A Novel Device for Peritoneal Dialysis Catheter Placement

V Chidanbaram-Nathan 1, P Atkinson 2, MS Karim 1
PMCID: PMC1963520

BACKGROUND

Peritoneal dialysis catheters are placed by Seldinger technique percutaneously or by open method surgically.1 The intraperitoneal part of the catheter needs to be in the pelvis to facilitate good function. Open surgical placement of the dialysis catheter involves use of a stilette on which the catheter is mounted to position it in the pelvis. We have devised a peritoneal dialysis catheter positioner (Fig. 1) that enables accurate, safe placement of the catheter in the pelvis.

Figure 1.

Figure 1

Peritoneal dialysis catheter positioner.

TECHNIQUE

The abdomen is opened through a 5–7-cm long median or paramedian incision. The peritoneum is opened just enough to introduce the positioner which, by gentle manipulation (Fig. 2), is brought to rest (Fig. 3) in the pelvis. When the positioner lies in front of the rectum, it just brings to view the window at the edge of the peritoneum. The dialysis catheter is introduced through the positioner until the tip of the catheter touches pelvis.

Figure 2.

Figure 2

Introduction and manipulation of the positioner.

Figure 3.

Figure 3

Positioner resting in the pelvis.

The catheter is held through the window in its position with a nontoothed dissecting forceps (Fig. 4). The positioner is withdrawn carefully (Fig. 5) and flow checked by flushing the tube (Fig. 6). The catheter is fixed to the peritoneal edge and tunnelled to the predetermined exit site.

Figure 4.

Figure 4

Catheter control.

Figure 5.

Figure 5

Withdrawing the positioner.

Figure 6.

Figure 6

Flushing the catheter.

The linea alba or the rectus sheath and abdominal wound is closed by the surgeon's preferred method. The positioner enables safe and accurate placement of the peritoneal dialysis catheter.

DISCUSSION

This positioner was made by the Engineering Department of St Bartholomew's Hospital in the 1980s and has been used by the authors in over 300 procedures. It can now be obtained from Bolton Surgicals, Sheffield, UK