Multicenter evaluation of the MicroScan Rapid Gram-Negative Identification Type 3 Panel.

S Bascomb 1, S L Abbott 1, J D Bobolis 1, D A Bruckner 1, S J Connell 1, S K Cullen 1, M Daugherty 1, D Glenn 1, J M Janda 1, S J Lentsch 1, D Lindquist 1, P B Mayhew 1, D M Nothaft 1, J R Skinner 1, G B Williams 1, J Wong 1, B L Zimmer 1
PMCID: PMC230005  PMID: 9316902

Abstract

The accuracy and performance of the revised MicroScan Rapid Gram-Negative Identification Type 3 Panel (Dade MicroScan Inc., West Sacramento, Calif.) were examined in a multicenter evaluation. The revised panel database includes data for 119 taxa covering a total of 150 species, with data for 12 new species added. Testing was performed in three phases: the efficacy, challenge, and reproducibility testing phases. A total of 405 fresh and stock gram-negative isolates comprising 54 species were tested in the efficacy phase; 96.8% of these species were identified correctly in comparison to the identification obtained either with the API 20E system (bioMérieux Vitek, Hazelwood, Mo.) or by the conventional tube method. The number of correctly identified isolates in the challenge phase, including new species added to the database, was 221 of 247, or 89.5%, in comparison to the number correctly identified by the conventional tube method. A total of 465 isolates were examined for intra- and interlaboratory identification reproducibility and gave an agreement of 464 of 465, or 99.8%. The overall reproducibility of each individual identification test or substrate was 14,373 of 14,384, or 99.9%. The new Rapid Gram-Negative Identification Type 3 Panel gave accurate and highly reproducible results in this multiple-laboratory evaluation.

Full Text

The Full Text of this article is available as a PDF (120.0 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Abbott S. L., Hanson D. F., Felland T. D., Connell S., Shum A. H., Janda J. M. Escherichia coli O157:H7 generates a unique biochemical profile on MicroScan conventional gram-negative identification panels. J Clin Microbiol. 1994 Mar;32(3):823–824. doi: 10.1128/jcm.32.3.823-824.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Brenner D. J., Grimont P. A., Steigerwalt A. G., Fanning G. R., Ageron E., Riddle C. F. Classification of citrobacteria by DNA hybridization: designation of Citrobacter farmeri sp. nov., Citrobacter youngae sp. nov., Citrobacter braakii sp. nov., Citrobacter werkmanii sp. nov., Citrobacter sedlakii sp. nov., and three unnamed Citrobacter genomospecies. Int J Syst Bacteriol. 1993 Oct;43(4):645–658. doi: 10.1099/00207713-43-4-645. [DOI] [PubMed] [Google Scholar]
  3. Farmer J. J., 3rd, Davis B. R., Hickman-Brenner F. W., McWhorter A., Huntley-Carter G. P., Asbury M. A., Riddle C., Wathen-Grady H. G., Elias C., Fanning G. R. Biochemical identification of new species and biogroups of Enterobacteriaceae isolated from clinical specimens. J Clin Microbiol. 1985 Jan;21(1):46–76. doi: 10.1128/jcm.21.1.46-76.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Feng P. Escherichia coli serotype O157:H7: novel vehicles of infection and emergence of phenotypic variants. Emerg Infect Dis. 1995 Apr-Jun;1(2):47–52. doi: 10.3201/eid0102.950202. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Manafi M., Kneifel W., Bascomb S. Fluorogenic and chromogenic substrates used in bacterial diagnostics. Microbiol Rev. 1991 Sep;55(3):335–348. doi: 10.1128/mr.55.3.335-348.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. O'Hara C. M., Miller J. M. Evaluation of the autoSCAN-W/A system for rapid (2-hour) identification of members of the family Enterobacteriaceae. J Clin Microbiol. 1992 Jun;30(6):1541–1543. doi: 10.1128/jcm.30.6.1541-1543.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. O'Hara C. M., Tenover F. C., Miller J. M. Parallel comparison of accuracy of API 20E, Vitek GNI, MicroScan Walk/Away Rapid ID, and Becton Dickinson Cobas Micro ID-E/NF for identification of members of the family Enterobacteriaceae and common gram-negative, non-glucose-fermenting bacilli. J Clin Microbiol. 1993 Dec;31(12):3165–3169. doi: 10.1128/jcm.31.12.3165-3169.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Pfaller M. A., Sahm D., O'Hara C., Ciaglia C., Yu M., Yamane N., Scharnweber G., Rhoden D. Comparison of the autoSCAN-W/A rapid bacterial identification system and the Vitek AutoMicrobic system for identification of gram-negative bacilli. J Clin Microbiol. 1991 Jul;29(7):1422–1428. doi: 10.1128/jcm.29.7.1422-1428.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Tenover F. C., Mizuki T. S., Carlson L. G. Evaluation of autoSCAN-W/A automated microbiology system for the identification of non-glucose-fermenting gram-negative bacilli. J Clin Microbiol. 1990 Jul;28(7):1628–1634. doi: 10.1128/jcm.28.7.1628-1634.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Trenholme G. M., Kaplan R. L., Karakusis P. H., Stine T., Fuhrer J., Landau W., Levin S. Clinical impact of rapid identification and susceptibility testing of bacterial blood culture isolates. J Clin Microbiol. 1989 Jun;27(6):1342–1345. doi: 10.1128/jcm.27.6.1342-1345.1989. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES