Effect of environmental conditions during heating on commercial spore strip performance.

G M Smith, M Kopelman, A Jones, I J Pflug
PMCID: PMC241960  PMID: 7125646

Abstract

Commercial biological indicator spore strips in glassine envelopes, produced by three manufacturers, were evaluated by fraction-negative procedures after being heated at 121.0 +/- 0.05 degrees C. Only one type of spore strip met the manufacturer's specifications. The strips of one manufacturer were further evaluated by fraction-negative and survivor curve-plate count procedures after being heated under several conditions (enclosed in glassine envelopes, in trypticase soy broth plus 0.0015% bromocresol purple, in Trypticase soy broth alone in Water for Injection, directly); Trypticase soy broth plus bromocresol purple and tryptic soy agar, respectively, were used as recovery media. The heating condition affected the D-value of the spore strip. Recovery procedures also had an effect; in all cases, the D-values obtained from the survivor curve tests were larger than those obtained from fraction-negative tests carried out under the same conditions. To determine if the differences in D-values between the two evaluation procedures were caused by the recovery media, we evaluated, by both methods, one type of spore strip heated directly and in glassine envelopes, using tryptic soy agar plus bromocresol purple and Trypticase soy broth plus 1.5% agar, respectively, as the recovery media. The survivor curve results showed that for both enclosed and unenclosed spore strips, there was a marked difference between the two recovery media; however, there was no difference when fraction-negative tests were used.

Full text

PDF
12

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bühlmann X., Gay M., Schiller I. Test objects containing Bacillus stearothermophilus spores for the monitoring of antimicrobial treatment in steam autoclaves. Results obtained with commercial preparations and with the authors' own test objects. Pharm Acta Helv. 1973 Apr;48(4):223–244. [PubMed] [Google Scholar]
  2. KELSEY J. C. The testing of sterilizers. 2. Thermophilic spore papers. J Clin Pathol. 1961 May;14:313–319. doi: 10.1136/jcp.14.3.313. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Mayernik J. J. Biological indicators for steam sterilization--a U.S.P. collaborative study. Bull Parenter Drug Assoc. 1972 Sep-Oct;26(5):205–211. [PubMed] [Google Scholar]
  4. Roberts T. A. Symposium on bacterial spores: VII. Recovering spores damaged by heat, ionizing radiations or ethylene oxide. J Appl Bacteriol. 1970 Mar;33(1):74–94. doi: 10.1111/j.1365-2672.1970.tb05235.x. [DOI] [PubMed] [Google Scholar]
  5. Selkon J. B., Sisson P. R., Ingham H. R. The use of spore strips for monitoring the sterilization of bottled fluids. J Hyg (Lond) 1979 Aug;83(1):121–125. doi: 10.1017/s0022172400025882. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Yokoya F., York G. K. Effect of several environmental conditions on the "thermal death rate" of endospores of aerobic, thermophilic bacteria. Appl Microbiol. 1965 Nov;13(6):993–999. doi: 10.1128/am.13.6.993-999.1965. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Applied and Environmental Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES