Volumatic spacer (pMDI + Volumatic) and the Diskus dry-powder inhaler. The corresponding placebo-corrected (PL-corrected) recovery times are also indicated. See Methods for further details
| FVC | FEV1 | isoFEF25-75 | |
|---|---|---|---|
| 200 µg salbutamol via pMDI + Volumatic | 6.58 (6.14, 7.42)* | 19.06 (18.20, 19.52)* | 12.50 (12.02, 13.38)* |
| PL corrected | 8.52 (8.12, 9.36)* | 32.10 (30.48, 33.36)* | 14.20 (13.50, 14.50)* |
| 400 µg salbutamol via pMDI + Volumatic | 2.24 (2.12, 2.45)*§ | 4.42 (4.10, 5.14)*§ | 2.46 (2.27, 3.05)*§ |
| PL corrected | 4.20 (4.07, 4.33)*§ | 7.40 (7.1, 8.10)*§ | 3.05 (2.45, 3.20)*§ |
| 200 µg salbutamol via Diskus | 18.43 (17.40, 19.45) | 35.27 (33.19, 37.35) | 29.24 (27.42, 31.06) |
| PL corrected | 29.50 (27.47, 31.53) | 42.15 (40.07, 44.23) | 44.38 (42.08, 47.00) |
| 400 µg salbutamol via Diskus | 12.38 (10.56, 14.18)§ | 18.42 (17.37, 19.46)§ | 13.07 (11.32, 14.42)§ |
| PL corrected | 16.50 (14.38, 19.12)§ | 30.15 (-29.10, 31.20)§ | 16.12 (14.02, 18.22)§ |
FVC, forced vital capacity; FEV1, forced expiratory volume in 1 s; isoFEF25–75, volume-adjusted forced expiratory flow rate between 25 and 75% of the FVC. Within each column
P < 0.01 comparing the same salbutamol dose administered via different inhalation devices
P < 0.01 comparing different salbutamol doses administered via the same device.
Figure 3.
Comparison of recovery times for forced vital capacity (FVC), forced expiratory volume in 1 s (FEV1) and volume-adjusted forced expiratory flow rate between 25 and 75% of the FVC (isoFEF25–75) performed on pooled data. Each bar represents the mean (95% confidence interval) FVC, FEV1, and isoFEF25–75 values (n = 18) obtained by pooling the corresponding values recorded during all salbutamol trials. *P < 0.01
The demonstration of dose–response relationships and the absence of a significant deviation from parallelism allowed us to evaluate the relative potencies between the two inhalation devices [24]. Assessment of relative potencies revealed that, in Diskus trials, twice the salbutamol dose was required to obtain, for each considered variable, recovery times equal to those achieved in pMDI + Volumatic trials (Table 2).
Table 2.
Bioequivalence between salbutamol administered via the pMDI+Volumatic and via the Diskus
| Variable | Salbutamol dose (µg) via Diskus equivalent to 200 µg via pMDI + Volumatic | Relative potency (pMDI + Volumatic vs Diskus) |
|---|---|---|
| FVC | 558 (537, 579) | 2.79 (2.68, 2.90) |
| FEV1 | 395 (388, 404) | 1.98 (1.94, 2.02) |
| IsoFEF25–75 | 404 (393, 415) | 2.02 (1.96, 2.07) |
Values are means ± 95% confidence interval. See Table 1 for abbreviations.
The s : b ratios for FVC, FEV1 and isoFEF25-75 were 0.11 (95% CI 0.09, 0.12), 0.05 (95% CI 0.04, 0.06) and 0.06 (95% CI 0.05, 0.07), respectively. Analysis of variance showed that s : b ratios for FEV1 and isoFEF25–75 were similar and significantly (P < 0.05) lower than that for FVC.
Discussion
The results show that, in patients with intermittent asthma, administration of salbutamol via the pMDI + Volumatic causes a more rapid reversal of MCh-induced bronchocostriction than via the Diskus. The results also indicate that it would require approximately twice the dose of salbutamol administered via the Diskus to obtain the same recovery times as those recorded when salbutamol is administered via the pMDI + Volumatic.
A demonstration of clinical bioequivalence and assessment of the relative potency of inhaled medications are required before the marketing of new types of delivery systems [6, 7]. Bioequivalence ensures that equal doses of two different formulations (e.g. powdered or aerosolized) of the same drug produce equivalent pharmacodynamic effects [6, 7]. The principal in vivo methods currently available for the assessment of bioequivalence between different formulations of the same drug are plasma pharmacokinetic studies, radioaerosol drug deposition studies, and pharmacodynamic clinical efficacy studies [6, 7]. Pharmacokinetic studies are of limited value for inhaled bronchodilators because the doses administered are small and the resulting serum concentrations are often too low to assay accurately; in addition, they may not correlate with the dose delivered to the lungs [6, 7]. Indeed, regulatory authorities do not consider pharmacokinetic studies reliable reflections of the amount of bronchodilator drug delivered to the lungs [27]. Radioaerosol studies provide useful information about lung deposition, but their ability to predict the clinical efficacy of β2-adrenoceptor agonists is controversial [7]. Pharmacodynamic clinical efficacy studies are considered the most useful tool for assessing both the effectiveness and bioequivalence of inhaled bronchodilators [6, 7]. Generally, such studies rely on the measurement of clinically relevant pharmacodynamic responses that are thought to reflect the amount of drug delivered to the lungs by the tested preparation and the reference preparation [6, 7]. Many of these studies have assessed bioequivalence between different formulations of β2-adrenoceptor agonists by evaluating their ability to prevent induced bronchoconstriction [13, 14, 26]. A potential disadvantage of this method is that it is time-consuming and requires considerable investigator expertise [8]; in addition, it has been questioned whether bronchoprotection represents a situation amenable to the clinical world [8]. Other pharmacodynamic clinical efficacy studies of bronchodilators, including salbutamol, are carried out by measuring bronchodilation in patients with spontaneous [2] or induced bronchoconstriction [3, 5]. Although relatively quick and easy to perform, evaluation of salbutamol-induced changes in airway calibre in patients with spontaneous bronchoconstriction may be difficult, due to the potentially confounding effects related to the variability of the levels of spontaneous airway narrowing, especially when such variability is not adequately accounted for in patient selection. With induced bronchoconstriction, a common problem arises from the magnitude of the provoked effect. Indeed, if the level of induced airway narrowing is mild, even low doses of β2-adrenoceptor agonists delivered by virtually any inhalation device may produce near maximal bronchodilator responses [8]. Therefore, in these experimental conditions, detection of differences in the bronchodilator response to different formulations (e.g. powdered and aerosolized) of the same agent may be difficult [15]. The bioassay used in our study is based upon the firmly established pharmacological principle of functional antagonism between muscarinic and β-adrenergic agonists, which, working through different receptors, produce opposite responses [17]. According to this principle, the airway response to β2-adrenergic agonists depends on the MCh dose [11, 16]; the greater the amount of MCh needed to cause bronchoconstriction, the greater the concentration of β2-adrenergic agonist needed to reverse it [17]. Conceivably, administration of higher MCh concentrations with subsequently more marked levels of bronchoconstriction probably increases the functional antagonism between MCh and salbutamol. A strong functional antagonism shifts the dose–response curve to the right and may produce differences that are not apparent in bronchodilator studies performed in patients with a mild level of MCh-induced bronchconstriction. Thus, we believe that the present clinical bioassay, based on the induction of a more marked level of bronchoconstriction (FEV1 fall −35%) offers a valuable alternative to conventional bronchodilator studies [2] or to studies using bronchoprotection [13, 14, 26].
Admittedly, such a level of induced bronchoconstriction raises important safety issues. However, reductions in FEV1 ranging from 30 to 75% of baseline have been achieved in several previous studies [28–30] involving asthmatic patients, and have been reported to be well tolerated and devoid of significant side-effects [28–30]. In keeping with this, the MCh-induced 35% fall in baseline FEV1 was always well tolerated by our patients, all having normal baseline FEV1 values (mean FEV1 100.4%, 95% CI 93.4, 107.4%). Thus, once the desired level of induced bronchoconstriction had been achieved, their FEV1 fell to approximately 75% of predicted, a value corresponding to only a mild level of airway obstruction [18]. Based on these considerations, we believe that the extent of MCh-induced bronchoconstriction employed in our study is safe and tolerable.
We compared the dose–response curves obtained in pMDI + Volumatic and Diskus trials by using the Finney linear bioassay statistical method [24]. This procedure calculates the salbutamol doses required by the two inhalation devices to achieve the same recovery times. We found that approximately twice the salbutamol dose administered via the Diskus is needed to obtain the same recovery times as those obtained with a single salbutamol dose inhaled via the pMDI + Volumatic. Thus, it seems logical to conclude that the pMDI + Volumatic is preferable for administration of aerosol medications, especially when the fastest bronchodilator response is required.
It has been shown that, in asthmatic patients, administration of bronchodilator drugs in a cumulative-dose regimen causes greater bronchodilator responses than with the equivalent single-dose regimen [31, 32]. In consequence, it could be argued that the differences in recovery times observed in our study are also due to differences in the scheme of salbutamol administration. We believe that this is unlikely in the light of the following considerations. First, at variance with studies [31, 32] in which the time elapsed from administration of multiple bronchodilator doses was at least 10 min, we have always performed multiple actuations within 30 s, i.e. probably in the absence of any prominent effect by the first administered dose [33]. Second, and more importantly, if one compares the recovery times of trials performed with 200 µg salbutamol via the pMDI + Volumatic (two actuations) and 400 µg salbutamol via the DPI (two actuations), i.e. in conditions where the number of actuations is not influential, the FEV1 response obtained with 400 µg via DPI is the same as that obtained with 200 µg salbutamol via the pMDI + Volumatic. Therefore, the key factor accounting for the overlapping recovery times despite twice as much the salbutamol dose must have been the inhalation device and not the drug administration scheme.
Recovery times for FVC turned out to be significantly shorter than those for FEV1 and isoFEF25–75 (Figure 3). We have previously shown [19] that, following a 35% fall in FEV1 induced by MCh, hyperinflation occurs, as documented by parallel reductions in both the inspiratory capacity and vital capacity. Hyperinflation could be promptly abolished by salbutamol administration [19]. Given the similar functional meaning of the slow and forced vital capacity in patients with airflow limitation [34], we believe that the reductions in FVC observed here after MCh-induced bronchoconstriction also reflect dynamic hyperinflation. It seems possible that even submaximal bronchodilation may have promptly abolished, at least to a considerable extent, the dynamic hyperinflation induced by MCh inhalation, thus explaining the shorter recovery times for FVC compared with those of flow-dependent variables such as the FEV1 and the isoFEF25–75.
In the present study we have also aimed to evaluate which pulmonary function variable was most sensitive in detecting differences in the recovery times observed in pMDI + Volumatic and Diskus trials. This was done by calculating the ratio of the variability of the response (commonly symbolized as s) to the slope of the dose–response curve (commonly symbolized as b). We found that, although all considered pulmonary function variables displayed high b values and low variability of the response, the s : b values for FEV1 and isoFEF25–75 were significantly lower than that calculated for FVC, thus suggesting that the former are more sensitive than FVC in assessing differences in the recovery times. The sensitivity of a given clinical bioassay study for evaluating the bioequivalence between different drugs or different inhalation devices is related to the slope of the dose–response (b) and the variability of the response (s) [25, 26]. Thus, we believe that our clinical bioassay has sufficiently steep dose–response slopes and low variability of the response to provide a clinically useful assessment of the bioequivalence of the two inhalation devices being compared and therefore sufficient sensitivity to identify which inhalation device delivers the greater amount of salbutamol to the lungs.
Bronchoactive drugs are optimally deposited in the small airways when the aerosol particle size is in the respirable range, that is, a mass median aerodynamic diameter <5 µ[35]. To the best of our knowledge, however, no studies have specifically compared lung deposition of bronchoactive drugs obtained by using the pMDI + Volumatic with that obtained with the Diskus inhaler. Pharmacokinetic and radioaerosol studies have shown a correlation between the amount of particles with a mass median aerodynamic diameter <5 µ (commonly referred as fine particles) and lung deposition [36]. It has also been shown that the proportion of fine particles obtained by using the pMDI + Volumatic is significantly higher than that obtained with the Diskus [37]. Furthermore, physiological studies have shown that low FEF25–75 values reflect significant obstruction in peripheral airways, even in patients with FEV1 values >80% predicted [21, 38]. Thus, we are confident that the rapid isoFEF25–75 improvements observed in salbutamol pMDI + Volumatic trials actually reflect peripheral airway dilation, possibly due to a higher percentage of fine particle produced by using the pMDI + Volumatic than the DPI. This may be important not only to reverse airway obstruction rapidly, but also to prevent future asthma attacks [39] and the progression to airway fibrosis and remodelling [40].
In conclusion, this study shows that salbutamol administered via the pMDI + Volumatic determines more rapid bronchodilation than via the Diskus, and that the pMDI + Volumatic is approximately twice as efficient as the Diskus inhaler in relative lung delivery of salbutamol in asthmatic patients with induced bronchoconstriction. However, it should be emphasized that large differences in drug delivery can be observed among spacers of different size and shape, as well as when the same device is used with different drugs. Furthermore, inhalation of sensitizing allergens may induce, in patients with allergic asthma, late asthmatic reactions [41] that begin 3–4 h after allergen inhalation, may last 24 h or longer, and are not easily reversed by doses of β-adrenergic agonists such as those employed in the present study. For these reasons, care should be taken in extrapolating our results to inhalation devices, bronchodilator drugs and clinical conditions other than those examined here.
Acknowledgments
This study was supported exclusively by research grants from the Ministero dell’Istruzione, dell’Università e della Ricerca of Italy, and received no support from pharmaceutical industries.
References
- 1.Global Initiative for Asthma. Global strategy for asthma management and prevention. Washington DC: National Institute of Health, National Heart, Lung and Blood Institute; 2002. Publication no. 02-3659. [Google Scholar]
- 2.Pauwels R, Lamont H, Hidinger K, Van Der Straeten M. Influence of an extension tube on the bronchodilator efficacy of terbutaline delivered from a metered dose inhaler. Respiration. 1984;45:61–6. doi: 10.1159/000194599. [DOI] [PubMed] [Google Scholar]
- 3.Fontana GA, Lavorini F, Chiostri M, Castellani W, Boddi V, Pistolesi M. Large and small airway responses to procaterol hydrochloride administered through different extension devices in asthmatic patients. J Aerosol Med. 1999;12:177–85. doi: 10.1089/jam.1999.12.177. [DOI] [PubMed] [Google Scholar]
- 4.Price AH, Clissold SP. Salbutamol in the 1980s. A reappraisal of its clinical efficacy. Drugs. 1989;38:77–122. doi: 10.2165/00003495-198938010-00004. [DOI] [PubMed] [Google Scholar]
- 5.Beach JR, Bromly CL, Avery AJ, Reid RW, Walters EH, Hendrick DJ. Speeds of action of single doses of formoterol and salbutamol compared with placebo in reversing methacholine-induced bronchoconstriction. Pulm Pharmacol. 1996;9:245–9. doi: 10.1006/pulp.1996.0031. [DOI] [PubMed] [Google Scholar]
- 6.Chrystyn H. Standard for bioequivalence of inhaled products. Clin Pharmacokinet. 1994;26:1–6. doi: 10.2165/00003088-199426010-00001. [DOI] [PubMed] [Google Scholar]
- 7.British Association for Lung Research. Determining equivalence of inhaled medications. Respir Med. 1995;89:253–61. doi: 10.1016/0954-6111(95)90085-3. [DOI] [PubMed] [Google Scholar]
- 8.Ahrens RC. On comparing inhaled beta adrenergic agonists. Ann Allergy. 1991;64:296–8. [PubMed] [Google Scholar]
- 9.Ahrens RC, Hendeles L, Clarke WR, Dockhorn RJ, Hill MR, Vaughan LM, Lux C, Han SH. Therapeutic equivalence of Spiros dry powder inhaler and Ventolin metered dose inhaler. A Bioassay Using Methacholine. Am J Respir Crit Care Med. 1999;160:1238–43. doi: 10.1164/ajrccm.160.4.9806101. [DOI] [PubMed] [Google Scholar]
- 10.Roux FJ, Grandordy B, Douglas JS. Functional and binding characteristics of long-acting β2-agonists in lung and heart. Am J Respir Crit Care Med. 1996;153:1489–95. doi: 10.1164/ajrccm.153.5.8630591. [DOI] [PubMed] [Google Scholar]
- 11.Jenne JW, Shaughnessy TK, Druz WS, Manfredi CJ, Vestal RE. In vivo functional antagonism between isoproterenol and bronchoconstrictants in dog. J Appl Physiol. 1987;63:812–9. doi: 10.1152/jappl.1987.63.2.812. [DOI] [PubMed] [Google Scholar]
- 12.Ahrens RC, Harris JB, Milavetz G, Annis L, Ries R. Use of bronchial provocation with histamine to compare the pharmacodynamics of inhaled albuterol and metaproterenol in patients with asthma. J Allergy Clin Immunol. 1987;79:876–82. doi: 10.1016/0091-6749(87)90235-1. [DOI] [PubMed] [Google Scholar]
- 13.Parameswaran KN, Inman MD, Ekholm BP, Morris MM, Summers E, O’Byrne PM, Hargreave FE. Protection against methacholine bronchocosntriction to assess relative potency of inhaled β2-agonist. Am J Respir Crit Care Med. 1999;106:354–7. doi: 10.1164/ajrccm.160.1.9812035. [DOI] [PubMed] [Google Scholar]
- 14.Creticos PS, Adams WP, Petty BG, Lewis LD, Singh GJ, Khattignavong AP, Molzon JA, Martinez MN, Lietman PS, Williams RL. A methacholine challenge dose–response study for development of a pharmacodynamic bioequivalence methodology for albuterol metered-dose inhalers. J Allergy Clin Immunol. 2002;110:713–20. doi: 10.1067/mai.2002.129036. [DOI] [PubMed] [Google Scholar]
- 15.Hendeles L, Beaty R, Ahrens R, Stevens G, Harman EM. Response to inhaled albuterol during nocturnal asthma. J Allergy Clin Immunol. 2004;113:1058–62. doi: 10.1016/j.jaci.2004.03.046. [DOI] [PubMed] [Google Scholar]
- 16.Elsasser S, Donna E, Demirozu CM, Danta I, Wanner A. Metaproterenol responsiveness after methacholine- and histamine-induced bronchoconstriction. Chest. 1996;110:617–23. doi: 10.1378/chest.110.3.617. [DOI] [PubMed] [Google Scholar]
- 17.Van den Brink FG. The model of functional interaction. I. Development and first check of a new model of functional synergism and antagonism. Eur J Pharmacol. 1973;22:270–8. doi: 10.1016/0014-2999(73)90026-5. [DOI] [PubMed] [Google Scholar]
- 18.American Thoracic Society. Lung function testing: selection of reference values and interpretative strategies. Am Rev Respir Dis. 1991;144:1202–18. doi: 10.1164/ajrccm/144.5.1202. [DOI] [PubMed] [Google Scholar]
- 19.Lavorini F, Geri P, Luperini M, Maluccio NM, Mariani L, Marmai C, Pistolesi M, Fontana GA. Clinical and functional responses to salbutamol inhaled via different devices in asthmatic patients with induced bronchoconstriction. Br J Clin Pharmacol. 2004;58:512–20. doi: 10.1111/j.1365-2125.2004.02185.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Quanjer PH, Tammeling GJ, Cotes JE, Pederson OF, Peslin R, Yernault JC. Lung volumes and forced ventilatory flows. Report Working Party: standardisation of lung function tests. European Community for Steel and Coal. Official statement of the European Respiratory Society. Eur Respir J. 1993;6(Suppl 16):5–40. [PubMed] [Google Scholar]
- 21.McFadden ER, Jr, Linden DA. A reduction in maximal mid-expiratory flow rate. A spirographic manifestation of airway disease. Am J Med. 1972;52:725–37. doi: 10.1016/0002-9343(72)90078-2. [DOI] [PubMed] [Google Scholar]
- 22.Olsen CR, Hale FC. A method for interpreting acute response to bronchodilators from the spirogram. Am Rev Respir Dis. 1968;98:301–2. doi: 10.1164/arrd.1968.98.2.301. [DOI] [PubMed] [Google Scholar]
- 23.Beach JR, Young CL, Stenton SC, Avery AJ, Walters EH, Hendrick DJ. A comparison of the speed of action of salmeterol and salbutamol in reversing methacholine-induced bronchoconstriction. Pulm Pharmacol. 1992;5:133–5. doi: 10.1016/0952-0600(92)90031-b. [DOI] [PubMed] [Google Scholar]
- 24.Finney DJ. Statistical methods in biological assay. 3. London: Charles Griffin & Co; 1978. [Google Scholar]
- 25.Stewart BA, Arhens RC, Carrier S, Frosolono M, Han SH, Milavetz G. Demonstration of in vivo bioequivalence of a generic albuterol metered-dose inhaler to Ventolin. Chest. 2000;117:714–21. doi: 10.1378/chest.117.3.714. [DOI] [PubMed] [Google Scholar]
- 26.Ahrens RC, Teresi ME, Han SH, Donnell D, Vanden Burgt JA, Lux CR. Asthma stability after oral prednisone. A clinical model for comparing inhaled steroid potency. Am J Respir Crit Care Med. 2001;164:1138–45. doi: 10.1164/ajrccm.164.7.2008112. [DOI] [PubMed] [Google Scholar]
- 27.Food and Drug Administration. Interim guidance for documentation of in vivo bioequivalence of albuterol aerosols (metered-dose inhalers) Rockville, MD: Division of Bioequivalence, Office of Generic Drugs. Food and Drug Administration; 1994. Publication no. 1-27. [Google Scholar]
- 28.Marks GB, Yates DH, Sist M, Ceyhan B, De Campos M, Scott DM, Barnes PJ. Respiratory sensation during bronchial challenge testing with methacholine, sodium metabisulphite, and adenosine monophosphate. Thorax. 1996;51:793–8. doi: 10.1136/thx.51.8.793. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Politiek MJ, Boorsma M, Aalbers R. Comparison of formoterol, salbutamol and salmeterol in methacholine-induced severe bronchoconstriction. Eur Respir J. 1999;13:988–92. doi: 10.1034/j.1399-3003.1999.13e10.x. [DOI] [PubMed] [Google Scholar]
- 30.van der Woude HJ, Boorsma M, Bergqvist PB, Winter TH, Aalbers R. Budesonide/formoterol in a single inhaler rapidly relieves methacholine-induced moderate-to-severe bronchoconstriction. Pulm Pharmacol Ther. 2004;17:89–95. doi: 10.1016/j.pupt.2003.11.001. [DOI] [PubMed] [Google Scholar]
- 31.Heimer D, Shim C, Williams MH., Jr The effect of sequential inhalations of metaproterenol aerosol in asthma. J Allergy Clin Immunol. 1980;66:75–7. doi: 10.1016/0091-6749(80)90141-4. [DOI] [PubMed] [Google Scholar]
- 32.Fishwick D, Bradshaw L, Macdonald C, Beasley R, Gash D, Bengtsson T, Bondesson E, Borgstrom L. Cumulative and single-dose design to assess the bronchodilator effects of β2-agonists in individuals with asthma. Am J Respir Crit Care Med. 2001;163:474–7. doi: 10.1164/ajrccm.163.2.2003027. [DOI] [PubMed] [Google Scholar]
- 33.Lulling J, Delwiche JP, Prignot J. Evaluation of onset of bronchodilating effect by interruption of the air flow method. Eur J Respir Dis. 1980;61:108–12. [PubMed] [Google Scholar]
- 34.Newton MF, O’Donnell DE, Forkert L. Response of lung volumes to inhaled salbutamol in a large population of patients with severe hyperinflation. Chest. 2002;121:1042–50. doi: 10.1378/chest.121.4.1042. [DOI] [PubMed] [Google Scholar]
- 35.Lipworth BJ. Targets for inhaled treatment. Respir Med. 2000;94:S13–S16. doi: 10.1016/s0954-6111(00)90118-5. [DOI] [PubMed] [Google Scholar]
- 36.Chrystyn H. Is inhalation rate important for a dry powder inhaler? Using the In-Check Dial to identify these rates. Respir Med. 2003;97:181–7. doi: 10.1053/rmed.2003.1351. [DOI] [PubMed] [Google Scholar]
- 37.Prime D, Grant AC, Slater Al, Woodhouse RN. A critical comparison of the dose delivery characteristics of four alternative inhalation devices delivering salbutamol. pressurized metered-dose inhaler, Diskus inhaler, Diskaler inhaler and Turbuhaler inhaler. J Aerosol Med. 1999;12:75–84. doi: 10.1089/jam.1999.12.75. [DOI] [PubMed] [Google Scholar]
- 38.Stãnescu D. Small airways obstruction syndrome. Chest. 1999;116:231–3. doi: 10.1378/chest.116.1.231. [DOI] [PubMed] [Google Scholar]
- 39.McFadden ER, Jr, Kiser R, DeGroot WJ. Acute bronchial asthma. Relations between clinical and physiological manifestation. N Engl J Med. 1973;288:221–5. doi: 10.1056/NEJM197302012880501. [DOI] [PubMed] [Google Scholar]
- 40.Evans DJ, Green M. Small airways: a time to revisit? Thorax. 1998;53:629–30. doi: 10.1136/thx.53.8.629. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.O’Byrne PM, Dolovich J, Hargreave FE. The late asthmatic response. Am Rev Respir Dis. 1987;136:741–51. doi: 10.1164/ajrccm/136.3.740. [DOI] [PubMed] [Google Scholar]



