Abstract
Indirect inhibition of an external temporary pacemaker by electrocautery is reported. Before induction of general anesthesia for a hemimaxillectomy and radical neck dissection, a temporary transvenous demand pacemaker was inserted into a patient with a first-degree atrioventricular block and complete left bundle-branch block. Although we provided common precautions to prevent electromagnetic interference by electrocautery, pacing failure still occurred. It was thought to be caused by current dispersing from the active electrocautery electrode. This case suggests that occipital placement of the electrocautery ground plate should be considered during neck surgery in a patient requiring a temporary pacemaker.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Greene L. F., Merideth J. Transurethral operations employing high frequency electrical currents in patients with demand cardiac pacemakers. J Urol. 1972 Sep;108(3):446–448. doi: 10.1016/s0022-5347(17)60767-5. [DOI] [PubMed] [Google Scholar]
- Lerner S. M. Suppression of a demand pacemaker by transurethral electrocautery. Anesth Analg. 1973 Sep-Oct;52(5):703–706. [PubMed] [Google Scholar]
- Nishino M., Itota Y., Konno K., Haseba S., Honda N. [Anesthetic management of patients with cardiac pacemaker]. Masui. 1976 Nov;25(12):1306–1311. [PubMed] [Google Scholar]
- Omote T., Namiki A., Dohi S., Takahashi T., Ohhori K. [A case of cardiac arrest caused by electrocautery during surgical pacemaker implanting (author's transl)]. Masui. 1981 Oct;30(10):1126–1129. [PubMed] [Google Scholar]
- Simon A. B. Perioperative management of the pacemaker patient. Anesthesiology. 1977 Feb;46(2):127–131. doi: 10.1097/00000542-197702000-00010. [DOI] [PubMed] [Google Scholar]
- Wynands J. E. Anesthesia for patients with heart block and artificial cardiac pacemakers. Anesth Analg. 1976 Sep-Oct;55(5):626–632. doi: 10.1213/00000539-197609000-00004. [DOI] [PubMed] [Google Scholar]
- Yamamura T., Furukawa K. [Basic electronics for clinical practice and laboratory. II. Electrical stimulation of the body. III. Medical equipment and electrical hazards]. Masui. 1983 May;32(5):590–598. [PubMed] [Google Scholar]
- Zaidan J. R. Pacemakers. Anesthesiology. 1984 Apr;60(4):319–334. doi: 10.1097/00000542-198404000-00008. [DOI] [PubMed] [Google Scholar]