Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1995 Apr;33(4):978–981. doi: 10.1128/jcm.33.4.978-981.1995

Controlled evaluation of BacT/Alert standard aerobic and FAN aerobic blood culture bottles for detection of bacteremia and fungemia.

M P Weinstein 1, S Mirrett 1, L G Reimer 1, M L Wilson 1, S Smith-Elekes 1, C R Chuard 1, K L Joho 1, L B Reller 1
PMCID: PMC228079  PMID: 7790471

Abstract

A new medium, FAN, designed to enhance the recovery of microorganisms, has been developed for the BacT/Alert blood culture system (Organon Teknika Corp., Durham, N.C.). We compared the yield and speed of detection of microorganisms in 6,847 adequately filled paired aerobic standard and FAN bottles at four university hospitals. Of 499 clinically significant microorganisms isolated from one or both bottles, significantly more Staphylococcus aureus isolates (P < 0.001), coagulase-negative staphylococci (P < 0.001), yeasts (P < 0.01), and all microorganisms combined (P < 0.001) were recovered from the FAN bottles. Overall, the speeds of detection of positive cultures did not differ between the two medium formulations; mean times to detection in the standard and FAN bottles were 16.1 and 16.0 h, respectively. When a subset of patients on antimicrobial therapy was evaluated, significantly enhanced yield from the FAN bottle was evident for staphylococci. Overall, the FAN bottle outperformed the standard aerobic BacT/Alert bottle.

Full Text

The Full Text of this article is available as a PDF (169.1 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Appelbaum P. C., Beckwith D. G., Dipersio J. R., Dyke J. W., Salventi J. F., Stone L. L. Enhanced detection of bacteremia with a new BACTEC resin blood culture medium. J Clin Microbiol. 1983 Jan;17(1):48–51. doi: 10.1128/jcm.17.1.48-51.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Appleman M. D., Swinney R. S., Heseltine P. N. Evaluation of the Antibiotic Removal Device. J Clin Microbiol. 1982 Feb;15(2):278–281. doi: 10.1128/jcm.15.2.278-281.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Bille J., Stockman L., Roberts G. D. Detection of yeasts and filamentous fungi in blood cultures during a 10-year period (1972 to 1981). J Clin Microbiol. 1982 Nov;16(5):968–970. doi: 10.1128/jcm.16.5.968-970.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Hall M. M., Mueske C. A., Ilstrup D. M., Washington J. A., 2nd Evaluation of a biphasic medium for blood cultures. J Clin Microbiol. 1979 Nov;10(5):673–676. doi: 10.1128/jcm.10.5.673-676.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Krisher K. K., Whyburn D. R., Koepnick F. E. Comparison of the BacT/Alert pediatric blood culture system, Pedi-BacT, with conventional culture using the 20-milliliter Becton-Dickinson supplemented peptone broth tube. J Clin Microbiol. 1993 Apr;31(4):793–797. doi: 10.1128/jcm.31.4.793-797.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Thorpe T. C., Wilson M. L., Turner J. E., DiGuiseppi J. L., Willert M., Mirrett S., Reller L. B. BacT/Alert: an automated colorimetric microbial detection system. J Clin Microbiol. 1990 Jul;28(7):1608–1612. doi: 10.1128/jcm.28.7.1608-1612.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Weinstein M. P., Mirrett S., Wilson M. L., Harrell L. J., Stratton C. W., Reller L. B. Controlled evaluation of BACTEC Plus 26 and Roche Septi-Chek aerobic blood culture bottles. J Clin Microbiol. 1991 May;29(5):879–882. doi: 10.1128/jcm.29.5.879-882.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Weinstein M. P., Mirrett S., Wilson M. L., Reimer L. G., Reller L. B. Controlled evaluation of 5 versus 10 milliliters of blood cultured in aerobic BacT/Alert blood culture bottles. J Clin Microbiol. 1994 Sep;32(9):2103–2106. doi: 10.1128/jcm.32.9.2103-2106.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Weinstein M. P., Reller L. B., Murphy J. R., Lichtenstein K. A. The clinical significance of positive blood cultures: a comprehensive analysis of 500 episodes of bacteremia and fungemia in adults. I. Laboratory and epidemiologic observations. Rev Infect Dis. 1983 Jan-Feb;5(1):35–53. doi: 10.1093/clinids/5.1.35. [DOI] [PubMed] [Google Scholar]
  10. Wilson M. L., Davis T. E., Mirrett S., Reynolds J., Fuller D., Allen S. D., Flint K. K., Koontz F., Reller L. B. Controlled comparison of the BACTEC high-blood-volume fungal medium, BACTEC Plus 26 aerobic blood culture bottle, and 10-milliliter isolator blood culture system for detection of fungemia and bacteremia. J Clin Microbiol. 1993 Apr;31(4):865–871. doi: 10.1128/jcm.31.4.865-871.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Wilson M. L., Harrell L. J., Mirrett S., Weinstein M. P., Stratton C. W., Reller L. B. Controlled evaluation of BACTEC PLUS 27 and Roche Septi-Chek anaerobic blood culture bottles. J Clin Microbiol. 1992 Jan;30(1):63–66. doi: 10.1128/jcm.30.1.63-66.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Wilson M. L., Weinstein M. P., Reimer L. G., Mirrett S., Reller L. B. Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems. J Clin Microbiol. 1992 Feb;30(2):323–329. doi: 10.1128/jcm.30.2.323-329.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)