SQUASH – Simple Qualitative User-Friendly Amplitude Sensor of Hand-Strength – A Non-Invasive Monitor During Carotid Endarterectomy Under Local Anaesthetic

R Joy 1, MA Chadwick 1, JD Arthur 1, RJ McCarthy 1, IC Currie 1
PMCID: PMC1964571

BACKGROUND

There is currently no objective, reliable and recordable method of measuring contralateral motor function during a carotid endarterectomy (CEA) under local anaesthesia. A major complication of CEA is cerebral hypoperfusion requiring shunting. We describe a novel technique to determine the need for shunting.

TECHNIQUE

SQUASH is a non-invasive technique, which allows assessment of contralateral upper limb grip strength during CEA under local anaesthesia without sedation (Fig. 1). The SQUASH system is shown in Figure 2. The voluntary grip strength is transduced to zero and the patient is instructed to grip the bag maximally and its trace calibrated for maximum ‘squash’ (Fig. 3). The patient is asked to squash (Fig. 4) during the procedure especially before and after the carotid cross clamping. Reduction in grip strength is used to determine the need for shunting. We noticed a decrease in the frequency and/or amplitude of the waveforms during episodes of hypoperfusion (Figs 5 and 6) in 2 patients out of 92 who were shunted and made a full recovery. There were no adverse outcomes intra-operatively using this technique.

Figure 1.

Figure 1

SQUASH bag.

Figure 2.

Figure 2

Schematic diagram of SQUASH system.

Figure 3.

Figure 3

Maximum SQUASH.

Figure 4.

Figure 4

Normal SQUASH.

Figure 5.

Figure 5

Abnormal SQUASH.

Figure 6.

Figure 6

Abnormal SQUASH.

Key points Figs 1 & 2

  • Fixed volume bag (100 ml normal saline), which fits the palm.

  • Drawing up needle to which extension pressure tubing with Luer lock is attached.

  • Reinforcing tape.

  • The ‘drawing up’ needle, attached to the CVP tubing at the Luer lock, is inserted into saline bag via the rubber seal and reinforced (we use Elastoplast).

  • Central venous pressure (CVP) transducer and non-expansile pressure tubing with Luer lock attachment.

  • The CVP tubing is filled from the bag and flushed to ensure no air bubbles are present.

  • Monitor with flow module (standard anaesthetic monitor).

DISCUSSION

CEA is potentially complicated by cerebral hypoperfusion resulting in irreversible neurological deficit. Intra-operative indicators of cerebral hypoperfusion or hypofunction like transcranial Doppler, EEG, infrared spectroscopy, shunt pressures or awake-patient monitoring can be used to determine the need for selective shunting. Alternatively, the SQUASH provides a visual representation of grip strength in real time and is a direct measure of cortical sensorimotor integrity, which indirectly measures cerebral perfusion. It is easy to assemble, inexpensive, re-usable and non-invasive. Whether criteria to consider shunt placement can be derived from the waveform requires future study. Until then, we believe this will remain a promising means of non-invasive cerebral perfusion monitoring during CEA under local anaesthesia.


Articles from Annals of The Royal College of Surgeons of England are provided here courtesy of The Royal College of Surgeons of England