Abstract
A bacterial spore test has been developed which enables the efficacy of the sterilizing cycle recommended by the British Pharmaceutical Codex (1973) for bottled fluids to be accurately monitored. During a 14-month period this test detected faults in 3.3% of the sterilizing cycles, representing five distinct episodes of sterilization failure that passed unnoticed by the conventional controls of physical measurements and sterility testing. There were no failures of sterilization as detected by conventional techniques which were not indicated by the spore test.
Full text
PDF




Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- KELSEY J. C. The testing of sterilizers. 2. Thermophilic spore papers. J Clin Pathol. 1961 May;14:313–319. doi: 10.1136/jcp.14.3.313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Phillips I., Eykyn S., Laker M. Outbreak of hospital infection caused by contaminated autoclaved fluids. Lancet. 1972 Jun 10;1(7763):1258–1260. doi: 10.1016/s0140-6736(72)90981-6. [DOI] [PubMed] [Google Scholar]
