Abstract
The MB/BacT automated system is designed for the isolation of mycobacteria from clinical specimens. It utilizes a colorimetric sensor and reflected light to continuously monitor the CO2 concentration in the culture medium. We compared its performance to that of the BACTEC 12B media for the radiometric BACTEC 460 instrument and that of solid culture media. Respiratory specimens and urine samples were decontaminated with 4% NaOH. The vials of the two instruments were inoculated with 500 microl of sample and two solid egg-based media at 200 microl each. All vials were incubated at 37 degrees C for 6 weeks. A total of 1,078 specimens (633 respiratory specimens, 78 cerebrospinal fluid specimens, 177 other body fluid specimens, 87 urine specimens, and 103 other types of specimens) were cultured in parallel. Mycobacteria could be identified from 73 (6.8%) specimens: 67 M. tuberculosis, 3 M. kansasii, 1 M. xenopi, 1 M. terrae, and 1 mixed M. avium with M. scrofulaceum. Of these, 63 (86.3%) specimens were positive with the MB/BacT system, 67 (91.8%) were positive with the BACTEC 460 instrument, and 58 (79.5%) were positive with the two egg-based media. MB/BacT cultures were positive on average after 17.5 (+/-6.4) days, BACTEC cultures with a growth index of >20 (mean, 200) were positive after 14.3 (+/-8.2) days, and egg-based media were positive after 24.2 (+/-7.5) days. Microorganisms other than mycobacteria contaminated 46 (4.3%) MB/BacT cultures and 31 (2.9%) BACTEC cultures, which had to be discarded. The MB/BacT system is a well-automated system for the detection of M. tuberculosis in clinical specimens without using radioactive reagents. Further trials are required to determine whether it is suitable for the culture of nontuberculous mycobacteria.
Full Text
The Full Text of this article is available as a PDF (144.0 KB).
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Badak F. Z., Kiska D. L., Setterquist S., Hartley C., O'Connell M. A., Hopfer R. L. Comparison of mycobacteria growth indicator tube with BACTEC 460 for detection and recovery of mycobacteria from clinical specimens. J Clin Microbiol. 1996 Sep;34(9):2236–2239. doi: 10.1128/jcm.34.9.2236-2239.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Böddinghaus B., Rogall T., Flohr T., Blöcker H., Böttger E. C. Detection and identification of mycobacteria by amplification of rRNA. J Clin Microbiol. 1990 Aug;28(8):1751–1759. doi: 10.1128/jcm.28.8.1751-1759.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- D'Amato R. F., Isenberg H. D., Hochstein L., Mastellone A. J., Alperstein P. Evaluation of the Roche Septi-Chek AFB system for recovery of mycobacteria. J Clin Microbiol. 1991 Dec;29(12):2906–2908. doi: 10.1128/jcm.29.12.2906-2908.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hoffner S. E. Improved detection of Mycobacterium avium complex with the Bactec radiometric system. Diagn Microbiol Infect Dis. 1988 May;10(1):1–6. doi: 10.1016/0732-8893(88)90121-6. [DOI] [PubMed] [Google Scholar]
- Huebner R. E., Good R. C., Tokars J. I. Current practices in mycobacteriology: results of a survey of state public health laboratories. J Clin Microbiol. 1993 Apr;31(4):771–775. doi: 10.1128/jcm.31.4.771-775.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Isenberg H. D., D'Amato R. F., Heifets L., Murray P. R., Scardamaglia M., Jacobs M. C., Alperstein P., Niles A. Collaborative feasibility study of a biphasic system (Roche Septi-Chek AFB) for rapid detection and isolation of mycobacteria. J Clin Microbiol. 1991 Aug;29(8):1719–1722. doi: 10.1128/jcm.29.8.1719-1722.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jorgensen J. H., Mirrett S., McDonald L. C., Murray P. R., Weinstein M. P., Fune J., Trippy C. W., Masterson M., Reller L. B. Controlled clinical laboratory comparison of BACTEC plus aerobic/F resin medium with BacT/Alert aerobic FAN medium for detection of bacteremia and fungemia. J Clin Microbiol. 1997 Jan;35(1):53–58. doi: 10.1128/jcm.35.1.53-58.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Middlebrook G., Reggiardo Z., Tigertt W. D. Automatable radiometric detection of growth of Mycobacterium tuberculosis in selective media. Am Rev Respir Dis. 1977 Jun;115(6):1066–1069. doi: 10.1164/arrd.1977.115.6.1066. [DOI] [PubMed] [Google Scholar]
- Morgan M. A., Horstmeier C. D., DeYoung D. R., Roberts G. D. Comparison of a radiometric method (BACTEC) and conventional culture media for recovery of mycobacteria from smear-negative specimens. J Clin Microbiol. 1983 Aug;18(2):384–388. doi: 10.1128/jcm.18.2.384-388.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mosca Adriana, D'Alagni Marina, Del Prete Raffaele, Simone Anna, De Santis Antonio, Miragliotta Giuseppe. Rapid recovery of Mycobacterium tuberculosis complex from clinical specimens using the BACTEC 9000 MB system, a new automated fluorimetric technique. Clin Microbiol Infect. 1997 Jun;3(3):352–355. doi: 10.1111/j.1469-0691.1997.tb00625.x. [DOI] [PubMed] [Google Scholar]
- Pfyffer G. E., Cieslak C., Welscher H. M., Kissling P., Rüsch-Gerdes S. Rapid detection of mycobacteria in clinical specimens by using the automated BACTEC 9000 MB system and comparison with radiometric and solid-culture systems. J Clin Microbiol. 1997 Sep;35(9):2229–2234. doi: 10.1128/jcm.35.9.2229-2234.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pfyffer G. E., Welscher H. M., Kissling P., Cieslak C., Casal M. J., Gutierrez J., Rüsch-Gerdes S. Comparison of the Mycobacteria Growth Indicator Tube (MGIT) with radiometric and solid culture for recovery of acid-fast bacilli. J Clin Microbiol. 1997 Feb;35(2):364–368. doi: 10.1128/jcm.35.2.364-368.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Roberts G. D., Goodman N. L., Heifets L., Larsh H. W., Lindner T. H., McClatchy J. K., McGinnis M. R., Siddiqi S. H., Wright P. Evaluation of the BACTEC radiometric method for recovery of mycobacteria and drug susceptibility testing of Mycobacterium tuberculosis from acid-fast smear-positive specimens. J Clin Microbiol. 1983 Sep;18(3):689–696. doi: 10.1128/jcm.18.3.689-696.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sewell D. L., Rashad A. L., Rourke W. J., Jr, Poor S. L., McCarthy J. A., Pfaller M. A. Comparison of the Septi-Chek AFB and BACTEC systems and conventional culture for recovery of mycobacteria. J Clin Microbiol. 1993 Oct;31(10):2689–2691. doi: 10.1128/jcm.31.10.2689-2691.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Styrt B. A., Shinnick T. M., Ridderhof J. C., Crawford J. T., Tenover F. C. Turnaround times for mycobacterial cultures. J Clin Microbiol. 1997 Apr;35(4):1041–1042. doi: 10.1128/jcm.35.4.1041-1042.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wilson M. L., Stone B. L., Hildred M. V., Reves R. R. Comparison of recovery rates for mycobacteria from BACTEC 12B vials, Middlebrook 7H11-selective 7H11 biplates, and Lowenstein Jensen slants in a public health mycobacteriology laboratory. J Clin Microbiol. 1995 Sep;33(9):2516–2518. doi: 10.1128/jcm.33.9.2516-2518.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wolinsky E. Conventional diagnostic methods for tuberculosis. Clin Infect Dis. 1994 Sep;19(3):396–401. doi: 10.1093/clinids/19.3.396. [DOI] [PubMed] [Google Scholar]
- Woods G. L., Fish G., Plaunt M., Murphy T. Clinical evaluation of difco ESP culture system II for growth and detection of mycobacteria. J Clin Microbiol. 1997 Jan;35(1):121–124. doi: 10.1128/jcm.35.1.121-124.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]