Blood gas analysis is a basic and useful laboratory test for the critical care of patients (Wiwanitkit 1999; Barthwal 2004). Arterial blood gas analysis is an essential investigation for assessing clinical oxygenation and acid-base status in critically ill patients (Wiwanitkit 1999; Barthwal 2004), providing information about ventilation, oxygenation, and acid-base status, the three closely interrelated physiological parameters that maintain pH homeostasis. The correct interpretation and application of arterial blood gas analysis requires knowledge of basic applied physiology in relation to these parameters (Wiwanitkit 1999; Barthwal 2004). Quality control of blood gas analysis is therefore important. In general, this quality control should follow the basic principles of good laboratory practice, namely, pre-analytical, focusing on proper specimen collection, handling, and transportation; analytical, focusing on internal quality control and external quality assessment; and post-analytical, comprising good validation and interpretation (Wiwanitkit 1999).
An important source of aberrations in blood gas analysis results are errors in the pre-analytical phase (Mollard 2000). Several deficiencies in pre-analytical variables in blood gas analysis have been identified, most of which are caused by negligence and which are easily corrected (Ancic and Munoz 1997). Specific requirements for storage and transport of specimens for blood gas analysis have been proposed (Burnett et al 1994). Delay in analysis can decrease oxygen partial pressure (PO2) and increase carbon dioxide partial pressure (PCO2) because of the metabolism of blood cells. Ice preservation is recommended; however, there is no reason to keep arterial blood in ice if the blood gas analysis is done within 30 minutes (Liss and Payne 1993). The effect of syringe material on collected blood in general clinical chemistry has also been reported, including the diffusion of chemicals across the tube (Hilty et al 1969; Scott et al 1971). For blood gas analysis, the classical method requires a glass syringe; however, the new plastic syringes have been developed to address the increasing problems of blood-borne transmitted diseases (Evers et al 1972).
Some reports indicate that glass syringes are superior to plastic syringes in preserving samples, especially for PO2 determination (Pretto and Rochford 1994; Deane et al 2004). However, there has been no specific explanation for this observation. Here, an explanation is attempted based on nanomaterial composition.
First, the size of the O2 molecule was calculated based on the chemical-bonding principle (Goldberg 1989). The size of one O2 molecule can be calculated by size of one O2 molecule = 2 × (O2− ion size) + (O-O bond length).
This equals 0.346 ([2 × 0.280] + 0.066) nm.
Then the pore size and pore density of the glass material (polymer of silicon dioxide, molecular weight = 28.09) and plastic (polypropylene, polymer of propylene, molecular weight = 132.16) were estimated. The pore size of glass and plastic material is equal to about 3–50 nm (pore radius = 1.5–25 nm) (Kin et al 1997) and 200–450 nm (pore radius = 100–225 nm) (Kin et al 1997), respectively. The pore densities of glass and plastic materials are equal to about 4 × 106 pores/cm2 (Diem and Lentner 1971; Kin et al 1997) and 2 × 108 pores/cm2 (Diem and Lentner 1971; Kin et al 1997), respectively. Therefore, the overall areas allowing diffusion, calculated by pore density × π × (pore radius)2, are estimated as 3 × 10−7−8 × 10−5 cm2/cm2 of glass and 6 × 10−2−3 × 10−1 cm2/cm2 of plastic, respectively.
At controlled temperature and other environmental factors, O2 seems to have a greater chance (4–150 times) to diffuse across plastic than glass, which could be a good explanation why a glass syringe can better preserve oxygen in a blood sample for blood gas analysis.
References
- Ancic P, Munoz F. Importance of pre-analytical variables in the quality of arterial blood gas determination. Rev Med Chil. 1997;125:1045–8. [PubMed] [Google Scholar]
- Barthwal MS. Analysis of arterial blood gases – a comprehensive approach. J Assoc Physicians India. 2004;52:573–7. [PubMed] [Google Scholar]
- Burnett RW, Covington AK, Fogh-Andersen N, et al. Recommendations on whole blood sampling, transport, and storage for simultaneous determination of pH, blood gases, and electrolytes. International Federation of Clinical Chemistry Scientific Division. J Int Fed Clin Chem. 1994;6:115–20. [PubMed] [Google Scholar]
- Deane JC, Dagleish MP, Benamou AE, et al. Effects of syringe material and temperature and duration of storage on the stability of equine arterial blood gas variables. Vet Anaesth Analg. 2004;31:250–7. doi: 10.1111/j.1467-2995.2004.00173.x. [DOI] [PubMed] [Google Scholar]
- Diem K, Lentner C. Scientific table. 7. Basle: CIBA GEIGY; 1971. [Google Scholar]
- Evers W, Racz GB, Levy AA. A comparative study of plastic (polypropylene) and glass syringes in blood-gas analysis. Anesth Analg. 1972;51:92–7. [PubMed] [Google Scholar]
- Goldberg DL. Bonding. In: Goldberg DL, editor. Chemistry. Singapore: McGraw Hill; 1989. pp. 95–118. [Google Scholar]
- Hilty H, Karendal B. Effect of syringe material on oxygen tension in stored blood. Acta Soc Med Ups. 1969;74:193–205. [PubMed] [Google Scholar]
- Kin PT, Wee A, Dan L, et al. Diagnostic and laboratory section. In: Kin PT, Wee A, Dan L, Guo J, editors. MIMS Medex (Medical Equipment, Diagnostics, and Supplies Index) Bangkok: TIMS Publication; 1997. pp. 1–148. [Google Scholar]
- Liss HP, Payne CP., Jr Stability of blood gases in ice and at room temperature. Chest. 1993;103:1120–2. doi: 10.1378/chest.103.4.1120. [DOI] [PubMed] [Google Scholar]
- Mollard JF. Pre-analytical precautions and appropriate materials for blood gas analysis. Ann Biol Clin (Paris) 2000;58:472–83. [PubMed] [Google Scholar]
- Pretto JJ, Rochford PD. Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures. Thorax. 1994;49:610–2. doi: 10.1136/thx.49.6.610. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Scott PV, Horton JN, Mapleson WW. Leakage of oxygen from blood and water samples stored in plastic and glass syringes. Br Med J. 1971;3:512–6. doi: 10.1136/bmj.3.5773.512. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wiwanitkit V. Quality control in blood gas analysis. Yasothon Medical Journal. 1999;2:160–6. [Google Scholar]