Optical coherence tomography for clinical detection and monitoring of glaucoma?

Christopher Bowd
PMCID: PMC1955676  PMID: 17576706

Short abstract

OCT and all imaging technologies should be used in conjunction with careful clinical examination when making diagnostic and treatment decisions


Optical coherence tomography (OCT) has been used clinically for the detection of glaucoma for over a decade; however, its clinical usefulness has not been particularly well documented. Although many studies have shown that OCT measurements can discriminate between healthy and glaucomatous eyes, the great majority of studies demonstrating this have had limited clinical value. This is because most analyses were based on measurements of retinal nerve fibre layer (RNFL) thickness provided as continuous variables, and a great deal of overlap between measurements in healthy and glaucomatous eyes was observed. This made it difficult to determine what values should be used clinically as limits for classifying eyes as diseased, and in most published studies, no limits were recommended. Nonetheless, the performance of OCT for detecting RNFL thinning in glaucoma has been impressive given its limited number of data points (current maximum 512 A‐scans) and relatively unsophisticated analyses (primarily measurements of global and local RNFL thickness).

Recently, a normative database has been added to the latest version of OCT (StratusOCT, Carl Zeiss Meditec, Dublin, California, USA), allowing comparisons of measurements of RNFL thickness with measurements from 328 healthy eyes, thus providing clinicians with potentially useful thresholds to aid in the determination of disease. Printouts from the StratusOCT display RNFL normative database information in the form of clinically helpful colour‐coded graphs and charts. This is a significant improvement in the value of this technology because the clinician is now provided with additional information about measurements, including whether they are within normal limits, borderline or outside of normal limits relative to a healthy population.

A small number of studies have investigated the ability of StratusOCT normative data‐based thresholds (ie, cut‐offs) for classifying healthy and diseased eyes.1,2,3,4,5 In this issue of, British Journal of Opthalmology Hood et al6 (see page 905) describe the success of a variety of thresholds (and combinations thereof) for classifying eyes as healthy or glaucomatous. Their results indicate that the StratusOCT normative database is useful for detecting known glaucoma (defined as eyes with standard automated perimetry (SAP) and multifocal visual evoked potential defects), with sensitivities as high as 95% at a specificity of 98%. The most important outcome of their study is the identification of specific thresholds that are probably effective for detecting glaucoma. The study by Hood et al validates the use of OCT and the StratusOCT normative database for the detection of glaucoma, and provides the clinician with suggested thresholds by which to classify eyes, while providing researchers with suggested thresholds to investigate using independent subject populations.

It will be interesting to determine the usefulness of these suggested thresholds for detecting early or suspected glaucoma. We recently showed that OCT measurements are, predictably, less effective at classifying eyes with early disease than classifying those with advanced disease (with disease severity defined as degree of visual field defect).7 For instance, the sensitivity (at fixed specificity = 85%) of OCT‐measured average RNFL thickness for detecting glaucomatous eyes with average‐sized discs decreased from 94% to 73% when the SAP Advanced Glaucoma Intervention Study score8 decreased from 9 to 0. This finding suggests that the reported success of classification of glaucomatous eyes by OCT has probably been overestimated relative to the population in which it holds the most clinical utility, early or suspicious (ie, “pre‐perimetric”) glaucoma. Although it is important to demonstrate that a new imaging instrument can discriminate between healthy eyes and those with known glaucoma from a technology validation standpoint, in clinical practice one does not need an imaging test to differentiate a patient with repeatable glaucomatous visual field defects from an individual with no suspicion of disease. The more interesting (and important) question is, can the StratusOCT technology detect disease in patients suspected of having glaucoma, such as those who do not yet have repeatable visual field loss on SAP? Currently, some evidence suggests that this is the case.2,9

In general, it is accepted that OCT measurements can discriminate between healthy and glaucomatous eyes. For this task, the performance of StratusOCT compared with confocal laser ophthalmoscopy and scanning laser polarimetry has been impressive.3,10 However, other aspects of OCT performance are yet to be investigated. Primarily, it is unclear whether the current version of OCT can effectively detect RNFL thinning over time in glaucomatous eyes. One reason this has not been well tested is because of a recent update in OCT technology that rendered measurements obtained with previous OCT versions incompatible with those from the current version.11 Incompatible technology upgrades limit the available follow‐up time for assessing glaucomatous change (StratusOCT was introduced in 2001 and glaucoma, specially when treated, is a slowly progressing disease). Furthermore, no progression algorithms are currently available for OCT beyond linear regression analysis of changes in measurements over time. The difference between clinically significant change and short‐term and/or long‐term measurement variability for OCT measurements has not been determined, although arbitrary cut‐offs based on image resolution have been suggested.12 Other imaging technologies have incorporated sophisticated progression analyses that are yet to be incorporated in the OCT (to be fair, these algorithms developed by other manufacturers have not been well tested). Finally, it is not clear whether the current image registration techniques used by OCT will allow sensitive detection of change.

Overall, although OCT can clearly identify thinner RNFLs in glaucomatous eyes compared with healthy eyes, the usefulness of the StratusOCT normative database has not been well established in patients suspected of having glaucoma (ie, the population of most interest). In addition, the ability of OCT measurements to detect glaucomatous change has not been demonstrated. Cutting‐edge advances in OCT technology13,14,15,16 such as high‐resolution spectral/Fourier domain OCT and OCT–confocal scanning laser hybrids are theoretically promising (and exciting), but it will be some time before these technologies have been scrutinised to the point that their clinical use can be recommended by evidence‐based studies. In general, it is important that OCT and all imaging technologies be used in conjunction with careful clinical examination when making diagnostic and treatment decisions for glaucoma.

Footnotes

Competing interests: None declared.

References

  • 1.Badlani V, Shahidi M, Shakoor A.et al Nerve fiber layer thickness in glaucoma patients with asymmetric hemifield visual field loss. J Glaucoma 200615275–280. [DOI] [PubMed] [Google Scholar]
  • 2.Kim T W, Zangwill L M, Bowd C.et al Retinal nerve fiber layer damage as assessed by optical coherence tomography in eyes with a visual field defect detected by frequency doubling technology perimetry but not by standard automated perimetry. Ophthalmology 2007. (E‐pub) http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd = Retrieve&db = PubMed&dapt = Citation&list_uids = 17239441 [DOI] [PubMed]
  • 3.Medeiros F A, Zangwill L M, Bowd C.et al Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucoma. Arch Ophthalmol 2004122827–837. [DOI] [PubMed] [Google Scholar]
  • 4.Shah N N, Bowd C, Medeiros F A.et al Combining structural and functional testing for detection of glaucoma. Ophthalmology 20061131593–1602. [DOI] [PubMed] [Google Scholar]
  • 5.Jeoung J W, Park K H, Kim T W.et al Diagnostic ability of optical coherence tomography with a normative database to detect localized retinal nerve fiber layer defects. Ophthalmology 20051122157–2163. [DOI] [PubMed] [Google Scholar]
  • 6.Hood D, Harizman N, Kanadani F N.et al Retinal nerve fiber thickness measured with optical coherence tomography (OCT) accurately detects confirmed glaucomatous damage. Br J Ophthalmol 200791905–907. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Medeiros F A, Zangwill L M, Bowd C.et al Influence of disease severity and optic disc size on the diagnostic performance of imaging instruments in glaucoma. Invest Ophthalmol Vis Sci 2006471008–1015. [DOI] [PubMed] [Google Scholar]
  • 8.Advanced Glaucoma Intervention Study 2. Visual field test scoring and reliability. Ophthalmology 19941011445–1455. [PubMed] [Google Scholar]
  • 9.Bagga H, Feuer W J, Greenfield D S. Detection of psychophysical and structural injury in eyes with glaucomatous optic neuropathy and normal standard automated perimetry. Arch Ophthalmol 2006124169–176. [DOI] [PubMed] [Google Scholar]
  • 10.Deleon‐Ortega J E, Arthur S N, McGwin G., Jret al Discrimination between glaucomatous and nonglaucomatous eyes using quantitative imaging devices and subjective optic nerve head assessment. Invest Ophthalmol Vis Sci 2006473374–3380. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Bourne R R, Medeiros F A, Bowd C.et al Comparability of retinal nerve fiber layer thickness measurements of optical coherence tomography instruments. Invest Ophthalmol Vis Sci 2005461280–1285. [DOI] [PubMed] [Google Scholar]
  • 12.Wollstein G, Schuman J S, Price L L.et al Optical coherence tomography longitudinal evaluation of retinal nerve fiber layer thickness in glaucoma. Arch Ophthalmol 2005123464–470. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Drexler W. Ultrahigh‐resolution optical coherence tomography. J Biomed Opt 200494774. [DOI] [PubMed] [Google Scholar]
  • 14.Fujimoto J G. Optical coherence tomography for ultrahigh resolution in vivo imaging. Nat Biotechnol 2003211361–1367. [DOI] [PubMed] [Google Scholar]
  • 15.Podoleanu A G, Dobre G M, Cucu R G.et al Sequential optical coherence tomography and confocal imaging. Opt Lett 200429364–366. [DOI] [PubMed] [Google Scholar]
  • 16.Romero‐Borja F, Venkateswaran K, Roorda A.et al Optical slicing of human retinal tissue in vivo with the adaptive optics scanning laser ophthalmoscope. Appl Opt 2005444032–4040. [DOI] [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group

RESOURCES