Abstract
Human temporomandibular joint loading causes pressurization and flow of interstitial fluid in its cartilaginous structures. This largely determines its load-bearing and maintenance capacity. It was hypothesized that during cyclical jaw movements normal pressure distribution dynamics would enable fluid to reach all necessary cartilage regions. This was tested qualitatively by analysis of local volumetric strain dynamics during jaw open–close movements predicted by a dynamic model of the human masticatory system. Finite-element analysis was performed in separate regions of the articular cartilage layers and articular disc. Heterogeneous patterns of dilatation and compression were predicted. Compression was found to be more dominant during jaw closing than opening. The pressure gradient in the superior layer of the articular disc was more mediolaterally orientated than in its inferior layer. The findings suggest that, where necessary, regionally the cartilage can imbibe fluid to protect the subchondral bone from impact loads effectively. In the disc itself presumably all areas receive regular refreshment of interstitial fluid.
Keywords: cartilage, finite-element modelling, jaw joint, strain
Introduction
During jaw movements the human temporomandibular joint is loaded. These loads lead to local deformations in its cartilaginous structures. As both articular cartilage (Mow et al. 1984) and the cartilaginous articular disc (Tanaka & van Eijden, 2003) are not completely incompressible, this causes local volumetric changes.
Cartilage can be considered to be a mixture of solid and fluid constituents (Shrive & Frank, 1998; Wong & Carter, 2003). Local differences in compression or dilatation can cause pressure gradients of the fluid, both within the cartilage, and between the cartilage and the joint cavity. These pressure gradients cause the incompressible fluid to flow through the cartilage matrix and between the cartilage and the joint cavity. This flow supports the exchange of nutrients and waste products within the tissue and their transport into and out of it (Mow et al. 1984). Furthermore, it has been suggested to be a major determinant for chondrocyte biosynthesis (Kim et al. 1995). Consequently, fluid flow is suggested to play a role in the maintenance of the cartilaginous structures of the temporomandibular joint.
Fluid is also a determinant for the mechanical properties of cartilage. Upon loading it pressurizes, making the tissue relatively incompressible. A part of the pressurized fluid moves away with prolonged loading, thereby transferring load to the compressible cartilage matrix. At high strain rates (beyond 0.15% s−1) more than 80% of the load is carried by fluid pressurization, whereas at low strain rates (below 0.005% s−1) the matrix bears more than 80% of it (Li & Herzog, 2004). The cartilage matrix can be treated as hyperelastic, and its stiffness increases with compression (Hasler et al. 1999). Consequently, there is a delicate balance between the relative amount of present fluid and the apparent stiffness of the cartilage.
The synovial fluid also serves as a lubricant between the cartilaginous structures (Tanaka & van Eijden, 2003). When it is pressurized in contacting layers, it carries most of the force down to the subchondral bone through hydrostatic pressure. Roughness of the cartilaginous surfaces will thus hardly contribute to friction between the two (Shrive & Frank, 1998).
For proper maintenance of cartilage it is necessary that nutrients reach every part of it regularly. If nutrients are supplied through fluid flow, the distribution of hydrostatic pressure must change accordingly. The fluid that has been squeezed out of temporarily pressurized locations may return (with fresh nutrients) if the pressure is then released. In structures such as the intervertebral disc a diurnal loading pattern allows for fluid exudation during the day and reimbibition during the night (Li & Herzog, 2004). As the masticatory system remains active during the night (Gallo et al. 1999), the applicability of such a diurnal mechanism for adequate fluid exchange in the cartilaginous structures in the temporomandibular joint is questionable. By contrast, the dynamics of hydrostatic pressure during jaw movement may serve as a mechanism to perform and maintain this function. Furthermore, the loaded region in the temporomandibular joint has been demonstrated to shift continuously during jaw movement (Gallo et al. 2000). This could serve as a mechanism to redistribute the imbibed fluid within the cartilaginous structures.
Pressure and flow of fluid in the cartilaginous structures of the temporomandibular joint during jaw movements cannot be measured. However, they can be predicted by mathematical modelling. To that end a dynamic model of the human masticatory system was developed that included the cartilaginous structures of the temporomandibular joint as finite-element (FE) models (Koolstra & van Eijden, 2005). The transient distribution of dilatation and compression in these cartilaginous structures was analysed during jaw open–close movements. The results were expected to give an indication as to whether this relatively simple movement pattern could contribute to maintain these structures. It was hypothesized that, in particular in the disc, the pressure distribution would be subject to such alterations that it could give rise to flow of fluid throughout the whole disc.
Materials and methods
The model
A biomechanical model of the human masticatory system (Fig. 1A) was constructed using MADYMO 6.1 (TNO Automotive, The Netherlands), a simulation program that combines the capabilities of multibody motion and FE modelling. It contained the skull and the mandible, which were modelled as rigid bodies. They articulated at two six-degrees-of-freedom temporomandibular joints. Twelve pairs of Hill-type muscles were able to move the mandible with respect to the skull (Koolstra & van Eijden, 2005). Their attachments, maximum forces, fibre lengths and sarcomere lengths (Table 1) had been obtained from eight human cadavers (van Eijden et al. 1995, 1996, 1997). The contractile characteristics had been shaped according to van Ruijven & Weijs (1990).
Table 1.
Moments of inertia | Mass (kg) | Ixx (kg m2) | Iyy (kg m2) | Izz (kg m2) |
---|---|---|---|---|
Lower jaw | 0.44 | 0.00086 | 0.00029 | 0.00061 |
Muscle | Muscle length (mm) | Max. force (N) | CE, optimum length (mm) | SE, length (mm) |
---|---|---|---|---|
Superficial masseter | 48.0 | 272.8 | 22.6 | 25.8 |
Deep anterior masseter | 29.5 | 73.8 | 21.8 | 17.1 |
Deep posterior masseter | 30.9 | 65.8 | 15.0 | 13.3 |
Anterior temporalis | 57.4 | 308.0 | 30.7 | 24.2 |
Posterior temporalis | 62.9 | 222.0 | 31.3 | 28.8 |
Medial pterygoid | 43.3 | 240.0 | 14.1 | 27.6 |
Superior lateral pterygoid | 29.1 | 38.0 | 21.5 | 9.4 |
Inferior lateral pterygoid | 27.2 | 112.8 | 22.3 | 9.0 |
Anterior digastric | 51.9 | 46.4 | 42.6 | 3.0 |
Geniohyoid | 48.5 | 38.8 | 35.3 | 5.4 |
Anterior mylohyoid | 21.8 | 63.6 | 24.0 | 0.0 |
Posterior mylohyoid | 44.8 | 21.2 | 39.7 | 0.0 |
CE, contractile element; SE, series elastic element.
No. of finite elements | Temporal cartilage | Articular disc | Condylar cartilage |
---|---|---|---|
right joint | 14 878 | 12 516 | 12 123 |
left joint | 14 166 | 12 660 | 12 233 |
Mooney-Rivlin constants | C1 | C2 |
---|---|---|
articular disc | 9.0 × 105 | 9.0 × 102 |
temporal and condylar cartilage | 4.5 × 105 | 4.5 × 102 |
Both temporomandibular joints contained two deformable articular cartilage layers of 0.5 mm (Hansson et al. 1977) connected to the temporal bone and the mandibular condyle, respectively. As the influence of the stiffness of the bony articular surfaces on the tension and deformation of the overlying articular cartilage layers can be considered to be negligible (Dar & Aspden, 2003) they were assumed to be rigid. Between the cartilaginous articular layers was a freely movable deformable cartilaginous articular disc (Fig. 1B). This was medially and laterally connected to the condylar neck with pairs of inextensible wires representing the lower part of the articular capsule and anteriorly to the attachment of the superior lateral pterygoid muscle to the condylar neck. In the present model the temporomandibular ligaments were not included as they have been demonstrated to be of negligible influence in symmetric jaw movements (Koolstra & van Eijden, 1997; Koolstra et al. 2001). Contact between the deformable cartilaginous structures was assumed to be frictionless. The joint reaction force was computed with a penalty-based contact algorithm. The geometry of the deformable structures had been obtained from the right temporomandibular joint of one cadaver (Beek et al. 2000,2001a). The left side joint was constructed as a mirror image. The volumes of the deformable structures were divided into tetrahedral finite elements with edges of maximally 0.25 mm (HyperMesh 6.0, Altair Engineering GmbH, Böblingen, Germany) as specified in Table 1.
In order to quantify the distribution of hydrostatic pressure in the temporal cartilage, the condylar cartilage and the articular disc, these structures were divided into, respectively, 18, 18 and 27 portions. In the articular disc an anterior, intermediate and posterior region were defined concomitant with its anterior band, intermediate zone and posterior band, respectively. They were subdivided into equally wide lateral, central and medial regions, excluding the lateralmost and medialmost 5% of the discs (Fig. 1D). Each region was subdivided into an upper, middle and lower layer of equal thickness. The temporal cartilage and the condylar cartilage were similarly subdivided (Fig. 1C, E). The regional boundaries coincided with those of the articular disc when the mouth was opened (Fig. 1B). Each region was subdivided in a superficial and deep layer of equal thickness, the latter layer being connected to the subchondral bone.
As the present analysis focused on the response within the relatively short time of an open–close cycle, the material properties of the cartilaginous structures were approximated according to the Mooney–Rivlin material model (Chen et al. 1998; Koolstra & van Eijden, 2005). Its constants (Table 1) were approximated from experiments on human temporomandibular joint discs according to Beek et al. (2003).
Simulations
The jaw was closed at the start of all simulations. From this position symmetrical jaw-open movements were simulated by a 100% activation of the digastric, geniohyoid, mylohyoid and lateral pterygoid muscles simultaneously. Thereafter, the jaw was closed by a 25% simultaneous activation of the masseter, medial pterygoid and temporalis muscles. Finally, a second jaw-open movement was performed. Activation of each muscle was defined as a function of time. These functions included activation and deactivation ramps of 45 and 75 ms, respectively, to incorporate activation dynamics (Winters & Stark, 1987). During jaw closing a force of 50 N was applied between the central incisors to simulate food resistance. This force is about 10 times the yield force of cheese (Gavião et al. 2004).
The results were analysed and visualized using HyperView 7.0 (Altair Engineering GmbH, Böblingen, Germany). Changes in volume in each of the elements in the cartilaginous structures on the right side were monitored as an indication of hydrostatic pressure. The total change in volume of each region served as a measure for volumetric strain.
Results
The muscle forces as a result of the applied muscle activation pattern, the force–length and force–velocity relationships, and the passive tensions are depicted in Fig. 2. These forces generated relatively fast jaw movements, but in a pathway similar to that of natural movements. The mandibular condyles moved forward along the articular eminence when the jaw opened to about 3 cm interincisal distance, and moved backward when the jaw closed. The articular disc followed the condylar displacements.
The predicted compressive volumetric strain in the temporal and condylar cartilage reached about 4 and 5.5%, respectively. In the articular disc it remained below 1.5% (Fig. 3). The largest compression was predicted when the teeth contacted each other after the jaw had been closed. In the temporal cartilage the compressed areas alternated between the posterior region (mandibular fossa) and the intermediate region (articular eminence) when the jaw was closed and opened, respectively. In the condylar cartilage compression was concentrated in the central area and was greater as the jaw was closed than when it was open. In both the temporal and the condylar cartilage compression in the surface layer was slightly less than in the deep layer. The medial regions of the temporal and condylar cartilage were relaxed when the jaw was open. They became compressed in the later phase of jaw closing. Their central and lateral regions remained compressed at all times. Consequently, the pressurized area shifted from lateral when the jaw was open to more medial during jaw closing (Fig. 3A, C).
Dilatation was predicted only in the articular disc (Fig. 3B). This occurred predominantly in the superior layer of its central part, where simultaneously the middle and inferior layers were compressed. The changes in volumetric strain were relatively small in the anterior band of the disc. In the posterior band, which was compressed during jaw closing and relaxed during jaw opening, they were far more variable. In the lateral portion of the intermediate zone compression decreased upon jaw closing in contrast to its medial portion.
The differences in volumetric strain of the temporal cartilage and the condylar cartilage are reflected in the difference in the predicted distribution of hydrostatic pressure between the upper and lower side of the articular disc (Fig. 4). Superiorly, large portions were under a considerable dilatational pressure. Inferiorly, these areas were more sparse. Superiorly, the pressurization had a strong mediolateral heterogeneity as dilatation on the medial side was accompanied by compression on the lateral side. This mediolateral heterogeneity was not present inferiorly where the heterogeneity had a more antero-posterior orientation.
Discussion
The present results suggest that during jaw movement there is a considerable heterogeneity in volumetric strain within and between the cartilaginous structures of the temporomandibular joint. Volumetric strain causes hydrostatic pressure of the interstitial fluid and pressure gradients lead to fluid flow. Although the direct relationship between volumetric strain and hydrostatic pressure is temporary as pressure equilibrates with time as a consequence of fluid flow, the predicted volumetric strain can be considered to be indicative for hydrostatic pressure and thereby for mechanical properties such as stiffness and friction (Shrive & Frank, 1998; Soltz & Ateshian, 1998; Tanaka & van Eijden, 2003; Wong & Carter, 2003). The strain gradients can be considered to be indicative for fluid flow potential and thus for the maintenance capacity of the cartilage (Maroudas, 1976; Mow et al. 1984; Kim et al. 1995).
Although the present model has some limitations, to our knowledge it is the first to enable dynamic analysis of volumetric strain in the temporomandibular joint during jaw movements controlled by muscle activation. The large deformations of the cartilaginous structures of the joint (Beek et al. 2001b) were successfully modelled with the Mooney–Rivlin material model. This material model treats the cartilage as nearly incompressible. This assumption is considered to be acceptable for the relatively fast movements in normal activity (Kim et al. 1995; Shrive & Frank, 1998). It has the consequence, however, that the predicted volumetric strains were relatively small. For loads of longer duration cartilage displays creep. This viscoelastic property is caused by exudation and imbibition of interstitial fluid. Therefore, for slowly changing loads, for instance during clenching, the material does no longer behaves as nearly incompressible, and the applied Mooney–Rivlin material model would be invalid. The modelled resistance to tension is not very different from the resistance to compression while experimentally obtained tensile and compressive material properties can differ by orders of magnitude, for instance as a result of fibre reinforcement (Tanaka & van Eijden, 2003). The predicted strain in the dilated regions is therefore presumably overestimated with respect to the regions under compression. The applied material models disregarded the viscoelastic properties of cartilage. Although various material models have been proposed to implement these characteristics for the articular disc of the temporomandibular joint (Beek et al. 2003; Donzelli et al. 2004; Allen & Athanasiou, 2006; Pérez del Palomar & Doblaré, 2006), each has different characteristics. Furthermore, none of them was applicable in the present dynamic model. The consequences of disregarding viscoelastic behaviour in the present analysis are discussed in the Appendix. They can be considered to be limited. For instance, the viscoelastic creep in the disc is performed with a much longer (up to 45 s) time constant (Tanaka & van Eijden, 2003) than the cycle time of jaw movements, and is presumably limited by surface consolidation (Wong & Carter, 2003). In a qualitative analysis, it has been demonstrated that the applied simplifications do not produce unrealistic results (Koolstra & van Eijden, 2005). Yet, for a quantitative analysis they have to be interpreted with caution. Although fluid flow, one of the major determinants for viscoelastic behaviour, itself was not predicted, its prerequisites are indicated by the predicted strains. The consequence of actual fluid flow is that the hydrostatic pressure gradients generated as a consequence of volumetric strain differences equilibrate with time.
As the cartilaginous structures in the temporomandibular joint have a complex shape and undergo large deformations during functioning, they have been modelled using tetrahedral finite elements. The accuracy of these elements is less sensitive to element shape distortion than, for instance, so-called brick-elements. If such distortions should occur this would lead to numerical instabilities, and the simulation would have to be aborted. Convergence tests have demonstrated that an element size of 0.5 mm was adequate for this model (Koolstra & van Eijden, 2005). The present resolution was 0.25 mm to accommodate a better definition in the predicted stress and strain distributions.
In the articular cartilage layers the predicted volumetric strains were almost exclusively compressive. In the articular disc both compression and dilatation were predicted. Dilatation occurred in its anterior region, and more posteriorly only in the superior layer while the middle and inferior layers remained compressed (Fig. 3B). In the anterior portions the strain is most probably caused by pull of the superior lateral pterygoid. The difference between the superior and the other layers can be attributed to the incongruent joint contact. The articular tubercle makes a smaller and deeper impression in the superior layer of the disc than the mandibular condyle makes in its inferior layer, thereby leaving more room for dilatation. Furthermore, the disc bends as it is pulled taut over the mandibular condyle. These predictions, however, do not indicate necessarily that the articular disc exudes less fluid in the upper joint cavity than in the lower one. Owing to the viscoelastic character of the disc (Donzelli et al. 2004; Allen & Athanasiou, 2006) the hydrostatic pressure equilibrates over time. As viscoelasticity was not implemented in the present model, the effects of fluid exudation and imbibition could not be estimated quantitatively. The predicted volumetric strain differences (Fig. 3) also affect the situation in the solid matrix. This matrix is fibre-reinforced (Tanaka & van Eijden, 2003), which improves its resistance to stretch. As stretch was predicted predominantly in the upper layer of the disc, the present results suggest that this part primarily benefits from fibre-reinforcement. Consequently, the predicted stretch may be overestimated.
Generally, there is more compression during jaw closing than during jaw opening. This implies that the cartilaginous structures in the joint imbibe and exude fluid during, respectively, jaw opening and closing. By contrast, the central and lateral regions of the intermediate zone of the articular disc and temporal cartilage are more compressed during jaw opening as they are primarily load bearing during this movement (Koolstra & van Eijden, 2005).
The predicted heterogeneity in volumetric strain can be related to a relatively rapid distribution of solutes through the different cartilaginous structures (Mow et al. 1984). Although low-molecular-weight nutrients and waste products may be distributed easily by diffusion, for large molecules diffusion is extremely limited (Maroudas, 1976). Therefore, a continuously changing pattern of pressure gradients is particularly beneficial for the distribution of high-molecular-weight solutes (Ferguson et al. 2004).
The differences in predicted strains between the articular cartilage layers and the articular disc are in agreement with the difference in stiffness between these structures. The articular disc of the temporomandibular joint may deform up to 30% when the joint is loaded (Beek et al. 2001b, 2003). The presently predicted volumetric strain was much less. This difference is in agreement with the finding that shear deformation in the disc predominates over compressive deformation (Koolstra & van Eijden, 2005), indicating that changes in shape are much larger than volumetric changes. If the jaw is moved with more muscle force, causing a more heavily loaded joint, compression and relaxation will also be greater.
At the inferior side of the articular disc compression is predicted in and around the intermediate zone while anteriorly and posteriorly of this area some dilatation can be observed (Fig. 4). The differences between compression and dilatation are the largest when the jaw is open. At the superior side of the disc, however, the alternation between compression and dilatation is more medio-laterally orientated. In the open position the intermediate zone was compressed laterally and in the closed position its central part was also compressed. This mediolateral shift is similar to the observations of Gallo et al. (2000). The predicted patterns suggest that pressure gradients in the superior and inferior layer of the disc are more or less cross-wise. Although a quantitative relationship between pressure gradients and the amount of internal fluid flow has not been assessed, these patterns suggest that no area of the disc is a priori excluded from a regular refreshment of interstitial fluid.
The presence of interstitial fluid allows cartilage to protect the subchondral bone from impact loading. This occurs in the jaw joint when the teeth strike each other, or when they meet something hard between them as they close. In Fig. 3 this is indicated by sudden volumetric strain changes about 100 ms after the jaw had started closing. In the regions where the largest effects of impact were observed during jaw opening a relaxation was predicted. Consequently, these regions have had the opportunity to imbibe fluid, suggesting that thereafter they are ready to deal with a possible impact effectively.
In conclusion, the present results indicate that the construction of the human temporomandibular joint permits its cartilaginous structures to regulate their mechanical properties effectively by imbibition, exudation and redistribution of fluid. Furthermore, they suggest that refreshment of this fluid can be performed during normal function, such that for proper maintenance the cartilaginous structures do not have to rely only on an alternation of average baseline loading and relaxation during the day and night, respectively.
Acknowledgments
We thank Dr G. E. J. Langenbach for his constructive comments on the manuscript. This research was institutionally supported by the Interuniversity Research School of Dentistry, through the Academic Centre for Dentistry Amsterdam (ACTA).
Appendix
The applicability of the Mooney–Rivlin material model was tested by analysing the volumetric strain in a cylindrical disc subjected to a large strain indentation. The geometry of the disc (radius = 3.49 mm, thickness = 2.5 mm) and the indenter (radius = 1.97 mm) was similar to that of Beek et al. (2003). The applied material model was either a Mooney–Rivlin model as presently applied or a linear viscoelastic model as described by Allen & Athanasiou (2006). In the latter the relaxation modulus (Er), the instantaneous modulus (Ei) and the coefficient of viscosity (µ) had been obtained from various regions of porcine temporomandibular joint discs. For the present test their averages, Er = 133 kPa, Ei = 829 kPa, µ = 5.7 MPa s, were used. The indenter applied a 1-Hz sinusoidal compression cycle to the disc with a maximum strain of 20%. The volumetric strain was assessed in three areas: the centre of the disc, halfway between the centre and the outer rim, and adjacent to the outer rim. They occupied a layer with a thickness of about 1 mm between the upper and lower surfaces.
Qualitatively the volumetric strain in these areas was marginally dependent on the applied material model (Fig. A1). The largest compressive strain was predicted in the central area whereas in the outer area dilatation was predicted, irrespective of the applied material model. By contrast, quantitatively there were large differences as the volumetric strains in the linear viscoelastic model were up to 20 times larger than when the Mooney–Rivlin model was applied. Because the interstitial fluid is barely compressible, this difference has to be attributed to flow of this fluid. In areas under compressive volumetric strain fluid has to be squeezed out whereas in areas with dilatational strain it must be imbibed. Consequently, the magnitude of the volumetric strain can be considered to be indicative for the hydrostatic pressure and thus for the dynamics of fluid flow. In the Mooney–Rivlin model the changes in volumetric strain in the central area were less than in the other areas than predicted by the linear viscoelastic model. This suggests that the consequences of high volumetric strain are underestimated in the Mooney–Rivlin model. However, it must be noted that the influence of the non-linear behaviour of cartilage when subjected to large deformations (Hasler et al. 1999) has not been taken into account.
References
- Allen KD, Athanasiou KA. Viscoelastic characterization of the porcine temporomandibular joint disc under unconfined compression. J Biomech. 2006;39:312–322. doi: 10.1016/j.jbiomech.2004.11.012. [DOI] [PubMed] [Google Scholar]
- Beek M, Koolstra JH, van Ruijven LJ, van Eijden TMGJ. Three-dimensional finite element analysis of the human temporomandibular joint disc. J Biomech. 2000;33:307–316. doi: 10.1016/s0021-9290(99)00168-2. [DOI] [PubMed] [Google Scholar]
- Beek M, Koolstra JH, van Ruijven LJ, van Eijden TMGJ. Three-dimensional finite element analysis of the cartilaginous structures in the human temporomandibular joint. J Dent Res. 2001a;80:1913–1918. doi: 10.1177/00220345010800101001. [DOI] [PubMed] [Google Scholar]
- Beek M, Aarnts MP, Koolstra JH, Feilzer AJ, van Eijden TMGJ. Dynamical properties of the human temporomandibular joint disc. J Dent Res. 2001b;80:876–880. doi: 10.1177/00220345010800030601. [DOI] [PubMed] [Google Scholar]
- Beek M, Koolstra JH, van Eijden TMGJ. Human temporomandibular joint disc cartilage as a poroelastic material. Clin Biomech. 2003;18:69–76. doi: 10.1016/s0268-0033(02)00135-3. [DOI] [PubMed] [Google Scholar]
- Chen J, Akyuz U, Xu L, Pidaparti RMV. Stress analysis of the human temporomandibular joint. Med Eng Phys. 1998;20:565–572. doi: 10.1016/s1350-4533(98)00070-8. [DOI] [PubMed] [Google Scholar]
- Dar FH, Aspden RM. A finite element model of an idealized diarthrodial joint to investigate the effects of variation in the mechanical properties of the tissues. Proc Instn Mech Engrs Part H: J Engineering Med. 2003;217:341–348. doi: 10.1243/095441103770802504. [DOI] [PubMed] [Google Scholar]
- Donzelli PS, Gallo LM, Spilker RL, Palla S. Biphasic finite element simulation of the TMJ disc from in vivo kinematic and geometric measurements. J Biomech. 2004;37:1787–1791. doi: 10.1016/j.jbiomech.2004.01.029. [DOI] [PubMed] [Google Scholar]
- van Eijden TMGJ, Koolstra JH, Brugman P. Architecture of the human pterygoid muscles. J Dent Res. 1995;74:1489–1495. doi: 10.1177/00220345950740080901. [DOI] [PubMed] [Google Scholar]
- van Eijden TMGJ, Koolstra JH, Brugman P. Three-dimensional structure of the human temporalis muscle. Anat Rec. 1996;246:565–572. doi: 10.1002/(SICI)1097-0185(199612)246:4<565::AID-AR17>3.0.CO;2-M. [DOI] [PubMed] [Google Scholar]
- van Eijden TMGJ, Korfage JAM, Brugman P. Architecture of the human jaw-closing and jaw-opening muscles. Anat Rec. 1997;248:464–474. doi: 10.1002/(sici)1097-0185(199707)248:3<464::aid-ar20>3.3.co;2-4. [DOI] [PubMed] [Google Scholar]
- Ferguson SJ, Ito K, Nolte L-P. Fluid flow and convective transport of solutes within the intervertebral disc. J Biomech. 2004;37:213–221. doi: 10.1016/s0021-9290(03)00250-1. [DOI] [PubMed] [Google Scholar]
- Gallo LM, Salis Gross SS, Palla S. Nocturnal masseter EMG activity of healthy subjects in a natural environment. J Dent Res. 1999;78:1436–1444. doi: 10.1177/00220345990780080901. [DOI] [PubMed] [Google Scholar]
- Gallo LM, Nickel JC, Iwasaki LR, Palla S. Stress-field translation in the healthy human temporomandibular joint. J Dent Res. 2000;79:1740–1746. doi: 10.1177/00220345000790100201. [DOI] [PubMed] [Google Scholar]
- Gavião MDB, Engelen L, van der Bilt A. Chewing behavior and salivary secretion. Eur J Oral Sci. 2004;112:19–24. doi: 10.1111/j.0909-8836.2004.00105.x. [DOI] [PubMed] [Google Scholar]
- Hansson T, Oberg T, Carlsson GE, Kopp S. Thickness of the soft tissue layers and the articular disc in the temporomandibular joint. Acta Odontol Scand. 1977;35:77–83. doi: 10.3109/00016357709055993. [DOI] [PubMed] [Google Scholar]
- Hasler EM, Herzog W, Wu JZ, Müller W, Wyss U. Articular cartilage biomechanics: theoretical models, material properties, and biosynthetic response. Crit Rev Biomed Eng. 1999;27:415–488. [PubMed] [Google Scholar]
- Kim YJ, Bonassar LJ, Grodzinsky AJ. The role of cartilage streaming potential, fluid flow and pressure in the stimulation of chondrocyte biosynthesis during dynamic compression. J Biomech. 1995;28:1055–1066. doi: 10.1016/0021-9290(94)00159-2. [DOI] [PubMed] [Google Scholar]
- Koolstra JH, van Eijden TMGJ. The jaw open-close movements predicted by biomechanical modeling. J Biomech. 1997;30:943–950. doi: 10.1016/s0021-9290(97)00058-4. [DOI] [PubMed] [Google Scholar]
- Koolstra JH, Naeije M, van Eijden TMGJ. The three-dimensional active envelope of jaw border movements and its determinants. J Dent Res. 2001;80:1908–1912. doi: 10.1177/00220345010800100901. [DOI] [PubMed] [Google Scholar]
- Koolstra JH, van Eijden TMGJ. Combined finite-element and rigid-body analysis of human jaw joint dynamics. J Biomech. 2005;38:2431–2439. doi: 10.1016/j.jbiomech.2004.10.014. [DOI] [PubMed] [Google Scholar]
- Li LP, Herzog W. Strain-rate dependence of cartilage stiffness in unconfined compression: the role of fibril reinforcement versus tissue volume change in fluid pressurization. J Biomech. 2004;37:375–382. doi: 10.1016/s0021-9290(03)00263-x. [DOI] [PubMed] [Google Scholar]
- Maroudas A. Transport of solutes through cartilage: permeability to large molecules. J Anat. 1976;122:335–347. [PMC free article] [PubMed] [Google Scholar]
- Mow VC, Holmes MH, Lai WM. Fluid transport and mechanical properties of articular cartilage: a review. J Biomech. 1984;17:377–394. doi: 10.1016/0021-9290(84)90031-9. [DOI] [PubMed] [Google Scholar]
- Pérez del Palomar A, Doblaré M. The effect of collagen reinforcement in the behaviour of the temporomandibular joint disc. J Biomech. 2006;39:1075–1085. doi: 10.1016/j.jbiomech.2005.02.009. [DOI] [PubMed] [Google Scholar]
- van Ruijven LJ, Weijs WA. A new model for calculating muscle forces from electromyograms. Eur J Appl Physiol. 1990;61:479–485. doi: 10.1007/BF00236071. [DOI] [PubMed] [Google Scholar]
- Shrive NG, Frank CB. Articular cartilage. In: Nigg BM, Herzog W, editors. Biomechanics of the Musculoskeletal System. Chichester: John Wiley and Sons; 1998. pp. 86–106. [Google Scholar]
- Soltz MA, Ateshian GA. Experimental and theoretical prediction of cartilage interstitial fluid pressurization at an impermeable contact interface in confined compression. J Biomech. 1998;31:927–934. doi: 10.1016/s0021-9290(98)00105-5. [DOI] [PubMed] [Google Scholar]
- Tanaka E, van Eijden TMGJ. Biomechanical behavior of the temporomandibular joint disc. Crit Rev Oral Biol Med. 2003;14:138–150. doi: 10.1177/154411130301400207. [DOI] [PubMed] [Google Scholar]
- Winters JM, Stark L. Muscle models: what is gained and what is lost by varying model complexity. Biol Cybern. 1987;55:403–420. doi: 10.1007/BF00318375. [DOI] [PubMed] [Google Scholar]
- Wong M, Carter DR. Articular cartilage functional histomorphology and mechanobiology: a research perspective. Bone. 2003;33:1–13. doi: 10.1016/s8756-3282(03)00083-8. [DOI] [PubMed] [Google Scholar]