. 1995 Jul;33(7):1966–1967. doi: 10.1128/jcm.33.7.1966-1967.1995

Errors arising from incorrect orientation of E test strips.

J M Hamilton-Miller 1, S Shah 1, T S Yam 1
PMCID: PMC228315  PMID: 7665686

Abstract

Major errors arise if E test strips are placed upside down. Asymmetric zones, or no zone at all, may result. MICs indicated by upside-down tests were almost always considerably higher than true values. This situation differs markedly from that for conventional testing, where orientation of disks is not important.