Introduction
Bronchodilator drugs, used alone or in conjunction with inhaled corticosteroids or other anti-inflammatories, are essential to the effective management of asthma [1, 2]. The three main categories of effective bronchodilators are inhaled β2-adrenoceptor agonists and anticholinergics, and methylxanthines, particularly oral theophylline. The incidence of asthma continues to rise, however, enhancing the need to develop new bronchodilator drugs and products and to evaluate their therapeutic efficacy, potency and cost-effectiveness in relation to existing treatments [3]. The short-acting β2-adrenoceptor agonist salbutamol, in a pressurized metered dose inhaler (pMDI), has long been the most widely used bronchodilator, and a key factor influencing the requirement for new products is the phasing out of chlorofluorocarbon (CFC) pMDI propellants to curtail damage to the stratospheric ozone layer [4]. The pharmaceutical industry is also responding to a decade of off-patent availability of effective asthma drugs including salbutamol and the anticholinergic ipratropium bromide, which has encouraged the development of novel generic bronchodilator/inhaler device combinations [5]. This combination of factors has led to a pressing need to develop the most economical and precise methods of evaluating efficacy and potential clinical equivalence of inhaled bronchodilators with the traditional pMDI and current alternatives in order to satisfy both clinicians and regulatory authorities.
Clinical equivalence of drugs relates to both potency and safety since the lowest effective dose must be used in order to minimize systemic side-effects. Measures of bronchodilator drug efficacy have in the past relied on direct measurement of the induced bronchodilatation by lung function testing. This often does not, however, allow potency discrimination between drug products, since a single dose can result in maximal bronchodilatation. This has led both clinicians and pharmaceutical professionals to develop more sensitive, repeatable, and therefore more discriminating methodologies employing alternative patient populations and alternative efficacy endpoints. Dose–response data are fundamental to the accurate measurement of both efficacy and systemic effects. Importantly, regulatory authorities are also addressing these issues [6–8].
This review focuses on β2-adrenoceptor agonist studies to highlight the discriminating measures of bronchodilator drug efficacy and potency which should enable a valid comparison of novel and traditional inhaled asthma treatments. (Measures and methodologies are summarized in Table 1.)
Table 1.
Comparison of methodologies for testing of bronchodilator efficacy.
| Methodology | Reference* | Population | Advantages | Disadvantages | Nominal rating |
|---|---|---|---|---|---|
| Spirometry – FEV1 | 13 | Mild to moderate reversible airways disease | Simple technique | Plateau effect | + |
| Single dose comparison | |||||
| Spirometry – FEV1 | 21 | Mild to moderate reversible airways disease | Simple technique | Variability of disease (baseline and responsiveness) | ++++ |
| Separate day dose–response comparison | No carry over effects | ||||
| Increased number of study days required | |||||
| Spirometry – FEV1 | 24 | Mild to moderate reversible airways disease | Simple technique | Plateau effect | ++ |
| Cumulative dose–response comparison | Single study day | Carry over effects | |||
| Spirometry – FEV1 | 30,32 | More severe reversible airways disease | Simple technique | Carry over effects | +++ |
| Cumulative dose–response comparison | Clinically relevant population | ||||
| Reduced chance of plateau effect | |||||
| Plethsymography – sGaw | 34 | Healthy volunteers | No variability due to disease state | Specialized equipment requirement | + |
| Potentially larger response | Not tested in target population | ||||
| Challenge – PC20/PD20 | 46,47 | Mild to moderate reversible airways disease | Dose–reponse more easily demonstrated | Complex technique requiring experience | ++++ |
| Clinical relevance of model | |||||
| Smaller number of patients than FEV1 | Patient safety | ||||
| Pharmacokinetic – drug levels in blood | 60 | Healthy volunteers and mild to moderate reversible airways disease | Early levels indicate relative lung dose | Invasive | + |
| Simple, requires no specialized lung function testing | Indirect measure of efficacy | ||||
| Small number of subjects | |||||
| Pharmacokinetic – drug levels in urine | 64,65 | Healthy volunteers and mild to moderate reversible airways disease | Early levels indicate relative lung dose | Indirect measure of efficacy | + |
| Simple, requires no specialized lung function testing | No information on lung distribution | ||||
| Non–invasive | |||||
| Small number of subjects | |||||
| Gamma scintigraphy | 67 | Healthy volunteers and mild to moderate reversible airways disease | Lung distribution information (confounded by 2 dimensional images) | Complex technique requiring experience | + |
| Specialized equipment requirement | |||||
| Small numbers of subject | Indirect measure of efficacy | ||||
| Long term clinical studies | 68 | Mild to moderate reversible airways disease | Clinically relevant population and treatment period | Variability issues | +++ |
Reference to an example of a study using such a design/methodology.
Spirometry – the major endpoint
Measurements of lung function are fundamental to both the diagnosis of asthma and the demonstration of treatment effects. Acute drug effects such as bronchodilatation may be measured spirometrically at intervals following drug inhalation to establish the peak and duration of effect. The forced expiratory volume in the first second (FEV1) which like other spirometric indices is affected by airway obstruction, is generally accepted as the most useful measure of airway patency since it is considered to be a repeatable index. Peak expiratory flow (PEF) measurement is a convenient way of monitoring longer-term drug effects on lung function during clinical trials (e.g. twice-daily self-testing by patients using a mini-Wright peak flow meter at home). PEF is more effort-dependent than the relatively prolonged FEV1 and although average changes in FEV1 and PEF are similar during bronchodilatation or constriction, the variability of PEF is greater. Both values are made more reliable by taking the best of at least three attempts, but accuracy is better assured during spirometry where a low forced vital capacity (FVC) reading, if this is inconsistent, indicates a poor effort which can validly be discounted. Rigorous adherence to standardized spirometry techniques is fundamental to the accurate use of these endpoints [9, 10].
When 25% of the vital capacity (VC) has been expired, the larger airways will be largely compressed and flow-limiting. From this point, the forced expiratory flow (FEF) can provide some information about resistance of the small airways, which depends on the elastic properties of the lung tissue and the magnitude of VC. FEF values between 25% and 75% of VC (e.g. FEF25, FEF50, FEF75, FEF25-75) can be calculated from spirometry and used to gain added insight into bronchodilator [11] and anti-inflammatory [12] drug effects. This may be particularly important in patients with a low bronchodilator reversibility of FEV1 who are not typically selected for Phase 3 drug studies.
Despite its limitations, FEV1 has been widely used to compare both different bronchodilator drugs and delivery devices, since the method is easy to perform and relatively noninvasive [13]. The sensitivity of FEV1 to differentiate in these comparative studies has, however, depended upon the selection of the dose used and, importantly, upon the population of patients. The acceptability of FEV1 as an efficacy endpoint, to both clinicians and regulatory authorities, is therefore limited to carefully designed discriminating studies. Some of these approaches are discussed below. It should be noted that this issue applies not just to spirometric indices but to other measures of bronchodilator efficacy, potency and safety.
Study design issues
Appropriate study design can greatly increase the discriminatory power of bronchodilator studies, avoiding a great deal of wasted effort and conflicting results [14]. This is recognized in current regulatory guidelines for demonstrating bioequivalence, although the standards for new drug/device comparability may not be clearly defined [7, 8, 15]. The most common study design flaw responsible for inconsistencies in past comparisons of inhaled β-adrenoceptor agonists has been identified as being the use of only a single dose of each product [14]. Most often this would be the standard therapeutic dose, which might or might not exert a maximum effect under the particular study conditions [16]. Two bronchodilators showing no statistically significant differences in such a single-dose comparison could reflect either a lack of statistical power due to insufficient sample size (type II error) or merely the fact that each drug was in excess of the amount required for maximum bronchodilatation in that specific patient population. Dose-equivalence problems are compounded if the two drugs also have different time courses, as when comparing short-acting (e.g. salbutamol) with long-acting (e.g. salmeterol) β2-adrenoceptor agonists [17].
Dose–response curves
To obviate the bronchodilator saturation effect which will bias the study outcome if the plateau of the individual dose–response curve has been attained, it is imperative to measure the responses to at least two, and preferably three or more, doses of each drug/device under test, establishing significant dose–response relationships for bronchodilator efficacy. Dose–response curves of the systemic drug effects are equally important for establishing full therapeutic equivalence [17]. Comparison of the ascending bronchodilator dose–response curves can determine the relative doses of each agent which produce similar bronchodilatatory effects in the lung [18]. The ratio of doses giving equivalent effects (potency ratio) indicates the relative potency of the two products. Employment of a standard reference product, often the brand leader such as Ventolin pMDI (GlaxoSmithKline, UK), further allows for indirect comparison between studies and new products in terms of their potencies relative to such a reference. However, the lowest manufactured dose of an inhaled bronchodilator will often produce a good response, and developing a full dose–response curve by including smaller, experimental doses is unrealistic [6]. Adequate statistical methods such as the Finney two-by-two bioassay are therefore recommended to validate dose–response comparative studies and provide relative potency values with confidence intervals [8, 15, 19]. It has been indicated by the US regulatory authorities that a valid and sensitive comparison of the relative potency ratio of two products can be determined not only by examination of the response ratio for linear dose response relationships but also from the dose ratio for nonlinear dose response curves [20]. The principles behind the recommendation of this type of analysis are shown in Figure 1.
Figure 1.
Calculation of relative potency ratio based on the dose and response scale. The use of the dose scale improves the sensitivity of the comparison. Figure adapted from Singh 1996 [20].
Single dose/separate day regimen
Bronchodilator dose–response studies generally employ a double-blind (often double-dummy) randomised, crossover protocol with appropriate washout periods between study days. Two or more doses of each drug, and placebo if included, may be administered each on a separate day, making a total of at least 4 study days [21]. This has the disadvantage that the participating asthmatic patients must ideally show the same baseline lung function on each occasion. Whilst the majority of patients with mild/moderate stable asthma will reproduce a baseline FEV1 ≥ 65% of predicted, they are less likely to maintain FEV1 reversibility ≥ 15%, so that the potential for improvement can be insufficient for discriminating measurements in some patients on certain days. Increasing patient numbers can reduce the potential impact of a variable baseline on study outcome. The sensitivity of FEV1 and other endpoints may be increased by expressing each drug-induced improvement from baseline as a percentage of the maximum improvement observed for that individual throughout the study, rather than a percentage of the predicted normal value. This approach reflects the degree to which the patient's ventilation returns to maximum achievable function after a bronchodilator treatment, giving steeper dose–response curves and emphasizing differences when the baseline obstruction is mild [22].
Cumulative dose regimen
An alternative method is the use of cumulative dosing which allows estimation of a bronchodilator dose–response within 1 day, thus reducing some of these confounding effects. Starting with a potentially submaximal dose of 100 µg of salbutamol, for example, this is followed by a second 100 µg at 30 min, and a further 200 µg at 60 min. Since each dose adds on to the residual effect of the preceding one, the dose–response is less clear-cut [6]. A comparison of cumulative and noncumulative techniques in the same subjects showed a significantly greater FEV1 response to isoprenaline following cumulative dosing [23]. This method is affected by timing, which should take account of the duration of action of each drug: for example, with short-acting β2-adrenoceptor agonists some of the early effects will have worn off before the last dose, making the overall dose uncertain. This makes for difficulties regarding bronchodilators with a different onset or duration of action, such as salbutamol and salmeterol [17]. However, cumulative dosing is a useful and convenient technique for comparison of drugs with a similar time course of action, or for the same drug in different formulations. If conducted in appropriate patients, this method reliably provides two key points – e.g. for salbutamol, the submaximal effect of 100 µg and the cumulative effect of 400 µg from the same baseline – and it has been used to demonstrate clinical efficacy in a number of studies [24].
Patient selection
This represents the second key factor to improving the discriminating power of studies based on the measurement of FEV1. Subject selection for bronchodilator testing has generally taken a cautious approach, avoiding asthmatic patients with more severely impaired lung function (< 65% predicted). However, this tends to limit the scope of the dose–response relationship, which may be particularly important when comparing drugs at dose levels used in clinical practice. The presence of more severe airway obstruction would shift the bronchodilator dose–response curve to the right, requiring higher doses of drug to achieve the same degree of improvement. Use of patients with severe airway obstruction, in addition to increasing the discriminatory potential of the study, also provides important information in a clinically relevant population.
Acute asthma
In a hospital-based study, patients manifested a greater response to a higher dose of salbutamol only when they had acute asthma [25]. Thus, 30 min after initial dosage, the mean FEV1 had improved by approximately 34% in patients treated with 2.5 mg of salbutamol via a jet nebuliser, as compared with only 6% improvement in patients receiving 0.27 mg of salbutamol via a pMDI with spacer. After 24 h, no difference was apparent in the recovery rate of these two groups, irrespective of the earlier scale of lung β2-receptor activation and bronchodilatation. Nebulisers, though variable, have been shown to deliver a similar proportion of the nominal dose of salbutamol to the lungs as the pMDI [26, 27]. Therefore, although different inhalers were employed, this study was able to distinguish a tenfold difference in the delivered dose of β2-adrenoceptor agonist during relatively severe airway obstruction. Hospitalized patients with acute bronchial obstruction (mean FEV1 37% of the predicted normal) have also been shown to respond equally well to cumulative doses of salbutamol totalling either 2 mg from the Turbuhaler DPI or 4 mg from a pMDI plus spacer [28]. This nominal dose-difference is in keeping with improved lung delivery of drug from the Turbuhaler in comparison with the pMDI [13, 29].
Patients with relatively impaired lung function (mean FEV1 58% of the predicted normal) have also been used outside the hospital setting, for example, to compare the FEV1 response to salbutamol delivered via the Ventolin pMDI and a novel DPI [30]. Since these patients again started with a relatively low baseline FEV1, a greater degree of improvement was required before they reached the plateau of their dose–response. A potentially confounding issue in utilizing patients with acute asthma is that there may be an improvement in lung function over time which is independent of bronchodilator therapy.
Nocturnal asthma
Patients with nocturnal asthma demonstrate increased severity of obstruction during the night, and ethical studies are possible if such volunteers are continually observed and drugs are compared on a single-blind (to the subject) basis [31]. The cumulative dosing method is suited to this situation, and the procedure has been used as a practical bioassay in a two-night crossover study demonstrating the therapeutic equivalence of a new salbutamol DPI with the Ventolin pMDI [32]. The mean baseline FEV1 of these subjects upon waking or being woken during the night was 43% ± 10% of predicted.
Whole-body plethysmography – an alternative endpoint
An alternative method of investigating bronchodilator drug efficacy is to record a flow-volume or pressure-volume loop using whole-body plethysmography. The specific airway conductance (sGaw=conductance or 1/airway resistance/lung volume) gives a measure of airway patency that is broadly independent of lung size and has been found more sensitive than FEV1 as a measure of response to bronchodilators in normal, nonasthmatic subjects [33]. Although it may be considered less repeatable than FEV1, more recent studies have demonstrated the use of sGaw as a sensitive pharmacodynamic endpoint able to establish the therapeutic equivalence of two salbutamol inhalers: a standard pMDI and a novel DPI [34]. This comparison was carried out in a preselected group of healthy volunteers who showed sGaw reversibility (≥ 30%) in the absence of significant changes in spirometry. The potential use of healthy volunteers is advantageous in reducing the effect of the variability in baseline airway calibre and bronchodilator responsiveness occurring in patients with asthma. On the other hand, the specialized equipment is expensive and necessitates thorough training of volunteers and operators. Whole-body plethysmography may be a useful technique which is more sensitive than spirometry, but it is less readily available and has yet to be considered in regulatory authority guidelines.
Challenge methods – a further option
Airway hyperresponsiveness (AHR) is a key feature of the asthmatic condition which persists during stable asthma and increases during exacerbations due to allergen exposure [35]. The persistence of AHR in patients with stable asthma has encouraged development of nonimmunological bronchoprovocation testing as an objective diagnostic tool [36]. A variety of bronchoconstrictor agents exist, including histamine, methacholine, sodium metabisulphite, adenosine monophosphate (AMP), cold air and hypertonic saline, but histamine and methacholine are the most reliable and widely validated [37]. Compared with normal subjects, the majority of whom show little response to inhaled bronchoconstrictors, patients with asthma demonstrate a reduction in the challenge dose of methacholine required to trigger a fall in FEV1 [35]. This suggests a potentially efficient method for comparisons of inhaled bronchodilator drug efficacy; however, the detailed conditions required for accurate discrimination remain undecided [8, 9, 37, 38].
Bronchoconstrictors
Bronchoconstrictor agents are routinely delivered in doubling doses by tidal breathing via a nebuliser or using a dosimeter device, which may help standardize dosage and improve repeatability of the results [39, 40]. The bronchoconstrictor dose–response curve is used to establish the concentration or dose of the agent required to provoke a 20% fall in FEV1 (PC20 or PD20). These endpoints are more sensitive but less repeatable than FEV1 measurement per se, and it should also be noted that challenge methods may be less acceptable to patients due to the potentially unpleasant bronchoconstrictive effects of the challenge agents. However, this diagnostic methodology provides a useful measure of therapeutic efficacy in clinical trials of anti-inflammatory treatments for asthma, when an increase in methacholine or histamine PC20/PD20 after some weeks of treatment indicates a reduction in AHR [41–44]. Increasingly, it is being applied in acute bronchodilator studies to thereby increase the discriminatory capacity of the bronchodilator or bronchoprotective dose–response curve [15, 22, 37–39, 45–48].
Methacholine challenge
Methacholine or histamine challenge testing is recognized and recommended by regulatory authorities [6, 8]. Methacholine has been used in a number of recent comparative studies in which discriminatory power was demonstrated in relatively small numbers of patients [15, 45, 48]. Parameswaran has demonstrated that for inhaled β2-adrenoceptor agonists in vivo therapeutic equivalence is adequately established, given repeatability and quality control of outcome measurements, by comparing relative potencies of both bronchodilation (requiring 12 subjects for a cumulative dose–response study) and bronchoprotection (requiring 18 subjects in a crossover challenge study) [48, 49]. In common with other methods, the inclusion of more severe patient populations would serve to increase the sensitivity of challenge testing as a measure of bronchodilator efficacy. Although patient admission to challenge tests is largely restricted to exclude those with a poor baseline lung function, the methodology of methacholine testing has been shown to be safe in asthmatic subjects with a mean baseline FEV1 only 45% of predicted [38, 50]. This would in turn ease problems of recruiting sufficient patients with the required degree of PC20 dose-responsiveness to one and two actuations of the reference salbutamol pMDI, as previously requested by regulatory authorities [7, 51].
Exercise challenge
Exercise-induced asthma (EIA) or bronchoconstriction (EIB) offers a more clinically relevant way of measuring AHR which can be applied to assessments of short- and long-acting bronchodilators, generally following a single-dose/separate day regimen [36]. Exercise challenge may be ethically preferable in paediatric studies, but it is less easily reproducible than methacholine or histamine challenge [14]. In common with other indirect provocations, exercise reflects more complex mechanisms of airway obstruction including cellular and neurogenic components [12, 37].
Surrogate measures
Lung deposition in vitro
Pharmacodynamic (PD) measurements of comparative bronchodilator efficacy are based on the relative number of receptors stimulated in the lung, which in turn is linked to the amount of drug reaching the appropriate regions of the lung. This opens up the possibility of using pharmacokinetic (PK) measurements as surrogates for the potentially more expensive and time-consuming PD methods [52, 53]. Lung delivery of inhaled drugs is related to the particle size (fine particle or respirable fraction) which is aerosolized by the particular inhaler device [29, 54]. In vitro measurement of the respirable fraction is limited by the ability of the laboratory apparatus to mimic airway anatomy and in vitro predictions of the fine particle fraction tend to overestimate lung deposition, although some studies have successfully correlated in vitro performance to in vivo efficacy [55, 56]. Current recommendations include the use of multistage impingers or impactors to characterize particle size distribution adequately. Measurements should be made at a range of flow rates and profiles to reflect clinical use of the inhaler device. Development of an apparatus with a more anatomically correct ‘throat’ is also encouraged and such laboratory studies have been conducted [57, 58].
Lung deposition in vivo
Two in vivo methods of assessing lung deposition are currently recognized by the British Association for Lung Research (BALR) – a PK method, using plasma or urine concentrations of drug to represent whole-lung deposition and absorption, and a gamma-scintigraphic method [52, 57]. These methodologies are currently only recognized as ‘supplementary’ by the regulatory authorities. The PK method is valid for drugs which are absorbed but not metabolized locally in the lung and undergo little gastrointestinal absorption. The latter can be blocked by coadministration of charcoal allowing determination of lung deposition, or lung absorption can be distinguished by timing of sample collections [59–62]. Plasma levels of salbutamol over the first 20 min after inhalation represent drug absorbed from the lungs [60]. Urinary salbutamol during the first 30 min, a noninvasive and simpler method, also represents drug absorbed from the lungs and indicates relative lung bioavailability when expressed as a proportion of the 24 h urinary excretion [62, 63]. A clear linear dose–response relationship has been demonstrated over 1–5 actuations of salbutamol (100 µg) pMDI [64]. This method assumes that the amount of drug absorbed rapidly is a good measure of the amount available for bronchodilation. Although the optimal particle size distribution for lung absorption may not be the same as for bronchodilation, PK parameters have been correlated with the methacholine challenge response in patients [65]. The chief disadvantage of the PK methods is that they lack the ability to detect differences in regional lung deposition; however, a relationship has been shown between lung deposition pattern and PK profile [66]. The other in vivo BALR-recommended method, gamma-scintigraphy, can assess differences in regional deposition and lung penetration (following inhalation of 99mTc-radiolabelled drug) by scanning the chest with a gamma camera to produce a computerized image. This method is the subject of a further review in the current series [67].
Clinical trials
Most comparisons of efficacy of inhaled bronchodilator products are concentrated on acute drug effects but formal, long-term (4- and 12-week) clinical efficacy and safety studies of short- and long-acting β2-adrenoceptor agonists are still required by regulatory authorities [8]. In addition to subjective and clinical assessments of asthma severity, the usual objective efficacy measurements are daily (morning and evening) PEF, and periodic spirometry (FEV1) performed prior to morning inhalation of the trial drug, comparator or placebo. A recent 12 week clinical trial has employed maximum percentage changes from baseline in FEV1, area under the curve and duration of bronchodilation, measured for 6 h following the first daily dose at 4 weekly intervals, to demonstrate therapeutic equivalence of a new salbutamol DPI with the standard salbutamol pMDI [68]. Such periodic assessment of spirometric changes can be weakened by baseline variation since, for example, an increase in baseline FEV1 during treatment produces an apparent diminution of bronchodilator effect [68]. Periodic tests for bronchodilator responsiveness must also take into account the duration of action of the trial drugs in order to obviate residual bronchodilatation carried over from the previous dose [69]. The potential for development of tolerance to β2-adrenoceptor agonists is another important consideration in longer-term clinical studies, especially where regular dosing is involved [70]. Assessment of protection against bronchoconstrictors is a more sensitive measure of β2-adrenoceptor agonist activity than bronchodilatation alone, therefore changes in acute β2-adrenoceptor agonist responsiveness during the clinical use of trial bronchodilators can be more precisely monitored by means of methacholine or histamine PC20 measurements [69, 70].
Conclusion
The burgeoning selection of new bronchodilator/inhaler devices calls for definitive standards by which to establish efficacy and safety relative to the current major treatment, the CFC-driven salbutamol pMDI. These standards are being set under regulatory authority guidance with a view to phasing out the environmentally unsafe CFC-pMDIs as soon as acceptable alternative treatments are established [7, 8, 71]. Cost and convenience are major considerations, which encourage the development of study designs reliant on short-term PD and surrogate measures rather than lengthy therapeutic studies. These alternative methodologies are often more sensitive, and therefore more discriminating for comparison of novel products and generics with the standard treatment. FEV1 remains the key measurement, and the provision of adequate bronchodilator dose–response curves can be improved by selecting more severely asthmatic volunteers, or by employing this spirometric endpoint within a methacholine or histamine challenge study design.
Development of more sensitive, repeatable, and therefore more discriminating, methodologies will enable a clinically valid ranking of new therapies, new delivery systems and traditional inhaled asthma treatments.
References
- 1.National Asthma Education and Prevention Program. Bethesda, MD: NIH; Expert Panel Report 2: Guidelines for the diagnosis and management of asthma; pp. 245–253. Publication number 97–4051. [Google Scholar]
- 2.British Thoracic Society guidelines on asthma management. Thorax. 1997;52(Suppl 1):S1–S21. [Google Scholar]
- 3.The International Study of Asthma and Allergies in Childhood (ISAAC) Steering Committee. Worldwide variation in prevalence of symptoms of asthma, allergic rhinoconjunctivitis, and atopic eczema: ISAAC. Lancet. 1998;351:1225–1232. [PubMed] [Google Scholar]
- 4.The Montreal Protocol. The Montreal protocol on substances that deplete the ozone layer. Final Act (Nairobi: UNEP 1987) Fed Reg. 1994;59:56276–56298. [Google Scholar]
- 5.Bell J. International development of portable inhalers. J Invest Allergol Clin Immunol. 1997;7:417–419. [PubMed] [Google Scholar]
- 6.Adams WP, Poochikian G, Taylor AS, Patel RM, Burke GP, Williams RL. Regulatory aspects of modifications to innovator bronchodilator metered dose inhalers and development of generic substitutes. J Aerosol Med. 1994;7:119–134. doi: 10.1089/jam.1994.7.119. [DOI] [PubMed] [Google Scholar]
- 7.Division of Bioequivalence. Interim guidance for documentation of in vivo bioequivalence of albuterol inhalation aerosols (metered dose inhalers) Rockville, MD: FDA; 1994. Office of Generic Drugs, Center for Drug Evaluation and Research, Food and Drug Administration. [Google Scholar]
- 8.Points to consider: clinical development programmes for MDI and DPI drug products. Rockville, MD: FDA; 1994. Division of Oncology and Pulmonary Drug Products, Center for Drug Evaluation and Research, Food and Drug Administration. [Google Scholar]
- 9.Standardised Lung Function Testing. Official Statement of the European Respiratory Society. Eur Respir J. 1993;6(Suppl 16) [PubMed] [Google Scholar]
- 10.American Thoracic Society. Standardisation of spirometry – update. Am Rev Respir Dis. 1987;136:1285–1298. doi: 10.1164/ajrccm/136.5.1285. [DOI] [PubMed] [Google Scholar]
- 11.Newhouse MT, Nantel NP, Chambers CB, Pratt B, Parry-Billings M. Clickhaler (a novel dry powder inhaler) provides similar bronchodilation to pressurized metered-dose inhaler, even at low rates. Chest. 1999;115:952–956. doi: 10.1378/chest.115.4.952. [DOI] [PubMed] [Google Scholar]
- 12.Oseid S, Mellbye E, Hem E. Effect of nedocromil sodium on exercise-induced bronchoconstriction exacerbated by inhalation of cold air. Scand J Med Sci Sports. 1995;5:88–93. doi: 10.1111/j.1600-0838.1995.tb00018.x. [DOI] [PubMed] [Google Scholar]
- 13.Borgström L, Derom E, Ståhl E, Wåhlin-Boll E, Pauwels R. The inhalation device influences lung deposition and bronchodilating effect of terbutaline. Am J Respir Crit Care Med. 1996;153:1636–1640. doi: 10.1164/ajrccm.153.5.8630614. [DOI] [PubMed] [Google Scholar]
- 14.Ahrens RC. On comparing inhaled beta adrenergic agonists. Ann Allergy. 1991;67:296–298. [PubMed] [Google Scholar]
- 15.Ahrens RC, Hendeles L, Clarke WR, et al. Therapeutic equivalence of Spiros dry powder inhaler and Ventolin metered dose inhaler: a bioassay using methacholine. Am J Respir Crit Care Med. 1999;160:1238–1243. doi: 10.1164/ajrccm.160.4.9806101. [DOI] [PubMed] [Google Scholar]
- 16.Silvasti M, Laurikainen K, Nieminen M, et al. Single dose comparison between a novel multiple dose powder inhaler and a conventional metered dose inhaler in asthmatic patients. Acta Ther. 1993;19:125–134. [Google Scholar]
- 17.Bennett JA, Tattersfield AE. Time course and relative dose potency of systemic effects from salmeterol and salbutamol in healthy subjects. Thorax. 1997;52:458–464. doi: 10.1136/thx.52.5.458. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Bennis J, Svedmyr N. A controlled comparison of salbutamol and terbutaline by IPPV in asthmatic patients: a dose–response study. Scand J Respir Dis. 1977;101(Suppl):113–117. [PubMed] [Google Scholar]
- 19.Finney DJ. Statistical methods in biological assay. 3. London: Charles Griffin and Co Ltd; 1978. [Google Scholar]
- 20.Singh GJP. Examples of assessment on the response scale and dose scale. 1996. FDA Advisory Committee for Pharmaceutical Science and Pulmonary-Allergy Drugs Advisory Committee Meeting, August 16.
- 21.Löfdahl C-G, Andersson L, Bondesson E, et al. Differences in bronchodilating potency of salbutamol in Turbuhaler® as compared with a pressurized metered-dose inhaler formulation in patients with reversible airway obstruction. Eur Respir J. 1997;10:2474–2478. doi: 10.1183/09031936.97.10112474. [DOI] [PubMed] [Google Scholar]
- 22.Blake KV, Hoppe M, Harman E, Hendeles L. Relative amount of albuterol delivered to lung receptors from a metered-dose inhaler and nebulizer solution. Bioassay by histamine bronchoprovocation. Chest. 1992;101:309–315. doi: 10.1378/chest.101.2.309. [DOI] [PubMed] [Google Scholar]
- 23.Britton J, Tattersfield A. Comparison of cumulative and non-cumulative techniques to measure dose–response curves for beta agonists in patients with asthma. Thorax. 1984;39:597–599. doi: 10.1136/thx.39.8.597. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Morice AH, Peake MD, Allen MB, Campbell JC, Parry-Billings M. Reproducibility of bronchodilator response for a reservoir dry powder inhaler following routine clinical use. J Asthma. 2000;37:81–87. doi: 10.3109/02770900009055431. [DOI] [PubMed] [Google Scholar]
- 25.Morley TF, Marmozsan E, Zappasodi SJ, Gordon R, Griesback R, Giudice JC. Comparison of beta-adrenergic agents delivered by nebulizer vs metered dose inhaler with InspirEase in hospitalized asthmatic patients. Chest. 1988;94:1205–1210. doi: 10.1378/chest.94.6.1205. [DOI] [PubMed] [Google Scholar]
- 26.Hess D, Horney D, Snyder T. Medication-delivery performance of eight small-volume, hand-held nebulizers: effects of diluent volume, gas flowrate, and nebulizer model. Respir Care. 1989;34:717–723. [Google Scholar]
- 27.Zainudin BMZ, Biddiscombe M, Tolfree SEJ, Short M, Spiro SG. Comparison of bronchodilator responses and deposition patterns of salbutamol inhaled from a pressurised metered dose inhaler, as a dry powder, and as a nebulised solution. Thorax. 1990;45:469–473. doi: 10.1136/thx.45.6.469. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Youngchaiyud P, Maranetra N, Nana A, et al. Good bronchodilating effect of a dry powder inhaler in the treatment of acute asthma. Eur Respir J. 1996;9(Suppl 23):54s. [Google Scholar]
- 29.Bisgaard H. Drug delivery from inhaler devices. Lung deposition, clinical effect, and cost-effectiveness vary. Br Med J. 1996;313:895–896. doi: 10.1136/bmj.313.7062.895. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Geoffroy P, Lalonde RL, Ahrens R, Clarke W, Hill MR, Vaughan LM, Grossman J. Clinical comparability of albuterol delivered by the breath-actuated inhaler (Spiros®) and albuterol by MDI in patients with asthma. Ann Allergy Asthma Immunol. 1999;82:377–382. doi: 10.1016/S1081-1206(10)63287-5. [DOI] [PubMed] [Google Scholar]
- 31.Marshik PL, Kifle Y, Hendeles L. Response to inhaled beta-agonists during nocturnal attacks of asthma: a bioassay for relative potency. Am J Respir Crit Care Med. 1997;155:A672. [Google Scholar]
- 32.Hendeles L, Ahrens R, Clarke W, et al. Dose–response of inhaled albuterol during nocturnal bronchospasm. J Allergy Clin Immunol. 1998;101:S10. doi: 10.1016/j.jaci.2004.03.046. [DOI] [PubMed] [Google Scholar]
- 33.Skinner C, Palmer KNV. Changes in specific airways conductance and forced expiratory volume in one second after a bronchodilator in normal subjects and patients with airways obstruction. Thorax. 1974;29:574–577. doi: 10.1136/thx.29.5.574. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Morice AH, Waterhouse JC, Peers EM, Parry-Billings M. Use of whole-body plethysmography to compare bronchodilator efficacy. Respiration. 1998;65:120–124. doi: 10.1159/000029242. [DOI] [PubMed] [Google Scholar]
- 35.Lötvall J, Inman M, O'Byrne P. Measurement of airway hyperresponsiveness: new considerations. Thorax. 1998;53:419–424. doi: 10.1136/thx.53.5.419. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Britton J. Symptoms and objective measures to define the asthma phenotype. Clin Exp Allergy. 1998;28(Suppl 1):2–7. doi: 10.1046/j.1365-2222.1998.0280s1002.x. [DOI] [PubMed] [Google Scholar]
- 37.Sterk PJ, Fabbri LM, Quanjer Ph H, et al. Official statement of the European Respiratory Society: report working party: standardization of lung function tests; European Community for steel and coal. Airway responsiveness: standardized challenge testing with pharmacological, physical and sensitizing stimuli in adults. Eur Respir J. 1993;6(Suppl 16):53–83. [PubMed] [Google Scholar]
- 38.Banik AN, Holgate ST. Problems and progress in measuring methacholine bronchial reactivity. Clin Exp Allergy. 1998;28(Suppl 1):15–19. doi: 10.1046/j.1365-2222.1998.0280s1015.x. [DOI] [PubMed] [Google Scholar]
- 39.Egbagbe E, Pavord ID, Wilding P, Thompson-Coon J, Tattersfield AE. Adenosine monophosphate and histamine induced bronchoconstriction: repeatability and protection by terbutaline. Thorax. 1997;52:239–243. doi: 10.1136/thx.52.3.239. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Cockcroft DW, Killian DN, Mellon JJA, Hargreave FE. Bronchial reactivity to inhaled histamine: a method and clinical survey. Clin Allergy. 1977;7:235–243. doi: 10.1111/j.1365-2222.1977.tb01448.x. [DOI] [PubMed] [Google Scholar]
- 41.Juniper EF, Frith PA, Hargreave FE. Airway responsiveness to histamine and methacholine: relationship of minimum treatment to control symptoms of asthma. Thorax. 1981;36:575–579. doi: 10.1136/thx.36.8.575. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Bootsma GP, Dekhuijzen PNR, Festen J, Mulder PGH, van Herwaarden CLA. Comparison of fluticasone propionate and beclomethasone dipropionate on direct and indirect measurements of bronchial hyperresponsiveness in patients with stable asthma. Thorax. 1995;50:1044–1050. doi: 10.1136/thx.50.10.1044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Kraan J, Koëter GH, Sluiter HJ, de Vries K. Changes in bronchial hyperreactivity induced by 4 weeks of treatment with antiasthmatic drugs in patients with allergic asthma: a comparison between budesonide and terbutaline. J Allergy Clin Immunol. 1985;76:628–636. doi: 10.1016/0091-6749(85)90786-9. [DOI] [PubMed] [Google Scholar]
- 44.Morice AH, Andrews B, Taylor M. Comparison of the effect on bronchial hyperresponsiveness of beclomethasone dipropionate administered via a novel multidose dry powder inhaler or a conventional pressurised metered dose inhaler. Respiration. 2000;67:298–305. doi: 10.1159/000029514. [DOI] [PubMed] [Google Scholar]
- 45.Newhouse MT, Pratt B, Clelland C, Tran C, Patel P. Protection against methacholine challenge: a method of establishing the dose response relationship of salbutamol delivered via dry powder inhaler (DPI) and metered dose inhaler (MDI) Thorax. 1999;54(Suppl 3):A30(P48). [Google Scholar]
- 46.Ahrens RC, Harris JB, Milavetz G, Annis L, Ries R. Use of bronchial provocation with histamine to compare the pharmacodynamics of inhaled albuterol and metaproterenol in patients with asthma. J Allergy Clin Immunol. 1987;79:876–882. doi: 10.1016/0091-6749(87)90235-1. [DOI] [PubMed] [Google Scholar]
- 47.Wong AG, O'Byrne PM, Lindbladh C, Inman MD, Ståhl E, Hargreave FE. Dose–response protective effect of salbutamol on methacholine airway responsiveness using pressurized metered dose inhalers and Turbuhalers. Can Respir J. 1998;5:119–123. doi: 10.1155/1998/865042. [DOI] [PubMed] [Google Scholar]
- 48.Parameswaran KN, Inman MD, Ekholm BP, et al. Protection against methacholine bronchoconstriction to assess relative potency of inhaled β2-agonist. Am J Respir Crit Care Med. 1999;160:354–357. doi: 10.1164/ajrccm.160.1.9812035. [DOI] [PubMed] [Google Scholar]
- 49.Parameswaran K. Concepts of establishing clinical bioequivalence of chlorofluorocarbon and hydrofluoroalkane β-agonists. J Allergy Clin Immunol. 1999;104:S243–S245. doi: 10.1016/s0091-6749(99)70039-4. [DOI] [PubMed] [Google Scholar]
- 50.Martin RJ. Methacholine challenge testing: safety of low starting FEV1; Asthma clinical research network (ACRN) Chest. 1997;112:53–56. doi: 10.1378/chest.112.1.53. [DOI] [PubMed] [Google Scholar]
- 51.Approved drug products with therapeutic equivalence evaluations. 13. Washington DC: Food and Drug Administration US Dept of Health and Human Services, US Govt Printing Office; 1993. pp. vii–viii. [Google Scholar]
- 52.Snell NJC. Equivalence testing – the special case of inhaled medications. Int J Pharm Med. 1998;12:245–246. [Google Scholar]
- 53.Webb J. Respir Drug Delivery VI. Vol. 6. Interpharm Press; 1998. Avoidance of Regulatory Delays for Inhalation Products – Alternative Approaches to Product Development and Regulatory Approval; pp. 457–459. Buffalo Grove. [Google Scholar]
- 54.Pauwels R, Newman S, Borgström L. Airway deposition and airway effects of antiasthma drugs delivered from metered-dose inhalers. Eur Respir J. 1997;10:2127–2138. doi: 10.1183/09031936.97.10092127. [DOI] [PubMed] [Google Scholar]
- 55.Zanen P, Go TL, Lammers J-WJ. The optimal particle size for β-adrenergic aerosols in mild asthmatics. Int J Pharm. 1994;107:211–217. [Google Scholar]
- 56.Padfield JM, Winterborn IK, Pover GM, Tattersfield A. Correlation between inertial impactor performance and clinical performance of a bronchodilator aerosol. J Pharm Pharmacol. 1983;35(Suppl):10P. [Google Scholar]
- 57.Snell NJC, Ganderton D. Respir Med. Vol. 93. 1999. Assessing lung deposition of inhaled medications. Consensus Statement from a Workshop of the British Association for Lung Research; pp. 123–133. [DOI] [PubMed] [Google Scholar]
- 58.Olsson B. Can in vitro measurements predict the lung deposition of dry-powders. J Aerosol Med. 1997;10:S13–S16. [Google Scholar]
- 59.Lipworth BJ. Pharmacokinetics of inhaled drugs. Br J Clin Pharmacol. 1996;42:697–705. doi: 10.1046/j.1365-2125.1996.00493.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Clark DJ, Lipworth BJ. Lung bioavailability of chlorofluorocarbon free, dry powder and chlorofluorocarbon containing formulations of salbutamol. Br J Clin Pharmacol. 1996;41:247–249. doi: 10.1111/j.1365-2125.1996.tb00191.x. [DOI] [PubMed] [Google Scholar]
- 61.Clark DJ, Gordon-Smith J, McPhate G, Clark G, Lipworth BJ. Lung bioavailability of generic and innovator salbutamol metered dose inhalers. Thorax. 1996;51:325–326. doi: 10.1136/thx.51.3.325. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Hindle M, Chrystyn H. Determination of the relative bioavailability of salbutamol to the lung following inhalation. Br J Clin Pharmacol. 1992;34:311–315. doi: 10.1111/j.1365-2125.1992.tb05921.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Hindle M, Peers EM, Parry-Billings M, Chrystyn H. Relative bioavailability of salbutamol to the lung following inhalation via a novel dry powder inhaler and a standard metered dose inhaler. Br J Clin Pharmacol. 1997;43:336–338. doi: 10.1046/j.1365-2125.1997.00564.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Tomlinson HS, Corlett SA, Chrystyn H. Effect of dose on the relative lung bioavailability of salbutamol. J Aerosol Med. 1995;8:127. [Google Scholar]
- 65.Chrystyn H, Allen MB, Corlett SA, Tomlinson HS. Simultaneous measurement of pharmacodynamic and pharmacokinetic parameters which can be used to evaluate the equivalence of inhaled salbutamol. Am J Respir Crit Care Med. 1998;157:A636. [Google Scholar]
- 66.Laube BL, Benedict GW, Dobs AS. Time to peak insulin level, relative bioavailability, and effect of site of deposition of nebulised insulin in patients with non-insulin-dependent diabetes mellitus. J Aerosol Med. 1998;11:153–173. doi: 10.1089/jam.1998.11.153. [DOI] [PubMed] [Google Scholar]
- 67.Warren S. The measurement of pulmonary drug delivery using gamma scintigraphy. Br J Clin Pharmacol. 2001 in press. [Google Scholar]
- 68.Nelson H, Kemp JP, Bieler S, Vaughan LM, Hill MR. Comparative efficacy and safety of albuterol sulfate Spiros inhaler and albuterol metered-dose inhaler in asthma. Chest. 1999;115:329–335. doi: 10.1378/chest.115.2.329. [DOI] [PubMed] [Google Scholar]
- 69.Tattersfield AE. Long-acting β2-agonists. Clin Exp Allergy. 1992;22:600–605. doi: 10.1111/j.1365-2222.1992.tb00175.x. [DOI] [PubMed] [Google Scholar]
- 70.Tattersfield AE. Tolerance to β-agonists. Bull Eur Physiopathol Respir. 1985;21:1S–5S. [PubMed] [Google Scholar]
- 71.Food and Drug Administration. Chlorofluorocarbon propellants in self-pressurized containers; determinations that uses are no longer essential; request for comments. Federal Register. 1997;62:10242–10246. [Google Scholar]

