Use of UV ParaLens adapter for detection of acid-fast organisms.

K V Patterson 1, C L McDonald 1, B F Miller 1, K C Chapin 1
PMCID: PMC227919  PMID: 7535314

Abstract

Auramine-stained mycobacterial smears from 136 clinical specimens were interpreted by using the UV ParaLens adapter (Beckton Dickinson), and results were compared with smear interpretations using a traditional fluorescent microscope and culture. The sensitivity and specificity of the ParaLens were 84 and 93%, respectively. Smears yielding discrepant results were overstained by the Kinyoun method. Overall, the sensitivity of auramine-stained smears interpreted with the UV ParaLens was comparable to that of Kinyoun-stained smears.

Full Text

The Full Text of this article is available as a PDF (156.6 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Makler M. T. Fluorescent microscope objective. Trans R Soc Trop Med Hyg. 1992 Jan-Feb;86(1):108–108. doi: 10.1016/0035-9203(92)90465-o. [DOI] [PubMed] [Google Scholar]
  2. Mattia A. R., Waldron M. A., Sierra L. S. Use of the UV ParaLens adapter as an alternative to conventional fluorescence microscopy for detection of Pneumocystis carinii in direct immunofluorescent monoclonal antibody-stained pulmonary specimens. J Clin Microbiol. 1993 Mar;31(3):720–721. doi: 10.1128/jcm.31.3.720-721.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Polsuwan C., Lumlertdaecha B., Tepsumethanon W., Wilde H. Using the UV ParaLens adapter on a standard laboratory microscope for fluorescent rabies antibody detection. Trans R Soc Trop Med Hyg. 1992 Jan-Feb;86(1):107–107. doi: 10.1016/0035-9203(92)90464-n. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES