Abstract
A method was devised to determine the precision of the all-glass impinger and the Andersen six-stage microbial impactor over a wide range of aerosol concentrations like those found in facilities which process solid waste. Simultaneous samples were collected inside a municipal solid-waste recovery system, and the data were treated statistically to estimate the precision of each air-sampling device. All-glass impingers yielded colony counts which indicated a linear relationship between samplers over an observed aerosol concentration of 1.1 X 10(3) to 2.8 X 10(7) colony-forming units per m3 of air. Impactors also yielded colony counts which indicated a linear relationship over an observed aerosol concentration range of 3.9 X 10(3) to 1.9 X 10(5) colony-forming units per m3 of air. The coefficients of variation for the all-glass impinger and the six-stage impactor in an environment with a high and variable dust level were determined to be 0.38 and 0.23, respectively.