Abstract
Introduction
Accurate determination of human epidermal growth factor receptor 2 (HER2) status is essential for optimal patient management with trastuzumab (Herceptin). However, standard guidelines do not specify a particular commercial kit, antibody or probe for testing, and discrepancies arise from variability between kits. The aim of this study was to compare the accuracy of four commercially available fluorescence/chromogenic in situ hybridisation (FISH/CISH) kits and validate one for the resolution of borderline immunohistochemistry (IHC) cases. The interpretation pitfalls, optimal threshold values, assay duration and complexity of each kit were also considered.
Methods
The Food and Drug Administration (FDA)-approved dual-probe FISH assay PathVysion was chosen as the 'gold standard' against which pharmDx (dual-probe) and INFORM (mono-probe) FDA-approved FISH kits and the SPoT-Light CISH kit were compared. Tumours were also evaluated by IHC with the FDA-approved HercepTest kit and a validated in-house IHC protocol. Fifty-five patients with invasive breast carcinoma were selected as a representative proportion of HER2 IHC 2+ cases.
Results
HER2 amplification was observed in 31% of tumours by PathVysion compared with 33% with pharmDx. The number of amplified tumours detected by INFORM and CISH varied with the threshold applied. Agreement was excellent between PathVysion and pharmDx (100%), good with SPoT-Light (89%; cutoff at least five signals per nucleus) and moderate with INFORM (76%; cutoff more than four signals per nucleus). Agreement with INFORM improved to 98% with a cutoff of at least six signals per nucleus.
Conclusion
With an appropriate cutoff, the INFORM kit was comparable to dual-probe FISH kits for evaluating HER2 status. We validate and recommend CISH as an appropriate assay for HER2 scoring that is easy to interpret and requires equipment readily found in, or that can be adapted to, all pathology laboratories. For borderline IHC cases, dual-probe FISH analysis remains the most useful protocol to apply.
Introduction
About 20 to 30% of breast cancer tumours are positive for amplification and overexpression of human epidermal growth factor receptor 2 (HER2), which is associated with poor prognosis and clinical outcome [1-3]. HER2 is targeted therapeutically by the humanised monoclonal antibody trastuzumab (Herceptin; F Hoffman-La Roche Ltd, Basel, Switzerland).
Trastuzumab is effective only in patients whose tumours are positive for HER2 gene amplification and/or protein overexpression [4-7]. Therefore, accurate HER2 testing is essential for selecting patients eligible for trastuzumab-based therapies. Originally, HER2 status was determined by Southern blotting [3,8], but the present methods of evaluation combine immunohistochemistry (IHC) and fluorescence/chromogenic in situ hybridisation (FISH/CISH). Inconsistencies arise in HER2 scoring because of the variety of commercial assays available, which use different antibodies and probes. Although the College of American Pathologists [9] and a UK-based study [10] recommend IHC and FISH for HER2 testing, neither specify a particular commercial kit.
In the present study, the Food and Drug Administration (FDA)-approved dual-probe FISH assay PathVysion (Abbott France SAS, Rungis, France) was used as 'gold standard' against which the FDA-approved dual-probe pharmDx (Dako France SAS, Trappes, France) and mono-probe INFORM (Ventana Medical Systems SA, Illkirch, France) FISH assays, plus the CISH assay SPoT-Light (Zymed Laboratories Inc., San Francisco, CA, USA) were compared. Invasive tumour specimens from 55 patients were evaluated for HER2 status with these different methods. This study is the first, to our knowledge, that compares three commercially available FISH analysis kits with CISH, and evaluates the respective thresholds of each assay.
Materials and methods
Samples
Tumour samples from 55 patients with invasive breast carcinoma were collected for evaluation between January 2001 and April 2003. All samples were fixed in neutral-buffered formalin and embedded in paraffin. Cases were selected to ensure a representative proportion of HER2 IHC 2+ cases, and blind assessments were made of the invasive tumour component of all samples by two independent pathologists.
The study was conducted in line with the Declaration of Helsinki and Good Clinical Practice, and with the approval of an independent Institutional Review Board.
Immunohistochemistry
Tumour samples were assessed for HER2 status with the FDA-approved HercepTest kit (Dako) and a standardised in-house laboratory technique with the rabbit anti-human HER2 polyclonal A0485 (1:400 dilution; Dako) for 32 minutes [11]. Tissue sections (3 μm) were deparaffinised and rehydrated, and antigens were retrieved for 40 minutes in citrate buffer (pH 6.1) at 95°C. IHC was performed with the NexES automated immunostaining system and diaminobenzidine detection kit (Ventana Medical Systems SA). Staining intensity was graded in accordance with the HercepTest protocol system as 0, 1+, 2+ or 3+. Samples scored as 0 or 1+ were considered negative for HER2 overexpression, 2+ was weak positive and 3+ was strong positive, with complete membrane staining of more than 10% of tumour cells.
Fluorescence in situ hybridisation
The FDA-approved PathVysion and pharmDx assays were used in accordance with the manufacturers' recommended protocols but with some minor modifications. The DNA probes, HER2-specific sequence probe (LSI HER2/neu), chromosome enumeration probe 17 (CEP17) and tissue sections were denatured for 5 minutes at 85°C (PathVysion) or 82°C (pharmDx) with a HYBrite instrument (Abbott France SAS). An additional wash in distilled water was performed before counterstaining and mounting with a solution of 4,6-diamidino-2-phenylindole (DAPI). As the intensity of fluorescence observed with the DAPI solution provided in the pharmDx kit was poor, the commercially available VECTASHIELD plus DAPI mounting medium (AbCys SA, Paris, France) was used instead for these samples. The results are reported as the ratio of average HER2:CEP17 signals per nucleus. Signal ratios of less than two were classified as non-amplified (NA), and ratios of two or more as amplified.
The INFORM assay, an automated system using the BenchMark automated staining system (Ventana Medical Systems SA), was used in accordance with the manufacturer's recommended protocols. Amplification was defined as more than four signals per nucleus, as stated in the manufacturer's guidelines. Additional thresholds were applied (more than five and at least six signals per nucleus) for comparison and optimisation. At least 30 nuclei were counted for signals in each section.
Chromogenic in situ hybridisation
CISH analysis of HER2 amplification was performed with the SPoT-Light assay in accordance with the manufacturer's protocol. The DNA probe (HER2 SPoT-Light) and sections were denatured at 94 to 95°C and hybridised overnight at 37°C with a HYBrite instrument. Amplification was defined with two thresholds – more than five or at least six signals per nucleus – detected in more than 50% of nuclei or when a large signal cluster was detected. At least 30 nuclei were counted per section.
Statistics
PathVysion was used as the 'gold standard' against which all protocols were compared, and agreement was quantified with pairwise kappa (κ) statistics. The analytical performance of these kits was also estimated for specificity and sensitivity. The pairwise χ2 test was used to assess overvaluation or undervaluation of HER2 status.
Results
HER2 status for all samples as determined by IHC, FISH and CISH are detailed in Table 1.
Table 1.
Case ID | PathVysion | pharmDx | INFORM | CISH | In-house | HercepTest | |||
>4 | >5 | ≥6 | >5 | ≥6 | |||||
1 | A | A | A | A | A | A | A | 3 | 3 |
2 | A | A | A | A | A | A | A | 3 | 3 |
3 | A | A | A | A | A | A | A | 3 | 3 |
4 | A | A | A | A | A | A | A | 3 | 3 |
5 | A | A | A | A | A | A | A | 3 | 3 |
6 | A | A | A | A | A | A | A | 3 | 3 |
7 | A | A | A | A | A | A | A | 3 | 3 |
8 | A | A | A | A | A | A | A | 3 | 3 |
9 | A | A | A | A | A | A | A | 3 | 3 |
10 | A | A | A | A | A | A | A | 3 | 3 |
11 | A | A | A | A | A | A | A | 3 | 3 |
12 | A | A | A | A | A | A | A | 3 | 3 |
13 | A | A | A | A | A | A | A | 3 | 3 |
14 | A | A | A | A | A | A | A | 3 | 3 |
15 | A | A | A | A | A | A | A | 3 | 3 |
16 | A | A | A | A | A | A | A | 3 | 3 |
17 | A | A | A | A | A | A | A | 2 | 3 |
18 | NA | NA | NA | NA | NA | NA | NA | 2 | 3 |
19 | NA | NA | NA | NA | NA | NA | NA | 2 | 2 |
20 | NA | NA | A | NA | NA | A | A | 3 | 3 |
21 | NA | NA | A | A | NA | A | NA | 3 | 3 |
22 | - | A | NA | NA | NA | NA | NA | 2 | 3 |
23 | - | NA | NA | NA | NA | NA | NA | 2 | 1 |
24 | NA | NA | A | NA | NA | A | A | 1 | 3 |
25 | NA | NA | A | A | A | A | NA | 2 | 3 |
26 | NA | NA | A | A | NA | A | A | 2 | 2 |
27 | NA | NA | A | NA | NA | NA | NA | 2 | 2 |
28 | NA | NA | A | A | NA | NA | NA | 2 | 2 |
29 | NA | NA | A | NA | NA | NA | NA | 2 | 1 |
30 | NA | NA | A | A | NA | NA | NA | 1 | 2 |
31 | NA | NA | NA | NA | NA | NA | NA | 1 | 1 |
32 | NA | NA | A | NA | NA | NA | NA | 0 | 0 |
33 | NA | NA | A | NA | NA | A | NA | 0 | 1 |
34 | NA | NA | A | NA | NA | NA | NA | 0 | 2 |
35 | NA | NA | A | NA | NA | NA | NA | 0 | 0 |
36 | NA | NA | NA | NA | NA | NA | NA | 0 | 1 |
37 | NA | NA | NA | NA | NA | NA | NA | 0 | 1 |
38 | NA | NA | NA | NA | NA | NA | NA | 2 | 3 |
39 | NA | NA | NA | NA | NA | NA | NA | 2 | 3 |
40 | NA | NA | NA | NA | NA | NA | NA | 2 | 2 |
41 | NA | NA | NA | NA | NA | NA | NA | 2 | 2 |
42 | NA | NA | NA | NA | NA | NA | NA | 2 | 2 |
43 | NA | NA | NA | NA | NA | NA | NA | 2 | 1 |
44 | NA | NA | NA | NA | NA | NA | NA | 2 | 0 |
45 | NA | NA | NA | NA | NA | NA | NA | 0 | 2 |
46 | NA | NA | NA | NA | NA | NA | NA | 0 | 2 |
47 | NA | NA | NA | NA | NA | NA | NA | 0 | 2 |
48 | NA | NA | NA | NA | NA | NA | NA | 0 | 1 |
49 | NA | NA | NA | NA | NA | NA | NA | 0 | 1 |
50 | NA | NA | NA | NA | NA | NA | NA | 0 | 1 |
51 | NA | NA | NA | NA | NA | NA | NA | 0 | 0 |
52 | NA | NA | NA | NA | NA | NA | NA | 0 | 0 |
53 | NA | NA | NA | NA | NA | NA | NA | 0 | 0 |
54 | NA | NA | NA | NA | NA | NA | NA | 0 | 0 |
55 | NA | NA | NA | NA | NA | NA | NA | 0 | 0 |
HER2, human epidermal growth factor receptor 2; A, amplified; NA, non-amplified; IHC, immunohistochemistry; FISH, fluorescence in situ hybridisation; CISH, chromogenic in situ hybridisation.
Immunohistochemistry
With the A0485 antibody-based in-house IHC protocol, 18 tumours (33%) were scored as IHC 3+, 17 (31%) as IHC 2+ and 20 (36%) as IHC 0/1+ (HER2 negative). In contrast, HercepTest identified 25 tumours (45%) as IHC 3+ and 12 (22%) as 2+ (HER2-positive); 18 (33%) were scored as IHC 0/1+ (HER2-negative). The κ coefficient between the two tests was 0.427 ± 0.083 (SD) and agreement was 58% if in-house IHC was considered as the reference standard. Concordance between the two protocols was 70% for HER2-negative IHC 0/1+ cases, 47% for IHC 2+ and 100% for HER2-positive IHC 3+ cases, if in-house IHC was considered as the reference standard. All IHC 3+ cases demonstrated complete membrane staining in at least 30% of tumour cells.
Fluorescence in situ hybridisation
Two cases could not be evaluated with the PathVysion kit but were assessed and graded successfully using pharmDx and INFORM. PathVysion identified 17 tumours (31%) as amplified, in comparison with 18 (33%) by pharmDx. INFORM had three thresholds applied to determine the optimal cutoff for evaluating positive HER2 amplification. These thresholds were more than four signals per nucleus as recommended by the manufacturer, more than five signals per nucleus as recommended for CISH analysis, and at least six signals per nucleus as recommended by Zymed for the SPoT-Light CISH kit used in this study. Using a cutoff of more than four signals per nucleus, 30 tumours (55%) were designated as amplified. This was decreased to 22 (40%) when the threshold was increased to more than five signals per nucleus. Application of the threshold of at least six signals per nucleus decreased the proportion of amplified samples to 18 (33%).
Chromogenic in situ hybridisation
Two different thresholds of more than five or at least six signals per nucleus were used to establish the optimal cutoff point for HER2 amplification by CISH. When the cutoff of more than five signals per nucleus was applied, 23 tumours (42%) were defined as amplified, and 32 (58%) as NA. In contrast, application of the threshold of at least six signals per nucleus decreased the number of amplified tumours to 20 (36%) and increased the number of NA tumours to 35 (64%).
Inter-assay agreement
The PathVysion assay kit was chosen as the 'gold standard' for comparisons. Agreement with PathVysion and other techniques used to determine HER2 status is shown in Table 2.
Table 2.
Comparison with PathVysion | κ coefficient* | Degree of agreement | Agreement (percentage) | Sensitivity (percentage) | Specificity (percentage) |
PharmDx | 1.000 | Excellent | 100 | 100 | 100 |
INFORM (at least six signals per nucleus) | 0.916 ± 0.058 | Excellent | 98 | 100 | 94 |
INFORM (at least five signals per nucleus) | 0.799 ± 0.085 | Good | 91 | 100 | 88 |
INFORM (more than four signals per nucleus) | 0.532 ± 0.111 | Moderate | 76 | 100 | 57 |
CISH (at least six signals per nucleus) | 0.876 ± 0.070 | Excellent | 94 | 100 | 94 |
CISH (at least five signals per nucleus) | 0.762 ± 0.091 | Good | 89 | 100 | 88 |
In-house IHC (positive 3+, negative 0, 1+, 2+) | 0.792 ± 0.089 | Good | 91 | 89 | 91 |
In-house IHC (positive 3+, 2+, negative 0, 1+) | 0.475 ± 0.111 | Moderate | 72 | 100 | 57 |
HercepTest (positive 3+, negative 0, 1+, 2+) | 0.645 ± 0.104 | Good | 83 | 89 | 80 |
HercepTest (positive 3+, 2+, negative 0, 1+) | 0.350 ± 0.119 | Poor | 64 | 94 | 49 |
HER2, human epidermal growth factor receptor 2; CISH, chromogenic in situ hybridisation; IHC, immunohistochemistry. (* errors given as ± SD)
There was excellent agreement (100%) between the dual-probe techniques, namely PathVysion and pharmDx. However, agreement between PathVysion and INFORM differed depending on the threshold applied. The lowest level of agreement between the two protocols (76%) occurred with a cutoff of more than four signals per nucleus. This increased to 91% with the threshold of more than five signals per nucleus, and to 98% with a cutoff of at least six signals per nucleus (Table 2).
Similarly, application of different thresholds to CISH altered agreement with PathVysion. An agreement of 89% with a cutoff of more than five signals per nucleus was increased to 94% by changing the cutoff value to at least six signals per nucleus (Table 2).
For IHC, general agreement varied for the HercepTest and in-house IHC protocol depending on whether IHC 2+ samples were classified as positive (grouped with IHC 3+ cases) or negative (grouped with IHC 0/1+ cases) (Table 2). When considering only HER2-negative IHC 0/1+ tumours, both protocols demonstrated 100% concordance with PathVysion (data not shown). Good agreement (89%) was demonstrated between the in-house IHC protocol and PathVysion for the evaluation of IHC 3+ tumours but HercepTest demonstrated poor concordance at 68% (data not shown).
Details of the unmatched in situ hybridisation cases
There were 14 cases that scored differently for HER2 amplification when using the various kits (Table 1). One case scored as amplified with pharmDx but NA with the INFORM and SPoT-Light assays when using all thresholds. The HER2:CEP17 ratio of 2.27 highlighted low amplification but this could not be confirmed with the alternative dual-probe kit (PathVysion) (Table 3). Six cases were scored as amplified (five with low amplification and one with presence of clusters) with the use of the CISH SPoT-Light assay (threshold more than five signals per nucleus) and NA with PathVysion. Four of these tumours had chromosome 17 polysomy, with a mean number of three signals per nucleus for CEP17 with PathVysion. The mean number of CISH HER2 signals per nucleus for the remaining two discrepant cases was 5.6 and clusters. When CISH was used in conjunction with the threshold of at least six signals per nucleus, only one case remained discrepant with the presence of clusters (Table 3).
Table 3.
Case ID | PathVysion | PharmDx™ | INFORM | CISH | ||||
HER2 signals per nucleus | CEP17 signals per nucleus | Ratio | HER2 signals per nucleus | CEP17 signals per nucleus | Ratio | HER2 signals per nucleus | HER2 signals per nucleus | |
20 | 5.6 | 5.3 (P) | 1.1 | 6.2 | 4.0 (P) | 1.6 | 4.9 | 6.4 |
21 | 2.2 | 2.5 | 0.9 | 5.1 | 3.3 (P) | 1.6 | 5.6 | 5.6 |
22 | - | - | - | 4.5 | 2.0 | 2.3 | 3.8 | 3.6 |
24 | 2.6 | 2.5 | 1.0 | 3.3 | 2.6 | 1.2 | 4.3 | Clusters |
25 | 3.7 | 3.8 (P) | 1.0 | 6.2 | 4.0 (P) | 1.6 | 7.7 | 5.5 |
26 | 4.4 | 2.8 | 1.6 | 6.1 | 4.2 (P) | 1.5 | 5.3 | 6.6 |
27 | 3.4 | 2.2 | 1.5 | 3.8 | 2.4 | 1.6 | 4.8 | 3.7 |
28 | 3.8 | 3.7 (P) | 1.0 | 4.0 | 3.4 (P) | 1.2 | 5.7 | 4.9 |
29 | 2.0 | 2.0 | 1.0 | 3.5 | 2.0 | 1.7 | 4.4 | 3.6 |
30 | 4.6 | 3.3 (P) | 1.4 | 5.7 | 3.7 (P) | 1.5 | 5.5 | 3.4 |
32 | 2.0 | 2.5 | 0.8 | 2.0 | 2.0 | 1.0 | 4.2 | 4.9 |
33 | 5.6 | 5.0 (P) | 1.1 | 5.2 | 3.5 (P) | 1.5 | 4.5 | 5.7 |
34 | 3.3 | 2.2 | 1.5 | 3.3 | 2.1 | 1.6 | 4.1 | 2.5 |
35 | 3.1 | 3.5 (P) | 0.9 | 2.8 | 2.7 | 1.0 | 4.2 | 3.3 |
HER2, human epidermal growth factor receptor 2; CEP17, chromosome enumeration probe 17; P, polysomy; CISH, chromogenic in situ hybridisation.
With the INFORM kit and a threshold of more than four signals per nucleus, comparisons with the 'gold standard' PathVysion identified an overvaluation of HER2 amplification in 13 specimens (amplified with INFORM and NA with PathVysion). Overvaluation was not random (P = 0.00031, paired-series χ2 test). No false negatives were observed. Chromosome 17 polysomy accounted for six of the unmatched cases considering PathVysion and eight considering PathVysion and pharmDx, but the remaining false positives could not be explained. Application of the following thresholds reduced the number of false positives and improved agreement with PathVysion: more than five signals per nucleus reduced the number of mismatches to five with only one case that could not be explained by a polysomy, and at least six signals per nucleus left only one unmatched case, which could be explained by chromosome 17 polysomy (Table 3). The cutoff points of more than five and at least six signals per nucleus improved agreement from 76% to 91% and 98%, respectively.
Technical and economic considerations
The minimum time requirements and direct costs for the different techniques are presented in Table 4. CISH was the most time-consuming method because it required immunodetection of the amplification product after hybridisation; however, excluding the in-house kit, it was the cheapest. INFORM was less time-consuming because the staining protocol was automated; PathVysion was the most expensive.
Table 4.
Technical stage | Stage presence or duration | ||||
PathVysion | PharmDx | INFORM | CISH | In-house | |
Deparaffinisation and rehydration | 40 min | 33 min | Overnight in the BenchMark | 25 min | 20 min |
Pretreatment | 67 min | 31 min | 21 min | 60 min | |
Digestion | 26 min | 16 min | 16 min | No | |
Fixation | 20 min | No | No | Fixation, blocking, incubation with primary and secondary antibodies and detection, all performed in NexES 90 min | |
Dehydration | No | 6 min | 10 min | ||
Drying | Yes | Yes | Yes | ||
Denaturation | 5 min | 5 min | 5 min | ||
Hybridisation | Overnight | Overnight | Overnight | ||
Washing | 4 min | 20 min | 4 min | 11 min | Washing, haematoxylin counterstain (15 sec), dehydration and mounting, 12 min |
Dehydration | No | 6 min | 3 min | No | |
Drying | 30 min | Fast | Fast | No | |
DAPI staining | Yes | Yes | Yes | No | |
Detection | No | No | No | 147 min | |
Total duration | Overnight + 3 h 12 min | Overnight + 1 h 57 min | Overnight + 7 min | Overnight + 3 h 55 min | 3 h 2 min |
Price (€ per test) | 131 | 85 | 75 | 56 | 4 |
CISH, chromogenic in situ hybridisation; DAPI, 4,6-diamidino-2-phenylindole.
Comparisons of the different in situ hybridisation protocols are listed in Table 5. CISH provides several technical advantages over the FISH protocols assessed in this study. First, as CISH uses a chromogenic peroxidase reaction to reveal amplification signals, samples can be assessed with a standard light microscope, allowing simultaneous viewing of staining and histopathology. Second, because the chromogenic labels are stable, samples can be stored easily at ambient temperature for long periods without signal degeneration. However, there are important disadvantages, namely the time associated with immunodetection of the amplification product and the lack of dual chromogenic probes HER2:CEP17, although SPoT-Light chromosome 17 is available.
Table 5.
Feature | PathVysion | PharmDx | INFORM | CISH |
Reagents | Preparation required | Ready to use | Ready to use | Ready to use |
Stability of reagents | pH requires checking | OK | OK | OK |
DAPI | Bright | Low intensity | Not provided | - |
Background | Orange fluorescence | Strong; green fluorescence | Faint; green fluorescence | - |
Stability of signal | Good | Rapid quenching of FITC signal | Good | Excellent |
HER2 probe | Spectrum Orange; clear signal but high background fluorescence | Texas red; strong clear signal | FITC; strong clear signal | Chromogenic; peroxidase-based immunodetection, clear signal |
CEP17 probe | FITC; very strong signal, with blurring of large signals | FITC; faint and transient | - | - |
Tissue morphology | Poor definition; control with H&E to view areas of interest | Very poor definition; control with H&E to view areas of interest | Poor definition; control with H&E to view areas of interest | Good definition, allowing simultaneous analysis of amplification and histopathology |
Microscope | Fluorescent | Fluorescent | Fluorescent | Normal light |
Storage | 1 year at -20°C | 1 year at -20°C | 1 year at -20°C | Long period at ambient temperature |
Threshold | OK | OK | CISH threshold must be applied and borderline cases must be evaluated with a dual-probe method (5–10) |
DAPI, 4,6-diamidino-2-phenylindole; HER2, human epidermal growth factor receptor 2; CEP17, chromosome enumeration probe 17; FITC, fluorescein isothiocyanate; H&E, haematoxylin and eosin staining; CISH, chromogenic in situ hybridisation.
Because FISH techniques use fluorophores to localise target DNA, definition of cell morphology is poor; the best is achieved with PathVysion. The key technical disadvantages associated with PathVysion are that the solutions are not ready to use, the reagent pH requires confirmation before use, and fluorescent HER2 probes labelled with Spectrum Orange are hard to detect above the high background fluorescence.
In contrast to the HER2 probe, the pharmDx CEP17 probe was conjugated with fluorescein isothiocyanate (FITC), and positive signals were transient and faint. The high background fluorescence associated with this kit created problems when evaluating the faint CEP17 signals.
The most significant technical disadvantage with INFORM is the lack of dual fluorescence probes HER2:CEP17, which would counter the level of false positives achieved with this kit at its recommended threshold.
Discussion
Trastuzumab has a proven survival benefit in the treatment of women with HER2-positive breast cancer. Accurate identification of HER2 status is essential for identifying patients who may benefit from HER2-targeted therapy. The protocols currently validated for use in routine diagnosis are IHC for identifying HER2 protein overexpression, and FISH or CISH for determining HER2 gene amplification. IHC and FISH are the protocols of choice in the routine diagnosis of HER2 status, providing reliable results. These protocols are reliable only if they are validated within individual laboratories. It is essential that protocols are adhered to, that quality-control and assurance programmes are implemented, and that diagnostic personnel have experience and regular training in score interpretation.
The UK best practice recommends a two-phase testing algorithm based on first-line IHC evaluation and second-line FISH assessment of borderline cases (IHC 2+) [10]. This correlates with recommendations from the College of American Pathologists if the concordance rates are more than 90% between IHC 3+ and FISH amplified, IHC 0 and FISH NA, and 95% between IHC 1+ and FISH NA. The in-house IHC was in accordance with the criteria above; the algorithm was therefore applied to all samples. Our in-house IHC technique has been calibrated with FISH within the framework of the Groupe d'Étude des Facteurs Pronostiques par Immunohistochimie dans le Cancer du Sein (GEFPICS) study [11]. This is not true for HercepTest, for which the agreement was weaker (68%) between IHC 3+ and PathVysion. However, HercepTest has been proposed as the standard IHC method for HER2 status evaluation, capable of overcoming the problems of standardisation and reproducibility between laboratories. In agreement with our results, several studies have demonstrated the excessive sensitivity of this kit [12-14].
The goal of this study was to evaluate the accuracy of four commercial in situ hybridisation kits (three FISH and one CISH) and validate a protocol for resolving borderline IHC cases. The PathVysion kit was chosen as the 'gold standard' because it has a proven record of sensitivity and specificity for HER2 gene amplification against solid matrix blotting techniques [8].
Concordance between this kit and pharmDx was 100% (the manufacturers report a concordance of 95%); the performance of both kits can therefore be regarded as equivalent. Differences arise in price, time and interpretation. PharmDx is cheaper and faster than PathVysion but interpretation is more difficult (except for highly amplified cases) because of strong background fluorescence and the transient nature of the FITC signal.
Agreement between PathVysion and CISH varied depending on the threshold applied. Excellent agreement (94%) was achieved with a cutoff of at least six signals per nucleus. All patients with amplification of the HER2 gene were detected (sensitivity of 100%); however, three false positives (specificity of 94%) were reported with CISH. Two of these were low-level amplifications, having chromosome 17 polysomy and requiring confirmation of diagnosis with one of the dual-probe FISH kits. This threshold recommended by Zymed for the SPoT-Light CISH kit accounted for the occurrence of aneuploid cell populations that can give three to five signals per nucleus, and the manufacturer advised that these be interpreted as NA. In addition, a small proportion of aneuploid cells may contain five to eight signals per nucleus because of DNA replication during the S and G2/M phases of the cell cycle and should also be regarded as NA.
Agreement between CISH and FISH in the literature varies between 84% and 100% [15-19]. One study reported 100% agreement between CISH and FISH when samples scoring more than two signals per nucleus with HER2 CISH were systematically controlled for by using a chromosome 17 CISH probe on adjacent sections [19]. CISH provides a good alternative to FISH in laboratories that lack access to fluorescence microscopy. However, we recommend that for borderline cases with CISH scores of more than five to ten signals per nucleus, diagnosis must be confirmed by using either a dual-probe FISH kit as used in this study, or an internal control such as the chromosome 17 SPoT-Light probe for CISH.
With the manufacturer's recommended threshold of at least four signals per nucleus, concordance between PathVysion and INFORM was only 76%, with a sensitivity of 100% but a specificity of 57% (13 false positives). Application of the thresholds of more than five or at least six signals per nucleus to the INFORM results increased concordance with PathVysion to 91% and 98%, respectively. The small amount of discordance was due to CEP17 polysomy, except in one unexplained case. As with CISH, we advise a control with a dual-probe FISH assay for borderline INFORM cases (five to ten signals per nucleus). Our results compare well with those recently published on an evaluation of three scoring methods for FISH: dual-probe HER2:CEP17 or single-probe HER2 assay with a threshold of more than four or more than six signals per nucleus [20]. This study also recommended a threshold of more than six signals per nucleus for FISH or a dual probe HER2:CEP17 FISH assay for optimal HER2 scoring. This was based on high concordance rates with IHC or mRNA/assay/nucleic acid sequence-based amplification and are in agreement with the recent American Society of Clinical Oncology/College of American Pathologists guidelines for HER2 testing [21]. The theoretical advantage of dual-probe FISH analysis is the ability to distinguish chromosomal aneuploidy from amplification by using a differentially labelled reference probe (CEP17). Although a single probe-based assay, the main benefit of INFORM over other assays is that it is an automated system.
Conclusion
We can draw several conclusions from this study. First, we can validate CISH as a promising alternative to FISH for determining gene amplification, because it is easy to interpret and the equipment can be readily found in most pathology departments. The potential of CISH as a HER2 status assay for routine practice has been the subject of an international multicentre ring study, to which we have contributed, to validate CISH against the current 'gold standards' [22]. Second, the automated mono-probe assay INFORM provides the opportunity for a more rapid evaluation of samples compared with CISH and the two dual-probe FISH assays evaluated in this study, provided that an appropriate threshold is applied. Despite the clear benefits attributed to these two approaches, we would advise that borderline CISH or single-probe FISH results are controlled for by using a dual-probe FISH assay. The ideal situation for pathologists in the future would be for FDA approval of CISH in routine diagnosis and for automation of the assay. The algorithm currently employed would then change to that presented in Figure 1.
Abbreviations
CEP17 = chromosome enumeration probe 17; CISH = chromogenic in situ hybridisation; DAPI = 4,6-diamidino-2-phenylindole; FDA = Food and Drug Administration; FISH = fluorescence in situ hybridisation; FITC = fluorescein isothiocyanate; HER2 = human epidermal growth factor receptor 2; IHC = immunohistochemistry; NA = non-amplified.
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
AC and FP-L participated in the design of the study, performed the blind scoring of CISH and FISH samples, and the interpretation and statistical analysis of the results, and wrote the manuscript. FM and NL performed blind scoring of the IHC samples. All authors read and approved the final manuscript.
Acknowledgments
Acknowledgements
The authors thank Maureen Lijour, Christiane Jullien and Matthieu Roche for their excellent technical assistance, and Georgina Collett for medical writing assistance during the preparation of this manuscript. We acknowledge the cooperation and support of Abbott, Zymed Laboratories, Ventana Medical Systems and Dako in the provision of technical information and reagents for this study. The study was supported by La Ligue Contre le Cancer du Puy de Dôme et du Cantal.
Contributor Information
Anne Cayre, Email: anne.cayre@cjp.fr.
Florence Mishellany, Email: florence.mishellany@cjp.fr.
Nicole Lagarde, Email: nicole.lagarde@chu-breast.fr.
Frédérique Penault-Llorca, Email: fpenault@cjp.fr.
References
- Ross JS, Fletcher JA. The HER-2/neu oncogene in breast cancer: prognostic factor, predictive factor, and target for therapy. Oncologist. 1998;3:237–252. [PubMed] [Google Scholar]
- Slamon DJ, Clark GM, Wong SG, Levin WJ, Ullrich A, McGuire WL. Human breast cancer: correlation of relapse and survival with amplification of the HER-2/neu oncogene. Science. 1987;235:177–182. doi: 10.1126/science.3798106. [DOI] [PubMed] [Google Scholar]
- Slamon DJ, Godolphin W, Jones LA, Holt JA, Wong SG, Keith DE, Levin WJ, Stuart SG, Udove J, Ullrich A, et al. Studies of the HER-2/neu proto-oncogene in human breast and ovarian cancer. Science. 1989;244:707–712. doi: 10.1126/science.2470152. [DOI] [PubMed] [Google Scholar]
- Cobleigh MA, Vogel CL, Tripathy D, Robert NJ, Scholl S, Fehrenbacher L, Wolter JM, Paton V, Shak S, Lieberman G, Slamon DJ. Multinational study of the efficacy and safety of humanized anti-HER2 monoclonal antibody in women who have HER2-overexpressing metastatic breast cancer that has progressed after chemotherapy for metastatic disease. J Clin Oncol. 1999;17:2639–2648. doi: 10.1200/JCO.1999.17.9.2639. [DOI] [PubMed] [Google Scholar]
- Vogel CL, Cobleigh MA, Tripathy D, Gutheil JC, Harris LN, Fehrenbacher L, Slamon DJ, Murphy M, Novotny WF, Burchmore M, et al. Efficacy and safety of trastuzumab as a single agent in first-line treatment of HER2-overexpressing metastatic breast cancer. J Clin Oncol. 2002;20:719–726. doi: 10.1200/JCO.20.3.719. [DOI] [PubMed] [Google Scholar]
- Marty M, Cognetti F, Maraninchi D, Snyder R, Mauriac L, Tubiana-Hulin M, Chan S, Grimes D, Antón A, Lluch A, et al. Randomized phase II trial of the efficacy and safety of trastuzumab combined with docetaxel in patients with human epidermal growth factor receptor 2-positive metastatic breast cancer administered as first-line treatment: the M77001 study group. J Clin Oncol. 2005;23:4265–4274. doi: 10.1200/JCO.2005.04.173. [DOI] [PubMed] [Google Scholar]
- Piccart-Gebhart MJ, Procter M, Leyland-Jones B, Goldhirsch A, Untch M, Smith I, Gianni L, Baselga J, Bell R, Jackisch C, et al. Trastuzumab after adjuvant chemotherapy in HER2-positive breast cancer. N Engl J Med. 2005;353:1659–1672. doi: 10.1056/NEJMoa052306. [DOI] [PubMed] [Google Scholar]
- Press MF, Slamon DJ, Flom KJ, Park J, Zhou J-Y, Bernstein L. Evaluation of HER-2/neu gene amplification and overexpression: comparison of frequently used assay methods in a molecularly characterized cohort of breast cancer specimens. J Clin Oncol. 2002;20:3095–3105. doi: 10.1200/JCO.2002.09.094. [DOI] [PubMed] [Google Scholar]
- Zarbo RJ, Hammond ME. Conference summary, strategic science symposium. Her-2/neu testing of breast cancer patients in clinical practice. Arch Pathol Lab Med. 2003;127:549–553. doi: 10.5858/2003-127-0549-CSSSS. [DOI] [PubMed] [Google Scholar]
- Ellis IO, Bartlett J, Dowsett M, Humphreys S, Jasani B, Miller K, Pinder SE, Rhodes A, Walker R. Best Practice No 176. Updated recommendations for HER2 testing in the UK. J Clin Pathol. 2004;57:233–237. doi: 10.1136/jcp.2003.007724. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vincent-Salomon A, MacGrogan G, Couturier J, Arnould L, Denoux Y, Fiche M, Jacquemier J, Mathieu MC, Penault-Llorca F, Rigaud C, et al. Calibration of immunohistochemistry for assessment of HER2 in breast cancer: results of the French multicentre GEFPICS study. Histopathology. 2003;42:337–347. doi: 10.1046/j.1365-2559.2003.01598.x. [DOI] [PubMed] [Google Scholar]
- Hammock L, Lewis M, Phillips C, Cohen C. Strong HER-2/neu protein overexpression by immunohistochemistry often does not predict oncogene amplification by fluorescence in situ hybridization. Hum Pathol. 2003;34:1043–1047. doi: 10.1053/S0046-8177(03)00409-X. [DOI] [PubMed] [Google Scholar]
- Roche PC, Ingle JN. Increased HER2 with U.S. Food and Drug Administration-approved antibody [letter] J Clin Oncol. 1999;17:434. doi: 10.1200/JCO.1999.17.1.434. [DOI] [PubMed] [Google Scholar]
- Tubbs RR, Pettay JD, Roche PC, Stoler MH, Jenkins RB, Grogan TM. Discrepancies in clinical laboratory testing of eligibility for trastuzumab therapy: apparent immunohistochemical false-positives do not get the message. J Clin Oncol. 2001;19:2714–2721. doi: 10.1200/JCO.2001.19.10.2714. [DOI] [PubMed] [Google Scholar]
- Arnould L, Denoux Y, MacGrogan G, Penault-Llorca F, Fiche M, Treilleux I, Mathieu MC, Vincent-Salomon A, Vilain MO, Couturier J. Agreement between chromogenic in situ hybridisation (CISH) and FISH in the determination of HER2 status in breast cancer. Br J Cancer. 2003;88:1587–1591. doi: 10.1038/sj.bjc.6600943. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gupta D, Middleton LP, Whitaker MJ, Abrams J. Comparison of fluorescence and chromogenic in situ hybridization for detection of HER-2/neu oncogene in breast cancer. Am J Clin Pathol. 2003;119:381–387. doi: 10.1309/P40P-2EAD-42PU-KDMG. [DOI] [PubMed] [Google Scholar]
- Isola J, Tanner M, Forsyth A, Cooke TG, Watters AD, Bartlett JM. Interlaboratory comparison of HER-2 oncogene amplification as detected by chromogenic and fluorescence in situ hybridization. Clin Cancer Res. 2004;10:4793–4798. doi: 10.1158/1078-0432.CCR-0428-03. [DOI] [PubMed] [Google Scholar]
- Tanner M, Gancberg D, Di Leo A, Larsimont D, Rouas G, Piccart MJ, Isola J. Chromogenic in situ hybridization. A practical alternative for fluorescence in situ hybridization to detect HER-2/neu oncogene amplification in archival breast cancer samples. Am J Pathol. 2000;157:1467–1472. doi: 10.1016/S0002-9440(10)64785-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhao J, Wu R, Au A, Marquez A, Yu Y, Shi Z. Determination of HER2 gene amplification by chromogenic in situ hybridization (CISH) in archival breast carcinoma. Mod Pathol. 2002;15:657–665. doi: 10.1038/modpathol.3880582. [DOI] [PubMed] [Google Scholar]
- Dal Lago L, Durbecq V, Desmedt C, Salgado R, Verjat T, Lespagnard L, Ma Y, Veys I, Di Leo A, Sotiriou C, et al. Correction for chromosome-17 is critical for the determination of true Her-2/neu gene amplification status in breast cancer. Mol Cancer Ther. 2006;5:2572–2579. doi: 10.1158/1535-7163.MCT-06-0129. [DOI] [PubMed] [Google Scholar]
- Wolff AC, Hammond MEH, Schwartz JN, Hagerty KL, Allred DC, Cote RJ, Dowsett M, Fitzgibbons PL, Hanna WM, Langer A, et al. American Society of Clinical Oncology/College of American Pathologists guideline recommendations for human epidermal growth factor receptor 2 testing in breast cancer. J Clin Oncol. 2007;25:118–145. doi: 10.1200/JCO.2006.09.2775. [DOI] [PubMed] [Google Scholar]
- van de Vijver M, Bilous M, Hanna W, Hofmann M, Kristel P, Penault-Llorca F, Rüschoff J. Chromogenic in-situ hybridisation for the assessment of HER2 status in breast cancer: an international validation ring study. Breast Cancer Res. 2007. [DOI] [PMC free article] [PubMed]