Abstract
The Schwarzschild-Villiger effect has been experimentally demonstrated with the optical system used in this laboratory. Using a photographic mosaic specimen as a model, it has been shown that the conclusions of Naora are substantiated and that the SV effect, in large or small magnitude, is always present in optical systems. The theoretical transmission error arising from the presence of the SV effect has been derived for various optical conditions of measurement. The results have been experimentally confirmed. The SV contribution of the substage optics of microspectrophotometers has also been considered. A simple method of evaluating a flare function f(A) is advanced which provides a measure of the SV error present in a system. It is demonstrated that measurements of specimens of optical density less than unity can be made with less than 1 per cent error, when using illuminating beam diameter/specimen diameter ratios of unity and uncoated optical surfaces. For denser specimens it is shown that care must be taken to reduce the illuminating beam/specimen diameter ratio to a value dictated by the magnitude of a flare function f(A), evaluated for a particular optical system, in order to avoid excessive transmission error. It is emphasized that observed densities (transmissions) are not necessarily true densities (transmissions) because of the possibility of SV error. The ambiguity associated with an estimation of stray-light error by means of an opaque object has also been demonstrated. The errors illustrated are not necessarily restricted to microspectrophotometry but may possibly be found in such fields as spectral analysis, the interpretation of x-ray diffraction patterns, the determination of ionizing particle tracks and particle densities in photographic emulsions, and in many other types of photometric analysis.
Full Text
The Full Text of this article is available as a PDF (1.4 MB).
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- DAVIES H. G., WILKINS M. H. F. Interference microscopy and mass determination. Nature. 1952 Mar 29;169(4300):541–541. doi: 10.1038/169541a0. [DOI] [PubMed] [Google Scholar]
- GLICK D., ENGSTROM A., MALMSTROM B. G. A critical evaluation of quantitative histo- and cytochemical microscopic techniques. Science. 1951 Sep 7;114(2958):253–258. doi: 10.1126/science.114.2958.253. [DOI] [PubMed] [Google Scholar]
- LISON L. Schwarzschild-Villiger effect in microspectrophotometry. Science. 1953 Oct 2;118(3066):382–383. doi: 10.1126/science.118.3066.382. [DOI] [PubMed] [Google Scholar]
- NAORA H. Microspectrophotometry and cytochemical analysis of nucleic acids. Science. 1951 Sep 14;114(2959):279–280. doi: 10.1126/science.114.2959.279. [DOI] [PubMed] [Google Scholar]
- NAORA H. Microspectrophotometry of cell nucleus stained by feulgen reaction. I. Microspectrophotometric apparatus without Schwarzschild-Villiger effect. Exp Cell Res. 1955 Apr;8(2):259–278. doi: 10.1016/0014-4827(55)90141-5. [DOI] [PubMed] [Google Scholar]
- NAORA H., SIBATANI A. Unreliability of the data hitherto reported on the desoxypentosenucleic acid content of cell nuclei determined by microspectrophotometry. Biochim Biophys Acta. 1952 Nov;9(5):582–584. doi: 10.1016/0006-3002(52)90216-3. [DOI] [PubMed] [Google Scholar]
- NAORA H. Schwarzschild-Villiger effect in microspectrophotometry. Science. 1952 Feb 29;115(2983):248–249. doi: 10.1126/science.115.2983.248. [DOI] [PubMed] [Google Scholar]
- ORNSTEIN L. The distributional error in microspectrophotometry. Lab Invest. 1952;1(2):250–265. [PubMed] [Google Scholar]
- Ornstein L., Pollister A. W. Schwarzschild-Villiger Effect in Microspectrophotometry. Science. 1952 Aug 22;116(3008):203–204. doi: 10.1126/science.116.3008.203. [DOI] [PubMed] [Google Scholar]
- PATAU K. Absorption microphotometry of irregular-shaped objects. Chromosoma. 1952;5(4):341–362. doi: 10.1007/BF01271492. [DOI] [PubMed] [Google Scholar]
- POGO A. O., FUNES J. R. C. An experimental study of flare in a new micro-spectrophotometer. Exp Cell Res. 1959 May;17(2):205–213. doi: 10.1016/0014-4827(59)90211-3. [DOI] [PubMed] [Google Scholar]
- POLLISTER A. W. Microspectrophotometry of fixed cells by visible light. Lab Invest. 1952;1(2):231–249. [PubMed] [Google Scholar]