Abstract
Background
Coxiella burnetii is the causative agent of Q-fever, a widespread zoonosis. Due to its high environmental stability and infectivity it is regarded as a category B biological weapon agent. In domestic animals infection remains either asymptomatic or presents as infertility or abortion. Clinical presentation in humans can range from mild flu-like illness to acute pneumonia and hepatitis. Endocarditis represents the most common form of chronic Q-fever. In humans serology is the gold standard for diagnosis but is inadequate for early case detection. In order to serve as a diagnostic tool in an eventual biological weapon attack or in local epidemics we developed a real-time 5'nuclease based PCR assay with an internal control system. To facilitate high-throughput an automated extraction procedure was evaluated.
Results
To determine the minimum number of copies that are detectable at 95% chance probit analysis was used. Limit of detection in blood was 2,881 copies/ml [95%CI, 2,188–4,745 copies/ml] with a manual extraction procedure and 4,235 copies/ml [95%CI, 3,143–7,428 copies/ml] with a fully automated extraction procedure, respectively. To demonstrate clinical application a total of 72 specimens of animal origin were compared with respect to manual and automated extraction. A strong correlation between both methods was observed rendering both methods suitable. Testing of 247 follow up specimens of animal origin from a local Q-fever epidemic rendered real-time PCR more sensitive than conventional PCR.
Conclusion
A sensitive and thoroughly evaluated real-time PCR was established. Its high-throughput mode may show a useful approach to rapidly screen samples in local outbreaks for other organisms relevant for humans or animals. Compared to a conventional PCR assay sensitivity of real-time PCR was higher after testing samples from a local Q-fever outbreak.
Background
Coxiella burnetii (C. burnetii) is an obligate intracellular, gram negative bacterium. It is the causative agent of Q-fever. Q-fever is a zoonosis with a worldwide distribution except New Zealand that affects different animal species and humans. Clinical presentation in humans ranges from mild flu-like symptoms to, sometimes, severe atypical pneumonia and hepatitis [1]. Convalescence can be slow and endocarditis is the most frequent and serious manifestation of chronic Q-fever [2]. In animals, primarily cattle, sheep, and goats, C. burnetii can cause abortion and infertility as it localizes in the female reproductive system. High doses of C. burnetii have been found in conception products of infected animals. The organism is shed in the urine, feces and milk of infected animals. In general infected animals remain asymptomatic. Instead they often serve as the source of infection for humans via infective aerosols or contaminated dust [3]. C. burnetii is very resistant to environmental conditions and can remain infectious for a considerable time outside the host cell. Recent outbreaks in France documented the high environmental stability of the organism when local Q-fever epidemics were observed weeks after lambing season [4]. Due to its high infectivity, environmental stability and its potential to cause severe disease in humans it is regarded as a category B biological weapon agent by the Centers for Disease Control and Prevention [5]. Its natural widespread availability and potential for aerosolized use makes it considerably suitable as a biological weapon. Proper administration of antibiotics can significantly reduce chronic Q-fever associated mortality making timely diagnosis of utmost importance. For diagnosing C. burnetii infection serology remains the method of choice as it is easy to establish and widely applicable. However, antibodies are detected only after 2–3 weeks from the onset of disease [6] making it too slow in selected clinical settings. A capture enzyme-linked immunosorbent assay (EIA) can be used for direct detection of C. burnetii [7]. However, its high limit of detection significantly reduces its reliability. Direct detection of C. burnetii is also possible by cell culture, but this requires biosafety level-three laboratories. Sensitivity of cell culture is sometimes low [8]. More recently, PCR has been successfully applied for the direct detection of C. burnetii in clinical specimens [9]. Though it appears to be highly sensitive, conventional PCR protocols remain time-consuming due to laborious post PCR processing, and they are prone to cross-contamination. Modern real-time PCR assays with in tube detection of amplicons decrease turn around time considerably [10]. In a bioterrorist event or in the case of local epidemics, masses of samples have to be expected. We have shown for other agents that real-time PCR provides the technical prerequisites for high throughput testing [11].
Here we describe a novel 5'nuclease (TaqMan) based real-time PCR assay for the rapid, sensitive and specific detection of C. burnetii. A mimic positive control monitors the reaction under the same conditions as applicable for C. burnetii, including use of the same primers. It identifies breaches in sensitivity in each single sample due to insufficient sample preparation, PCR inhibition or inherent failure of the PCR itself. To further facilitate high-throughput application a fully automated extraction procedure using the BioRobot M48 (Qiagen, Hilden, Germany) was evaluated and compared to an established manual sample preparation method.
Results
Coxiella burnetii real-time PCR
Since the transposase gene of C. burnetii is present in approximately 20 copies per cell it was chosen as the target sequence. Using Primer Express software two primer pairs and one 5'nuclease minor groove binder (MGB) probe were selected. Prepared nucleic acid from a cultured C. burnetii (Nine mile RSA493) strain was used to optimize the assay. The 86 bp amplicon of the transposase gene was cloned into E. coli plasmids (pCoxquant), and used for sensitivity determination. Single copies were detected in limiting dilution series on an occasional basis. Constantly positive results were obtained at a concentration of 15 copies per PCR reaction or more.
To monitor the sensitivity of the assay in each single reaction an internal control mimic DNA was constructed next [12]. The probe binding site of pCoxquant was deleted and replaced by an alternative sequence to yield plasmid pCoxmimic. Its corresponding 5'nuclease probe contained the dye VIC for detection in a separate channel of the real-time PCR instrument. One copy of the internal control mimic pCoxmimic was occasionally and 20 copies constantly detectable. Cross talk into the wild type channel was not observed. To exclude that the internal control affected the amplification efficiency of C. burnetii detection, 15 copies/reaction of pCoxquant were amplified in the presence of increasing numbers of pCoxmimic. The cycle threshold (Ct) values for pCoxquant and pCoxmimic were recorded separately for each target gene (Figure 1C). The simultaneous amplification of up to 100 copies of internal control did not influence the Ct for pCoxquant. Only from 700 copies of internal control onward, variation and delay in Ct for pCoxquant occurred. A concentration of 20 copies of pCoxmimic per reaction was chosen as a working concentration for all further experiments, in order to detect even slight drops in assay sensitivity. The exact limit of detection of real-time PCR was determined next. Human EDTA blood was spiked with plasmid pCoxquant in five different concentrations. EDTA blood of each concentration was first extracted manually by means of a QIAmp DNA Mini Kit (Qiagen) in duplicates, and each duplicate was tested in replicates of 4 (5 × 2 × 4 = 40 reactions). Prior to extraction the lysis buffer was spiked with pCoxmimic at a concentration corresponding to 20 copies per reaction. The observed proportions of positive results in each concentration were subjected to probit regression analysis (Figure 1A). 2,881 copies per ml were calculated to be detectable at ≥95% chance (95%CI, 2,188–4,745). This corresponded to 14 copies/PCR reaction.
For automated extraction, the Biorobot M48 (Qiagen) was evaluated exactly in the same manner (Figure 1B). 4,235 (95%CI, 3,143–7,428) copies per ml were calculated to be detectable with ≥95% chance, corresponding to 21 copies per PCR.
The internal control pCoxmimic was detected in all samples including the negative controls. Specificity was evaluated by testing a number of viral and bacterial pathogens that might be present in human blood samples. None of the tested pathogens reacted positive with the C. burnetii real-time PCR assay (see Methods section). To further assess the specificity 35 different C. burnetii strains were subjected to the new real-time assay. As expected all strains reacted positive (Table 1).
Table 1.
C. burnetii strain | Geographical origin | Host species |
Nine Mile USA | USA | Tick |
Priscilla USA | USA | Goat |
Scurry USA | USA | Human |
Dugway USA | USA | Rodent |
Z 2775 | Germany | Cattle |
Pohlheim | Germany | Sheep |
Max | Germany | Sheep |
Tiho 1 | Germany | unknown |
Hardthof/90 | Germany | Cattle |
Frankfurt | Germany | Cattle |
Z 104/94 | Germany | Sheep |
München | Germany | Sheep |
OSH-1 | Germany | Cattle |
Bru 180 | Germany | Cattle |
Wdk 1188 | Germany | Sheep |
Zeckenpool 11 | Germany | Tick |
Namibia | Namibia | Goat |
F-2 | France | Human |
F-4 | France | Human |
R1140 | Russia | Human |
CS-Florian | Slovakia | Human |
CS-Bud | Slovakia | Human |
CS-KL 4 | Slovakia | Tick |
CS-Dayer | Slovakia | Tick |
Utvinis | Romania | Human |
Stancia | Romania | Human |
Brasov | Romania | Human |
Balaceanu | Romania | Human |
J-3 | Japan | Cattle |
Henzerling | Italy | Human |
CS-R | Italy | Human |
Herzberg | Greece | Human |
Andelfingen | Switzerland | Cattle |
Soyta | Switzerland | Cattle |
Boren | unknown | Cattle |
Real-time PCR versus conventional PCR
To further assess the performance of real-time PCR we finally compared the assay to a conventional PCR. A total of 247 purified specimens were tested which were sampled during a follow-up study after a local Q-fever epidemic in North-Rhine Westphalia. Results are shown in Table 2 and Figure 2B. The mean Ct value for samples positive in both assays was 31.31 +/- 2.15 [29.16; 33.46], which equals 2.7 × 103 copies per PCR (see below). Samples that were positive by real-time PCR only yielded a mean Ct of 37.42 +/- 0.81 [36.61; 38.23], which equals 48 copies per PCR. Means were significantly different (t-test, p ≤ 0.05).
Table 2.
Real-time PCR positive | Real-time PCR negative | total | |
PCR positive | n = 38 | n = 5 | n = 43 |
PCR negative | n = 32 | n = 172 | n = 204 |
total | n = 70 | n = 177 |
Quantitative real-time PCR
As an example of applicability the novel real-time PCR assay was evaluated in combination with the manual as well as the automated extraction procedure on a panel of samples. Seventy-two homogenized tissue samples were purified using both extraction methods as described above. After purification of samples with the manual extraction method, 27/72 (37.5%) samples yielded a positive result by real-time PCR. With automated extraction 31/72 (43%) samples tested positive. The mean Ct value in positive samples for each procedure was 28.2 for the manual extraction and 27.65 for the automated procedure, respectively (t-test, p = 0.8028). Ct values correlated proportionally with DNA concentrations, and showed good accordance between manual and automatic methods (Figure 1D, correlation coefficient = 0.975).
For absolute quantification a calibration curve was generated by limiting dilution series of plasmid pCoxquant. The amplification efficiency calculated from the calibration curve slope was 0.98, indicating optimal PCR reaction condition. Assuming 20 copies of target gene per genome, concentrations of C. burnetii per milliliter of homogenized sample ranged from 9.4 × 104 genomes per mL to 3.2 × 1010 genomes per ml (470-1.6 × 108 genomes/reaction).
Cell culture has long been regarded the gold standard for diagnosing C. burnetii infection. To re-assess its performance, all 72 samples were also subjected to cell culture. Only 13 of 72 (18%) samples yielded a positive result by cell culture. All cell culture positive samples yielded a positive real-time PCR result irrespective of the purification method. Again the bacterial concentration per ml sample was determined and copy numbers of 27 real-time PCR positive samples were plotted in increasing increments and compared to cell culture results (Figure 2A). Only samples with >1 × 108 genomes/ml of homogenized sample yielded a positive result by cell culture.
Discussion
Recent experience has shown that in the event of a C. burnetii epidemic masses of samples are to be expected [13]. We describe here a C. burnetii real-time PCR assay in combination with manual as well as fully automated extraction procedures for high-throughput use. The assay is capable of detecting single copies of C. burnetii transposase gene. This target region was chosen because it is highly conserved among C. burnetii strains and present at 10–30 copies per bacterium, making it an appropriate target for a diagnostic assay [14-16]. However, despite the assay was validated on a range of C. burnetii isolates, strains lacking elements of the transposase gene have been described [17]. For our calculations of sensitivity we assumed 20 copies per genome as present in our Nine Mile RSA 493 reference strain [16]. At 2,881 DNA copies/ml blood, the analytical sensitivity of our assay was close to what is possible due to mathematical limitations, according to the probit model [18]. This sensitivity, corresponding to about 144 genomes per ml, was as high as with commercial PCR systems available for other bacterial agents and should be compatible with the reliable detection of bacteremia [19]. Of note, since the number of copies of the C. burnetii transposase gene can vary greatly different sensitivities have to be expected with other strains.
Due to increased hybridization properties and lower background fluorescence we decided to use a minor groove binder (MGB) probe in combination with a non-fluorescent-quencher (NFQ) [20]. The use of a NFQ improves the signal-to-noise ratio and decreases spectral overlap compared to other fluorescent quencher dyes e.g. TAMRA [21]. This feature facilitates double dye detection, which allows implementation of a stable internal control system. The internal control was detected in all negative samples irrespective of the extraction procedure, although only a low amount of internal control mimic was added (20 copies plasmid DNA per reaction). Since the internal control is almost identical to the C. burnetii target gene including additional positive controls is not necessary. The advantages of such controls have already been described and implemented in various assays [22,23].
Besides the PCR itself, the extraction procedure constitutes another crucial factor in molecular diagnostics. It is the lengthiest part of the whole procedure and automation is highly desirable especially in a high-throughput setting. The most valuable advantage of robotic systems is their speed when numerous samples have to be analyzed [24]. However, there is little data on the performance of automated systems in diagnostic settings. In this study the limit of detection of the automated procedure was only 1.5 times higher than that of the manual protocol (4,235 vs. 2,881 DNA copies/ml blood). Assuming 20 copies per bacterium (Nine mile RSA493 strain) the limit of detection for the automated procedure was around 210 C. burnetii bacteria per ml (1 bacterium per PCR reaction). High sensitivity and good quantitative correlation with other extraction methods is in concordance with our earlier study on the automated extraction of B. anthracis [11].
The novel assay yielded more positive results than conventional PCR upon testing of a large panel of stored samples of animal origin. However, it should be noted that 2.5 times more template was added to the real-time PCR reaction. Because PCR negative/real-time PCR positive samples had significantly lower Ct values than samples positive in both methods, improved analytical sensitivity can be assumed for real-time PCR. Of note 5 conventional PCR positive samples remained repeatedly negative upon real-time PCR testing. All of these 5 samples were stored for approximately 3 years at -20°C and reported to be weakly positive only by conventional PCR. Due to prolonged storage it seems likely that some DNA degradation had happened thus preventing positive results by real-time PCR.
Critically, classical isolation of C. burnetii by cell culture showed rather disappointing results. Henning and Sting have already reported similar results for cell culture [25]. The strain of C. burnetii as well as its source might account for the here observed rather low number of conventional cell culture positive results. More sensitive isolation methods like shell vial technique or inoculation of specimens into guinea pigs or mice were not available in our laboratory. By quantitative PCR we could show that rather high bacterial loads are necessary for cell culture to become positive. For this reason and because of the associated biological risks, cell culture is not the method of choice in the first run. However, it remains an essential tool for further pathogen characterization, and allows isolation of related agents which would go undetected by PCR.
Conclusion
In summary, this assay provides a homogenous tool for diagnostics in human and veterinary medicine. Since C. burnetii infection is still not well understood, quantitative PCR may yield new insights into the pathogenesis of the disease. Another future field of application of high throughput assays may comprise the screening of bulk milk samples, an issue which may be addressed by food safety authorities.
Methods
Reference strain
A cultured C. burnetii (Nine mile RSA493) strain was kindly donated by D. Raoult, Unite de Rickettsies, Universite de la Mediterranee, Marseille, France.
Samples
To evaluate two different extraction procedures 53 specimens were tested, which were sampled during a local Q-fever epidemic among sheep in North-Rhine Westphalia/Germany in 2003. All samples were placenta material. In addition 19 specimens collected independently by the Friedrich-Loeffler-Institute (FLI) from different other sources were analyzed. For specificity testing a panel of 35 C. burnetii strains was tested. In addition, 27 different bacteria and viruses which might be present in clinical samples were obtained from the American Type Culture Collection (Manassas, VA, USA), the German Collection of Microorganisms and Cell Cultures (DSMZ, Braunschweig, Germany) or our in house strain collection. For an additional part of the study 247 aliquots of previously purified DNA samples from a follow-up examination of placenta material after the local epidemic of Q-fever in North-Rhine Westphalia were available.
Cell culture
Isolation of C. burnetii was performed using Buffalo Green Monkey (BGM) cells. Cells were propagated in 25 cm2 plastic flasks with in UltraCulture medium (BioWitthaker, Walkersville, Maryland, USA) without supplements. Material of each sample (1 g) was homogenised using sterile mortar, sand, and cell culture medium. The supernatants were filtered through membrane filters (Minisart™ Sartorius, Göttingen, Germany) with pore diameters of 0.2 μm. A volume of 0.5 ml homogenate per flask was inoculated. Cell cultures were examined weekly by phase-contrast microscopy for inclusion bodies.
Sample preparation
Samples were first homogenized for cell culture as described above. For conventional PCR assays nucleic acids were prepared using the Puregene DNA Blood Isolation Kit (Gentra Systems, Minneapolis, USA). One mL of the homogenized sample was centrifuged at 12,000 g for 10 min. The pellet was resuspended in 600 μl lysis buffer and incubated at 80°C for 5 min. Further steps were as recommended by the manufacturer.
Conventional PCR
A conventional PCR with primers amplifying a 448 bp product of the transposase gene of C. burnetii (IS1111) was conducted. Amplification was carried out in a total reaction volume of 50 μl with 1× PCR Buffer, 2 mM MgCl2, 200 μM of each dNTP and 0.5 units of Taq Polymerase (Bioline, Luckenwalde, Germany), 0.5 μM of primer CoxP4 (ttaaggtgggctgcgtggtgatgg, nt positions 222–245 in GenBank accession M80806) (TIB-Molbiol, Berlin, Germany), 0.5 μM of primer CoxM9 (gcttcgtcccggttcaacaattgc, nt 669–648) (TIB-Molbiol) and 2 μl of total DNA. Thermal cycling involved 94°C for 9 min, followed by 5 cycles of 94°C 30 s, 75°C to 67°C 30 s with 2 °C decrements per cycle, 77°C 30 s; and 37 cycles of 94°C 30 s, 65°C 30 s, 77°C 30 s with a final elongation step at 77°C 2 min. PCR products were visualized by gel electrophoresis on a standard 1% agarose gel stained with ethidium bromide.
Sample preparation/real-time PCR assay
Total genomic DNA was extracted manually using the QIAmp DNA Minikit (Qiagen, Hilden, Germany) from aliquots of homogenized samples as described above. One hundred microliters of sample was added to 100 μl of buffer ATL (Qiagen). Twenty microliters of Qiagen Proteinase K were added, and incubated at 56°C for 30 min. Next 200 μl of buffer AL (Qiagen) was added, which was previously spiked with the C. burnetii internal control plasmid at a final concentration of 20 copies per PCR reaction. Further recommendations of the manufacturer were followed (200 μl elution volume). Alternatively, genomic DNA was extracted by an automated procedure using the BioRobot M48 (Qiagen) in combination with a MagAttract DNA Mini Kit (Qiagen). One hundred microliters of sample were added to 100 μl buffer G2 (Qiagen) to which the internal control was previously added at a concentration corresponding to 20 copies per PCR reaction. Ten microliters of Proteinase K solution (Qiagen) were added and incubated at 56°C for 30 min. Further procedure was as recommended by the manufacturer, setting the elution volume to 200 μl.
Plasmid standard (pCoxquant)
An 86 bp insert of the transposase gene of C. burnetii was ligated into plasmid vectors and cloned in E. coli by means of a pCR2.1-TOPO TA cloning kit (Invitrogen, Karlsruhe, Germany). Plasmids were purified with a NucleoSpin Plasmid kit (Macherey Nagel, Düren, Germany) and sequenced using the BigDye 3.1 terminator cycle sequencing chemistry (Applied Biosystems, Weiterstadt, Germany) on an automatic ABI 377 DNA sequencer (Applied Biosystems). Plasmid DNA content was measured spectrophotometrically.
Internal control plasmid (pCoxmimic)
The C. burnetii plasmid pCoxquant was used to generate an internal control plasmid. Gene splicing by overlap extension was carried out using primers IcS (atcgttcgttgagcgattagcagttgccaatttaaatcgtgatgccggat) and IcAs (aactgctaatcgctcaacgaacgatgcaaggttgatgcttatcgggctatc) to introduce an alternative probe-binding site at nucleotide positions 1266–1293 (GenBank accession M80806). Resulting constructs of the correct length were cloned by means of a pCR 2.1-TOPO TA cloning kit (Invitrogen) and processed as described above.
Oligonucleotide design
Using primer Express software with default settings for 5'nuclease (TaqMan) minor groove binder (MGB) probes (Applied Biosystems) two possible primer combinations and one probe were evaluated. GenBank accession M80806 served as the query sequence. The optimal primer/probe combination was experimentally determined by checkerboard titration first. In a next set of experiments individual primer concentration was optimized and finally the magnesium concentration adjusted.
Real-time 5'nuclease PCR
A 25 μl reaction volume contained 5 μl of DNA, 4 mM MgCl2, 1× Platinum Taq polymerase reaction buffer (Invitrogen), 200 μM of each dNTP, 0.8 μM primer CoxbS (gatagcccgataagcatcaac, nt position 1241–1261, GenBank accession M80806) (TIB-Molbiol), 0.8 μM primer CoxbAs (gcattcgtatatccggcatc, nt 1326–1307) (TIB-Molbiol), 0.3 μM probe CoxbMGB (tcatcaaggcaccaat, nt 1272–1287), 0.2 μM probe prCoxmutant (atcgttcgttgagcgattagcagtt) and 1 unit of Taq DNA polymerase. 5'nuclease probe CoxbMGB was labeled with 5'FAM and a 3' minor groove binder non-fluorescent quencher (Applied Biosystems), 5'nuclease probe prCoxmutant was labeled with 5'VIC and 3'Black Hole Quencher (Eurogentec, Seraing, Belgium). Cycling conditions in an ABI Prism 7000 machine (Applied Biosystems) were: 95°C/2 min, and 45 cycles of 95°C/15 sec, 60°C/30 sec. Data were analyzed with the Sequence detector software V 2.1 (Applied Biosystems).
Specificity panel
Bacillus cereus (ATCC 4313), Bacillus subtilis (ATCC 6633), Candida albicans (ATCC 10231), Chlamydophila pneumoniae (CWL-029H), Enterococcus faecalis (in house reference strain), Epstein-Barr virus (patient isolate), Escherichia coli (ATCC 25922), Hepatitis B virus (INSTAND ref. material 11019), Hepatitis C virus (1st International WHO NAT standard), Human cytomegalovirus (INSTAND ref. material 15005), Herpes simplex virus type 1 (INSTAND ref. material 13017), Human immunodefieciency virus type 1 (NL-43), Mycobacterium tuberculosis (in house reference strain), Orientia tsutsugamushi (in house reference strain), Plasmodium falciparum (patient isolate), Pseudomonas aerogenosa (in-house reference strain), Rickettsia conorii (in house reference strain), Rickettsia prowazeckii (in house reference strain), Rickettsia rickettsii (in house reference strain), Salmonella enteritidis (in house reference strain), Shigella sonnei (in house reference strain), Staphylococcus aureaus (ATCC 13565), Streptococcus pneumoniae (ATCC 6305), Streptococcus pyogenes (ATCC 19615)
Statistical analysis
Different input concentrations of C. burnetii plasmid DNA pCoxquant were tested to calculate the predicted proportion of positive results in replicate tests using probit analysis as a non-linear regression model. Statgraphics plus version 5.1 software was used for probit analysis (Manugistics, Dresden, Germany).
Authors' contributions
MP established the real-time PCR assay, coordinated the study and drafted the manuscript. JK, SG-F and MPE prepared and analyzed samples of the follow up examination. SK participated in performing real-time PCR experiments and provided the cross reactivity panel. DF and HM provided the specificity panel, performed experiments and drafted the manuscript. KH was responsible for cultivation of C. burnetii isolates and helped to design the study. CD participated in the design of the study, provided technical and financial support and helped to draft the manuscript. All authors read and approved the final manuscript.
Acknowledgments
Acknowledgements
This work was supported by grants from the Bundesministerium für Gesundheit und Soziales (No. 325-4539-85/3) and the Bundesamt für Bevölkerungsschutz und Katastrophenhilfe (BBK-F-440-00-1).
We thank Britta Liedigk for excellent technical assistance.
Contributor Information
Marcus Panning, Email: panning@virology-bonn.de.
Jochen Kilwinski, Email: Jochen.Kilwinski@SVUA-Arnsberg.NRW.DE.
Susanne Greiner-Fischer, Email: greiner-fischer@t-online.de.
Martin Peters, Email: peters@SVUA-ARNSBERG.NRW.DE.
Stefanie Kramme, Email: kramme@microbiology-bonn.de.
Dimitrios Frangoulidis, Email: DimitriosFrangoulidis@Bundeswehr.org.
Hermann Meyer, Email: Hermann1Meyer@Bundeswehr.org.
Klaus Henning, Email: Klaus.Henning@wus.bfav.de.
Christian Drosten, Email: drosten@virology-bonn.de.
References
- Raoult D, Marrie T, Mege J. Natural history and pathophysiology of Q fever. Lancet Infect Dis. 2005;5:219–226. doi: 10.1016/S1473-3099(05)70052-9. [DOI] [PubMed] [Google Scholar]
- Gami AS, Antonios VS, Thompson RL, Chaliki HP, Ammash NM. Q fever endocarditis in the United States. Mayo Clin Proc. 2004;79:253–257. doi: 10.4065/79.2.253. [DOI] [PubMed] [Google Scholar]
- Maurin M, Raoult D. Q fever. Clin Microbiol Rev. 1999;12:518–553. doi: 10.1128/cmr.12.4.518. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tissot-Dupont H, Amadei MA, Nezri M, Raoult D. Wind in November, Q fever in December. Emerg Infect Dis. 2004;10:1264–1269. doi: 10.3201/eid1007.030724. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Madariaga MG, Rezai K, Trenholme GM, Weinstein RA. Q fever: a biological weapon in your backyard. Lancet Infect Dis. 2003;3:709–721. doi: 10.1016/S1473-3099(03)00804-1. [DOI] [PubMed] [Google Scholar]
- Fournier PE, Raoult D. Comparison of PCR and serology assays for early diagnosis of acute Q fever. J Clin Microbiol. 2003;41:5094–5098. doi: 10.1128/JCM.41.11.5094-5098.2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thiele D, Karo M, Krauss H. Monoclonal antibody based capture ELISA/ELIFA for detection of Coxiella burnetii in clinical specimens. Eur J Epidemiol. 1992;8:568–574. doi: 10.1007/BF00146378. [DOI] [PubMed] [Google Scholar]
- Musso D, Raoult D. Coxiella burnetii blood cultures from acute and chronic Q-fever patients. J Clin Microbiol. 1995;33:3129–3132. doi: 10.1128/jcm.33.12.3129-3132.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fenollar F, Fournier PE, Raoult D. Molecular detection of Coxiella burnetii in the sera of patients with Q fever endocarditis or vascular infection. J Clin Microbiol. 2004;42:4919–4924. doi: 10.1128/JCM.42.11.4919-4924.2004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Harris RJ, Storm PA, Lloyd A, Arens M, Marmion BP. Long-term persistence of Coxiella burnetii in the host after primary Q fever. Epidemiol Infect. 2000;124:543–549. doi: 10.1017/S0950268899003763. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Panning M, Kramme S, Petersen N, Drosten C. High throughput screening for spores and vegetative forms of pathogenic B. anthracis by an internally controlled real-time PCR assay with automated DNA preparation. Med Microbiol Immunol (Berl) 2007;196:41–50. doi: 10.1007/s00430-006-0029-7. [DOI] [PubMed] [Google Scholar]
- Courtney BC, Smith MM, Henchal EA. Development of internal controls for probe-based nucleic acid diagnostic assays. Anal Biochem. 1999;270:249–256. doi: 10.1006/abio.1999.4099. [DOI] [PubMed] [Google Scholar]
- Porten K, Rissland J, Tigges A, Broll S, Hopp W, Lunemann M, van Treeck U, Kimmig P, Brockmann SO, Wagner-Wiening C, Hellenbrand W, Buchholz U. A super-spreading ewe infects hundreds with Q fever at a farmers' market in Germany. BMC Infect Dis. 2006;6:147. doi: 10.1186/1471-2334-6-147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lorenz H, Jager C, Willems H, Baljer G. PCR detection of Coxiella burnetii from different clinical specimens, especially bovine milk, on the basis of DNA preparation with a silica matrix. Appl Environ Microbiol. 1998;64:4234–4237. doi: 10.1128/aem.64.11.4234-4237.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Willems H, Thiele D, Frolich-Ritter R, Krauss H. Detection of Coxiella burnetii in cow's milk using the polymerase chain reaction (PCR) Zentralbl Veterinarmed B. 1994;41:580–587. doi: 10.1111/j.1439-0450.1994.tb00267.x. [DOI] [PubMed] [Google Scholar]
- Klee SR, Tyczka J, Ellerbrok H, Franz T, Linke S, Baljer G, Appel B. Highly sensitive real-time PCR for specific detection and quantification of Coxiella burnetii. BMC Microbiol. 2006;6:2. doi: 10.1186/1471-2180-6-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Denison AM, Thompson HA, Massung RF. IS1111 insertion sequences of Coxiella burnetii: characterization and use for repetitive element PCR-based differentiation of Coxiella burnetii isolates. BMC Microbiol. 2007;7:91. doi: 10.1186/1471-2180-7-91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smieja M, Mahony JB, Goldsmith CH, Chong S, Petrich A, Chernesky M. Replicate PCR testing and probit analysis for detection and quantitation of Chlamydia pneumoniae in clinical specimens. J Clin Microbiol. 2001;39:1796–1801. doi: 10.1128/JCM.39.5.1796-1801.2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Navarro E, Segura JC, Castano MJ, Solera J. Use of real-time quantitative polymerase chain reaction to monitor the evolution of Brucella melitensis DNA load during therapy and post-therapy follow-up in patients with brucellosis. Clin Infect Dis. 2006;42:1266–1273. doi: 10.1086/503035. [DOI] [PubMed] [Google Scholar]
- Kutyavin IV, Afonina IA, Mills A, Gorn VV, Lukhtanov EA, Belousov ES, Singer MJ, Walburger DK, Lokhov SG, Gall AA, Dempcy R, Reed MW, Meyer RB, Hedgpeth J. 3'-minor groove binder-DNA probes increase sequence specificity at PCR extension temperatures. Nucleic Acids Res. 2000;28:655–661. doi: 10.1093/nar/28.2.655. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Afonina IA, Reed MW, Lusby E, Shishkina IG, Belousov YS. Minor groove binder-conjugated DNA probes for quantitative DNA detection by hybridization-triggered fluorescence. Biotechniques. 2002;32:940–4, 946-9. doi: 10.2144/02324pf01. [DOI] [PubMed] [Google Scholar]
- Pham DG, Madico GE, Quinn TC, Enzler MJ, Smith TF, Gaydos CA. Use of lambda phage DNA as a hybrid internal control in a PCR-enzyme immunoassay to detect Chlamydia pneumoniae. J Clin Microbiol. 1998;36:1919–1922. doi: 10.1128/jcm.36.7.1919-1922.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rosenstraus M, Wang Z, Chang SY, DeBonville D, Spadoro JP. An internal control for routine diagnostic PCR: design, properties, and effect on clinical performance. J Clin Microbiol. 1998;36:191–197. doi: 10.1128/jcm.36.1.191-197.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Knepp JH, Geahr MA, Forman MS, Valsamakis A. Comparison of automated and manual nucleic acid extraction methods for detection of enterovirus RNA. J Clin Microbiol. 2003;41:3532–3536. doi: 10.1128/JCM.41.8.3532-3536.2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Henning K, Sting R. [Definitive ability of Stamp-staining, antigen-ELISA, PCR and cell culture for the detection of Coxiella burnetii] Berl Munch Tierarztl Wochenschr. 2002;115:381–384. [PubMed] [Google Scholar]